Message ID: 12385
Entry time: Tue Aug 9 13:53:57 2016
In reply to: 12365
Reply to this: 12389
|
Author: |
babbott |
Type: |
Update |
Category: |
SEI |
Subject: |
long Guralp EX cable repaired on the D-sub side |
|
|
I checked out the cable that I took from you, and all of the connections looked right. The only thing I did notice was that some of the soldered wires on the 37-pin connector had gotten hot enough to melt their insulation, and potentially short together. I cut off that connector, and left it on your desk to check out. I put on a new connector, and checked the pinout. If the Guralps still doesn't work, we'll have to check out other possibilities. |