40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log  Not logged in ELOG logo
Entry  Wed Nov 5 01:58:54 2014, ericq, Update, LSC, 3F RFPD RF spectra Nov52014_3fPD_DRMIspectra.zipREFL165_DRMIspectrum.pngREFL33_DRMIspectrum.png
    Reply  Wed Nov 5 11:09:44 2014, Koji, Update, LSC, 3F RFPD RF spectra 
       Reply  Wed Nov 5 22:25:42 2014, ericq, Update, LSC, 3F RFPD RF spectra REFL165_demod_DRMIspectrum.png
          Reply  Thu Nov 6 01:58:55 2014, Koji, Update, LSC, 3F RFPD RF spectra 
             Reply  Thu Nov 6 11:49:58 2014, ericq, Update, LSC, 3F RFPD RF spectra 
                Reply  Thu Nov 6 14:41:49 2014, Koij, Update, LSC, 3F RFPD RF spectra 
                   Reply  Fri Nov 7 02:21:12 2014, ericq, Update, LSC, 3F RFPD RF spectra 
                      Reply  Fri Nov 7 14:41:18 2014, ericq, Update, LSC, 3F RFPD RF spectra 165_ampSaturation.png165_HPatPD.png
                         Reply  Fri Nov 7 16:15:53 2014, Jenne, Update, LSC, 3F RFPD RF spectra 
                            Reply  Sat Nov 8 11:35:05 2014, rana, Update, LSC, 3F RFPD RF spectra 
                            Reply  Mon Nov 10 18:11:57 2014, ericq, Update, LSC, 3F RFPD RF spectra 165signals.pdf
Message ID: 10686     Entry time: Fri Nov 7 16:15:53 2014     In reply to: 10685     Reply to this: 10689   10692
Author: Jenne 
Type: Update 
Category: LSC 
Subject: 3F RFPD RF spectra 

I have found an SHP-150, but no SHP-175's (also, several 200's, and a couple of 500's).

Why do you say the SHP-150 isn't enough?  The blue peak at 10*fmod in your plot looks like it's at -12 dBm.  -13 dB on top of that will leave that peak at -25 dBm.  That should be enough to keep us from saturation, right?  It's not a lot of headroom, but we can give it a twirl until a 175-er comes in.  

Koji also suggests putting in a 110 MHz notch, combined with either an SHP-150 or SHP-175, although we'll have to measure the combined TF to make sure the notch doesn't spoil the high pass's response too much.

Quote:

165_HPatPD.png


Minicircuits' SHP-150 only has 13dB suppression at 110MHz, which would not be enough either. SHP-175 has 31dB suppression at 110MHz and 0.82dB at 160MHz, maybe this is what we want.

 

ELOG V3.1.3-