Entry  Mon Jul 26 16:23:03 2021, Aidan, Misc, Flood, Lab flooded from broken pipe leading to sump room IMG_3881.MOV
    Reply  Wed Aug 25 11:28:56 2021, Aidan, Misc, Flood, Lab flooded from broken pipe leading to sump room 
       Reply  Thu Aug 26 09:00:39 2021, Aidan, Misc, Flood, Lab flooded from broken pipe leading to sump room 20210825_-_TCS_Lab_Flood_-_compressed.pdf
Message ID: 248     Entry time: Wed Aug 25 11:28:56 2021     In reply to: 245     Reply to this: 249
Author: Aidan 
Type: Misc 
Category: Flood 
Subject: Lab flooded from broken pipe leading to sump room 

11:29AM - Lab has flooded again this morning. I'm calling PMA. Looks to be the same issue as before.

Quote:

 

Koji: QIL/TCS entrance flooding. Check your lab

Anchal: Can someone take a look at CTN too?

Koij: TCS needs more people @aidan

Koji: CTN ok

Aidan: On my way

Shruti: Cryo seems fine

Aidan: There was a leak in a pipe in the wall of B265A. It was coming from the building air conditioner condensation overflow. Facilities has fixed the pipe and is working on clean-up

 