Better elog tomorrow - notes for now:
REFL165 for PRMI has been "a champ" (quote from Q). We're able to sit on ALS at average arm powers of 30ish. Nice.
Some ALSfool work - measured cancellation almost as good as single arm.
One time transitioned CARM -> normalized REFL55I
Many times did DARM -> normalized AS55Q, see lots of noise at 39ish Hz - may be coupling from MICH??
Arm ASC loops helped improve dark port contrast.
Note to selfs: Need to make sure DTT templates have correct freq ordering - must be small freqs to large freqs. |