40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  TCS elog, Page 2 of 5  Not logged in ELOG logo
    Reply  Thu Dec 10 22:23:47 2009, Not Aidan, Laser, Laser, Test 
Yes.


    
        
            Quote:
        
        
            
Entry  Tue Aug 17 16:31:55 2010, Aidan, Things to Buy, Laser, Bought a laser diode from Thorlabs for HWS 
http://www.thorlabs.com/thorProduct.cfm?partNumber=CPS180
I bought this laser diode from Thorlabs today to try the current modulation trick Phil and I discussed last Friday. 
That is:
    Reply  Mon Aug 23 08:43:16 2010, Aidan, Things to Buy, Laser, Bought a laser diode from Thorlabs for HWS 
It arrived on Friday.


    
        
            Quote:
        
        
            
Entry  Fri Feb 18 14:53:50 2011, Phil, Laser, Laser, LTG laser delivering specified power 
I got the LTG CO2 laser to deliver 50.02W as measured by the Thorlabs 200W power head today. This required running the Glassman HV supply at full power
(30.0kV, 31.1mA), tweaking the end grating and output coupler alignments, and cleaning the ZnSe Brewster windows on the laser tubes, and it only lasted
a few seconds before dropping back to ~48W, but the laser delivered the specified power. In the factory it delivered 55W at the 10.6 micron line I am using
Entry  Thu Feb 24 10:23:31 2011, Christopher Guido, Laser, Laser, LTG initial noise LTG_InitialTest.pdf
Cheryl Vorvick, Chris Guido, Phil Willems
Attached is a PDF with some initial noise testing. There are 5 spectrum plots (not including the PreAmp spectrum) of the laser. The first two are
with V_DC around 100 mV, and the other three are with V_DC around 200 mV. (As measured with the 100X gain preamplifier, so ideally 1 and 2 mV actual) We
Entry  Thu May 26 11:56:46 2011, Aidan, Electronics, Laser, Green Laser Pointer beam profile green_laser_pointer.bmpgreen_laser_pointer.pdfScreen_shot_2011-05-26_at_12.28.20_PM.png
This measurement was made with the Thorlabs DCC1545M-GL camera with an RG850 3mm long-pass filter over the CCD.
The beam radius (w) is 191 pixels, where the beam intensity = exp[-2 (x/w)^2 ]
The pixel size is 5.2um. Hence the beam size is 993.2um, which is basically near enough to 1mm radius.
Entry  Thu Feb 8 15:42:13 2018, Aidan, Laser, Laser, Low power CO2 laser [400mW] beam size measurement L3_CO2_laser_beam_size.mL3_CO2_Laser_154507-154935.pdf
I did a beam size/beam propagation measurement of the low power CO2 laser (Access Laser L3, SN:154507-154935)

 
Entry  Tue Nov 29 09:13:49 2011, rana, Misc, LIGO 3G, Switching from CO2 to shorter wavelength solid state laser 
 Around a year ago, Phil and I discussed the possibility of using an OPO to possibly generate our own laser beam at ~2 microns for TCS. This was
to avoid all of the usual hassle of the 10 micron CO2 laser.
As it turns out, the 1.5-3 micron range doesn't have enough absorption in fused silica: the absorption depth would be basically the whole thickness
Entry  Thu Jun 7 00:23:11 2012, Aditi Mittal, Misc, LIGO 3G, Summary June 5 and June 6, 2012 
 June 5
-Discussed the actual project outline 
-Installed Comsol on the system
Entry  Thu Jun 7 23:24:56 2012, Aditi Mittal, Misc, LIGO 3G, Summary June 7, 2012 
-Created a COMSOL model for variation of temperature in two mass system.
-Used the above model for cryogenic conditions.
-checked it analytically.
Entry  Fri Jun 8 23:51:13 2012, Aditi Mittal, Misc, LIGO 3G, Summary June 8, 2012 
-Created a COMSOL model for cryogenically shielded test mass with compensation plate.
-Analyzed the behavior of the model in different size configurations.
Entry  Mon Jun 11 20:53:31 2012, Aditi Mittal, Misc, LIGO 3G, Summary June 11, 2012 
 -Continued with the same cryogenic model created and varied the length of  outer shield and studied the temperature variation inside.
-Compared the temperature difference given by COMSOL with manually calculated one.
Entry  Thu Jun 14 05:37:30 2012, Aditi Mittal, Misc, LIGO 3G, Summary June 13, 2012 
-Derived formula for manual calculation of temperature due to total influx.
-Compared the results by COMSOL and by the formula.
Entry  Mon Jun 18 23:42:39 2012, Aditi Mittal, Misc, LIGO 3G, Cryogenic Shielding 
 -Read about blue team design for maximum power budget.
-Read third generation talks to get a better understanding of the work. 
Entry  Thu Jun 21 11:16:27 2012, Aditi Mittal, Misc, LIGO 3G, Summary June 19 and 20, 2012 temp_of_test_mass.png
 -Attented LIGO orientation meeting and safety session.
-Prepared 3 week report
 
