Message ID: 13016     Entry time: Sat May 27 10:26:28 2017
Author: Kaustubh 
Type: Update 
Category: General 
Subject: Transimpedance Calibration 

Using Alberto's paper LIGO-T10002-09-R titled "40m RF PDs Upgrade", I calibrated the vertical axis in the bode plots I had obtained for the two PDs ET-3010 and ET-3040.

I am not sure whether the values I have obtained are correct or not(i.e. whether the calibration is correct or not). Kindly review them.

EDIT: Attached the formula used to calculate transimpedance for each data point and the values of other paramaters.

EDIT 2: Updated the plots by changing the conversion for gettin ghte ratio of the transfer functions from 10^(y/10) to 10^(y/20).

Attachment 1: ET-3040_test_transimpedance.pdf  38 kB  Uploaded Wed May 31 12:33:45 2017  | Hide | Hide all
ET-3040_test_transimpedance.pdf
Attachment 2: ET-3010_test_transimpedance.pdf  38 kB  Uploaded Wed May 31 12:33:45 2017  | Hide | Hide all
ET-3010_test_transimpedance.pdf
Attachment 3: Formula_for_Transimpedance.pdf  61 kB  Uploaded Wed May 31 12:34:47 2017  | Hide | Hide all
Formula_for_Transimpedance.pdf