  121   Tue Mar 8 11:30:26 2011 AidanComputingHartmann sensorHartmann sensor code changes and NTP server

I've made the following changes to the Hartmann sensor code and to the machine running the HWS.

  • Machine name is now princess_sparkle (10.0.1.26)
  • I set up ntpd on that machine to sync the clock to GPS - roughly.
  • I added a MATLAB function (store_current_centroids.m) to the Hartmann sensor that saves the centroids and peak intensities to file in GPS labeled files
  • ~/Hartmann_Sensor_Data/centroids/<GPSTIME1>/<GPSTIME2>/<GPSTIME>_<name>.mat

GPSTIME1 = floor(GPSTIME/4E4)*4E4

GPSTIME2 = floor(GPSTIME/2E2)*2E2

I had to add a line in the run_acquire_auto.m script to accommodate this new function and I had to add a function that calculates the peak intensities to the HS_Centroids.m class.

  123   Tue Mar 8 18:48:00 2011 AidanComputingHartmann sensorHWS code is running and recording centroids

The Hartmann sensor is running continuously and is now recording data to file. The formatting has changed slightly with the data now stored in structures called store_measurement every 200s in files in the following way:

  • store_measurement(ii).centroids - the ii-th centroids
  • store_measurement(ii).intensities - the ii-th intensity list
  • store_measurement(ii).time - the time of the ii-th measurement

The files are stored in ~/Hartmann_Sensor_Data/centroids/<GPSTIME rounded to nearest 4E4 seconds>/<GPSTIME rounded to nearest 2E2 seconds>_<subname>.mat

 

  125   Wed Mar 9 01:00:12 2011 Peter Veitch, Won KimComputingHartmann sensorControl of frame rate usign external trigger

We managed to successfully apply frame rate control via external trigger from a pulse generator.

We supplied 5V pulse train when connected to the optocoupler load, and connected to pins 1 and 2 of external trigger (on the frame grabber board) for using camera 0 (which is the case for us).

Then made the following changes to the config file dalsa_1m60.cfg;

MODE_CNTL_NORM: A0 (previously this value would have been 00)

user_timeout: 0    (this line should be added)

Then I saved the new config file as dalsa_1m60_et.cfg

Next, I loaded the new config using initcam command, then set the exposure mode to be 3. This can be done either using serial_cmd directly or using HS_Camera method set_exposure_mode.

In exposure mode 3, the exposure time is set by the time separation between the falling edges of the pulses, and the camera sets the expousure time to be the maximum value possible (as specified in the camera manual). 

Then I took 10 images using take command, and verified that the frame rate is equal to the frequency of the pulse. We tested 1 Hz and 2 Hz pulse trains, and the frame grabber recoded 1 frames per sec and 2 frames per sec respectively.

We could not yet test the frequency values < 1 Hz as pulse generator we used could not go under 1 Hz.

(EDIT)

We used another pulse generator to test pulse frequencies under 1 Hz, and verified that external trigger mode still works.
 

  128   Mon Mar 28 13:00:50 2011 AidanLaserHartmann sensorTo do: Check the polarization from the SLED
  129   Wed Mar 30 12:55:54 2011 AidanLaserHartmann sensorPrism modulation experiment

I've set up a quick experiment to modulate the angle of the Hartmann sensor probe beam at 10mHz and to monitor the measured prism. The beam from the SLED is collimated by a lens and this is incident on a galvo mirror. The reflection travels around 19" and is incident on the HWS. When the galvo mirror is sent a 1.1Vpp sine wave, then beam moves around +/- 0.5" on the surface of the Hartmann sensor, giving around 50mrad per Vpp.

The galvo is currently being sent a 0.02Vpp sine wave at 10mHz.

  130   Thu Mar 31 11:27:02 2011 AidanLaserHartmann sensorPrism modulation experiment

I changed the drive amplitude on the function generator to 0.05Vpp and have measured the angle of deflection by bouncing a laser off the laser mirror and projecting it 5.23m onto the wall. The total displacement of the spot was ~3.3mm +/- 0.4mm, so the amplitude of the angular signal is 1.6mm/5.23m ~ 3.1E-4 radians. The Hartmann Sensor should measure a prism of corresponding magnitude.

The frequency is still 10mHz.

Quote:

I've set up a quick experiment to modulate the angle of the Hartmann sensor probe beam at 10mHz and to monitor the measured prism. The beam from the SLED is collimated by a lens and this is incident on a galvo mirror. The reflection travels around 19" and is incident on the HWS. When the galvo mirror is sent a 1.1Vpp sine wave, then beam moves around +/- 0.5" on the surface of the Hartmann sensor, giving around 50mrad per Vpp.

The galvo is currently being sent a 0.02Vpp sine wave at 10mHz.

 

  131   Fri Apr 1 02:41:55 2011 WonComputingHartmann sensorexposure time and reproducibility of centroids

 Here is a brief and preliminary summary of rms of centroid displacements calculated at a number of different exposure time values. To get the results I did the following for each value of exposure time:

1. Take a set of images. I took 2000 images for shorter exposure times, 1000 images with exposure times greater than 1 second, and 200 images with exposure time 4.4 second. I tried to keep the maximum pixel count to be roughly the same (about 2430 plus/minus 40).

2. Obtain the centroids for each image frames in a set. I saved centroids as an array of n by m by 2, where n is the number of image frames that I took, m is the number of centroids in each frame, and 2 for x and y coordinates.

Then I iterated through the centroid sets to calculate total rms, using various N_av values. If N_av = 100; then reference centroids were obtained by averaging the centroids of first 50 and the last 50 frames, and remaining 100 frames are averaged to get the other (or non-reference) centroids. I think this method gives a better view of the centroid reproducibility than fixing the number of reference centroids to be, say, first 1000 and last 1000 frames and varying the number of frames to be averaged for non-reference centroid.

