ID |
Date |
Author |
Type |
Category |
Subject |
52
|
Thu Jun 17 22:03:51 2010 |
James K | Misc | Hartmann sensor | SURF Log -- Day 2, Getting Started |
For Thursday, June 17:
Today I attended a basic laser safety training orientation, the second Introduction to LIGO lecture, a Summer Research Student Safety Orientation, and an Orientation for Non-Students living on campus (lots of mandatory meetings today). I met with Dr. Willems and Dr. Brooks in the morning and went over some background information regarding the project, then in the afternoon I got an idea of where I should progress from here from talking with Dr. Brooks. I read over the paper "Adaptive thermal compensation of test masses in advanced LIGO" and the LIGO TCS Preliminary Design document, and did some further reading in the Brooks thesis.
I'm making a little bit of progress with accessing the Hartmann lab computer with Xming but got stuck, and hopefully will be able to sort that out in the morning and progress to where I want to be (I wasn't able to get much further than that, since I can't access the Hartmann computer in the lab currently due to laser authorization restrictions). I'm currently able to remotely open an X terminal on the server but wasn't able to figure out how to then be able to log in to the Hartmann computer. I can do it via SSH on that terminal, of course, but am having the same access restrictions that I was getting when I was logging in to the Hartmann computer via SSH directly from my laptop (i.e. I can log in to the Hartmann computer just fine, and access the camera and framegrabber programs, but for the vast majority of the stuff on there, including MATLAB, I don't have permissions for some reason and just get 'access denied'). I'm sure that somebody who actually knows something about this stuff will be able to point out the problem and point me in the right direction fairly quickly (I've never used SSH or the X Window system before, which is why it's taking me quite a while to do this, but it's a great learning experience so far at least).
Goals for tomorrow: get that all sorted out and learn how to be able to fully access the Hartmann computer remotely and run MATLAB off of it. Familiarize myself with the camera program. Set the camera into test pattern mode and use the 'take' programs to retrieve images from it. Familiarize myself with the 'take' programs a bit and the various options and settings of them and other framegrabber programs. Get MATLAB running and use fread to import the image data arrays I take with the proper data representation (uint16 for each array entry). Then, set the camera back to recording actual images, take those images from the framegrabber and save them, then import them into MATLAB. I should familiarize myself with the various settings of the camera at this stage, as well.
--James |
51
|
Thu Jun 17 07:40:07 2010 |
James K | Misc | Hartmann sensor | SURF Log -- Day 1, Getting Started |
For Wednesday, June 16:
I attended the LIGO Orientation and first Introduction to LIGO lecture in the morning. In the afternoon, I ran a few errands (got keys to the office, got some Computer Use Policy Documentation done) and toured the lab. I then got Cygwin installed on my laptop along with the proper SSH packets and was successfully able to log in to and interact with the Hartmann computer in the lab through the terminal, from the office. I have started reading relevant portions of Dr. Brooks' thesis and of "Fundamentals of Interferometric Gravitational Wave Detectors" by Saulson. |
50
|
Wed Jun 16 11:47:11 2010 |
Aidan | Misc | Hartmann sensor | Spot displacement maps - temperate sensitivity tests - PRISM |
I think that the second plot is just showing PRISM and converting it to its radial components. This is due to the fact that the sign of the spot displacement on the LHS is the opposite of the sign of the spot displacement on the RHS. For spherical or cylindrical power, the sign of the spot displacement should be the same on both the RHS and the LHS.
I've attached a Mathematica PDF that illustrates this.
Quote: |
Results of initial measurement of temperature sensitivity of Hartmann sensor
"Cold" images were taken at the following temperature:
| before | 32.3 | 45.3 | 37.0 |
| after | 32.4 | 45.6 | 37.3 |
"Hot" Images were taken at the following temperature:
| before | 36.5 | 48.8 | 40.4 |
| after | 36.4 | 48.8 | 40.4 |
"before" are the temperatures just before taking 5000 images, and "after" are
the temperatures just after taking the images. First column is the temperature
measured using the IR temp sensor pointed at the heat sink, the second column the
camera temperature, and the third column the sensor board temperature.
Temperature change produced by placing a "hat" over the top of the HP assembly and the top of the heatsinks.
Averaged images "cool" and "hot" were created using 200 frames (each).
Aberration parameter values are as follows:
Between cool and hot images (cool spots - hot spots)
p: 4.504320557133363e-005
al: 0.408248504544179
phi: 0.444644135542724
c: 0.001216006036395
s: -0.002509569677737
b: 0.054773177423349
be: 0.794567342929695
a: -1.030687344054648
Between cool images only
p: 9.767143368103721e-007
al: 0.453972584677992
phi: -0.625590459774765
c: 2.738206187344315e-004
s: 1.235384158257808e-006
b: 0.010135170457321
be: 0.807948378729832
a: 0.256508288049258
Between hot images only
p: 3.352035441252169e-007
al: -1.244075541477539
phi: 0.275705676833192
c: -1.810992355666772e-004
s: 7.076678388064736e-005
b: 0.003706221758158
be: -0.573902879552339
a: 0.042442307609231
Attached are two contour plots of the radial spot displacements, one between
cool and hot images, and the other between cool images only. The color
bars roughly indicate the values of maximum and minimum spot
displacements.
Possible causes:
1. anisotropy of the thermal expansion of the invar foil HP caused by the rolling
2. non-uniform clamping of the HP by the clamp plate
3. vertical thermal gradient produced by the temperature raising method
4. buckling of the HP due to slight damage (dent)
|
|
Attachment 1: Prism_as_radial_vector.pdf
|
|
49
|
Tue Jun 15 16:30:10 2010 |
Peter Veitch | Misc | Hartmann sensor | Spot displacement maps - temperate sensitivity tests |
Results of initial measurement of temperature sensitivity of Hartmann sensor
"Cold" images were taken at the following temperature:
| before | 32.3 | 45.3 | 37.0 |
| after | 32.4 | 45.6 | 37.3 |
"Hot" Images were taken at the following temperature:
| before | 36.5 | 48.8 | 40.4 |
| after | 36.4 | 48.8 | 40.4 |
"before" are the temperatures just before taking 5000 images, and "after" are
the temperatures just after taking the images. First column is the temperature
measured using the IR temp sensor pointed at the heat sink, the second column the
camera temperature, and the third column the sensor board temperature.
Temperature change produced by placing a "hat" over the top of the HP assembly and the top of the heatsinks.
Averaged images "cool" and "hot" were created using 200 frames (each).
Aberration parameter values are as follows:
Between cool and hot images (cool spots - hot spots)
p: 4.504320557133363e-005
al: 0.408248504544179
phi: 0.444644135542724
c: 0.001216006036395
s: -0.002509569677737
b: 0.054773177423349
be: 0.794567342929695
a: -1.030687344054648
Between cool images only
p: 9.767143368103721e-007
al: 0.453972584677992
phi: -0.625590459774765
c: 2.738206187344315e-004
s: 1.235384158257808e-006
b: 0.010135170457321
be: 0.807948378729832
a: 0.256508288049258
Between hot images only
p: 3.352035441252169e-007
al: -1.244075541477539
phi: 0.275705676833192
c: -1.810992355666772e-004
s: 7.076678388064736e-005
b: 0.003706221758158
be: -0.573902879552339
a: 0.042442307609231
Attached are two contour plots of the radial spot displacements, one between
cool and hot images, and the other between cool images only. The color
bars roughly indicate the values of maximum and minimum spot
displacements.
