Finally I reallized what is killing the ETMY oplev laser. Wrong power supply, it was driving the HeNe laser by 600V higher voltage than recommended. Power supply 101T-2300Vdc replaced by 101T-1700Vdc ( Uniphase model 1201-1, sn 2712420 )
The laser head 1103P, sn P947049 lived for 120 days and it was replaced by sn P964431 New laser output 2.8 mW, quadrant sum 19,750 counts
|