Entry  Wed Oct 30 12:13:14 2019, Jon, Misc, Delivery Note, New lab hardware IMG_3782.jpgIMG_3783.jpgIMG_3784.jpgIMG_3785.jpg
Ian and I moved some new hardware into the lab, shown in the below photos. It is from the shipment of loaned
equipment recently returned by Whitman College. 

Attachment 1:
Entry  Tue Nov 30 11:01:19 2010, Aidan, Computing, General, New workstation added in TCS Lab. New Static IP 
I added a workstation at 10.0.1.26 in the TCS lab.
Entry  Mon Apr 4 13:13:23 2011, Aidan, Things to Buy, Delivery Note, Newfocus 5102 mirrors and Firewire extension cable have arrived receipt_mirrors.pdf
 See attached delivery note ...
Entry  Thu Apr 15 01:47:47 2010, Won Kim, Computing, Hartmann sensor, Notes on installing EDT PCIe4 DV frame grabber 
* EDT PCIe4 DV frame grabber: installation notes for linux system 

(Note)
Entry  Mon Apr 18 15:06:53 2011, Aidan, Things to Buy, Hartmann sensor, Ordered 2" optics from Newport 
 Given that the HWS requires several 2" optics to handle the big beam size, I've ordered the following items from Newport:

    2x 2" 50/50 beam splitter: 20B20BS.2
    6x 2" NIR mirrors: 
Entry  Thu May 6 21:32:26 2010, Won Kim, Computing, Hartmann sensor, Peak detection and centroiding code detect_peaks_uml.m
 
 
Attached is .m file of the custom function that I wrote and used to automatically detect peaks in a
    Reply  Tue May 11 10:32:00 2010, Aidan, Computing, Hartmann sensor, Peak detection and centroiding code - review test_centroid_code.m
This looks really efficient! However, I think there's a systematic error in the calculation. I tested it on some simulated data and it had trouble getting
the centroids exactly right. I need to better understand the functions that are called to get an idea of what might be the problem.
 
