40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  TCS elog  Not logged in ELOG logo
Message ID: 32     Entry time: Thu May 6 10:34:38 2010
Author: Aidan 
Type: Computing 
Category: Hartmann sensor 
Subject: EPICS and MEDM screen for Hartmann sensor - part 2 

I added the camera parameters to EPICS and the MEDM screen. These are available as channels now in EPICS and eventually there will be a python script that writes the EPICS value to those channels, but right now it is just a python script that reads the values off the Dalsa camera.

I updated the channels in /cvs/cds/caltech/chans/daq/C4TCS.ini so that these are saved to the daq and I also restarted the daq daemon.

The python script that gets the camera parameters is here: scripts/Dalsa1M60/GetCameraParameters.py and the script that writes the parameters to the EPICS channels is here scripts/dalsa_to_epics.py.

These are attached as is C4TCS.ini and HWS.db which defines the new channels.

Attachment 1: dalsa_to_epics.py  3 kB  Uploaded Thu May 6 11:40:40 2010  | Show | Hide all | Show all
Attachment 2: GetCameraParameters.py  16 kB  Uploaded Thu May 6 11:40:54 2010  | Show | Hide all | Show all
Attachment 3: HWS.db  927 Bytes  Uploaded Thu May 6 11:41:22 2010  | Hide | Hide all | Show all
record(ai,"C4:TCS-HWS_TEMP_DIGITIZER")
record(ai,"C4:TCS-HWS_TEMP_SENSOR")
record(ai, "C4:TCS-HWS_TAP1GAIN")
record(ai, "C4:TCS-HWS_TAP2GAIN")
record(ai, "C4:TCS-HWS_PRETRIGGER")
record(ai, "C4:TCS-HWS_DATA_MODE")
record(ai, "C4:TCS-HWS_BINNING_MODE")
record(ai, "C4:TCS-HWS_GAIN_MODE")
record(ai, "C4:TCS-HWS_OUTPUT_CONFIG")
record(ai, "C4:TCS-HWS_EXPOSURE_MODE")
... 27 more lines ...
Attachment 4: C4TCS.ini  428 Bytes  Uploaded Thu May 6 11:41:52 2010  | Show | Hide all | Show all
ELOG V3.1.3-