Possible causes:
1. anisotropy of the thermal expansion of the invar foil HP caused by the rolling
2. non-uniform clamping of the HP by the clamp plate
3. vertical thermal gradient produced by the temperature raising method
4. buckling of the HP due to slight damage (dent) |
Attachment 1: spot_displacements_same_temp_0611.jpg
|
|
Attachment 2: spot_displacements_diff_temp_0611.jpg
|
|
48
|
Thu May 27 17:49:02 2010 |
Aidan | Electronics | Hartmann sensor | Hartmann sensor cooling fins added - base piece added |
Back to Configuration 1 again - this time the fins were bolted very securely to the camera.
7:25 PM - [about 2 hours later] - Digitizer = 39.7C, Sensor = 31.4C, Ambient = 19.0C
|
Attachment 1: HWS_CONFIG1-tight.jpg
|
|
47
|
Thu May 27 15:42:06 2010 |
Aidan | Electronics | Hartmann sensor | Hartmann sensor with just the base piece |
I switched in just the base piece of the Hartmann sensor. The cooling fins are removed. I bolted the camera securely to the base plate and I bolted the plate securely to the table.
5:00PM - (Digitizer = 41.9C, Sensor = 33.8C, Ambient = 19.3C)
|
Attachment 1: HWS_CONFIG4.jpg
|
|
46
|
Thu May 27 13:18:51 2010 |
Aidan | Electronics | Hartmann sensor | Removed Cooling fins from Hartmann sensor |
I removed the cooling fins from the Hartmann sensor to see what steady-state temperature it reached without any passive cooling elements. I also dropped the set-point temperature for the lab to help keep for getting too hot.
After nearly 3 hours the temperature is:
(Digitizer: 54.3C, Sensor: 46.6C, Ambient: 19.6C)
|
Attachment 1: HWS_CONFIG3.jpg
|
|
45
|
Thu May 27 08:25:37 2010 |
Aidan | Electronics | Hartmann sensor | Hartmann sensor cooling fins added - base piece removed |
8:10AM - I removed the base plate from the Hartmann sensor. I want to know what steady-state temperature the HWS achieves without the plate.
The photo below shows the current configuration.
11:22AM - (Digitizer - 52.2C, Sensor - 43.8C, Ambient - 21.8C) |
Attachment 1: HWS_CONFIG2.jpg
|
|
44
|
Wed May 26 14:58:04 2010 |
Aidan | Misc | Hartmann sensor | Hartmann sensor cooling fins added |
14:55 - Mindy stopped by with the copper heater spreaders and the cooling fins for the Hartmann sensor. We've set them all up and have turned on the camera to see what temperature above ambient is achieves.
17:10 - Temperature of the HWS with no active cooling( Digitizer = 44.1C, Sensor = 36.0C, Ambient = 21.4C)
|
Attachment 1: HWS_CONFIG1.jpg
|
|
43
|
Wed May 26 06:47:02 2010 |
Aidan | Laser | SLED | Switched off SLED - 6:40AM |
|
42
|
Mon May 24 19:17:32 2010 |
Aidan | Laser | Hartmann sensor | Replaced Brass Plate with Invar Hartmann Plate |
I just replaced the brass Hartmann plate with the Invar one. The camera was off during the process but has been turned on again. The camera is now warming up again. I've manually set the temperature in the EPICS channels by looking at the on-board temperature via the serial communications.
I also made sure the front plate was secured tightly. |
41
|
Thu May 20 17:08:36 2010 |
Aidan | Electronics | SLED | SLED module - and driver - LIGO D1000892 - and Hartmann sensor |
Verified that the test-point for the current limit pot on the driver (Wavelength Electronics - LDTC 0520) was at 0.5V. Driver is set to INTERNAL set point at the moment. This is down about 10% below the current limited point.
Voltage across TP7 and TP9 = 0.970V = LD Current Mon
Voltage across TP2 and TP3 = 0.017V = LD P Mon
--- Hartmann sensor ---
-set the sampling rate on the CCD to 16HZ. With the current alignment and intensity this gives as maximum intensity of around 3850 out of 4095. Thus the pixels are not saturated.
- centroid_image located some of the spots - see attached image of spots where those located by the algorithm and circled. I need to play with the threshold level and spot_radius to get this to work properly.
|
Attachment 1: 2010-05-20_hartmann_image_and_located_spot.jpg
|
|
40
|
Thu May 20 10:44:13 2010 |
Aidan | Electronics | SLED | QPhotonics 980nm source power |
dV = 0.385V
Responsivity= 0.65A/W
Transimpedance = 1.5E4 V/A
therefore power= 0.385V / (1.5E4 * 0.65 V/W) = 40uW
|
39
|
Thu May 20 08:20:54 2010 |
Aidan | Computing | Hartmann sensor | Centroiding algorithm and code to generate simulated data |
Here's a copy of an email I distributed today that describes the centroid and simulation code I wrote.
Hi Won,
I've written some code that generates an image of Gaussian spots and provides you with the coordinates of the centers used to generate those spots. There is the facility to turn on i) photo-electron shot noise, ii) random displacement of the nominal positions of the centers from a regular array and iii) 12-bit digitization to more accurately model the output from a CCD.
I've included an example routine that calls this function and then centroids those spots using a variant of your centroiding algorithm.
You should be able to use this to generate reliable simulated data to test versions of your centroiding algorithm.
Cheers,
Aidan.
Attached files:
1. test_spot_generation_and_centroiding.m - the example routine. Run this first
2. generate_simulated_spots.m - the function to generate the simulated spots in an image and as a set of positions
3. centroid_image.m - the function to centroid an image |
Attachment 1: test_spot_generation_and_centroiding.m
|
% example usage of generate_simulated_spots and centroid_image
clear all
close all
%% example 1 -
%----------------------------------------------------------
npixels = 1024; % the number of pixels in the image
... 143 more lines ...
|
Attachment 2: generate_simulated_spots.m
|
function output = generate_simulated_spots(npixels, digitizeFLAG, ...
IntensityNoiseFLAG, positionNoiseFLAG)
%
% a function to generate an image of spots to centroid and to provide the original
% locations of the spots.
%
% input
% -----
% npixels - pixels in output image
% digitizeFLAG: 0 - floating point array is output
... 160 more lines ...
|
Attachment 3: centroid_image.m
|
function centroids = centroid_image(image, centroids)
%
% This function centroids a supplied image. It returns a centroids structure
%
% 'centroids' structure
% --------------------
% centroids.image_background_level - the background intensity of an image
% with no illumination on it
% .spot_radius - the radius of a hartmann spot
% .spot_threshold_level - the minimum intensity of pixels used to
... 95 more lines ...
|
38
|
Tue May 18 09:33:44 2010 |
Aidan | Computing | EPICS | Added defocus and other Hartmann sensor channels to EPICS and DAQ |
I've added the following channels to the HWS softIoc in /cvs/cds/caltech/target/softIoc/HWS.db
EPICS and DAQ restart procedure
- Kill the existing softIoc. Use a "
ps -e | grep softIoc" command to determine the process id.
- After editing the HWS.db file restart the softIoc with the following command:
[controls@hartmann softIoc]$ /cvs/opt/epics-3.14.10-RC2-i386/base/bin/linux-x86/softIoc -S HWS.cmd &
[3] 11280
[controls@hartmann softIoc]$ dbLoadRecords "HWS.db"
iocInit
Starting iocInit
############################################################################
## EPICS R3.14.10- $R3-14-10-RC2$ $2008/10/10 15:01:51$
## EPICS Base built Oct 28 2009
############################################################################
iocRun: All initialization complete
3. Edit the /cvs/cds/caltech/chans/daq/C4TCS.ini file and kill the daqd process on fb1. It should restart automatically.
Done!
|
37
|
Mon May 17 19:41:13 2010 |
Aidan | Computing | Frame Grabber | C code that calls MATLAB engine and centroiding algorithms |
This is an amended version of simple_take.c.