Entry  Fri Aug 9 20:33:11 2019, Jon, Electronics, Ring Heater, Power-regulated heater ready to use IMG_3576.jpgIMG_3577.jpg
Despite some considerable time spent, I was not able to get the Omega SCR controllers
working. The first unit definitely arrived damaged. None of its LED indicator lights ever functioned, despite those on the second controller
working fine under the same setup. I tried swapping in the second controller, but it has no voltage output and a red LED is illuminated which, according
Entry  Fri Apr 1 09:51:45 2011, Aidan, Computing, Hartmann sensor, Prism measurement Hartmann_Sensor_prism_measurement_2011-03-31.pdfHartmann_Sensor_Prism_measurement_times_series_2011-03-31.pdf
 I analyzed the results from the prism experiment. The time series and spectra of the prism are attached.
Conclusions to follow ...
Entry  Wed Mar 30 12:55:54 2011, Aidan, Laser, Hartmann sensor, Prism modulation experiment 
I've set up a quick experiment to modulate the angle of the Hartmann sensor probe beam at 10mHz and to monitor the measured prism. The beam from the
SLED is collimated by a lens and this is incident on a galvo mirror. The reflection travels around 19" and is incident on the HWS. When the galvo
mirror is sent a 1.1Vpp sine wave, then beam moves around +/- 0.5" on the surface of the Hartmann sensor, giving around 50mrad per Vpp.
    Reply  Thu Mar 31 11:27:02 2011, Aidan, Laser, Hartmann sensor, Prism modulation experiment 
I changed the drive amplitude on the function generator to 0.05Vpp and have measured the angle of deflection by bouncing a laser off the laser mirror
and projecting it 5.23m onto the wall. The total displacement of the spot was ~3.3mm +/- 0.4mm, so the amplitude of the angular signal is 1.6mm/5.23m ~
3.1E-4 radians. The Hartmann Sensor should measure a prism of corresponding magnitude.
Entry  Wed May 5 18:45:51 2010, Aidan, Computing, Hartmann sensor, Python code to interface the Dalsa1M60 and export the temperature to EPICS dalsa_to_epics.pyVerifyTemperature.pydalsa_1m60.cfg
Python script
I wrote a Python script, ~/scripts/dalsa_to_epics.py that reads the temperature off the camera using
serial_cmd vt and then it writes this to the EPICS channels using ezcawrite.
Entry  Fri Sep 2 14:41:48 2011, Aidan, Computing, Hartmann sensor, QFLD-950-3S long term test finished 
I ran a test of the HWS with the QFLD-950-3S for 5 days. The test was terminated as we need to disconnect all the cabling and tidy up all the computers
in the lab.
Entry  Tue Aug 6 19:22:28 2019, Jon, Electronics, Computing, QIL NFS server set up 
I've started building the NFS server for the QIL cymac. It's sitting on the workbench in the TCS lab next to the rack. Please don't move
it or any of the parts behind it.
Entry  Thu May 20 10:44:13 2010, Aidan, Electronics, SLED, QPhotonics 980nm source power 
dV = 0.385V
Responsivity= 0.65A/W
Transimpedance = 1.5E4 V/A
Entry  Thu Jan 18 10:09:50 2018, Jon Richardson, Laser, General, RIN Measurement of 400 mW CO2 Laser in Progress 
The 400 mW CO2 laser on the Hartmann table is currently configured for a measurement of its relative intensity noise. It is aligned to a TCS CO2P photodetector
powered by a dual DC power supply beside the light enclosure. I got some data last night with the laser current dialed back for low output power (0.5-10 mW
incident), but still need to analyze it. In the meantime please don't remove parts from the setup, as I may need to repeat the measurement with better
    Reply  Mon Jan 22 10:39:32 2018, Jon Richardson, Laser, General, RIN Measurement of 400 mW CO2 Laser in Progress SRM_CO2_RIN_v1.pdf
Attached for reference is the RIN measurement from the initial data.




Quote:



The 400 mW CO2 laser on the Hartmann table is currently configured for a measurement of its relative intensity
Entry  Wed Aug 4 09:57:38 2010, Aidan, James, Computing, Hartmann sensor, RMS measurements with Hartmann sensor rms_analyze_centroids_aidan.pdfRMS_WonCode.pngRMS_WonCodeLessPrism.png
[INCOMPLETE ENTRY]
We set up the Hartmann sensor and illuminated it with the output from the fiber-coupled SLED placed about 1m away. The whole arrangement was covered
with a box to block out ambient light. The exposure time on the Hartmann sensor was adjusted so that the maximum number of counts in a pixel was about
    Reply  Wed May 30 18:50:05 2018, awade, Computing, DAQ, Rebooted fb4 
Time was still off by nine days as of yesterday.  I tried rebooting remotely to see if time would correct to system clock.  It didn't and
fb4 hung.

