40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  40m Log  Not logged in ELOG logo
Message ID: 10282     Entry time: Mon Jul 28 17:25:32 2014
Author: Harry 
Type: Update 
Category: General 
Subject: Fiber Mode With Collimators 

 Purpose

We want a measurement of the fiber modes at either end, with the collimators, because these will be the modes that we'll be trying to match in order to couple light into the fibers, for FOL and/or future projects.

Measurement

In order to measure these modes, I used the beam profiler (Thorlabs BP 209-VIS) to take measurements of the beam diameter (cut off at 13.5% of the amplitude) along the optical axis, for each of the fiber ends.

The ends are arbitrarily labelled End 1 and End 2.

For each measurement, the fibers were coupled to roughly 30%, or 25mW at the output.

Regarding the issue of free rotation in the collimator stages: while End 1 was relatively stable, End 2 tended to move away from its optimal coupling position. In order to correct for this, I chose a position where coupling was good, and repositioned the stage to that coordinate (124 degrees) before taking each measurement.

The data were then entered into A La Mode, which gave waist measurements as follows:

End 1--- X Waist: 197um at Z = 4.8mm       Y Waist: 190um at Z = 13.6mm

End 2--- X Waist: 192um at Z = 7.4mm       Y Waist: 190um at Z = 6.0mm

end1Profiles.pngend2Profiles.png

A La Mode code is attached in .zip file

Moving Forward

These are the types of profiles that we will hopefully be matching the PSL and AUX lasers to, for use in frequency offset locking.

More characterization of the fibers is to follow, including Polarization Extinction Ratio.

We also hope to be testing the overall setup soon.

 

 

Attachment 3: FiberModeWCollimators.zip  14 kB
ELOG V3.1.3-