Message ID: 1250
Entry time: Fri Jan 23 14:00:02 2009
In reply to: 1248
|
Author: |
Yoichi |
Type: |
Update |
Category: |
PSL |
Subject: |
PMC transmission is down |
|
|
Quote: | The PMC transmission is going down.
I have not relocked the PMC yet. |
I tweaked the alignment to the PMC.
The transmission got back to 2.65. But it is still not as good as it was 3 days ago (more than 3).
It is interesting that the PMC transmission is inversely proportional to the NPRO output.
My theory is that the increased NPRO power changed the heat distribution inside the power amplifier.
Thus the output mode shape changed and the coupling into the PMC got worse.
MOPA output shows a peak around Jan-21, whereas the NPRO power was still climbing up.
This could also be caused by the thermal lensing decreasing the amplification efficiency. |
|
|