Entry  Mon Mar 27 20:41:21 2017, rana, gautam, Update, PSL, PMC DAQ assay for feed-forward integration TEK00000.PNG20170327_194931.jpg20170327_204554.jpg
    Reply  Wed Mar 29 11:41:19 2017, gautam, Update, PSL, PMC DAQ assay for feed-forward integration C1ALS_PSLnamespace.pngPMCcheckout.pdfD980352-A-40m_151119.pdf
       Reply  Thu Mar 30 00:16:09 2017, gautam, Update, PSL, PSL NPRO PZT calibration 
          Reply  Mon Apr 3 17:25:13 2017, gautam, Update, PSL, PSL NPRO PZT calibration PDHerr.pdfNPROcalib.pdf
             Reply  Tue Apr 4 17:27:58 2017, rana, Update, PSL, PSL NPRO PZT calibration 
       Reply  Thu Mar 30 18:11:01 2017, rana, Update, PSL, PMC DAQ assay for feed-forward integration 
          Reply  Mon Apr 3 23:07:09 2017, gautam, Update, PSL, PMC DAQ assay for feed-forward integration 
             Reply  Tue Apr 11 00:38:37 2017, gautam, Update, PSL, PMC demod moved off servo board PMC_Ctrl_spec.pdf
                Reply  Wed Apr 12 00:36:53 2017, gautam, Update, PSL, PMC demod moved off servo board PMC_digitalSpec.pdfPMC_DAQ_spectra.pdf
                   Reply  Tue Apr 18 01:01:03 2017, gautam, Update, PSL, PMC OLTF measured, DAQ channels calibrated PMCspectra_calibrated.pdfPMC_MEDMs.png
                      Reply  Tue Apr 18 16:10:00 2017, gautam, Update, PSL, PMC OLTF measured, DAQ channels calibrated PMC_OLTF_170418.pdf
                         Reply  Tue Apr 18 23:37:15 2017, rana, Update, PSL, PMC OLTF measured, DAQ channels calibrated 
                         Reply  Wed Apr 19 15:13:30 2017, gautam, Update, PSL, PMC/MCL multicoherence PMC_MCL_multicoherence.pdf
Message ID: 12947     Entry time: Wed Apr 19 15:13:30 2017     In reply to: 12945
Author: gautam 
Type: Update 
Category: PSL 
Subject: PMC/MCL multicoherence 

I used a one hour stretch of data from last night to look at coherence between the PMC control signal and MCL, to see if the former can be used as a witness channel in some frequency band for MCL stabilization. Here is a plot of the predicted subtraction and coherence, made using EricQs pynoisesub code. I had thought about adopting the greedy channel ranking algorithm that Eric has been developing for noise subtraction in site data, but since I am just considering 3 witness channels, I figured this straight up comparison between different sets of witness channels was adequate. Looks like we get some additional coherence with MCL by adding the PMC control signal to the list of witness channels, there is about a factor of a few improvement in in the 1-2Hz band...  

Attachment 1: PMC_MCL_multicoherence.pdf  286 kB  | Hide | Hide all
PMC_MCL_multicoherence.pdf
ELOG V3.1.3-