Message ID: 10392     Entry time: Thu Aug 14 19:33:00 2014
Author: Jenne 
Type: Configuration 
Category: IOO 
Subject: Moved MC2 spot 

Last night, and again just now, I used the ./MC2_spot_[direction] scripts to center the MC2 spot on the trans QPD.  The MCWFS handled overall alignment to correct for the fact that the ratios in the script aren't perfect.  When I was finished, I ran the MC WFS relief script from the WFS screen.  Last night, and again today, things had drifted until the yaw spot was more than 0.5 counts off.