Just manually restarted the box.  Now dataviewer is showing a 'Time Now'  of 5 Jan 1980.   
Entry  Thu Aug 1 15:20:39 2019, Edita Bytyqi, Lab Infrastructure, Optics, Reflector Mount and DC Supply 20190731_170920.jpg20190801_145836.xcf20190801_145836.jpg20190801_150635.jpg20190801_145842.jpg
Yesterday, we were able to take some data using the 120 V DC power supply. The reflectors cut at the focal point and radius were both tested; the semi-circle
cut proved to give a better focus, likely because roughly half the heat is lost using the focal-point reflectors. For upcoming tests, the semicircle reflectors
will be used. We varied the surface shine by using the dull and reflective side of Al foil, as well as using the machined Al itself. The best result was
Entry  Thu May 27 13:18:51 2010, Aidan, Electronics, Hartmann sensor, Removed Cooling fins from Hartmann sensor HWS_CONFIG3.jpg
I removed the cooling fins from the Hartmann sensor to see what steady-state temperature it reached without any passive cooling elements. I also dropped
the set-point temperature for the lab to help keep for getting too hot.
After nearly 3 hours the temperature is:
Entry  Mon May 24 19:17:32 2010, Aidan, Laser, Hartmann sensor, Replaced Brass Plate with Invar Hartmann Plate 
I just replaced the brass Hartmann plate with the Invar one. The camera was off during the process but has been turned on again. The camera is now warming
up again. I've manually set the temperature in the EPICS channels by looking at the on-board temperature via the serial communications.
 I also made sure the front plate was secured tightly.
Entry  Wed Oct 11 19:14:16 2017, Jon Richardson, Computing, Electronics, Replaced TCS Monitor 
I replaced the dead 24" monitor on the work bench, which is connected to the video multiplexer. Mike Pedraza was kind enough to bring us us
a new one and take away the old one.
Entry  Thu Jul 22 11:02:59 2010, Aidan, Computing, General, Restarted hartmann machine 
hartmann had started responding to requests to log-in with the a request to change the password. Attempts to change the password proved
unsuccessful. I tried to access the machine locally to change the password but I couldn't the display started, so I had to reboot it.
 
Entry  Fri Jul 30 10:04:54 2010, Aidan, Computing, Hartmann sensor, Restarted the HWS EPICS channels 
 Restarted the HWS EPICS channels on hartmann with the following command:




/cvs/opt/epics-3.14.10-RC2-i386/base/bin/linux-x86/softIoc -S HWS.cmd &



 
Entry  Fri Sep 2 21:41:01 2011, Aidan, Electronics, Hartmann sensor, Restarting long term test of QFLD-950-3S 
9:40PM PDT - I've just restarted the long term measurement of the Hartmann sensor noise with the QFLD-950-3S.
 
Entry  Fri Feb 12 21:00:06 2010, Aidan, Electronics, Ring Heater, Ring heater step function response - time series silver_watlow_heater_step_function_response_2010-02-12.pdf
Hideously slow internet at airport is making me write a brief entry. This is the times series of the hesilver watlow heater radiative response to a step
function.
Laso United airlines are a bit cheap ....
Entry  Thu Feb 11 18:04:08 2010, Aidan, Laser, Ring Heater, Ring heater time constant 
I've been looking to see what the time constant of the ring heater is. The attached plot shows the voltage measured by the photodiode in response to
the heater turning on and off with a period of 30 minutes.
The time constant looks to be on the order of 600s.
Entry  Thu Feb 11 21:46:23 2010, Aidan, Electronics, Ring Heater, Ring heater time constant measurement - start time 
After leaving the ring heater off for several hours I turned on a 40V, 0.2A supply at a gps time of 949 988 700
The channel recording the PD response is C2:ATF-TCS_PD_HGCDTE_OUT.
However, there is a delay between the time at which something is supposed to be recorded and the time at which it is recorded. I looked
Entry  Fri Feb 12 11:39:28 2010, Aidan, Electronics, Ring Heater, Ring heater transfer function watlow_heater_transfer_fn.jpg
I applied a step function to the silver WATLOW heater and measured the response with the photodiode. The power spectrum of the derivative of the PD response
is attached. The voltage isn't calibrated, but that's okay because right now we're just interested in the shape of the transfer function. It looks like
a single pole around 850uHz. The noise floor is too great above 4 or 5 mHz to say anything about the transfer function.
Entry  Thu Feb 4 19:45:56 2010, Aidan, Misc, Ring Heater, Ring heater transfer function - increasing collection area low-emissivity-tube.jpg
I mounted the thinner Aluminium Watlow heater inside a 14" long, 1" inner diameter cylinder. The inner surface was lined with Aluminium foil
to provide a very low emissivity surface and scatter a lot of radiation out of the end. ZEMAX simulations show this could increase the flux on a PD by
60-100x. 
Entry  Thu Feb 4 14:05:59 2010, Aidan, Electronics, Ring Heater, Ring heater transfer function measurement 240mHz-5Hz 
I've been trying to measure the ring heater transfer function (current to emitted power) by sweeping the supply voltage and measuring the emitted power
with a photodector positioned right next to the ring heater.
Last night the voltage was sweeping with a 1000mV setting on the SR785 which was fed into the Voltage Control of the Kepco Bipolar Operational
    Reply  Thu Feb 4 15:26:37 2010, Aidan, Electronics, Ring Heater, Ring heater transfer function measurement 240mHz-5Hz 


