Yoichi and me found that the transfar function of the current shunt changed with the current of PA.
We changed PA current and fixed the unstability of ISS.
Now, laser power is stabilized finely, with band of about 1 Hz.
Yoich will post the stabilized noise spectrum.
There looks to be some non-linear relation between PA current and the TF of current shunt.
It had changed from the TF which we measured yesterday, so it might change again.
I try to write scripts to sweep PA current and measure the laser power and its rms automatically.
It will be apply for auto-adjustment of PA current.
Attached files are the transfar function of the current shunt with changing PA.
They have difference in lower frequency.
|