40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log  Not logged in ELOG logo
Message ID: 11015     Entry time: Thu Feb 12 15:21:37 2015
Author: ericq 
Type: Update 
Category: Computer Scripts / Programs 
Subject: netgpib updates 

I've fixed the gpib scripts for the SR785 and AG4395A to output data in the same format as expected by older scripts when called by them. In addition, there are now some easier modes of operation through the measurement scripts SRmeasure and AGmeasure. These are on the $PATH for the main control room machines, and live in scripts/general/netgpib


Case 1: I manually set up a measurement on the analyzer, and just want to download / plot the data. 

Make sure you have a yellow prologix box plugged in, and can ping the address it is labeled with. (i.e. 'vanna'). Then, in the directory you want to save the data, run:

SRmeasure -i vanna -f mydata --getdata --plot

This saves mydata_(datetime).txt and mydata_(datetime).pdf in the current directory. 

In all cases, AGmeasure has the identical syntax. If the GPIB address is something other than 10, specifiy it with -a, but this is rarely the case.


Case 2: I want to remotely specify a measurement

Rather than a series of command line arguments, which may get lost to the mists of time, I've set the scripts up to use parameter files that serve as arguments to the scripts. 

Get the templates for spectrum and TF measurements in your current directory by running

SRmeasure --template

Set the parameters with your text editor of choice, such as frequency span, filename output, whether to create a plot or not, then run the measurement:

SRmeasure SR785template.yml


Case 3: I want to compare my data with previous measurements

In the template parameter files, there is an option 'plotRefs', that will automatically plot the data from files whose filenames start with the same string as the current measurement.

If, in the "#" commented out header of the data file, there is a line that contains "memo:" or "timestamp:", it will include the text that follows in the plot legend. 


There are also methods to remotely trigger an already configured measurement, or remotely reset an unresponsive instrument. Options can be perused by looking at the help in SRmeasure -h

I've tested, debugged, and used them for a bit, but wrinkles may remain. They've been svn40m committed, and I also set up a separate git repository for them at github.com/e-q/netgpibdata