Author: Ignacio 
Type: Update 
Category: General 
Subject: More clear accelerometer huddle tests results 

I generated the following plots from the two sets of huddle test data we have for the accelerometers. 

Old data: 6 accelerometers, no cables clamped, no box, 55 mins worth of data.

New data: 3 accelerometers, cables clamped, foam box put on placed and completely sealed, 20 mins worth of data.

I made sure to use the same Impuse response time (6 sec) and sampling frequency (256 Hz), as well as every other parameter for the calculations.

  

 

Top left: The resultant self noise curve using the new data, there is definitely and improvement in the 0.5-2 Hz band. 

Top right: Resultant self noise using the old data, for the first set of three accelerometers.

Bottom left: Old data result for the remaining three accelerometers.

Bottom right: Old data result, using all six accelerometers as witnesses instead.

Attachment 1: new_data.png  60 kB  | Hide | Hide all
new_data.png
Attachment 2: new_data.png  60 kB  | Hide | Hide all
new_data.png
Attachment 3: old_data_1.png  59 kB  | Hide | Hide all
old_data_1.png
Attachment 4: old_data_1.png  59 kB  | Hide | Hide all
old_data_1.png
Attachment 5: old_data_2.png  60 kB  | Hide | Hide all
old_data_2.png