40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log  Not logged in ELOG logo
Entry  Mon Jan 23 15:16:52 2017, gautam, Update, Optical Levers, ETMY Oplev HeNe needs to be replaced ETMY_Oplev.pdf
    Reply  Mon Jan 23 17:24:26 2017, Steve, Update, Optical Levers, ETMY Oplev HeNe is running hot ETMY_oplev__lasers.png
       Reply  Tue Jan 24 17:52:15 2017, Craig, Update, Optical Levers, ETMY Oplev HeNe is replaced 6x
          Reply  Wed Mar 1 08:28:04 2017, Steve, Update, Optical Levers, oplev laser RIN RIN_1103P.pngRIN_HeNe.png
             Reply  Thu Mar 2 13:59:04 2017, Steve, Update, Optical Levers, oplev laser RIN 1103P@2d.png
                Reply  Thu Mar 2 17:58:45 2017, rana, Update, Optical Levers, oplev laser RIN 
                   Reply  Fri Mar 3 17:24:21 2017, Steve, Update, Optical Levers, oplev laser RIN 3march17.pdf
                      Reply  Wed Mar 8 15:28:37 2017, Steve, Update, Optical Levers, oplev laser RIN RIN.jpgRIN_1103P_rotated.pdf
                      Reply  Mon Apr 10 09:49:32 2017, Steve, Update, Optical Levers, oplev laser RIN test planning 
                         Reply  Mon Apr 10 14:21:57 2017, rana, Update, Optical Levers, oplev laser RIN test planning 
       Reply  Tue May 30 13:36:58 2017, Steve, Update, Optical Levers, ETMY Oplev HeNe is replaced oplevETMY120d.png
          Reply  Tue Jan 15 18:16:00 2019, gautam, Update, Optical Levers, ETMY Oplev HeNe needs replacement OLRIN.pdfOLsums.png
             Reply  Fri Jan 18 12:52:07 2019, gautam, Update, Optical Levers, BS/PRM Oplev HeNe replaced OLRIN_20190118.pdf
Message ID: 12932     Entry time: Mon Apr 10 09:49:32 2017     In reply to: 12866     Reply to this: 12934
Author: Steve 
Type: Update 
Category: Optical Levers 
Subject: oplev laser RIN test planning 

We are planning to test 3 identical 1103Ps RIN with  continous temp monitoring and control later.

Selected  temp sensor   Platinum RTD 1PT100KN1515CLA   or           RTD-830 

Temp controller  with analoge output 0-10Vdc, CNi854  and external dc pulse driven  relay

Temperature Measurement Comparison Chart
Criteria Thermocouple RTD Thermistor
Temp Range -267°C to 2316°C -240°C to 649°C -100°C to 500°C
Accuracy Good Best Good
Linearity Better Best Good
Sensitivity Good Better Best
Cost Best Good Better

Order placed 4-12-17 for sensor  RTD-830,  controller CNi8-5-4 and relay  SSRL240DC25 = ~$500.

Still need: fuse, fuse housing, on/off switch, female  AC receptical, chassy box and AC power cord.

 

ELOG V3.1.3-