Thu Dec 4 17:42:23 2008, caryn, Update, IOO, drum modes observable without excitation 
|
Fri Dec 5 09:29:59 2008, rana, Update, IOO, drum modes observable without excitation
|
Fri Dec 5 21:31:11 2008, Yoichi, Update, IOO, drum modes observable without excitation
|
|
Message ID: 1176
Entry time: Thu Dec 4 17:42:23 2008
Reply to this: 1179
|
Author: |
caryn |
Type: |
Update |
Category: |
IOO |
Subject: |
drum modes observable without excitation |
|
|
So, the mode cleaner was evidently aligned better and now the drum modes are observable using DTT.
The Lock-In was set to 27.8kHz and the drum mode frequencies were previously observed to be 28.039kHz(MC2), 28.222kHz(MC3) and 28.221kHz(MC1). So, we might expect peaks at ~239Hz,421Hz,422Hz.
Peaks have been observed around the expected frequencies in channel IOO-MC-DRUM1.
Note that it is possible to resolve the separate MC1 and MC3 peaks which are so close together.