Entry  Fri Aug 1 01:57:24 2014, Koji, Summary, IOO, MC auto locker 
    Reply  Fri Aug 1 08:55:34 2014, Koji, Summary, IOO, MC auto locker MC.png
Message ID: 10317     Entry time: Fri Aug 1 01:57:24 2014     Reply to this: 10319
Author: Koji 
Type: Summary 
Category: IOO 
Subject: MC auto locker 

To make MC auto locker running correctly, mcdown and mcup were revised

I tried it by unlocking MC several times. It seems OK. Let's see how it works.


Nominal gains for locking (to be taken care by mcdown)

C1:IOO-MC_REFL_GAIN
was 16 and is 19 now.

C1:IOO-MC_VCO_GAIN
was 9 and is 9 now too.

C1:PSL-FSS_MGAIN
was missing and now +13

C1:PSL-FSS_FASTGAIN
was +23.5 and is now +20.0

Nominal gains for operation ( to be taken care by mcup.

C1:IOO-MC_REFL_GAIN
was 19 and is 19 now too.

C1:IOO-MC_VCO_GAIN
was 25 and now uses ezcastep (ezcastep C1:IOO-MC_VCO_GAIN=9 +1,16 -s 0.1)

C1:PSL-FSS_MGAIN
C1:PSL-FSS_FASTGAIN

ezcawrite C1:PSL-FSS_MGAIN `ezcaread -n C1:PSL-STAT_FSS_NOM_C_GAIN`
ezcawrite C1:PSL-FSS_FASTGAIN `ezcaread -n C1:PSL-STAT_FSS_NOM_F_GAIN`

 

C1:PSL-STAT_FSS_NOM_C_GAIN`  is +18
C1:PSL-STAT_FSS_NOM_F_GAIN`   is +20

ELOG V3.1.3-