Datasets of centroids are labelled as spcdet_I_t, where spcdet stands for "same (maximum) pixel count (with) differen exposure times", I the value of the current that drives the light source, and t the exposure time that I used.

Here are the plots:

spcdet_28mA_28ms.pngspcdet_23mA_210ms.png

spcdet_21mA_455ms.pngspcdet_20mA_650ms.png

spcdet_19mA_900ms.pngspcdet_18mA_1290ms.png

spcdet_17mA_1800ms.pngspcdet_16mA_2430ms.png

spcdet_14mA_4400ms.png

It can be seen from these plots that the benefit of averaging multiple frames quickly diminish once we go over 1 second. I am investigating if there is any way to improve the reproducibility while using the same sets of images.

Issues that need further investigation:

  1. Effect of pixels with unusually high pixel count. Dark images that we took show that, with longer exposure times, not only overall dark noise increase (and become less uniform) but also several pixels show unusually high pixel count (even higher than 2000), without a light source on. More investigation is needed to determine how much this affects the centroids calculation and to devise an way to deal with it.

  2. Extra/Duplicate centroids. As exposure time increased, I observed that duplicate centroids start to appear, i.e., HS_Centroids##centroids had duplicate entries. The number of duplicate entries increased as exposure time increased. I believe this is due to the images getting noisier as exposure time increases. So After taking initial reference centroids, I removed duplicate centroid entries before calculating rms. I am thinking about adding a method to do this in HS_Centroids class.

In addition, there were one 'false' centroid when the exposure time was 4.4 seconds. For now I chose to manually remove it myself before calculating rms.

  132   Fri Apr 1 09:51:45 2011 AidanComputingHartmann sensorPrism measurement

 I analyzed the results from the prism experiment. The time series and spectra of the prism are attached.

Conclusions to follow ...

Attachment 1: Hartmann_Sensor_prism_measurement_2011-03-31.pdf
Hartmann_Sensor_prism_measurement_2011-03-31.pdf
Attachment 2: Hartmann_Sensor_Prism_measurement_times_series_2011-03-31.pdf
Hartmann_Sensor_Prism_measurement_times_series_2011-03-31.pdf
  137   Sun Apr 17 21:55:51 2011 AidanLaserHartmann sensorHartmann sensor prism/displacement test

I've set up an experiment to test the HWS intensity distribution displacement measurement code. Basically the beam from a SLED is just reflecting off a galvo mirror onto the HWS. The mirror is being fed a 0.02Vpp *10 gain, 10mHz sinewave from the function generator.

The experimental setup is shown below.

I hacked the HWS code to export the Gaussian X and Y centers to Seidel Alpha and Beta channels in EPICS (C4:TCS-HWSX_PSC_ALPHA, C4:TCS-HWSX_PSC_BETA)

Attachment 1: HWS_prims.jpg
HWS_prims.jpg
  139   Mon Apr 18 15:06:53 2011 AidanThings to BuyHartmann sensorOrdered 2" optics from Newport

 Given that the HWS requires several 2" optics to handle the big beam size, I've ordered the following items from Newport:

  • 2x 2" 50/50 beam splitter: 20B20BS.2
  • 6x 2" NIR mirrors: 5122
  • 8x 2" Ultima mirror mounts: U200-A2K
  144   Tue May 10 00:55:08 2011 WonComputingHartmann sensorMatlab Compiler and Matlab Compiler Runtime

 

I have spent some time with Matlab Compiler and Matlab Compiler Runtime (MCR). I could only get my hands on 2008b version so far, but I believe 2009b version will work in the same way.

 

Below is a set of notes based on my experience so far.

 

 

Matlab compiler installation

1. Copy the toolbox archive files to the folder where matlab is installed. To me this is /usr/matlab_2008b/ (may need to do as root or use sudo). Two files to be copied are tbx.compiler.common and tbx.compiler.glnxa64.

sudo cp tbx.compiler.* /usr/matlab_2008b/

2. Execute the install script.

cd /usr/matlab_2008b
sudo ./install


3. Enter the file installation key that came with the matlab compiler files.

4. Leave the root directory as it was.

5. Finish the installation by activating the toolbox with the license file.



Build a standalone application using matlab compiler

I created a folder called matlab_project as a place to put compiled applications, and matlab_predep as a place to put files to be deployed.

Once Matlab Compiler is installed, it can be launched from the matlab console by typing deploytool. Then I proceeded as below:

1. Create a new project by clicking the New icon (the first one from the left).

2. Choose a standalone application.

3. Click main file and Go to menu Project -> Add Files (or right_click on "main file" icon).

4. Choose the matlab file hello.m (in my case, from matlab_predep folder). hello.m could be, for example, a simple script like

function hello
     disp('Hello!');
end

5. Click the Build icon (third from the right).

When the process finishes, inside matlab_project folder I found a file called "hello.prj" and a folder called "hello". Insider the hello folder was two folders "distrib" and "src".

 

 

Install Matlab Compiler Runtime

1. Run MCRinstaller.bin as root. I found this file in the folder /usr/matlab_2008b/toolbox/compiler/deploy/glnxa64

2. Go along with the default options unless desired otherwise.

3. Add following to .bashrc, below the entries regarding EPICS. $LD_LIBRARY_PATH should now be aware of EPICS as well as MCR related paths. 

export MCR_ROOT=/opt/MATLAB/MATLAB_Compiler_Runtime
export LD_LIBRARY_PATH=$LD_LIBRARY_PATH:$MCR_ROOT/v79/runtime/glnxa64:$MCR_ROOT/sys/os/glnxa64:$MCR_ROOT/v79/sys/java/jre/glnxa64/jre1.6.0/lib/amd64/native_threads:$MCR_ROOT/v79/sys/java/jre/glnxa64/jre1.6.0/lib/amd64/server:$MCR_ROOT/v79/sys/java/jre/glnxa64/jre1.6.0/lib/amd64
export XAPPLRESDIR=$MCR_ROOT/v79/X11/app-defaults

4. Run the revised .bashrc by typing 'source ~/.bashrc'. Next time the user logs in, doing this won't be necessary.

 


Run the built application

1. Go to distrib folder to find the executable application file (e.g., hello). You will also find the executable shell script (e.g., run_hello.sh), the role of which is basically to set up environment variables and run the application, in case the environment variables are not globally set up.