The files below are all in the directory /opt/EDTpdv/hartmann/src
- simple_hartmann.c - the C code to access the frame grabber, retrieve an image, load the MATLAB engine and pass the image to MATLAB for centroiding
- centroid_image.m - the MATLAB routine that centroids the image
- get_defocus.m - the MATLAB function that determines the defocus in the centroids
- build_simple_hartmann.sh - a shell script I wrote that contains the compile and link options to build the thing correctly
|
Attachment 1: simple_hartmann.c
|
/**
* @file
* An example program to show usage of EDT PCI DV library to acquire and
* optionally save single or multiple images from devices connected to EDT
* high speed digital video interface such as the PCI DV C-Link or PCI DV
* FOX / RCX.
*
* Provided as a starting point example for adding digital video acquisition
* to a user application. Includes optimization strategies that take
* advantage of the EDT ring buffer library subroutines for pipelining image
... 521 more lines ...
|
Attachment 2: centroid_image.m
|
function centroids = centroid_image(image, centroids)
%
% This function centroids a supplied image. It returns a centroids structure
%
% 'centroids' structure
% --------------------
% centroids.image_background_level - the background intensity of an image
% with no illumination on it
% .spot_radius - the radius of a hartmann spot
% .spot_threshold_level - the minimum intensity of pixels used to
... 98 more lines ...
|
Attachment 3: get_defocus.m
|
function defocus = get_defocus(centroids)
%
% a function to extract the defocus of the gradient field
%
% 'centroids' structure
% --------------------
% centroids.image_background_level - the background intensity of an image
% with no illumination on it
% .spot_radius - the radius of a hartmann spot
% .spot_threshold_level - the minimum intensity of pixels used to
... 56 more lines ...
|
Attachment 4: build_simple_hartmann.sh
|
#!/bin/bash
gcc -O2 -I/opt/EDTpdv -I/apps/matlab_R2008b/extern/include -I/apps/matlab_R2008b/simulink/include -DMATLAB_MEX_FILE -c -D_GNU_SOURCE -fexceptions -I/apps/matlab_R2008b/extern/include -DMX_COMPAT_32 -O -DNDEBUG simple_hartmann.c
gcc -O2 -I/opt/EDTpdv -I/apps/matlab_R2008b/extern/include -I/apps/matlab_R2008b/simulink/include -DMATLAB_MEX_FILE -c -D_GNU_SOURCE -fexceptions -I/apps/matlab_R2008b/extern/include -DMX_COMPAT_32 -O -DNDEBUG /apps/matlab_R2008b/extern/src/mexversion.c
gcc -O2 -O -I/opt/EDTpdv -o /opt/EDTpdv/hartmann/bin/simple_hartmann simple_hartmann.o mexversion.o -L/opt/EDTpdv -lpdv -lpthread -lm -ldl -Wl,-rpath-link,/apps/matlab_R2008b/bin/glnxa64 -L/apps/matlab_R2008b/bin/glnxa64 -leng -lmx -lstdc++
|
36
|
Thu May 13 16:54:46 2010 |
Aidan | Computing | Hartmann sensor | Running MATLAB programs in C on CentOS - only use R2008b for less hassle |
After much effort trying to get a MATLAB routine to compile in C I discovered the following pieces of information.
1. CentOS will not install a gcc compiler more recent than 4.1.2 with yum install. This is circa 2007. If you want a more recent compiler it must be installed manually.
2. To compile and link C programs that call the MATLAB engine, they must be compiled in MATLAB using the mex command. Every version of MATLAB after R2008b requires the gcc compiler 4.2.3.
3. Building gcc 4.2.3 takes a lot more than 1 hour of compile time. I accidentally killed the build process and gave it up as a lost cause.
|
35
|
Tue May 11 10:32:00 2010 |
Aidan | Computing | Hartmann sensor | Peak detection and centroiding code - review |
This looks really efficient! However, I think there's a systematic error in the calculation. I tested it on some simulated data and it had trouble getting the centroids exactly right. I need to better understand the functions that are called to get an idea of what might be the problem.
Quote: |
Attached is .m file of the custom function that I wrote and used to automatically detect peaks in a Hartmann image,
and calculate the centroid corrdinates of each of those peaks.
A simple example of its usage, provided that myimage is a two-dimensional image array obtained from the camera, is
radius = 10;
peak_positions = detect_peaks_uml(myimage,radius);
no_of_peaks = length(peak_positions);
centroids_array = zeros(no_of_peaks);
for k = 1:no_of_peaks
centroids_array(k,1) = peak_positions(k).WeightedCentroid(1);
centroids_array(k,2) = peak_positions(k).WeightedCentroid(2);
end
I chose my value of radius by looking at spots in a sample image and counting the number of pixels across a peak. It may be
more useful to automatically obtain a value for the radius. I may run some tests to see how different choices of radius
affect the centroid calculations.
I may also need to add some error checking and/or image validating codes, but so far I have not encountered any problems.
Please let me know if anyone needs more explanation!
Won
|
|
Attachment 1: test_centroid_code.m
|
% generate an blank image
imarr = zeros(1024, 1024);
% get the background and intensity levels
bckgrd = 700.0;
I1 = 56000.0;
% get the 2D coordinate arrays
x = 1:1024;
... 61 more lines ...
|
34
|
Thu May 6 21:32:26 2010 |
Won Kim | Computing | Hartmann sensor | Peak detection and centroiding code |
Attached is .m file of the custom function that I wrote and used to automatically detect peaks in a Hartmann image,
and calculate the centroid corrdinates of each of those peaks.
A simple example of its usage, provided that myimage is a two-dimensional image array obtained from the camera, is
radius = 10;
peak_positions = detect_peaks_uml(myimage,radius);
no_of_peaks = length(peak_positions);
centroids_array = zeros(no_of_peaks);
for k = 1:no_of_peaks
centroids_array(k,1) = peak_positions(k).WeightedCentroid(1);
centroids_array(k,2) = peak_positions(k).WeightedCentroid(2);
end
I chose my value of radius by looking at spots in a sample image and counting the number of pixels across a peak. It may be
more useful to automatically obtain a value for the radius. I may run some tests to see how different choices of radius
affect the centroid calculations.
I may also need to add some error checking and/or image validating codes, but so far I have not encountered any problems.
Please let me know if anyone needs more explanation!
Won |
Attachment 1: detect_peaks_uml.m
|
function ctr = detect_peaks_uml(image,radius)
% Usage example:
% positions = detect_peaks_uml(myimage,10);
%
% total number of peaks detected: length(positions.WeightedCentroid)
% access the coordinates of the nth peak:
% positions(n).Weightedcentroid(1), positions(n).WeightedCentroid(2)
weighted_image = image .^ 2;
background = imopen(weighted_image,strel('disk',radius));
... 11 more lines ...
|
33
|
Thu May 6 12:32:11 2010 |
Aidan | Computing | Hartmann sensor | dalsa_to_epics Python script crashed ... |
Here's the error:
Traceback (most recent call last):
File "./dalsa_to_epics.py", line 81, in ?
stdout = subprocess.PIPE)
File "/usr/lib64/python2.4/subprocess.py", line 550, in __init__
errread, errwrite)
File "/usr/lib64/python2.4/subprocess.py", line 916, in _execute_child
errpipe_read, errpipe_write = os.pipe()
OSError: [Errno 24] Too many open files
[2]+ Exit 1 ./dalsa_to_epics.py (wd: ~/scripts)
(wd now: /cvs/users/abrooks/advLigo/HWS)
|
32
|
Thu May 6 10:34:38 2010 |
Aidan | Computing | Hartmann sensor | EPICS and MEDM screen for Hartmann sensor - part 2 |
I added the camera parameters to EPICS and the MEDM screen. These are available as channels now in EPICS and eventually there will be a python script that writes the EPICS value to those channels, but right now it is just a python script that reads the values off the Dalsa camera.
