Wed Nov 18 22:26:33 2015, yutaro, Update, Optical Levers, Calibration of oplevs for ITMY/ETMY    
|
Wed Nov 18 23:40:01 2015, rana, Update, Optical Levers, Calibration of oplevs for ITMY/ETMY
|
Thu Nov 19 16:06:54 2015, yutaro, Update, Optical Levers, Calibration of oplevs for ITMY/ETMY
|
Tue Dec 1 11:26:23 2015, yutaro, Update, Optical Levers, Calibration of oplevs for ITMX/ETMX   
|
Thu Dec 3 06:15:38 2015, rana, Update, Optical Levers, Calibration of oplevs for ITMX/ETMX
|
Thu Dec 3 10:05:07 2015, yutaro, Update, Optical Levers, Calibration of oplevs for ITMX/ETMX
|
Fri Dec 11 16:16:36 2015, yutaro, Update, Optical Levers, Calibration of oplevs for ITMX/ETMX
|
Mon Dec 14 23:49:03 2015, ericq, Update, Optical Levers, Calibration of oplevs for ITMX/ETMX
|
|
Message ID: 11881
Entry time: Mon Dec 14 23:49:03 2015
In reply to: 11875
|
Author: |
ericq |
Type: |
Update |
Category: |
Optical Levers |
Subject: |
Calibration of oplevs for ITMX/ETMX |
|
|
Quote: |
Based on calibration measurement I have done (elog 11785, 11831), I updated calibration factors of oplevs on medm screen as follows. Not to change loop gain oplev servo, I also changed oplev servo gain.
|
After making sure that the upper UGFs were properly in place, I saved these settings to the SDF files. Thanks Yutaro! |