2. Once in that folder, run the application by simply typing ./ and its name (e.g., ./hello). If the library environment variable is set up as in the step 3 of "Install Matlab Compiler Runtime", the application will be run and you should see the hello message as the output.


Build an application that uses HS classes

I wrote a simple script test_HS.m that takes and saves 10 images using the camera, averages the images and finds centroids. Thus the script requires the class files HS_Base, HS_Camera, HS_Image, and HS_Centroids.

I added those four class files by right-clicking Other Files (found below Main function) then choosing Add File, then clicked the build icon.
 

  145   Wed May 11 09:07:03 2011 AidanComputingHartmann sensorChanged ownership of /opt/EDTpdv

 I changed the ownership of /opt/EDTpdv to controls with the command:

controls@princess_sparkle:/opt/EDTpdv$ sudo chown controls EDTpdv/

 

  146   Wed May 11 18:38:47 2011 AidanComputingHartmann sensortest_HS binary

 From Won: (the zip file is also on the SVN /users/won/compiled_code/test_HS.zip)

 

Attached is test_HS.zip file, that contains

- test_HS.prj: project file created by Matlab Compiler. This file is not
 required to run the application but I included it just in case someone's
 insterested.

- test_HS folder contains two subfolders src and distrib, each of which
 contains the standalone application test_HS.

Usage: test_HS <path to the image folder>, for example

 test_HS ~/test_images/

Make sure you create the folder prior to running the application, and the
folder name ends with "/". Running test_HS will take and save 10 images
using the camera (provided the frame grabber applications are installed in
/opt/EDTpdv), averages those 10 images and find centroids, then plots the
result.

As I put in the eLOG, one needs MCRInstaller.bin and run it to install MCR
(probably 2008b 64bit version to test my files). If there are difficulties
getting MCRInstaller, let me know.

Won
<test_HS.zip>

 

Attachment 1: test_HS.zip
  147   Wed May 11 18:44:54 2011 AidanComputingHartmann sensorMatlab Compiler and Matlab Compiler Runtime

Installing MCR

I located the MCRInstaller on our distribution of MATLAB on the Ubuntu machine (/MATLAB_R2009b/toolbox/compiler/deploy/glnxa64/MCRInstaller.bin). I ran the installer, as root,and followed the default options to install it. Next I updated the .bashrc file to include the necessary pointers to various libraries:

 

export LD_LIBRARY_PATH=/home/controls/base-3-14-11/lib/linux-x86_64:/MATLAB_R2009b/runtime/glnxa64:/MATLAB_R2009b/bin/glnxa64
export MCR_ROOT=/opt/MATLAB/MATLAB_Compiler_Runtime
export LD_LIBRARY_PATH=$LD_LIBRARY_PATH:$MCR_ROOT/v79/runtime/glnxa64:$MCR_ROOT/sys/os/glnxa64:$MCR_ROOT/v79/sys/java/jre/glnxa64/jre1.6.0/lib/amd64/native_threads:$MCR_ROOT/v79/sys/java/jre/glnxa64/jre1.6.0/lib/amd64/server:$MCR_
ROOT/v79/sys/java/jre/glnxa64/jre1.6.0/lib/amd64
export XAPPLRESDIR=$MCR_ROOT/v79/X11/app-defaults
 
 
Running test_HS binary from Adelaide on Ubuntu distribution

 

I've downloaded the test_HS binary from the SVN and added the ~/test_images/ directory as recommended by Won. I then ran the code by entering ./test_HS ~/test_images/

 

The code ran successfully through the serial_cmd access and the image acquisition process and only crashed when it tried to access the variable mes_message. This indicates a run-time error, not a compilation error. If you examine lines 751 and 752 of HS_Camera.m you can see the typo (mes_meesage vs mes_message) in the code that is the source of the error:

751                 mes_meesage = ['Intensity too high: ',fobj.name];
752                 cam.inform_messenger('ImageNotValid',mes_message);

Here's the output:


 

controls@princess_sparkle:~/Hartmann_Sensor_SVN/users/won/compiled_code/test_HS/test_HS/distrib$ ./test_HS ~/test_images/
The camera is accessible.


G E N E R A L   C A M E R A   S E T T I N G S:

Camera Model No.:               DS-22-01M60-11E
Camera Serial No.:              04437062
Sensor Serial No.:              0411218

Tap 1 Gain:                     0
Tap 2 Gain:                     0

Firmware Design Rev.:           03-81-00070-03  Sep 30 2004
DSP Design Rev.:                17.3

Pretrigger:                     0      
Video Mode:                     Normal Operating Mode
Data Mode:                      12 bit 
Binning Mode:                   1x1

Gain Mode:                      1x Output Gain Mode
Output Configuration:           2 Tap
Exposure Control:               enabled
Exposure Mode:                  2      

SYNC Frequency:                 8 Hz
Exposure Time:                  123646.66 uSec

OK>

executing /opt/EDTpdv/take -s 1 -l 10 -f /home/controls/test_images/test ...
??? Undefined function or variable "mes_message".

Error in ==> HS_Camera>HS_Camera.read_raw at 752



Error in ==> HS_Camera>HS_Camera.read_from_folder at 668



Error in ==> HS_Camera>HS_Camera.read_from_fg at 721



Error in ==> HS_Camera>HS_Camera.read_images at 590



Error in ==> test_HS at 9


 
MATLAB:UndefinedFunction
 

 

Quote:

 
Build an application that uses HS classes

I wrote a simple script test_HS.m that takes and saves 10 images using the camera, averages the images and finds centroids. Thus the script requires the class files HS_Base, HS_Camera, HS_Image, and HS_Centroids.