I updated the channels in /cvs/cds/caltech/chans/daq/C4TCS.ini so that these are saved to the daq and I also restarted the daq daemon.
The python script that gets the camera parameters is here: scripts/Dalsa1M60/GetCameraParameters.py and the script that writes the parameters to the EPICS channels is here scripts/dalsa_to_epics.py.
These are attached as is C4TCS.ini and HWS.db which defines the new channels. |
Attachment 1: dalsa_to_epics.py
|
#!/usr/bin/python
# Import the Dalsa1M60 packzge
import Dalsa1M60, subprocess
# define the serial command location
serial_cmd_location = '/opt/EDTpdv/serial_cmd'
# start a loop that continually gets the temperatures
getTemperatures = 1
... 75 more lines ...
|
Attachment 2: GetCameraParameters.py
|
#!/usr/bin/python
# NAME
# GetCameraParameters - a module for getting the Dalsa 1M60 parameters
#
# PACKAGE
# Part of the Dalsa1M60 python package
#
# SYNOPSIS
# GetCameraParameters( serial_cmd_location )
... 412 more lines ...
|
Attachment 3: HWS.db
|
record(ai,"C4:TCS-HWS_TEMP_DIGITIZER")
record(ai,"C4:TCS-HWS_TEMP_SENSOR")
record(ai, "C4:TCS-HWS_TAP1GAIN")
record(ai, "C4:TCS-HWS_TAP2GAIN")
record(ai, "C4:TCS-HWS_PRETRIGGER")
record(ai, "C4:TCS-HWS_DATA_MODE")
record(ai, "C4:TCS-HWS_BINNING_MODE")
record(ai, "C4:TCS-HWS_GAIN_MODE")
record(ai, "C4:TCS-HWS_OUTPUT_CONFIG")
record(ai, "C4:TCS-HWS_EXPOSURE_MODE")
... 27 more lines ...
|
Attachment 4: C4TCS.ini
|
[default]
dcuid=4
datarate=16
gain=1.0
acquire=1
ifoid=0
datatype=4
slope=1.0
offset=0
units=NONE
... 14 more lines ...
|
31
|
Wed May 5 18:45:51 2010 |
Aidan | Computing | Hartmann sensor | Python code to interface the Dalsa1M60 and export the temperature to EPICS |
Python script
I wrote a Python script, ~/scripts/dalsa_to_epics.py that reads the temperature off the camera using serial_cmd vt and then it writes this to the EPICS channels using ezcawrite. See attached. It is now running continuously in the background as dalsa_to_epics.
Dalsa1M60 baud rate
Also I accessed the menu of the 1M60 and changed the baud rate to 115200 using sbr 115200. Then I edited the dalsa_1m60.cfg file to set the baud rate to 115200 in that file. Finally, I changed the settings on the camera so that it will boot with the new baud rate when it is turned off and on again - this was with wus in the camera menu.
All the files are attached.
~/scripts/dalsa_to_epics.py
~/scripts/Dalsa1M60/VerifyTemperature.py
/opt/EDTpdv/camera_config/dalsa_1m60.cfg |
Attachment 1: dalsa_to_epics.py
|
#!/usr/bin/python
# Import the Dalsa1M60 packzge
import Dalsa1M60, subprocess
# define the serial command location
serial_cmd_location = '/opt/EDTpdv/serial_cmd'
# start a loop that continually gets the temperatures
getTemperatures = 1
... 18 more lines ...
|
Attachment 2: VerifyTemperature.py
|
#!/usr/bin/python
# part of the Dalsa1M60 package
# a module for verifying the temperature of the Dalsa 1M60
#
# The serial command 'vt' is sent to the camera. The camera responds as follow
s
# > vt
# Camera Temperature on Digitzer Board: 47.2 Celsius
# Camera Temperature on Sensor Board: 39.4 Celsius
... 65 more lines ...
|
Attachment 3: dalsa_1m60.cfg
|
#
# CAMERA_MODEL "Dalsa 1m60 config file (freerun)"
#
# camera name/description
#
camera_class: "Dalsa"
camera_model: "1M60"
camera_info: "12 bit dual channel camera link"
... 51 more lines ...
|
30
|
Wed May 5 09:04:01 2010 |
Aidan | Computing | Hartmann sensor | Added /home/controls/scripts/modules directory to PYTHONPATH on hartmann |
I added the following line to ~/.bashrc
export PYTHONPATH=/home/controls/scripts/modules:/usr/local/lib/python
This adds the above directory to PYTHONPATH and allows those modules in that directory to be access from anywhere.
|
29
|
Tue May 4 13:35:13 2010 |
Aidan | Computing | Hartmann sensor | Hartmann temperature channels in frame builder |
I've added the digitizer and sensor board temperature readings from the HWS to the frames. This was done in the following way
1. Create a new file /cvs/cds/caltech/chans/daq/C4TCS.ini - with the channels in it - see below
2. open /cvs/cds/caltech/target/fb1/master
3. add a line that includes the C4TCS.ini file when the frame builder starts
4. restart frame-builder by killing the daq daemon - kill <process id for daqd> (this is the only thing that needs to be entered as it will automatically restart)
C4TCS.ini
[default]
dcuid=4
datarate=16
gain=1.0
acquire=1
ifoid=0
datatype=4
slope=1.0
offset=0
units=NONE
[C4:TCS-HWS_TEMP_SENSOR]
[C4:TCS-HWS_TEMP_DIGITIZER]
|
28
|
Tue May 4 10:30:07 2010 |
Aidan | Computing | Hartmann sensor | EPICS and MEDM screen for Hartmann sensor |
I added the Dalsa 1M60 temperature measurements to EPICS. The break down is as follows:
|
Digitizer Board Temperature |
Sensor Board Temperature |
Dalsa 1M60 menu command |
vt |
vt |
Response from 1M60
|
Camera Temperature on Digitizer Board: 47.2 Celsius |
Camera Temperature on Sensor Board: 39.4 Celsius |
Menu accessed via |
MATLAB: unix('/opt/EDTpdv/serial_cmd vt') |
MATLAB: unix('/opt/EDTpdv/serial_cmd vt') |
Temperature stored in |
MATLAB: local variable called DBtemp (from the numerical sub-string) |
MATLAB: local variable called SBtemp (from the numerical sub-string) |
EPICS channel written via |
MATLAB: unix(['ezcawrite {channel-name} ' num2str(DBtemp)]) |
MATLAB: unix(['ezcawrite {channel-name} ' num2str(SBtemp)]) |
EPICS channel defined in |
HWS.db |
HWS.db |
Channel name |
C4:TCS-HWS_TEMP_DIGITIZER |
C4:TCS-HWS_TEMP_SENSOR |
I added a softIoc called HWS to /cvs/cds/caltech/target/softIoc. It added the channels following channels: C4:TCS-HWS_TEMP_DIGITIZER and C4:TCS-HWS_TEMP_SENSOR. The ioc (input/output controller) is run with the following command:
/cvs/opt/epics-3.14.10-RC2-i386/base/bin/linux-x86/softIoc HWS.cmd
although this doesn't execute it in the background. The MATLAB routine /home/controls/matlab_scripts/read_dalsa_temperature_write_to_epics.m is run continuously to access the serial port, get the temperature data and to write it to the EPICS channels. These were then available to read in the Hartmann sensor MEDM screen which is shown below. Also shown is a StripTool monitoring the temperatures. I had just turned off a fan that was cooling the 1M60 which is why the temperature is rising.
