40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log  Not logged in ELOG logo
Entry  Fri Sep 4 20:42:14 2015, gautam, rana, Update, CDS, Checkout of the Wenzel dividers TEK00000.PNGTEK00001.PNGTEK00002.PNG
    Reply  Tue Sep 29 03:14:04 2015, gautam, Update, CDS, Frequency divider box IMG_0014.JPGIMG_0015.JPG
       Reply  Fri Oct 9 19:54:58 2015, gautam, Update, CDS, Frequency divider box - installation in 1X2 rack IMG_0027.JPGtime_seris_25MHz.pdf
          Reply  Mon Oct 12 17:04:02 2015, gautam, Update, CDS, Frequency divider box - further tests calibration.pdfsystematics.pdf
             Reply  Wed Oct 14 17:40:50 2015, gautam, Update, CDS, Frequency divider box - further tests time_series_input_signals.pdfcalibration_20151012.pdfsystematics_20151012.pdf
                Reply  Tue Oct 20 17:36:01 2015, gautam, Update, CDS, Frequency counting with moving average 
                   Reply  Fri Oct 23 18:36:48 2015, gautam, Update, CDS, Frequency counting - workable setup prepared Yscan.pdf
                      Reply  Fri Oct 23 19:27:19 2015, Koji, Update, CDS, Frequency counting - workable setup prepared 
                         Reply  Sat Oct 24 12:34:43 2015, gautam, Update, CDS, Frequency counting - workable setup prepared Yscan.pdfFrequency_readout.pdf
                      Reply  Thu Nov 5 03:04:13 2015, gautam, Update, CDS, Frequency counting - systematics and further changes Systematic_error.pdfsystematics_origin.pdf
Message ID: 11712     Entry time: Sat Oct 24 12:34:43 2015     In reply to: 11710
Author: gautam 
Type: Update 
Category: CDS 
Subject: Frequency counting - workable setup prepared 

Sorry for the confusion - I did mean Green beat frequency, and I had neglected the factor of 2 in my earlier calculations. However, the fit parameter "c" in my fit was actually the half-width at half maximum and not the full width at half maximum. After correcting for both these errors (new fit is Attachment #1, where I have now accounted for the factor of 2, and the X axis is the IR beat frequency), I don't think the numbers change too much. It could be that the frequency counter wasn't reading out the frequency correctly, but looking at a time series plot of the frequency counter readout (Attachment #2), and my earlier trials, I don't think this is the case (38 MHz is a frequency at which I don't expect much systematic error - also, the offset was stepped from -3 to 3 over 45 seconds). 

The revised numbers:

Fitted linewidth = 2*c = 10.884 kHz +/- 2 Hz (95% C.I.)

FSR of Y arm (from elog 9804) = 3.9468 MHz +/- 1.1 kHz

=> Y arm Finesse = FSR/fitted linewidth =  362.6 +/- 0.5

Total round trip loss = 2*pi/Finesse = 0.0173

 

Attachment 1: Yscan.pdf  112 kB  | Hide | Hide all | Show all
Yscan.pdf