I added those four class files by right-clicking Other Files (found below Main function) then choosing Add File, then clicked the build icon.
 

 

  148   Tue May 17 16:08:02 2011 AidanComputingHartmann sensorWrite speed of the frame grabber to file

 The attached file shows the output of the command. The maximum average frame rate is 57.2Hz when the nominal frame rate was 58Hz:

/opt/EDTpdv/take -f max_frame_rate_image -l 120 -N 4 -d > max_frame_rate_data.txt

 

 

 

 

 

 

 

Attachment 1: max_frame_rate_data.txt
reading image from Dalsa 1M60 12 bit dual channel camera link
width 1024 height 1024 depth 12  total bytes 2097152 
writing 1024x1024x12 raw file to max_frame_rate_image0000.raw (actual size 2097152)
writing 1024x1024x12 raw file to max_frame_rate_image0001.raw (actual size 2097152)
writing 1024x1024x12 raw file to max_frame_rate_image0002.raw (actual size 2097152)
writing 1024x1024x12 raw file to max_frame_rate_image0003.raw (actual size 2097152)
writing 1024x1024x12 raw file to max_frame_rate_image0004.raw (actual size 2097152)
writing 1024x1024x12 raw file to max_frame_rate_image0005.raw (actual size 2097152)
writing 1024x1024x12 raw file to max_frame_rate_image0006.raw (actual size 2097152)
writing 1024x1024x12 raw file to max_frame_rate_image0007.raw (actual size 2097152)
... 117 more lines ...
  149   Wed May 18 18:52:12 2011 AidanComputingHartmann sensorTest of position modulation algorithm

 I measured the prism and displacement of the Gaussian beam on the Hartmann sensor. The beam pointing was modulated at 10mHz using a galvo mirror as illustrated in Attachment 1. The galvo was around 680mm from the Hartmann sensor. The amplitude of the prism modulation was approximately 1E-5 radians. The displacement of the beam was measured using a new algorithm that tries to fit a parabola to the logarithm of the intensity of each Hartmann spot. The amplitude of the displacement modulation was measured at around 42 microns: corresponding to around 6E-5 radians (=42um/680mm).

To resolve the discrepancy between the prism and displacement measurements, I removed the Hartmann plate to simply get a Gaussian beam on the CCD (bottom right image in Figures 2 & 3 - the beam is slightly clipped and there is a ghost beam in the center - I'm not yet certain where this is coming from). I measured the Gaussian beam displacement directly by fitting a Gaussian to the mean horizontal cross-section of the intensity distribution (top right plot in Figures 2 & 3). Using this technique the measured displacement on the CCD had an amplitude of around 0.7 +/- 0.05 pixels = 8.4 +/- 0.6 microns, corresponding to a prism of 12.5E-6 radians (seen in top left plot in Figures 2 and 3). This indicates that there is an error in the Gaussian fitting algorithm using the Hartmann sensor data.

 The second plot simply shows the position modulation of the beam as I increased the amplitude of the signal going to the galvo. 

 

Voltage (Vpp) Displacement on CCD (pixels)
0.01 0.325
0.012 0.400
0.014 0.487
0.015 0.541
0.016 0.576
0.019 0.708
0.02 0.763
0.023 0.887
0.028 1.194
0.034 1.803
0.041 3.053
0.049 7.107

 

Attachment 1: galvo_mirror_experiment.jpg
galvo_mirror_experiment.jpg
Attachment 2: 2011-05-18_gaussian_beam_with_galvo_constant.pdf
2011-05-18_gaussian_beam_with_galvo_constant.pdf
Attachment 3: 2011-05-18_gaussian_beam_with_galvo.pdf
2011-05-18_gaussian_beam_with_galvo.pdf
  151   Mon May 30 10:53:00 2011 AidanLaserHartmann sensorDefocus vs time

 I've had the output from a fiber projected about 400mm onto the Hartmann sensor for around 5 days now. (The divergence angle from the fiber is around 86 mrad).

I played around with the temperature of the lab to induce some defocus changes in the Hartmann sensor. The system is mostly linear, but there are relatively frequent jumps in the defocus of the order of 1E-4 m^-1. This may be due to a number of things - the Hartmann plate may be moving, the fiber holder may be shifting back and forth, there may be some issue with the source wavelength shifting.

  • A change in defocus, dS, of around 1E-4 m^-1 from a point source at 400mm from the Hartmann plate (dS/dx = -1/x^2), corresponds to a change in the end position of the fiber of around 16 microns. Seems a little big ... unless it's not secured properly ...
  • Also the defocus vs temperature slope is different from what James measured last year. There is an additional -4E-5 m^-1 K^-1 due to the expansion of the stainless steel table moving the point source further or closer to the Hartmann plate. That leaves about -3E-4 m^-1 K^-1 for the Hartmann sensor. This is roughly twice what James measured last year. 

Sun 30th May 2011 - 11:40AM - the z-axis control on the NewFocus 9091 fiber coupling mount was not tightened. I tightened that to secure the control.

Attachment 1: HWS_defocus_vs_temperature.pdf
HWS_defocus_vs_temperature.pdf
  152   Mon May 30 20:11:27 2011 AidanLaserHartmann sensorHartmann sensor lever arm calibration

 I ran through the procedure to calibrate the lever arm of the Hartmann sensor. The beam from a 632.8nm HeNe laser was expanded to approximately 12mm diameter and injected into a Michelson interferometer. The Hartmann sensor was placed at the output port of the Michelson.