|
Attachment 1: Screenshot-C4HWS_medm_21.adl_(edited).png
|
|
Attachment 2: Screenshot-StripTool_Graph_Window.png
|
|
Attachment 3: HWS.db
|
record(ai,"C4:TCS-HWS_TEMP_DIGITIZER")
record(ai,"C4:TCS-HWS_TEMP_SENSOR")
|
Attachment 4: HWS.cmd
|
dbLoadRecords "HWS.db"
iocInit
|
Attachment 5: read_dalsa_temperature_write_to_epics.m
|
% get the temperature off the 1M60
% written by Aidan Brooks. 22nd Apr 2010
% define aliases
ezcawrite = '/cvs/opt/apps/Linux/gds/bin/ezcawrite';
ii = 1;
while ii == 1
[s, r] = unix('/opt/EDTpdv/serial_cmd vt');
... 54 more lines ...
|
27
|
Tue May 4 09:18:15 2010 |
Aidan | Computing | Hartmann sensor | Added aliases and icons for EPICS commands and dataviewer etc. to hartmann |
I updated the .bashrc file in controls@hartmann to include aliases for the ezca EPICS commands and a few others. Details shown below:
Also added launchers to the top panel for MATLAB, sitemap, dataviewer and StripTool. The icons for the launchers are located in:
/cvs/users/ops/ligo-launchers/icons
Changes to .bashrc
alias dv="/cvs/opt/apps/Linux/dataviewer/dataviewer"
alias StripTool = "/cvs/opt/apps/Linux/medm/bin/StripTool"
alias medm="/cvs/opt/apps/Linux/medm/bin/medm"
alias sitemap='medm -x /cvs/cds/caltech/medm/c2/atf/C2ATF_MASTER.adl'
# EPICS aliases
alias ezcademod="/cvs/opt/apps/Linux/gds/bin/ezcademod"
alias ezcaread="/cvs/opt/apps/Linux/gds/bin/ezcaread"
alias ezcaservo="/cvs/opt/apps/Linux/gds/bin/ezcaservo"
alias ezcastep="/cvs/opt/apps/Linux/gds/bin/ezcastep"
alias ezcaswitch="/cvs/opt/apps/Linux/gds/bin/ezcaswitch"
alias ezcawrite="/cvs/opt/apps/Linux/gds/bin/ezcawrite" |
26
|
Mon May 3 17:43:48 2010 |
Aidan | Computing | Frame Grabber | Successful image capture with EDT frame grabber |
I noticed that when i ran /opt/EDTpdv/camconfig and selected camera 331, which appeared to be closest to the Dalsa Pantera 1M60 camera, the software loaded the configuration file pantera11m4fr.cfg.
I tried to locate which entry in the camconfig list corresponded to the dalsa_1m60.cfg configuration file, but none of them seemed to. I couldn't select any entry and get it to report that it was using the 1m60 config file.
Next I noticed that there were 659 configuration files in the /opt/EDTpdv/camera_config directory but only 460 configuration options in camconfig. This seemed like 1/3 of the config files were somehow not formatted correctly, including,possibly the 1M60 config file.
By editing the pantera11m4fr.cfg I verified that the name of the camera, as it appears in the camconfig program, is the second line in the configuration file. For that file it was:
# CAMERA_MODEL "Dalsa Pantera 12 bit single channel camera link"
where the first line is just a single hash. The dalsa_1m60.cfg file did not have a name formatted in the same way as above: it was originally as shown below:
# Dalsa 1m60 config file (freerun)
so i changed the name in that configuration file to the following and it was suddenly available in the list when ./camconfig was run
# CAMERA_MODEL "Dalsa 1m60 config file (freerun)"
I selected that camera (number 53 in the list). Once this was done I ran pdv_flshow/pdvshow again the image that was displayed from the camera appeared to be correctlty demodulated.
Actually, the very first time i ran pdvshow the image was demodulated correctly but it appeared that the origin was offset and then the image wrapped around a little at the edges. However, every successive time I've run pdvshow since then I've had a perfectly demodulated image.
I ran some test patterns by changing the video mode using the serial communications menu in the camera. I also illuminated the Hartmann sensor with a torch/flashlight and got some spot patterns - see attached images.
Also, I've attached the dalsa_1m60.cfg file.
|
Attachment 1: 20100503_dalsa1m60_configuration_notes.txt
|
Configuring HWS to get image in CentOS
----------
9:34AM - Dalsa 1m60 turned on
----
$ /opt/EDTpdv
$ ./serial_cmd
%%this starts the serial communications device in the EDT FG but it isn't configured.
... 123 more lines ...
|
Attachment 2: 2010-05-03_dalsa1m60_image_test_pattern_and_spots.tif
|
|
Attachment 3: 2010-05-03_dalsa1m60_image_test_pattern_right_side.tif
|
|
Attachment 4: dalsa_1m60.cfg
|
#
# CAMERA_MODEL "Dalsa 1m60 config file (freerun)"
#
# camera name/description
#
camera_class: "Dalsa"
camera_model: "1M60"
camera_info: "12 bit dual channel camera link"
... 39 more lines ...
|
25
|
Mon May 3 17:42:20 2010 |
Aidan | Computing | EPICS | EPICS install by Alex |
Alex Ivanov came in on Friday and demonstrated his EPICS kung-fu. His EPICS knig-fu is strong.
We fixed the IP address of the Hartmann machine, renamed it hartmann, and mounted the cvs drives from the frame builder. - including the EPICS base from that machine. In principle, with a new softIoc, this should have been enough to run EPICS on the hartmann machine. However, whilst the softIoc would start, it wouldn't broadcast any channels. Eventually we figured out that this was because of the Windows Virtualization adding another IP address to the hartmann machine (revealed with /sbin/ifconfig). So we removed the virtualization system and then EPICS seemed to broadcast much better.
The minutia of install isshown in the history files for the controls and root users - attached.
|
Attachment 1: history.txt
|
1 cd
2 mkdir -p rpmbuild/{BUILD,RPMS,SOURCES,SPECS,SRPMS}
3 echo '%_topdir %(echo $HOME)/rpmbuild' > .rpmacros
4 ls
5 cd rpmbuild/
6 rpm -i http://mirror.centos.org/centos/5/updates/SRPMS/kernel-2.6.18.15.1.el5.src.rpm 2
7 cd ..
8 rpm -i http://mirror.centos.org/centos/5/updates/SRPMS/kernel-2.6.18.15.1.el5.src.rpm 2>&1 | grep -v mockb
9 rpm -i http://mirror.centos.org/centos/5/updates/SRPMS/kernel-2.6.18-164.15.1.e15.src.rpm 2>&1 | grep -v mockb
10 rpm -i http://mirror.centos.org/centos/5/updates/SRPMS/kernel-2.6.18-164.15.1.e15.src.rpm 2>&1 | grep -v mockb
... 183 more lines ...
|
Attachment 2: history_root.txt
|
1 yum
2 yum install gcc
3 yum install make
4 yum install tk
5 yum install tcl
6 yum install mm
7 yum install kernel
8 yum install source
9 yum install include
10 yum install kernel-source
... 797 more lines ...
|
24
|
Thu Apr 22 08:22:18 2010 |
Aidan | Computing | Frame Grabber | from the manual install.pdf |
Quote: |
Regarding the installation of EDT software, I overlooked a note from the install.pdf file.
