Fri Nov 14 19:20:42 2008, Yoichi, Update, PSL, Reference cavity ring down
|
Fri Nov 14 20:35:47 2008, rana, Update, PSL, Reference cavity ring down
|
Fri Nov 14 22:40:51 2008, Yoichi, Update, PSL, Reference cavity ring down
|
Mon Nov 17 15:07:06 2008, Yoichi, Update, PSL, Reference cavity ring down
|
|
Message ID: 1138
Entry time: Fri Nov 14 22:40:51 2008
In reply to: 1137
Reply to this: 1140
|
Author: |
Yoichi |
Type: |
Update |
Category: |
PSL |
Subject: |
Reference cavity ring down |
|
|
Quote: |
To make the DEI pulser make a fast pulse on the EO shutter EOMs, we had to make sure:
1) the cable had a high voltage rated dielectric. cheap dielectrics show the 'corona'
effect, especially when there is a bend in the cable.
|
I'll check it with Bob.
Quote: |
2) the EO has to have a resistor on it to prevent ringing due to the impedance mismatch.
|
Did you use a shunt or series resistor ?
If shunt, I guess it has to have a huge heat sink.
Actually, DEI says the pulser does not require any external shunt/series resistors or impedance-matching network.
Looks like it is not true ...
Quote: |
3) We needed ~3.5 kV to get the EO shutter crystal to flip the light by 90 deg.
|
Yes, I adjusted the voltage to maximize the power change and it was about 3.5kV. |