  1. I tilted one of the mirrors of the interferometer to induce a prism between the two beams at the output. This created about 135 vertical fringes on the CCD.
  2. With the Hartmann plate removed, I recorded the interference pattern and took its 2D FFT. There was a peak in the Fourier transform about 134 pixels from the DC level. 
  3. This next part is questionable ...  I centroided the frequencies around the peak in the FFT to try to determine the spatial frequency of the fringes to better than the bandwidth of 1/1024 pixels^-1 (yes, they're strange units). This is, probably acceptable at improving the accuracy of the frequency measurement if it is known that the signal is a generated by a pure sine wave in the spatial domain - this is not an unreasonable assumption for the output of an interferometer. Anyway, the peak fluctuated around 133.9 units from DC by around +/- 0.1 units.
  4. The prism between the two beams is the measured spatial frequency (measured as 133.9 oscillations across the CCD) multiplied by the wavelength and divided by the width of the CCD (= 1024 x 12um). In other words, the prism is the ratio of the wavefront change across the CCD divided by the diameter of the CCD (= 6.895 +/- 0.005 mrad)
  5. Next, I inserted the Hartmann plate, blocked one of the beams and recorded the spot pattern. I then blocked the other beam and unblocked the first and recorded another spot pattern. 
  6. The mean displacement between the spot patterns was calculated. Due to a fairly noisy intensity distribution (the 2" mirrors were AR coated for 700-1000nm and hence there were some stray beams), the mean displacement was relatively noisy - about 5.60 pixels with a standard deviation of around 0.3 pixels and a standard error of around 0.01 pixels ( = 67.2 um)
  7. The lever arm is equal to the mean displacement of the spots divided by the prism. In this case, 9.75 +/- 0.02 mm
  8. I removed the Hartmann plate and confirmed that the FFT of the fringes from the IFO still had a peak at 133.9 +/- 0.1 units. It did.

 

  154   Fri Sep 2 14:41:48 2011 AidanComputingHartmann sensorQFLD-950-3S long term test finished

I ran a test of the HWS with the QFLD-950-3S for 5 days. The test was terminated as we need to disconnect all the cabling and tidy up all the computers in the lab.

  155   Fri Sep 2 21:41:01 2011 AidanElectronicsHartmann sensorRestarting long term test of QFLD-950-3S

9:40PM PDT - I've just restarted the long term measurement of the Hartmann sensor noise with the QFLD-950-3S.

 

  184   Wed Apr 26 15:29:25 2017 AidanElectronicsHartmann sensorLong term test of fiber launcher

I've started a long-term measurement of the HWS fiber-launcher. I'm interested in seeing how stable the output is. The HWS is currently running in the following configuration:

  • SLED: QFLD-635-1S (635nm, 1mW) - current driver set to 38.02mA, thermistor set point = 10.778kOhm
  • SLED coupled to patch cable
  • Patch cable coupled to Newport fiber launcher FPR1-C1A which contains a FPH-CA4 fiber optic chuck (FC/APC connector). All grub screws are tighten on this (two for holding chuck, one for roll motion, one for X translation lock and one for Y translation lock)
  • Output of fiber hits a 100m focal length lens (SPX043AR.16 BBAR 0.65 - 1.0um) placed approximately 100mm from the fiber output
  • Output of lens is projected approximately 125mm onto the HWS where it is centered on the HWS plate

The HWS is currently running at 57Hz. The HWS code is running on HWS (10.0.1.167). It is the same as the site code with some modifications to determine information about the Gaussian beam envelope. The following data is written to file on the HWS machine in files containing 10,000 cycles. Each cycle (or row) the following data is recorded:

  • GPS TIME
  • PRISM_X
  • PRISM_Y
  • SPHERICAL_POWER
  • CYLINDRICAL_POWER 
  • Envelope peak intensity
  • Envelope X centroid
  • Envelope Y centroid
  • Envelope X width
  • Envelope Y width

These are saved to files on the HWS machine: ~/framearchive/C4/HWSlongterm/<GPSTIME>_CIT_HWS.txt

  186   Mon Jun 5 16:41:59 2017 AidanLaserHartmann sensorHWS long term measurement results.

The data from the long-term measurement of the HWS is presented here. The beam envelope moves by, at most, about 0.3 pixels, or around 3.6 microns. The fiber-launcher is about 5" away from the HWS. Therefore, the motion corresponds to around 30 micro-radians (if it is a tilt). The beam displacement is around 4 microns.

The optical properties change very little over the full 38 days (about 2 micro-radians for tilt and around 2 micro-diopters for spherical power).

The glitches are from when the SLED drivers were turned off temporarily for other use (with the 2004nm laser).

Attachment 1: HWS_envelope_long_term.pdf
HWS_envelope_long_term.pdf
Attachment 2: HWS_optical_parameters_long_term.pdf
HWS_optical_parameters_long_term.pdf
Attachment 3: HWS_envelope_long_term_mm.pdf
HWS_envelope_long_term_mm.pdf
Attachment 4: IMG_9836.JPG
IMG_9836.JPG
  190   Wed Oct 11 20:19:06 2017 Jon RichardsonLab InfrastructureHartmann sensorInitial Characterization of the SRM Adaptive Wavefront Control

Today I made the first characterization measurements of the mocked-up adaptive wavefront control system planned for the signal recycling mirrors.

The Setup

Inside the light-tight enclosure on the center table, I've assembled and aligned a 10.2 micron CO2 projector which provides a heating beam of up to 150 mW incident on an SRM-like test optic. A co-aligned 633 nm probe beam and Hartmann wavefront sensor are used to measure the resulting thermal lens. I've written and installed new software on the machine hws (10.0.1.167) for viewing the wavefront distortion in real time, as shown in the below screenshot. This viewer is launched from the terminal via the command $stream_gradient_CIT

Screenshot of a thermal lens forming on the test optic, as viewed from the real-time software.
To the right is the induced optical path difference across the surface of the test optic, and to the left is its gradient.  

There is also a second utility program for displaying the raw Hartmann sensor CCD image in real time, which is useful for aligning the probe beam. It is launched by the terminal command $stream_intensity_CIT

Initial Thermal Lens Measurements

Lens Formation Time Scale

First, I made a time-resolved measurement of the thermal lens formation on the test optic at maximum heating beam power (150 mW). The lens appears to reach steady-state after 30 s of heating. When the heating beam is turned off, the lens decays on a very similar time scale.

