Wed Nov 18 22:26:33 2015, yutaro, Update, Optical Levers, Calibration of oplevs for ITMY/ETMY    
|
Wed Nov 18 23:40:01 2015, rana, Update, Optical Levers, Calibration of oplevs for ITMY/ETMY
|
Thu Nov 19 16:06:54 2015, yutaro, Update, Optical Levers, Calibration of oplevs for ITMY/ETMY
|
Tue Dec 1 11:26:23 2015, yutaro, Update, Optical Levers, Calibration of oplevs for ITMX/ETMX   
|
Thu Dec 3 06:15:38 2015, rana, Update, Optical Levers, Calibration of oplevs for ITMX/ETMX
|
Thu Dec 3 10:05:07 2015, yutaro, Update, Optical Levers, Calibration of oplevs for ITMX/ETMX
|
Fri Dec 11 16:16:36 2015, yutaro, Update, Optical Levers, Calibration of oplevs for ITMX/ETMX
|
Mon Dec 14 23:49:03 2015, ericq, Update, Optical Levers, Calibration of oplevs for ITMX/ETMX
|
|
Message ID: 11831
Entry time: Tue Dec 1 11:26:23 2015
In reply to: 11785
Reply to this: 11842
11875
|
Author: |
yutaro |
Type: |
Update |
Category: |
Optical Levers |
Subject: |
Calibration of oplevs for ITMX/ETMX |
|
|
With the same method as reported in elog 11785, I calibrated oplevs for ITMX/ETMX.
RESULTS




According to this measurement, ratio of the calibration factor derived with this measurement (NEW) and the calibration factor for now (OLD), i.e. NEW/OLD was:
ETMX_PIT: 4.470
ETMX_YAW: 2.5970
ITMX_PIT: (-)1.1102
ITMX_YAW: 1.8173
NOTE
The calibration factors of the oplevs for ETMY/ITMY are NOT UPDATED YET. I updated on Dec 11, 2015
|
|
|
|
|
|
|
|
|