Entry  Tue Oct 4 19:31:17 2011, kiwamu, Update, SUS, PRM and BS oplev laser died oplevs.png
    Reply  Wed Oct 5 11:33:25 2011, steve, Update, SUS, PRM and BS oplev laser replaced 
       Reply  Mon Apr 20 16:42:48 2015, steve, Update, SUS, PRM and BS oplev laser replaced newOplevLaser.png
          Reply  Fri Apr 24 23:40:15 2015, rana, Update, SUS, PRM and BS oplev laser replaced BSOL.pdf
             Reply  Sun Apr 26 15:34:34 2015, Jenne, Update, SUS, PRM oplev centered 
             Reply  Tue Jun 9 10:57:12 2015, Steve, Update, SUS, BS oplev laser tuned 100dBStrend0425.png18dBSdrift.png
Message ID: 11227     Entry time: Mon Apr 20 16:42:48 2015     In reply to: 5620     Reply to this: 11246
Author: steve 
Type: Update 
Category: SUS 
Subject: PRM and BS oplev laser replaced 

The laser below is dead. JDSU 1103P, SN P845655 lived for 3.5 years.

Quote:

 

JDSU 1103P died after 4 years of service. It was replaced with new identical head of 2.9 mW output. The power supply was also changed.

The return spots of 0.04 mW  2.5 mm diameter on qpds are BS  3,700 counts and PRM 4,250 counts.

 

It was replaced by JDSU P/N 22037130,( It has a new name for 1103P Uniphase ) sn P919639 of mfg date 12-2014

Beam shape at 5 m nicely round. Output power 2.8 mW of 633 nm

BS spot size on qpd ~1 mm &  60 micro W

PRM spot size on qpd ~1 mm & 50 micro W

Attachment 1: newOplevLaser.png  57 kB  | Show | Show all