The gist of it is that if the scripts do not run, then remount the CD-ROM by typing the
following:
mount /mnt/cdrom -o remount,exec
which will then allow the scripts to be run. The directory /mnt/cdrom should be changed if
the cdrom is mounted somewhere else. (The note can be found in the page 1 of the file
install.pdf.)
Unfortunately I don't have linux installed at the moment so I cannot test this. My computer was
reinstalled with Windows XP, the previous CentOS system being wiped out. However if this works,
then there is probably no need to copy the files to the hard drive.
|
I saw this and tried it when i was installing, but I had more flexibility when I copied the files directly to the hard drive.
|
23
|
Thu Apr 22 08:20:51 2010 |
Aidan | Computing | Hartmann sensor | Installed MATLAB and Windows XP Virtualization on Hartmann machine |
I installed a Windows XP virtualization on the Hartmann machine. It can be accessed from the desktop, or by running virt-manager at the command line. Once the virtualization manager starts the virtualization of Windows needs to be started. It runs quite slowly.
I also installed MATLAB on this machine in /apps/. TThis was intended to be /apps/MATLAB/ but apparently the install program doesn't add a top directory called MATLAB as you might expect. I had to run a yum install libXp because it was complaining that "/apps/bin/glnxa64/MATLAB: error while loading shared libraries: libXp.so.6: cannot open shared object file: No such file or directory" |
22
|
Thu Apr 22 01:48:33 2010 |
Won Kim | Computing | Frame Grabber | from the manual install.pdf |
Regarding the installation of EDT software, I overlooked a note from the install.pdf file.
The gist of it is that if the scripts do not run, then remount the CD-ROM by typing the
following:
mount /mnt/cdrom -o remount,exec
which will then allow the scripts to be run. The directory /mnt/cdrom should be changed if
the cdrom is mounted somewhere else. (The note can be found in the page 1 of the file
install.pdf.)
Unfortunately I don't have linux installed at the moment so I cannot test this. My computer was
reinstalled with Windows XP, the previous CentOS system being wiped out. However if this works,
then there is probably no need to copy the files to the hard drive. |
21
|
Wed Apr 21 06:49:51 2010 |
Aidan | Computing | Frame Grabber | Installing CentOS 5.3 and the EDT frame-grabber - Part 1 |
Yesterday, I installed CentOS 5.3 on the Gateway GT5482 machine that housed the EDT frame-grabber.
- I installed CentOS 5.3 with all the default options
- As recommended by the README.lnx_pkg_reqs, I tried and failed to install the "Development Tools", "Development Libraries" and the "X Software Development" using the Add/Remove Software.
- I copied the entire install CD to ~/fgdriver on the hard disk.
- Installed the following packages at the command line
> yum install gcc
> yum install make
> yum install tk
> yum install kernel
I tried to run ~/fgdriver/linux.go at this point to install the EDT driver, but the installation failed about halfway through with the message "problem making the driver module". An investigation revealed that this was the due to the failure of ~/fgdriver/linux/module/makefile. I tried running that makefile separately to build the driver module and it crashed with the message: Can't find /lib/modules/2.6.18-128.el5/source/include/linux/mm.h. I concluded that the kernel source code wasn't installed
- Added "Development Libraries" with Add/Remove Software
- Ran the following command lines
> yum install kernel-devel
> yum install kernel-xen-devel
And then I followed the instructions at the link: http://wiki.centos.org/HowTos/I_need_the_Kernel_Source
from: > yum install rpm-build redhat-rpm-config unifdef
to: > rpm -i http://mirror.centos.org/centos/5/updates/SRPMS/kernel-2.6.18-164.15.1.el5.src.rpm 2>&1 | grep -v mockb
and at the latter point the rpm build pissed and moaned that it couldn't find the file kernel-2.6.18-164.15.1.el5.src.rpm
However, some combination of the above must have worked. I rebooted the computer and logged in again as root. At this point the install script ~/fgdriver/linux.go ran from start to finish without complaining. A quick test of the resulting /opt/EDTpdv/camconfig and then /opt/EDTpdv/serial_cmd showed that I could access the Dalsa 1M60 camera through the frame grabber.
|
20
|
Tue Apr 20 18:05:24 2010 |
Aidan | Computing | Hartmann sensor | Images off the Dalsa Camera in CentOS |
I installed CentOS on the machine with the EDT frame-grabber. I then installed the frame-grabber software from the CD.
In the /opt/EDTpdv/ directory the camconfig program was run and I entered "331" to start the frame-grabber and run with the Dalsa 1M60 settings ... this was necessary to get the frame grabber running, but didn't seem to force pdvshow, installed at a later point, to use this configuration file. At this point I could access the camera menu with the serial_cmd program.
After some effort, which will be detailed shortly, I managed to finally get the pdv_show GUI program compiled and installed. I found that trying to run that program with the dalsa_1m60.cfg configuration file resulted in a segmentation fault.
However, when I ran it with the default Dalsa configuration file, pantera11m4fr.cfg, and selected "Continuous Exposure" I got a stream of illuminated pixels on the screen. It was clear that the display was displaying the pixels coming back from the camera in the wrong way (for instance, trying to load a 1024x1024 image into a 1440x900 array), however, by changing the frame rate on the camera to 20Hz and waving my hand around in front of the camera I was able to modulate the intensity of the hash of pixels being displayed. This means that the frame-grabber is successfully getting data - it just isn't interpreting it correctly yet.
Here are a couple of images from pdv_show (hit Alt+PrtScrn to get a screenshot of the active window):
1. Screenshot-PCI_DV_Display.png - the image on the computer with the camera running unobscured
2. Screenshot-PCI_DV_Display-1.png - the image on the computer with me covering the camera with my hand.
3. -opt-EDTpdv.png - the camera parameters at the time of this test (running serial_cmd)
|
Attachment 1: Screenshot-PCI_DV_Display.png
|
|
Attachment 2: Screenshot-PCI_DV_Display-1.png
|
|
Attachment 3: -opt-EDTpdv.png
|
|
19
|
Thu Apr 15 01:47:47 2010 |
Won Kim | Computing | Hartmann sensor | Notes on installing EDT PCIe4 DV frame grabber |
* EDT PCIe4 DV frame grabber: installation notes for linux system
(Note)
Main issue I encountered was the fact that most of the shell scripts
did not run by simply entering them. It's bit strange because if you
do ls -al to view the file lists they are made executable. So it's
possible that others don't encounter the same kind of problems as I
did.
However, if one executes the command "./linux.go", for example, and
receives the message saying
bash: ./linux.go: /bin/sh: bad interpreter: Permission denied
then one may follow the steps I took as below.
1. Make a folder to put the content of CD, for example:
mkdir ~/fgdriver
2. Copy the content of the CD-ROM to the folder.
3. Go to the folder.
cd ~/fgdriver
4. Change or check the mode of the following script files (using the
command chmod) to be executable (using "chmod a+x filename"):
linux.go
~/fgdriver/linux/EDTpdv/installpdv (this one should already be
executable)
~/fgdriver/linux/EDTpdv/pdv/setup.sh
5. run ./linux.go and choose DV by clicking it.
(Note)
I am assuming that the programming language Tcl is already installed
in the machine. CentOS 5.4 that I have installed came with Tcl. If Tcl
is not installed, I think that linux.go will run cli_startmenu.sh
instead (located in the same directory as linux.go). So make sure
cli_startmenu.sh is executable (see step 4).