Lens Strength v. Incident Heating Power

Second, I measured the thermal lens strength as a function of incident heating beam power, which I measured via a power meter placed directly in front of the test optic. Below is the approximate maximum optical path difference induced at several heating beam powers.

Incident Power (mW) Maximum Optical Path Difference (nm)
150 +/- 15 35 +/- 5
85 +/- 2 25 +/- 5
50 +/- 4 15 +/- 5
10 +/- 1 5 +/- 5

The above optical path differences are approximate and were read-off from the live display. I recorded Hartmann sensor frame data during all of these measurements and will be analyzing it further.

  193   Mon Oct 23 19:02:42 2017 Jon RichardsonLaserHartmann sensorMitigated Heating Beam Losses

There were known to be huge (65%) heating beam power losses on the SRM AWC table, somewhere between the CO2 laser and the test optic. Today I profiled the setup with a power meter, looking for the dominant source of losses. It turned out to be a 10" focusing lens which had the incorrect coating for 10.2 microns. I swapped this lens with a known ZnSe 10" FL lens (Laser Research Optics LX-15A0-Z-ET6.0) and confirmed the power transmittance to be >99%, as spec'd. There is now ~310 mW maximum reaching the test optic, meaning that the table losses are now only 10%.

Using a single-axis micrometer stage I also made an occlusion measurement of the heating beam radius just in front of the test optic. I moved the 10" focusing lens back three inches away from the test optic to slightly enlarge the beam size. In this position, I measure a beam radius of 3.5+/-0.25 mm at 1.5" in front of the test optic (the closest I can place the power meter). The test optic is approximately 20" from the 10" FL lens, so the beam has gone through its waist and is again expanding approaching the test optic. I believe that at the test optic, the beam is very close to 4 mm.

  200   Mon Oct 30 17:13:32 2017 Jon RichardsonLaserHartmann sensorWrite-Up of CO2 Projector Measurements

For archive purposes, attached is a write-up of all the HWS measurements I've made to date for the SRM CO2 projector mock-up.

Attachment 1: awc_srm_actuator_v1.pdf
awc_srm_actuator_v1.pdf awc_srm_actuator_v1.pdf awc_srm_actuator_v1.pdf awc_srm_actuator_v1.pdf awc_srm_actuator_v1.pdf awc_srm_actuator_v1.pdf awc_srm_actuator_v1.pdf
  208   Fri May 11 12:32:21 2018 AidanLaserHartmann sensorHWS green LED fiber launcher

 [Aidan, Marie]

We tested the output of the fiber launcher D1800125-v3. We were using a 6mm spacer in the SM1 lens tube and 11mm spacer in the SM05 lens tube and the 50 micron core fiber.

The output of the fiber launcher was projected directly onto the CCD. Images of these are attached (coordinates are in pixels where 100 pixels = 1.2mm)

There is a lot of high-spatial frequency light on the output. It looks like there is core and cladding modes in addition to a more uniform background. There was an indication that we could clear up these annular modes with an iris immediately after the fiber launcher but I didn't get any images. We're going to test this next week when we get an SM1 mountable iris.

Attachment 1: LED_test_01_20180511.pdf
LED_test_01_20180511.pdf
  209   Fri May 11 16:18:47 2018 AidanLaserHartmann sensorHWS green LED fiber launcher

And here's the output of the fiber launcher when I fixed it at 313mm from the camera, attached an iris to the front and slowly reduced the aperture of the iris.

The titles reflect the calculated second moment of the intensity profiles (an estimate of the equivalent Gaussian beam radius). The iris is successful in spatially filtering the central annular mode at first and then the outer annular mode. 

We'll need to determine the optimum diameter to get good transmission spatially without sacrificing too much power.

Quote:

 [Aidan, Marie]

We tested the output of the fiber launcher D1800125-v3. We were using a 6mm spacer in the SM1 lens tube and 11mm spacer in the SM05 lens tube and the 50 micron core fiber.

The output of the fiber launcher was projected directly onto the CCD. Images of these are attached (coordinates are in pixels where 100 pixels = 1.2mm)

There is a lot of high-spatial frequency light on the output. It looks like there is core and cladding modes in addition to a more uniform background. There was an indication that we could clear up these annular modes with an iris immediately after the fiber launcher but I didn't get any images. We're going to test this next week when we get an SM1 mountable iris.

 

Attachment 1: LED_test_02_20180511.pdf
LED_test_02_20180511.pdf
  210   Wed May 16 16:47:29 2018 AidanLaserHartmann sensorHWS green LED fiber launcher - D1800125-v5_SN01

Here is the output from D1800125-v5_SN01.

Quote:

And here's the output of the fiber launcher when I fixed it at 313mm from the camera, attached an iris to the front and slowly reduced the aperture of the iris.

The titles reflect the calculated second moment of the intensity profiles (an estimate of the equivalent Gaussian beam radius). The iris is successful in spatially filtering the central annular mode at first and then the outer annular mode. 

We'll need to determine the optimum diameter to get good transmission spatially without sacrificing too much power.

Quote:

 [Aidan, Marie]

We tested the output of the fiber launcher D1800125-v3. We were using a 6mm spacer in the SM1 lens tube and 11mm spacer in the SM05 lens tube and the 50 micron core fiber.

The output of the fiber launcher was projected directly onto the CCD. Images of these are attached (coordinates are in pixels where 100 pixels = 1.2mm)

There is a lot of high-spatial frequency light on the output. It looks like there is core and cladding modes in addition to a more uniform background. There was an indication that we could clear up these annular modes with an iris immediately after the fiber launcher but I didn't get any images. We're going to test this next week when we get an SM1 mountable iris.

