40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log  Not logged in ELOG logo
Entry  Wed Aug 9 12:35:47 2017, gautam, Update, ALS, Fiber ALS IMG_3942.JPG
    Reply  Mon Aug 14 16:24:09 2017, gautam, Update, ALS, Fiber ALS IMG_7471.JPGIMG_7472.JPGIMG_7473.JPGIMG_7474.JPG
       Reply  Wed Aug 16 20:24:23 2017, gautam, Update, ALS, Fiber ALS IMG_7476.JPGIMG_7477.JPGIMG_7478.JPGIMG_7479.JPG
          Reply  Wed Aug 23 17:22:36 2017, gautam, Update, ALS, Fiber ALS - reinstalled 
             Reply  Fri Aug 25 15:54:14 2017, gautam, Update, ALS, Fiber ALS noise measurement IMG_7497.JPGFOL_schematic.pdf20170825_IR_ALS.pdf
                Reply  Fri Aug 25 17:11:07 2017, rana, Update, ALS, Fiber ALS noise measurement 
                   Reply  Tue Sep 26 19:10:13 2017, gautam, Update, ALS, Fiber ALS setup neatened IMG_7605.JPG
                Reply  Sun Aug 27 11:57:31 2017, rana, Update, ALS, Fiber ALS noise measurement 
                   Reply  Tue Aug 29 02:08:39 2017, gautam, Update, ALS, Fiber ALS noise measurement 
                      Reply  Fri Sep 1 19:15:40 2017, gautam, Update, ALS, Fiber ALS noise measurement IR_ALS_noise.pdf
Message ID: 13255     Entry time: Fri Aug 25 17:11:07 2017     In reply to: 13254     Reply to this: 13333
Author: rana 
Type: Update 
Category: ALS 
Subject: Fiber ALS noise measurement 

Is it better to mount the box in the PSL under the existing shelf, or in a nearby PSL rack?

Quote:

 

Further characterization needs to be done, but the results of this test are encouraging. If we are able to get this kind of out of loop ALS noise with the IR beat, perhaps we can avoid having to frequently fine-tune the green beat alignment on the PSL table. It would also be ideal to mount this whole 1U setup in an electronics rack instead of leaving it on the PSL table

 