Entry  Thu Jun 21 18:35:44 2012, Aditi Mittal, Misc, LIGO 3G, Summary June 21, 2012 
 -Updated 3 week progress report with new additions and deletions.
-Attended LIGO lecture which was very interesting and full of information.
Entry  Tue Jun 26 17:55:44 2012, Aditi Mittal, Misc, LIGO 3G, Summary June 26, 2012 
- Discussed the further project with Dr. Brooks.
-Tried to derive formula for the test mass inside cryogenic shield(infinitely long shield from one side)
 
Entry  Fri Jun 29 15:51:24 2012, Aditi Mittal, Misc, LIGO 3G, Summary June 28 and 29, 2012 pipelength_power.xlsx6week_plan.doc
 -Discussed the project outline for next 6 weeks.
-made a write up for the tasks. (attached)
-Analyzed the variation of temperature of the test mass with input power for different lengths of the shield. 
Entry  Mon Apr 12 08:55:37 2010, Aidan, Computing, Hartmann sensor, EDT frame grabber is here 
 The EDT PCIe4 DV C-Link frame grabber arrived this morning. There is a CD of drivers and software with it that I'll back up to the wiki or 40m
svn sometime soon.
Entry  Mon Apr 12 17:25:01 2010, Aidan, Electronics, Hartmann sensor, Fiber-Camera Link demonstration hartmann_sensor.JPGFiber_Camera_Link_1.jpgFiber_Camera_Link_3.jpg
 I installed the EDT PCIe4 DV C-Link frame grabber in a spare Windows XP PC and connected the Dalsa 1M60 camera directly to it via the CameraLink
cable. In this configuration I was able to access the menu system in the camera using the supplied serial_cmd.exe
routine.
Entry  Thu Apr 15 01:47:47 2010, Won Kim, Computing, Hartmann sensor, Notes on installing EDT PCIe4 DV frame grabber 
* EDT PCIe4 DV frame grabber: installation notes for linux system 

(Note)
Entry  Tue Apr 20 18:05:24 2010, Aidan, Computing, Hartmann sensor, Images off the Dalsa Camera in CentOS Screenshot-PCI_DV_Display.pngScreenshot-PCI_DV_Display-1.png-opt-EDTpdv.png
 I installed CentOS on the machine with the EDT frame-grabber. I then installed the frame-grabber software from the CD.
In the /opt/EDTpdv/ directory the camconfig program was run and I entered "331" to start
the frame-grabber and run with the Dalsa 1M60 settings ... this was necessary to get the frame grabber running, but didn't seem to force 
Entry  Thu Apr 22 08:20:51 2010, Aidan, Computing, Hartmann sensor, Installed MATLAB and Windows XP Virtualization on Hartmann machine 
I installed a Windows XP virtualization on the Hartmann machine. It can be accessed from the desktop, or by running virt-manager at the command line. Once the virtualization manager starts the virtualization of Windows needs to be started. It runs quite
slowly.
Entry  Tue May 4 09:18:15 2010, Aidan, Computing, Hartmann sensor, Added aliases and icons for EPICS commands and dataviewer etc. to hartmann 
I updated the .bashrc file in controls@hartmann to include aliases for the ezca EPICS commands and a few others. Details shown below:
Also added launchers to the top panel for MATLAB, sitemap, dataviewer and StripTool. The icons for the launchers are located in:
/cvs/users/ops/ligo-launchers/icons
Entry  Tue May 4 10:30:07 2010, Aidan, Computing, Hartmann sensor, EPICS and MEDM screen for Hartmann sensor Screenshot-C4HWS_medm_21.adl_(edited).pngScreenshot-StripTool_Graph_Window.pngHWS.dbHWS.cmdread_dalsa_temperature_write_to_epics.m
I added the Dalsa 1M60 temperature measurements to EPICS. The break down is as follows:

    
        
             
            Digitizer Board Temperature
            Sensor Board Temperature
        
        
  