 

 

Attachment 1: HWS_LED_D1800125_V5_sn01.pdf
HWS_LED_D1800125_V5_sn01.pdf
Attachment 2: HWS_LED_D1800125_V5_sn01.m
% get the beam size from the HWS ETM source D1800125-v5_sn01

[out,r] = system('tar -xf HWS*.tar');

% load the files
dist = [1,10,29,51,84,105,140,180,240,295,351,435,490,565]; % beam propagation distance
files = dir('*.raw');
close all

... 49 more lines ...
Attachment 3: HWS_D1800125-v5_sn01_characterize_20180516.tar
  211   Fri May 18 16:32:37 2018 AidanLaserHartmann sensorHWS green LED fiber launcher - D1800125-v5_SN01

Title was wrong - this is actually config [12,2,4,125]

Quote:

Here is the output from D1800125-v5_SN01.

Quote:

And here's the output of the fiber launcher when I fixed it at 313mm from the camera, attached an iris to the front and slowly reduced the aperture of the iris.

The titles reflect the calculated second moment of the intensity profiles (an estimate of the equivalent Gaussian beam radius). The iris is successful in spatially filtering the central annular mode at first and then the outer annular mode. 

We'll need to determine the optimum diameter to get good transmission spatially without sacrificing too much power.

Quote:

 [Aidan, Marie]

We tested the output of the fiber launcher D1800125-v3. We were using a 6mm spacer in the SM1 lens tube and 11mm spacer in the SM05 lens tube and the 50 micron core fiber.

The output of the fiber launcher was projected directly onto the CCD. Images of these are attached (coordinates are in pixels where 100 pixels = 1.2mm)

There is a lot of high-spatial frequency light on the output. It looks like there is core and cladding modes in addition to a more uniform background. There was an indication that we could clear up these annular modes with an iris immediately after the fiber launcher but I didn't get any images. We're going to test this next week when we get an SM1 mountable iris.

 

 

 

  214   Thu Jun 21 17:12:38 2018 AidanLaserHartmann sensorTwo inch PBS from Edmund Optics - effect on ETM HWS transmission

I'm considering the 86-711 2" 532nm PBS from Edmund Optics for the ETM HWS at the sites.

https://www.edmundoptics.com/optics/polarization/linear-polarizers/532nm-50mm-diameter-high-energy-laser-line-polarizer/

The effect on the transmission through the system, compared to the THorlabs PBS, is shown in the attached plot.

Conclusion: it looks almost as effective as the Thorlabs PBS with the added benefit of being 2" in diameter.

Attachment 1: PBS_Thorlabs_v_EdmundOptics.pdf
PBS_Thorlabs_v_EdmundOptics.pdf
  224   Tue Dec 4 16:57:08 2018 AidanComputingHartmann sensorUpdated GIT version of HWS code

I changed the HWS code to the new git.ligo HWS version.

  • Object files are in ~/hws
  • scripts are in ~/hws-server
  • utilities are still in old git repo moved to ~/.HWS_code_temporary_home

I've set up some symbolic links to these directories to mimic the old directory structure, so ..

  • ~/pyHWS links to new object file directory
  • ~/pyHWS/scripts lnks to new scripts directory
  • ~/pyHWS/utilities links to old utilities directory
  156   Tue Nov 29 09:13:49 2011 ranaMiscLIGO 3GSwitching from CO2 to shorter wavelength solid state laser

 Around a year ago, Phil and I discussed the possibility of using an OPO to possibly generate our own laser beam at ~2 microns for TCS. This was to avoid all of the usual hassle of the 10 micron CO2 laser.

As it turns out, the 1.5-3 micron range doesn't have enough absorption in fused silica: the absorption depth would be basically the whole thickness of the optics and this is not so useful when trying to correct surface heating.

During my recent trip to JILA, Jan Hall mentioned to me that it should be possible to operate instead at ~5 microns, where laser technology may be solid state and where we can use Si:As detectors instead of the inefficient HgCdTe ones which we use now.

JWST, in partnerships with industry, have developed some Si:As detectors:  http://www.jwst.nasa.gov/infrared.html   

Some internet searching shows that there are now several laser technologies for the mid-IR or MWIR range. Some are <1 W, but some are in the ~10 W range.

Of course, its possible that we'll switch to Silicon substrates, in which case we need to re-evaluate the goals and/or existence of TCS.

  159   Thu Jun 7 00:23:11 2012 Aditi MittalMiscLIGO 3GSummary June 5 and June 6, 2012

 June 5

-Discussed the actual project outline 

-Installed Comsol on the system

-Learned the basics of Comsol with the help of tutorials available on 40m wiki

and others.

 

June 6

-Made few simple models in Comsol 

-Studied LIGO GWADW slides for a better understanding of the project.

-Setup SVN to access remote repository.

 

 

 

  161   Thu Jun 7 23:24:56 2012 Aditi MittalMiscLIGO 3GSummary June 7, 2012

-Created a COMSOL model for variation of temperature in two mass system.

-Used the above model for cryogenic conditions.

-checked it analytically.

  163   Fri Jun 8 23:51:13 2012 Aditi MittalMiscLIGO 3GSummary June 8, 2012

-Created a COMSOL model for cryogenically shielded test mass with compensation plate.

-Analyzed the behavior of the model in different size configurations.

  165   Mon Jun 11 20:53:31 2012 Aditi MittalMiscLIGO 3GSummary June 11, 2012

 -Continued with the same cryogenic model created and varied the length of  outer shield and studied the temperature variation inside.

-Compared the temperature difference given by COMSOL with manually calculated one.

  167   Thu Jun 14 05:37:30 2012 Aditi MittalMiscLIGO 3GSummary June 13, 2012

-Derived formula for manual calculation of temperature due to total influx.

-Compared the results by COMSOL and by the formula.

  170   Mon Jun 18 23:42:39 2012 Aditi MittalMiscLIGO 3GCryogenic Shielding

 -Read about blue team design for maximum power budget.

-Read third generation talks to get a better understanding of the work. 

  173   Thu Jun 21 11:16:27 2012 Aditi MittalMiscLIGO 3GSummary June 19 and 20, 2012

 -Attented LIGO orientation meeting and safety session.

