1064 nm Semiconductor Laser Fiber Distribution System and Mirror Tomography
Below threshold these Semiconductor Fabry-Perot lasers have an axial mode structure with a spacing of about a THz. As you turn up the current to above threshold the first mode to oscillate saturates the gain down on all the modes and only it oscillates. The laser I have here in my office (a backup for the one you have at the 40 meter) has a wavelength of 1064.9 nm at 70 Degrees C. We should be able to temperature tune it down to 1064.3 nm although this could be a bit tedious the first time we do it. The specifications claim a "spectrum width" of 1.097 nm which I believe is the temperature tuning range. I don’t know what the line width is but it will be single frequency and we shouldn’t have mode hoping problems. So we should be able to use it in the “Mirror Tomography” experiment. You might want to use some sort of polarization diversity to avoid the problems of fiber polarization drift.
There have been 2 student projects on using the fiber distributed PD frequency response at1064 nm laser.
“Automated Photodiode Frequency Response Measurement System,” Alexander Cole - T1300618
“Final Report: Automated Photodiode Frequency Response Measurement System for Caltech 40m lab,” Nichin Sreekantaswamy - P140021
I’ll look up a few more references and add include them in the next elog.
Eric
|