Entry  Mon Mar 13 16:47:41 2017, Steve, Update, PEM, Y-arm air conditon belts replaced 
    Reply  Mon Mar 27 16:01:55 2017, Steve, Update, PEM, X arm AC set to 68F 
Message ID: 12909     Entry time: Mon Mar 27 16:01:55 2017     In reply to: 12884
Author: Steve 
Type: Update 
Category: PEM 
Subject: X arm AC set to 68F 

 

The X arm air conditioner was not regulating properly. The arm temp was warmer than usual. I requested thermistor calibration.

The mechanic reset the thermostate to 68F last week.  It was 70-71F before.

 

The ETMX oplev laser now running 4 C lower at 30 C inside the enclousure.

The ETMX optical table top is 5 C cooler  at 21 C

The ETMX concrete wall temp 20 C at 9am with flow bench on.

ETMY conrete wall  temp 23 C at 9am