Message ID: 12732     Entry time: Wed Jan 18 12:34:21 2017     In reply to: 12723     Reply to this: 12757
Author: ericq 
Type: Summary 
Category: IOO 
Subject: MCL / MCF / Calibration 
Quote:

In the filter banks there were various version of a 'dewhite' filter. They were all approximately z=150, p=15, g =1 @ DC, but with ~1% differences. I don't trust their provenance and so I've enforced symmetry and fixed their names to reflect what they are (150:15).

The filters were made in response to a measurement of the pentek whitening boards in 2015 (ELOG 11550), but this level of accuracy probably isn't important.