Entry  Tue May 4 13:35:13 2010, Aidan, Computing, Hartmann sensor, Hartmann temperature channels in frame builder 
 I've added the digitizer and sensor board temperature readings from the HWS to the frames. This was done in the following way
1. Create a new file /cvs/cds/caltech/chans/daq/C4TCS.ini - with the channels in it - see below
2.  open /cvs/cds/caltech/target/fb1/master
Entry  Wed May 5 09:04:01 2010, Aidan, Computing, Hartmann sensor, Added /home/controls/scripts/modules directory to PYTHONPATH on hartmann 
 I added the following line to ~/.bashrc
 
export PYTHONPATH=/home/controls/scripts/modules:/usr/local/lib/python
Entry  Wed May 5 18:45:51 2010, Aidan, Computing, Hartmann sensor, Python code to interface the Dalsa1M60 and export the temperature to EPICS dalsa_to_epics.pyVerifyTemperature.pydalsa_1m60.cfg
Python script
I wrote a Python script, ~/scripts/dalsa_to_epics.py that reads the temperature off the camera using
serial_cmd vt and then it writes this to the EPICS channels using ezcawrite.
Entry  Thu May 6 10:34:38 2010, Aidan, Computing, Hartmann sensor, EPICS and MEDM screen for Hartmann sensor - part 2 dalsa_to_epics.pyGetCameraParameters.pyHWS.dbC4TCS.ini
I added the camera parameters to EPICS and the MEDM screen. These are available as channels now in EPICS and eventually there will be a python script
that writes the EPICS value to those channels, but right now it is just a python script that reads the values off the Dalsa camera.
I updated the channels in /cvs/cds/caltech/chans/daq/C4TCS.ini so that these are saved to the
Entry  Thu May 6 12:32:11 2010, Aidan, Computing, Hartmann sensor, dalsa_to_epics Python script crashed ... 
Here's the error:
 
 Traceback (most recent call last):
Entry  Thu May 6 21:32:26 2010, Won Kim, Computing, Hartmann sensor, Peak detection and centroiding code detect_peaks_uml.m
 
 
Attached is .m file of the custom function that I wrote and used to automatically detect peaks in a
    Reply  Tue May 11 10:32:00 2010, Aidan, Computing, Hartmann sensor, Peak detection and centroiding code - review test_centroid_code.m
This looks really efficient! However, I think there's a systematic error in the calculation. I tested it on some simulated data and it had trouble getting
the centroids exactly right. I need to better understand the functions that are called to get an idea of what might be the problem.
 
Entry  Thu May 13 16:54:46 2010, Aidan, Computing, Hartmann sensor, Running MATLAB programs in C on CentOS - only use R2008b for less hassle 
 After much effort trying to get a MATLAB routine to compile in C I discovered the following pieces of information.
1. CentOS will not install a gcc compiler more recent than 4.1.2 with yum install. This is circa 2007. If you want a more recent compiler it must be installed manually.
Entry  Thu May 20 08:20:54 2010, Aidan, Computing, Hartmann sensor, Centroiding algorithm and code to generate simulated data test_spot_generation_and_centroiding.mgenerate_simulated_spots.mcentroid_image.m
 Here's a copy of an email I distributed today that describes the centroid and simulation code I wrote.
Hi Won,
Entry  Mon May 24 19:17:32 2010, Aidan, Laser, Hartmann sensor, Replaced Brass Plate with Invar Hartmann Plate 
I just replaced the brass Hartmann plate with the Invar one. The camera was off during the process but has been turned on again. The camera is now warming
up again. I've manually set the temperature in the EPICS channels by looking at the on-board temperature via the serial communications.
 I also made sure the front plate was secured tightly.
Entry  Wed May 26 14:58:04 2010, Aidan, Misc, Hartmann sensor, Hartmann sensor cooling fins added HWS_CONFIG1.jpg
14:55 -  Mindy stopped by with the copper heater spreaders and the cooling fins for the Hartmann sensor. We've set them all up and have turned on
the camera to see what temperature above ambient is achieves.
17:10 - Temperature of the HWS with no active cooling( Digitizer = 44.1C, Sensor = 36.0C, Ambient = 21.4C)
Entry  Thu May 27 08:25:37 2010, Aidan, Electronics, Hartmann sensor, Hartmann sensor cooling fins added - base piece removed HWS_CONFIG2.jpg
 8:10AM - I removed the base plate from the Hartmann sensor. I want to know what steady-state temperature the HWS achieves without the plate.
The photo below shows the current configuration.
11:22AM - (Digitizer - 52.2C, Sensor - 43.8C, Ambient - 21.8C)
Entry  Thu May 27 13:18:51 2010, Aidan, Electronics, Hartmann sensor, Removed Cooling fins from Hartmann sensor HWS_CONFIG3.jpg
I removed the cooling fins from the Hartmann sensor to see what steady-state temperature it reached without any passive cooling elements. I also dropped
the set-point temperature for the lab to help keep for getting too hot.
After nearly 3 hours the temperature is:
Entry  Thu May 27 15:42:06 2010, Aidan, Electronics, Hartmann sensor, Hartmann sensor with just the base piece HWS_CONFIG4.jpg
I switched in just the base piece of the Hartmann sensor. The cooling fins are removed. I bolted the camera securely to the base plate and I bolted the
plate securely to the table.
 5:00PM - (Digitizer = 41.9C, Sensor = 33.8C, Ambient = 19.3C)
Entry  Thu May 27 17:49:02 2010, Aidan, Electronics, Hartmann sensor, Hartmann sensor cooling fins added - base piece added HWS_CONFIG1-tight.jpg
 Back to Configuration 1 again - this time the fins were bolted very securely to the camera.
 7:25 PM - [about 2 hours later] - Digitizer = 39.7C, Sensor = 31.4C, Ambient = 19.0C
 