    
        
         
  
            Quote:
            
        
        
            
            I've
Entry  Tue Nov 14 19:08:00 2017, Jon Richardson, Computing, Network architecture, Rsync Server for Automatic LDAS Backups 
I implemented an access point for LDAS to pull data from the TCS lab EPICS frame archive (fb4:/frames) via rsync. The setup is analogous to what is already
running at the 40m for automated backups. Here are the implementation details in case we want to replicate this in other W. Bridge labs.

Two lab machines are needed, the frame builder machine (fb4; 10.0.1.156) and a second machine to handle the network interfacing with the outside
Entry  Thu May 13 16:54:46 2010, Aidan, Computing, Hartmann sensor, Running MATLAB programs in C on CentOS - only use R2008b for less hassle 
 After much effort trying to get a MATLAB routine to compile in C I discovered the following pieces of information.
1. CentOS will not install a gcc compiler more recent than 4.1.2 with yum install. This is circa 2007. If you want a more recent compiler it must be installed manually.
Entry  Mon Aug 9 10:58:37 2010, Aidan, Laser, SLED, SLED 15-day trend SLED_superlum_long_term_test_0005A_annotated_15-day_result.pdf
 Here's a plot of the 15-day output of the SLED.
Currently there is an 980nm FC/APC fiber-optic patch-cord attached to the SLED. It occurred to me this morning that even though the patch cord
is angle-cleaved, there may be some back-reflection than desired because the SLED output is 830nm (or thereabouts) while the patch cord is rated for 980nm.
Entry  Sat Jul 31 11:54:20 2010, Aidan, Computing, SLED, SLED Test Day 5 - Re-tuned current set-point control voltage 
Main Points

    Re-set SLED current set-point control voltage to 0.111V
    SLED current set-point voltage drops by about 5mV when the SLED
is dis-engaged
    Resetting was around 11:45AM PDT 31st-Jul-2010

I turned off the SLED for 10s and reset the current set-point
Entry  Wed Jul 28 18:32:12 2010, Aidan, Electronics, SLED, SLED long term test - Day 2 SLED_superlum_long_term_test_0003A.pdf
Here's the data from the last 2 1/2 days of running the SLED. The decrease in photo-current measured by the on-board photo-detector is consistent with
the decrease in the current set-point and the delivered current, but it is not clear why these should be changing.
Strictly speaking I should add some analysis that shows that delta_PD_voltage_measured = delta_I_set_measured * [delta_PD_voltage/delta_I_set (I_set)]_calculated
    Reply  Thu Jul 29 10:09:19 2010, Aidan, Electronics, SLED, SLED long term test - Day 2 superlum_SLED_ATR.pdf
I've attached the Acceptance Test Report data from SUPERLUM for this SLED. I've also determined the expected percentage decrease in power/photo-current
per mA drop in forward current.
The measured decrease in forward current over the last 2 1/2 days is around 1.4mA from around 111mA. The expected drop in power is thus (4.5% per
Entry  Thu May 20 17:08:36 2010, Aidan, Electronics, SLED, SLED module - and driver - LIGO D1000892 - and Hartmann sensor 2010-05-20_hartmann_image_and_located_spot.jpg
Verified that the test-point for the current limit pot on the driver (Wavelength Electronics - LDTC 0520) was at 0.5V. Driver is set to INTERNAL set
point at the moment. This is down about 10% below the current limited point.
Voltage across TP7 and TP9 = 0.970V = LD Current Mon
Entry  Thu Jun 17 07:40:07 2010, James K, Misc, Hartmann sensor, SURF Log -- Day 1, Getting Started 
 For Wednesday, June 16:
 