6. Choose default installation directory and start installation
(Note)
In my first attempt to install the files, the installation message
window hung after displaying many lines of "........". That was
because the file setup.sh was not made executable (see step 4). So I
made setup.sh executable, ran linux.go again, then I could see further
messages flowing through (basically compiling c source files). I'm not
sure whether others will enounter the exactly same problem though.
7. After the installation completes, go to the /opt/EDTpdv folder.
cd /opt/EDTpdv
8. Final Step: Make edt_load and edt_unload executable. (See step 4)
(Note)
Most of the other executables we need for running the frame
grabber/camera should already be executable at this point; but somehow
in my installation the above two files were not made executable. I
again do not know whether others will experience the same
problem. Since there are lots of executables generated when
installation completes, I advise that, whenever a certain command does
not run, one should check if that command file is executable or not.
----
Please let me know if you find any parts of the above confusing. I will
do my best to clarify. |
18
|
Mon Apr 12 17:25:01 2010 |
Aidan | Electronics | Hartmann sensor | Fiber-Camera Link demonstration |
I installed the EDT PCIe4 DV C-Link frame grabber in a spare Windows XP PC and connected the Dalsa 1M60 camera directly to it via the CameraLink cable. In this configuration I was able to access the menu system in the camera using the supplied serial_cmd.exe routine.
PC --> Frame-Grabber --> Camera-Link Cable --> Dalsa 1M60: works OK
Next, I attached the RCX C-Link: Fiber to Camera Link converters to either end of a 300' fiber, plugged them into the PC and the Dalsa 1M60 and then supplied them with 5V of power. Once again, I was able to access the on-board menu system in the camera (as the attached screen-capture shows). I also did a quick-test using the in-built video display program and verified that I could get an image from the camera - by waving around my hand in front of the CCD I was able to modulate the light in the image on the computer. This, therefore, demonstrates that the camera can be easily accessed and run at a distance of at least 300' via optical fiber.
PC --> Frame-Grabber --> RCX C-Link --> 300' optical fiber --> RCX C-Link --> Dalsa 1M60: works OK
The attached images:
hartman_sensor.JPG: a screencap of the Dalsa 1M60 on-board menu system captured with the C-Link to fiber connector running
Fiber_Camera_Link_1.jpg: A RCX C-Link and one end of the 300' fiber connected to the Dalsa 1M60
Fiber_Camera_Link_3.jpg: A RCX C-Link and the other end of the 300' fiber connected to the PC
|
Attachment 1: hartmann_sensor.JPG
|
|
Attachment 2: Fiber_Camera_Link_1.jpg
|
|
Attachment 3: Fiber_Camera_Link_3.jpg
|
|
17
|
Mon Apr 12 08:55:37 2010 |
Aidan | Computing | Hartmann sensor | EDT frame grabber is here |
The EDT PCIe4 DV C-Link frame grabber arrived this morning. There is a CD of drivers and software with it that I'll back up to the wiki or 40m svn sometime soon. |
16
|
Fri Feb 12 21:00:06 2010 |
Aidan | Electronics | Ring Heater | Ring heater step function response - time series |
Hideously slow internet at airport is making me write a brief entry. This is the times series of the hesilver watlow heater radiative response to a step function.
Laso United airlines are a bit cheap .... |
Attachment 1: silver_watlow_heater_step_function_response_2010-02-12.pdf
|
|
15
|
Fri Feb 12 11:39:28 2010 |
Aidan | Electronics | Ring Heater | Ring heater transfer function |
I applied a step function to the silver WATLOW heater and measured the response with the photodiode. The power spectrum of the derivative of the PD response is attached. The voltage isn't calibrated, but that's okay because right now we're just interested in the shape of the transfer function. It looks like a single pole around 850uHz. The noise floor is too great above 4 or 5 mHz to say anything about the transfer function.
|
Attachment 1: watlow_heater_transfer_fn.jpg
|
|
14
|
Thu Feb 11 21:46:23 2010 |
Aidan | Electronics | Ring Heater | Ring heater time constant measurement - start time |
After leaving the ring heater off for several hours I turned on a 40V, 0.2A supply at a gps time of 949 988 700
The channel recording the PD response is C2:ATF-TCS_PD_HGCDTE_OUT.
However, there is a delay between the time at which something is supposed to be recorded and the time at which it is recorded. I looked at the GPS clock and it read that time when I started the heater voltage. If you play the channel back in dataviewer you see the temperature start to increase around 80s BEFORE the heater current was switched on. This needs to be calibrated away!!! |
13
|
Thu Feb 11 18:04:08 2010 |
Aidan | Laser | Ring Heater | Ring heater time constant |
I've been looking to see what the time constant of the ring heater is. The attached plot shows the voltage measured by the photodiode in response to the heater turning on and off with a period of 30 minutes.
The time constant looks to be on the order of 600s. |
12
|
Mon Feb 8 17:44:38 2010 |
Aidan | Electronics | Pre-amplifier | replace Pot with fixed Resistor |
Quote: |
|
|
|
|
|
|
Preamp for Bulls eye detector |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
It was felt that the Pot used at the input stage to remove offset added Noise |
|
|
|
|
|
|
|
To test this the Pot was replaced with a fixed resistor and the offset removed at the second stage |
|
|
|
|
|
Noise was measured after the first stage and at the monitor point first with the pot and then with the pot replaced with a Resistor |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
First stage gain =1+500/10 |
test point 1 gain = 51 |
|
|
|
|
|
|
|
|
|
|
second stage gain=10K/1K |
test point 2 gain = 510 |
|
|
|
1K Pot (R19) is present |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Chan #1 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
dbVrms/Hz |
|
|
|
nV/Hz |
|
|
|
Referred Input Noise |
nV/Hz |
|
|
|
|
|
|
|
|
|
|
|
|
|
gain = 51 |
|
|
|
|
|
|
200Hz |
100Hz |
50Hz |
|
200Hz |
100Hz |
50Hz |
|
200Hz |
100Hz |
50Hz |
|
|
|
Test Point #1 |
-141.1 |
-140.0 |
-136.8 |
|
88.1 |
100.0 |
144.5 |
|
1.7 |
2.0 |
2.8 |
|
|
|
|
|
|
|
|
|
|
|
|
gain = 510 |
|
|
|
|
|
Test Point #2 |
-119.4 |
-120.4 |
-118.4 |
|
1071.5 |
955.0 |
1202.3 |
|
2.1 |
1.9 |
2.4 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Pot replaced with Resistor (R4) |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Chan #1 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
dbVrms/Hz |
|
|
|
nV/Hz |
|
|
|
RIN |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
gain = 50 |
|
|
|
|
|
|
200Hz |
100Hz |
50Hz |
|
200Hz |
100Hz |
50Hz |
|
200Hz |
100Hz |
50Hz |
|
|
|
Test Point #1 |
-142.7 |
-142.7 |
-141.9 |
|
73.7 |
73.3 |
80.8 |
|
1.4 |
1.4 |
1.6 |
|
|
|
|
|
|
|
|
|
|
|
|
gain = 500 |
|
|
|
|
|
Test Point #2 |
-122.0 |
-121.1 |
-120.7 |
|
794.3 |
881.0 |
922.6 |
|
1.6 |
1.7 |
1.8 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
When the Pot was replaced with R4, the offset was removed with the Pot at the second gain stage |
|
|
|
|
|
|
R4 was not a thin film metal resistor |
|
|
|
|
|
|
|
|
|
|
|
Just a note: this board was for the QPD not the Bull's eye detector.