Entry  Tue Jun 15 16:30:10 2010, Peter Veitch, Misc, Hartmann sensor, Spot displacement maps - temperate sensitivity tests spot_displacements_same_temp_0611.jpgspot_displacements_diff_temp_0611.jpg
Results of initial measurement of temperature sensitivity of Hartmann sensor
"Cold" images were taken at the following temperature:
| before | 32.3 | 45.3 | 37.0 |
    Reply  Wed Jun 16 11:47:11 2010, Aidan, Misc, Hartmann sensor, Spot displacement maps - temperate sensitivity tests - PRISM Prism_as_radial_vector.pdf
I think that the second plot is just showing PRISM and converting it to its radial components. This is due to the fact that the sign of the spot displacement
on the LHS is the opposite of the sign of the spot displacement on the RHS. For spherical or cylindrical power, the sign of the spot displacement should
be the same on both the RHS and the LHS.
Entry  Thu Jun 17 07:40:07 2010, James K, Misc, Hartmann sensor, SURF Log -- Day 1, Getting Started 
 For Wednesday, June 16:
 
I attended the LIGO Orientation and first  Introduction to LIGO lecture in the morning.
Entry  Thu Jun 17 22:03:51 2010, James K, Misc, Hartmann sensor, SURF Log -- Day 2, Getting Started 
For Thursday, June 17:
Today I attended a basic laser safety training orientation, the second Introduction to LIGO lecture, a Summer Research Student Safety Orientation,
and an Orientation for Non-Students living on campus (lots of mandatory meetings today). I met with Dr. Willems and Dr. Brooks in the morning and went
Entry  Sat Jun 19 17:31:46 2010, James K, Misc, Hartmann sensor, SURF Log -- Day 3, Initial Image Analysis 7x

For Friday, June 18:
(note that I haven't been working on this stuff all of Saturday or anything, despite posting it now. It was getting late on Friday
Entry  Tue Jun 22 00:21:47 2010, James K, Misc, Hartmann sensor, Surf Log -- Day 4, Hartmann Spot Flickering Investigation 7x
 I started out the day by taking some images from the CCD with the
OLED switched off, to just look at the pattern when it's dark. The images looked like this:
Entry  Tue Jun 22 22:30:24 2010, James K, Misc, Hartmann sensor, SURF Log -- Day 5, more Hartmann image preliminary analysis 13x
 Today I spoke with Dr. Brooks and got a rough outline of what my experiment for the next few weeks will entail. I'll be getting more of the details
and getting started a bit more, tomorrow, but today I had a more thorough look around the Hartmann lab and we set up a few things on the optical table.
The OLED is now focused through a microscope to keep the beam from diverging quite as much before it hits the sensor, and the beam is roughly aligned to
    Reply  Wed Jun 23 06:49:48 2010, Aidan, Misc, Hartmann sensor, SURF Log -- Day 5, more Hartmann image preliminary analysis 
Nice work!
 


    
        
            
Entry  Wed Jun 23 22:57:22 2010, James K, Misc, Hartmann sensor, SURF Log -- Day 6, Centroiding 10x
 So in addition to taking steps towards starting to set stuff up for the experiment in the lab, I spent a good deal of the day figuring out how
to use the pre-existing code for finding the centroids in spot images. I spent quite a bit of time trying to use an outdated version of the code that didn't
work for the actual captured images, and then once I was directed towards the right version I was hindered for a little while by a bug.
Entry  Fri Jun 25 00:11:13 2010, James K, Misc, Hartmann sensor, SURF Log -- Day 7, SLED Beam Characterization BeamData.xlsxbeam_pdfs.zipbeamdata.mgsbeam.m
 
BACKGROUND:

In order to conduct future optical experiments with the SLED and to be able to predict the behavior of the beam as it propagates across the
ELOG V3.1.3-