- Updated the circuit diagrams:
IMC WFS Demodulator Board, Rev. 40m https://dcc.ligo.org/LIGO-D1600503
IMC WFS Whitening Board, Rev. 40m https://dcc.ligo.org/LIGO-D1600504
- Measured the noise levels of the whitening board, demodboard, and nominal free running WFS signals.
- IMC WFS demod phases for 8ch adjusted
Injected an IMC PDH error point offset (@1kHz, 10mV, 10dB gain) and adjusted the phase to have no signal in the Q phase signals.
- The WFS2 PITCH/YAW matrix was fixed
It was found that the WFS heads were rotated by 45 deg (->OK) in CW and CCW for WFS1 and 2, respectively (oh!), while the input matrices were identical! This made the pitch and yaw swapped for WFS2. (See attachment)
- Measured the TFs MC1/2/3 P/Y actuation to the error signals |