|
11
|
Mon Feb 8 10:45:50 2010 |
Steve O'Connor | Electronics | Pre-amplifier | replace Pot with fixed Resistor |
|
|
|
|
|
|
Preamp for Bulls eye detector |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
It was felt that the Pot used at the input stage to remove offset added Noise |
|
|
|
|
|
|
|
To test this the Pot was replaced with a fixed resistor and the offset removed at the second stage |
|
|
|
|
|
Noise was measured after the first stage and at the monitor point first with the pot and then with the pot replaced with a Resistor |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
First stage gain =1+500/10 |
test point 1 gain = 51 |
|
|
|
|
|
|
|
|
|
|
second stage gain=10K/1K |
test point 2 gain = 510 |
|
|
|
1K Pot (R19) is present |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Chan #1 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
dbVrms/Hz |
|
|
|
nV/Hz |
|
|
|
Referred Input Noise |
nV/Hz |
|
|
|
|
|
|
|
|
|
|
|
|
|
gain = 51 |
|
|
|
|
|
|
200Hz |
100Hz |
50Hz |
|
200Hz |
100Hz |
50Hz |
|
200Hz |
100Hz |
50Hz |
|
|
|
Test Point #1 |
-141.1 |
-140.0 |
-136.8 |
|
88.1 |
100.0 |
144.5 |
|
1.7 |
2.0 |
2.8 |
|
|
|
|
|
|
|
|
|
|
|
|
gain = 510 |
|
|
|
|
|
Test Point #2 |
-119.4 |
-120.4 |
-118.4 |
|
1071.5 |
955.0 |
1202.3 |
|
2.1 |
1.9 |
2.4 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Pot replaced with Resistor (R4) |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Chan #1 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
dbVrms/Hz |
|
|
|
nV/Hz |
|
|
|
RIN |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
gain = 50 |
|
|
|
|
|
|
200Hz |
100Hz |
50Hz |
|
200Hz |
100Hz |
50Hz |
|
200Hz |
100Hz |
50Hz |
|
|
|
Test Point #1 |
-142.7 |
-142.7 |
-141.9 |
|
73.7 |
73.3 |
80.8 |
|
1.4 |
1.4 |
1.6 |
|
|
|
|
|
|
|
|
|
|
|
|
gain = 500 |
|
|
|
|
|
Test Point #2 |
-122.0 |
-121.1 |
-120.7 |
|
794.3 |
881.0 |
922.6 |
|
1.6 |
1.7 |
1.8 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
When the Pot was replaced with R4, the offset was removed with the Pot at the second gain stage |
|
|
|
|
|
|
R4 was not a thin film metal resistor |
|
|
|
|
|
|
|
|
|
|
|
10
|
Fri Feb 5 12:42:19 2010 |
Steve | Electronics | Pre-amplifier | Changes to board |
Test entry |
9
|
Thu Feb 4 19:45:56 2010 |
Aidan | Misc | Ring Heater | Ring heater transfer function - increasing collection area |
I mounted the thinner Aluminium Watlow heater inside a 14" long, 1" inner diameter cylinder. The inner surface was lined with Aluminium foil to provide a very low emissivity surface and scatter a lot of radiation out of the end. ZEMAX simulations show this could increase the flux on a PD by 60-100x.
There was 40V across the heater and around 0.21A being drawn. The #9005 HgCdTe photo-detector was placed at one end of the cylinder to measure the far-IR. (Bear in mind this is a 1mmx1mm detector in an open aperture of approximately 490 mm^2), The measured voltage difference between OFF and the steady-state ON solution, after a 5000x gain stage, was around 270mV. This corresponds to 0.054mV at the photo-diode. Using the responsivity of the PD ~= 0.05V/W then this corresponds to about 10mW incident on the PD.
|
Attachment 1: low-emissivity-tube.jpg
|
|
8
|
Thu Feb 4 15:26:37 2010 |
Aidan | Electronics | Ring Heater | Ring heater transfer function measurement 240mHz-5Hz |
Quote:
|
I've been trying to measure the ring heater transfer function (current to emitted power) by sweeping the supply voltage and measuring the emitted power with a photodector positioned right next to the ring heater.
Last night the voltage was sweeping with a 1000mV setting on the SR785 which was fed into the Voltage Control of the Kepco Bipolar Operational Power Supply/Amplifier which was biased around 10V.
The results are very, very strange. The magnitude of the transfer function decreases at lower frequency. I'll post the data just as soon as I can (ASCII dumps 13 and 14 on the disk from the SR785).
The circuit looks like this:
SR785 drive ----> Amplifier ----> Ring Heater : Photodetector ---> SR560 (5000x gain) ----> SR785 input
|
This is wrong. It turns out the SR785 was wired up incorrectly. |
7
|
Thu Feb 4 14:05:59 2010 |
Aidan | Electronics | Ring Heater | Ring heater transfer function measurement 240mHz-5Hz |
I've been trying to measure the ring heater transfer function (current to emitted power) by sweeping the supply voltage and measuring the emitted power with a photodector positioned right next to the ring heater.
Last night the voltage was sweeping with a 1000mV setting on the SR785 which was fed into the Voltage Control of the Kepco Bipolar Operational Power Supply/Amplifier which was biased around 10V.
The results are very, very strange. The magnitude of the transfer function decreases at lower frequency. I'll post the data just as soon as I can (ASCII dumps 13 and 14 on the disk from the SR785).
The circuit looks like this:
SR785 drive ----> Amplifier ----> Ring Heater : Photodetector ---> SR560 (5000x gain) ----> SR785 input
|
6
|
Fri Jan 29 10:02:15 2010 |
Aidan | Computing | DAQ | New DAQ ordered |
On the advice of Ben Abbott, I've ordered the Diamond Systems Athena II computer w/DAQ, as well as an I/O board, solid state disk and housing for it. The delivery time is 4-6 weeks.
Diamond Systems Athena II
|
5
|
Tue Dec 29 17:50:57 2009 |
Aidan | Computing | DAQ | VME crate has proper boot settings |
We fixed the start-up settings on the VME crate to look for a TCS startup file on fb0. The settings on the Baja 4700 are now: |
Attachment 1: VME_tcs_boot_settings.jpg
|
|
4
|
Tue Dec 29 16:05:09 2009 |
Frank | Computing | DAQ | booting VME crates from fb1 |
http://nodus.ligo.caltech.edu:8080/AdhikariLab/514 |
3
|
Mon Dec 28 14:48:29 2009 |
Aidan | Computing | DAQ | VME crate has a "new" CPU - needs to be configured |
I installed a recycled VME crate in the electronics rack. It currently has a Baja 4700E CPU card in it - and this needs to be configured. We also have the following cards, which are not plugged in right now.
1. ICS-110A-32 Analogue-to-Digital Converter - the jumpers need to be set on this to give it a unique memory address in the VME bus.
2. D000186 LIGO-type Anti Image card.
The CPU card needs to be configured to search it's OS binaries on the network (in this case we're going to store them on the framebuilder in Rana's lab). These settings are accessed by plugging a serial cable into the front of the card and using a terminal window to access the menu system. There are some screen caps of this below. As the card is reset we get the Start-up screen and then we can either do nothing (and a full boot will take place) or we can press a key and access the menu. From there we can restart the boot process by entering "@" or we can change the boot settings by entering "c". These are shown below:
|
Attachment 1: VME_boot_02.jpg
|
|
Attachment 2: VME_boot_01.JPG
|
|
Attachment 3: VME_boot_03.jpg
|
|