  251   Wed May 11 19:46:21 2022 Aidan, Radhika, JordanLab InfrastructureVacuum chamberLeak cleaning on IR labs vacuum chamber - suspects

[Aidan, Jordan, Radhika]

Radhika and Jordan identified some particulates (hair and flecks of foil) on the O-ring on the IR labs dewer. Additionally, we saw a scratch in the O-ring groove and a nick on the metal of the base of the dewer where it meets the O-ring. All were in the leaky vicinity previously identified by the He testing.

We set up a cradle to hold the dewer while we are working on it. Still needs vertical supports.

R & J replaced the O-ring with a new one with Crytox applied. 

Attachment 1: IMG_8812.jpg
IMG_8812.jpg
Attachment 2: IMG_8799.jpg
IMG_8799.jpg
Attachment 3: IMG_8789.jpg
IMG_8789.jpg
Attachment 4: IMG_8805.jpg
IMG_8805.jpg