40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log  Not logged in ELOG logo
Entry  Thu Nov 5 02:18:32 2015, gautam, Update, LSC, FSR and linewidth measurements with phase tracker 
    Reply  Fri Nov 6 15:56:00 2015, gautam, Update, LSC, FSR and linewidth measurements with phase tracker Xcalib.pdfYcalib.pdfY_scan_log.pdf2015-11-05_phase_tracker_calib.dat.zip2015-11-04_y_arm_scan.dat.zip
       Reply  Mon Nov 9 11:34:51 2015, yutaro, Update, LSC, FSR and linewidth measurements with phase tracker plt.pngfitresult_and_code.zip
          Reply  Mon Nov 9 14:24:36 2015, yutaro, Update, LSC, FSR and linewidth measurements with phase tracker 
             Reply  Mon Nov 9 15:59:06 2015, ericq, Update, LSC, FSR and linewidth measurements with phase tracker 
                Reply  Mon Nov 9 16:58:59 2015, gautam, Update, LSC, FSR and linewidth measurements with phase tracker Y_scan.pdfmodDepth.pdfMatlab_code.zip
                   Reply  Tue Nov 10 02:34:28 2015, gautam, Update, LSC, Updated interpretation of peaks Y_scan.pdfmodDepth.pdf
                      Reply  Tue Nov 10 11:06:02 2015, yutaro, Update, LSC, Updated interpretation of peaks HOMlocation.pnghomfit.png
                         Reply  Tue Nov 10 11:40:03 2015, Koji, Update, LSC, Updated interpretation of peaks 
                            Reply  Tue Nov 10 16:34:00 2015, yutaro, Update, LSC, Updated interpretation of peaks homfit2.png
                               Reply  Fri Nov 13 17:33:39 2015, gautam, Update, LSC, g-factor measurements findGFactor.zip
                      Reply  Tue Nov 10 11:41:56 2015, Koji, Update, LSC, Updated interpretation of peaks 
       Reply  Fri Nov 13 15:48:16 2015, gautam, Update, LSC, Phase tracker calibration using Rubidium standard 6x
Message ID: 11747     Entry time: Tue Nov 10 11:40:03 2015     In reply to: 11746     Reply to this: 11749
Author: Koji 
Type: Update 
Category: LSC 
Subject: Updated interpretation of peaks 

What is the uncertainty of your RoC estimation?

One measurement of the ETMY ROC was 57.6m, but we trust another measured value of 60.26m than the other.
The value is always dependent on the spotposition on the mirror and how the ROC is calculated from the mirror phase map (e.g. spotsize, averaging method).
So I don't think this is a huge deviation from the spec.