-Prepared 3 week report

 

Attachment 1: temp_of_test_mass.png
temp_of_test_mass.png
  175   Thu Jun 21 18:35:44 2012 Aditi MittalMiscLIGO 3GSummary June 21, 2012

 -Updated 3 week progress report with new additions and deletions.

-Attended LIGO lecture which was very interesting and full of information.

  176   Tue Jun 26 17:55:44 2012 Aditi MittalMiscLIGO 3GSummary June 26, 2012

- Discussed the further project with Dr. Brooks.

-Tried to derive formula for the test mass inside cryogenic shield(infinitely long shield from one side)

 

 

 

  179   Fri Jun 29 15:51:24 2012 Aditi MittalMiscLIGO 3GSummary June 28 and 29, 2012

 -Discussed the project outline for next 6 weeks.

-made a write up for the tasks. (attached)

-Analyzed the variation of temperature of the test mass with input power for different lengths of the shield. 

Attachment 1: pipelength_power.xlsx
Attachment 2: 6week_plan.doc
  1   Fri Nov 6 20:09:47 2009 AidanLaserLaserTest

 Does this work?

  2   Thu Dec 10 22:23:47 2009 Not AidanLaserLaserTest

Yes.

Quote:

 Does this work?

  90   Tue Aug 17 16:31:55 2010 AidanThings to BuyLaserBought a laser diode from Thorlabs for HWS

http://www.thorlabs.com/thorProduct.cfm?partNumber=CPS180

I bought this laser diode from Thorlabs today to try the current modulation trick Phil and I discussed last Friday. 

That is:

  1. Accept that there will be interference fringes on the Hartmann sensor probe beam with a laser diode source (especially if the probe beam is the retro-reflection from a Michelson interferometer with a macroscopic arm length difference)
  2. Modulate the current of the laser diode source to vary its wavelength by a few hundreds on nm. Do this on a time scale that is much faster than the exposure time for a Hartmann sensor measurement
  3. The contrast of the interference fringes should average out and the exposure should appear to be the sum of two incoherent beams.

 

 

 

  93   Mon Aug 23 08:43:16 2010 AidanThings to BuyLaserBought a laser diode from Thorlabs for HWS

It arrived on Friday.

Quote:

http://www.thorlabs.com/thorProduct.cfm?partNumber=CPS180

I bought this laser diode from Thorlabs today to try the current modulation trick Phil and I discussed last Friday. 

That is:

  1. Accept that there will be interference fringes on the Hartmann sensor probe beam with a laser diode source (especially if the probe beam is the retro-reflection from a Michelson interferometer with a macroscopic arm length difference)
  2. Modulate the current of the laser diode source to vary its wavelength by a few hundreds on nm. Do this on a time scale that is much faster than the exposure time for a Hartmann sensor measurement
  3. The contrast of the interference fringes should average out and the exposure should appear to be the sum of two incoherent beams.

 

 

 

 

  107   Fri Feb 18 14:53:50 2011 PhilLaserLaserLTG laser delivering specified power

I got the LTG CO2 laser to deliver 50.02W as measured by the Thorlabs 200W power head today. This required running the Glassman HV supply at full power (30.0kV, 31.1mA), tweaking the end grating and output coupler alignments, and cleaning the ZnSe Brewster windows on the laser tubes, and it only lasted a few seconds before dropping back to ~48W, but the laser delivered the specified power. In the factory it delivered 55W at the 10.6 micron line I am using now- I checked it with the CO2 laser spectrum analyzer- so there is more work to do.

  110   Thu Feb 24 10:23:31 2011 Christopher GuidoLaserLaserLTG initial noise

Cheryl Vorvick, Chris Guido, Phil Willems

Attached is a PDF with some initial noise testing. There are 5 spectrum plots (not including the PreAmp spectrum) of the laser. The first two are with V_DC around 100 mV, and the other three are with V_DC around 200 mV. (As measured with the 100X gain preamplifier, so ideally 1 and 2 mV actual) We did one spectrum (at each power level)  with no attempt of noise reduction and one spectrum with the lights off and a make shift tent to reduce air flow. The 5th plot is at 200mv with the tent and the PZT on. (The other 4 have the PZT off).

 

The second plot is just the spectrums divided by their respectives V_DC to get an idea of the RIN.

Attachment 1: LTG_InitialTest.pdf
LTG_InitialTest.pdf LTG_InitialTest.pdf
  150   Thu May 26 11:56:46 2011 AidanElectronicsLaserGreen Laser Pointer beam profile

This measurement was made with the Thorlabs DCC1545M-GL camera with an RG850 3mm long-pass filter over the CCD.

The beam radius (w) is 191 pixels, where the beam intensity = exp[-2 (x/w)^2 ]

The pixel size is 5.2um. Hence the beam size is 993.2um, which is basically near enough to 1mm radius.

 

Attachment 1: green_laser_pointer.bmp
Attachment 2: green_laser_pointer.pdf
green_laser_pointer.pdf
Attachment 3: Screen_shot_2011-05-26_at_12.28.20_PM.png
Screen_shot_2011-05-26_at_12.28.20_PM.png
  207   Thu Feb 8 15:42:13 2018 AidanLaserLaserLow power CO2 laser [400mW] beam size measurement

I did a beam size/beam propagation measurement of the low power CO2 laser (Access Laser L3, SN:154507-154935)

 

Attachment 1: L3_CO2_laser_beam_size.m
% 400mW CO2 laser beam propagation measurement
% measurements of Access Laser L3 CO2 output power (at about 30% PWM)
% SN: 154507-154935
%
% Aidan Brooks, 8-Feb-2018

xposn = 10.5:-0.5:5.0;
dataIN = [25	113.2	113.2	112.7	112.3	110.2	108.6	98.2	74.6	40.6	13.5	2.3	0.3
50	114.5	114.5	114.9	115	114.9	112.1	100	74.2	38.8	12.5	2	-0.1
... 62 more lines ...