Message ID: 10351     Entry time: Fri Aug 8 12:39:19 2014     In reply to: 10343     Reply to this: 10354
Author: ericq 
Type: Summary 
Category: IOO 
Subject: MC servo analysis 

I have measured the current boosted MC CLG below 100kHz with an SR785. Swept sine only could get me down to 10kHz, but I was able to get down to 5kHz with a noise-injection measurement. 

MCloopAug8.pdf

I am attaching the SR785 outputs, which are in dB and Degrees. Additionally I pruned the areas of bad coherence out of these, and merged them to provide data files for the CLG and OLG in Real,Imaginary format.

Attachment 1: mcLoopAug8.zip  64 kB