I attended the LIGO Orientation and first  Introduction to LIGO lecture in the morning.
Entry  Thu Jun 17 22:03:51 2010, James K, Misc, Hartmann sensor, SURF Log -- Day 2, Getting Started 
For Thursday, June 17:
Today I attended a basic laser safety training orientation, the second Introduction to LIGO lecture, a Summer Research Student Safety Orientation,
and an Orientation for Non-Students living on campus (lots of mandatory meetings today). I met with Dr. Willems and Dr. Brooks in the morning and went
Entry  Sat Jun 19 17:31:46 2010, James K, Misc, Hartmann sensor, SURF Log -- Day 3, Initial Image Analysis 7x

For Friday, June 18:
(note that I haven't been working on this stuff all of Saturday or anything, despite posting it now. It was getting late on Friday
Entry  Tue Jun 22 22:30:24 2010, James K, Misc, Hartmann sensor, SURF Log -- Day 5, more Hartmann image preliminary analysis 13x
 Today I spoke with Dr. Brooks and got a rough outline of what my experiment for the next few weeks will entail. I'll be getting more of the details
and getting started a bit more, tomorrow, but today I had a more thorough look around the Hartmann lab and we set up a few things on the optical table.
The OLED is now focused through a microscope to keep the beam from diverging quite as much before it hits the sensor, and the beam is roughly aligned to
    Reply  Wed Jun 23 06:49:48 2010, Aidan, Misc, Hartmann sensor, SURF Log -- Day 5, more Hartmann image preliminary analysis 
Nice work!
 


    
        
            
Entry  Wed Jun 23 22:57:22 2010, James K, Misc, Hartmann sensor, SURF Log -- Day 6, Centroiding 10x
 So in addition to taking steps towards starting to set stuff up for the experiment in the lab, I spent a good deal of the day figuring out how
to use the pre-existing code for finding the centroids in spot images. I spent quite a bit of time trying to use an outdated version of the code that didn't
work for the actual captured images, and then once I was directed towards the right version I was hindered for a little while by a bug.
Entry  Fri Jun 25 00:11:13 2010, James K, Misc, Hartmann sensor, SURF Log -- Day 7, SLED Beam Characterization BeamData.xlsxbeam_pdfs.zipbeamdata.mgsbeam.m
 
BACKGROUND:

In order to conduct future optical experiments with the SLED and to be able to predict the behavior of the beam as it propagates across the
Entry  Tue Jul 6 21:57:19 2010, James Kunert, Misc, Hartmann sensor, SURF Log -- SLED fiber output temporal analysis 6x
In the previous log, I describe the direct measurement of the fiber output beam using the Hartmann sensor with the plate removed. In order to analyze
how these properties might change as a function of time, we left the camera running over the holiday weekend, Dr. Brooks having written a bash script which
took images from the sensor every 500 seconds. This morning I wrote a MATLAB script to automatically analyze all of these images and plot the fit parameters
Entry  Thu Jul 1 09:40:13 2010, James Kunert, Misc, Hartmann sensor, SURF Log 8 -- more SLED characterization 8x
As I started setting up my next experiment, I noticed that the beam size from the SLED appeared to be larger than expected from previous analysis. It
was therefore necessary to conduct further experiments to characterize the divergence angle of the beam.
First, I set up the photodetector attached to an SLED and mounted a razor blade on a translational stage, in the same manner as done previously.
Entry  Thu Jul 15 20:06:37 2010, James K, Misc, Hartmann sensor, SURF Log: Thermally Induced Defocus Experiments pydefoc.mpyanalyze.mpyimportsingle.mpyimportM.mframesumexport2.py
A quick write-up on recent work can be found at: Google
Docs
 