40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log  Not logged in ELOG logo
Entry  Fri Dec 26 17:45:21 2014, Steve, Update, VAC, vac pressure rose to 1.3 mTorr noN2.pngbacktoVacNormal.png
    Reply  Sun Dec 28 07:29:16 2014, Steve, Update, VAC, vacuum is normal 3d.png
    Reply  Tue Feb 3 15:20:13 2015, Steve, Koji, Update, VAC, Pneumatic pressure is being read again N2pressureReadingIsBack.png
    Reply  Sun Mar 22 18:42:40 2015, Steve, Update, VAC, vac pressure rose to 1.2mTorr NoN2.pngPressureReadingWorks.png
       Reply  Mon Mar 23 10:36:55 2015, ericq, Update, VAC, Pressure watch script 
          Reply  Fri Apr 24 17:30:32 2015, Jenne, Update, VAC, Pressure watch script broken 
             Reply  Sat Apr 25 18:50:47 2015, ericq, Update, VAC, Pressure watch script broken 
                Reply  Sat Apr 25 22:17:49 2015, rana, Update, CDS, MXstream restart script working (beta) 
                   Reply  Tue May 19 17:03:34 2015, Koji, Update, CDS, MXstream restart script working (beta) 
                      Reply  Tue May 19 18:38:33 2015, rana, Update, CDS, MXstream restart script working (beta) 
          Reply  Mon Apr 27 14:37:55 2015, Steve, Update, VAC, N2 pneumatic pressure watch  
             Reply  Thu Apr 30 16:30:25 2015, Steve, Update, VAC, N2 pneumatic pressure watch set up 
          Reply  Tue Jan 31 14:14:29 2017, rana, Update, VAC, Pressure watch script 
       Reply  Sun May 15 19:48:08 2016, Steve, Update, VAC, run out of N2 noN2.png
          Reply  Tue Jun 14 10:13:34 2016, Steve, Update, VAC, temporary N2 supply line afterN2work.png
             Reply  Fri Jun 17 15:22:31 2016, gautam, Update, VAC, N2 supply line restored after retiling c0vac_06172016.png
Message ID: 11264     Entry time: Thu Apr 30 16:30:25 2015     In reply to: 11260
Author: Steve 
Type: Update 
Category: VAC 
Subject: N2 pneumatic pressure watch set up 

We have 2 transduser PX303-3KG5V   http://www.omega.com/pressure/pdf/PX303.pdf They  will be installed on the out put of the N2 cylinders to read the supply pressure.

I will order one DC power supply  http://www.omega.com/pptst/PSU93_FPW15.html      PSU-93

One full cylinder pressure is ~ 2400 PSI max so two of them will give us ~9Vdc

The email reminder should be send at 1000 PSI  =  1.8 V

 

 

 

 