40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log  Not logged in ELOG logo
Entry  Mon Aug 11 21:03:48 2014, ericq, rana, Summary, IOO, MC demod measurement 
    Reply  Mon Aug 11 22:07:31 2014, Koji, Summary, IOO, MC demod measurement 
    Reply  Mon Aug 11 23:32:54 2014, ericq, Summary, IOO, MC demod measurement MCDemod.pdfdemodLP.txt
Message ID: 10363     Entry time: Mon Aug 11 21:03:48 2014     Reply to this: 10364   10365
Author: ericq, rana 
Type: Summary 
Category: IOO 
Subject: MC demod measurement 

We measured the TF of the MC Demod board today.

We set the Marconi to +3dBm and drove the PD IN port of the demod board, starting at 29.5 MHz. Then we looked at the beat signal amplitude in the output of the demod board. So this is a transfer function but with mag only. Plots from Q below.

Rana took the demod board out and took pictures of it. Inside, the post mixer low pass is a SCLF-5 from mini-circuits. This has a lot of cutoff down low. Since the purpose of this filter is only to cutoff the 2f-1f and the 3f-2f products, we need to have a lot of attenuation at 29.5 MHz. One day, we may want to re-instate that notch for the (3*f1- f_MC) beat frequency, but for now we want stability.

So, I recommend that we (Steve) get 3 each of the SCLF-10 and SCLF-10.7 from Mini-Circuits Tuesday morning. Maybe we can put them into a spare board?

Also, we should probably remove the 140kHz:70kHz lead filter which is in the MC servo board. Its out of date. I think it would be fine for us to get a 7-15 kHz UGF for the CM servo and the MC can basically do that already. Mainly we want to fix the high frequency shape to get more stability.

After the measurements and photos, we had to reset the MCWFS offsets to get the WFS to not break the lock. Seems very sensitive to offsets. Hopefully Andres will give us a new Gouy phase telescope.

ELOG V3.1.3-