  251   Wed May 11 19:46:21 2022 Aidan, Radhika, JordanLab InfrastructureVacuum chamberLeak cleaning on IR labs vacuum chamber - suspects

[Aidan, Jordan, Radhika]

Radhika and Jordan identified some particulates (hair and flecks of foil) on the O-ring on the IR labs dewer. Additionally, we saw a scratch in the O-ring groove and a nick on the metal of the base of the dewer where it meets the O-ring. All were in the leaky vicinity previously identified by the He testing.

We set up a cradle to hold the dewer while we are working on it. Still needs vertical supports.

R & J replaced the O-ring with a new one with Crytox applied. 

Attachment 1: IMG_8812.jpg
IMG_8812.jpg
Attachment 2: IMG_8799.jpg
IMG_8799.jpg
Attachment 3: IMG_8789.jpg
IMG_8789.jpg
Attachment 4: IMG_8805.jpg
IMG_8805.jpg
  252   Fri May 20 16:42:47 2022 StephenThings to BuyOrderingUp to Air valve for IR Labs Cryostat

Purchased from Lesker, for venting of IR Labs Cryostat:

 - P/N C35103000 (KF10 Valve for up-to-air)

 - P/N QF25XQF10 (KF25-KF10 Reducer)

 - P/N QF25-100-T (KF25 Tee)

(to be connected in between the Chamber Tee and the Gate Valve)

Should be in next week!

  253   Thu Jun 23 10:53:17 2022 JCLab InfrastructureGeneralSeat For Photodiode Testing

[JC, Chub, Radhika]

Chub and I ordered a few parts from McMaster in order build a handrail-like stopper to keep the dewar from falling over. We also cut off the excess 8020 which was leaning over the table to fit. To hold down the support for the Dewar, Radhika and I decided to use C-clamps from the EE shop.

Attachment 1: 62AE8A4B-F817-4A03-A2F1-AE48B97FB935.jpg
62AE8A4B-F817-4A03-A2F1-AE48B97FB935.jpg
Attachment 2: IMG_1392_2.jpg
IMG_1392_2.jpg
ELOG V3.1.3-