Message ID: 237     Entry time: Thu Aug 1 15:20:39 2019
Author: Edita Bytyqi 
Type: Lab Infrastructure 
Category: Optics 
Subject: Reflector Mount and DC Supply 

Yesterday, we were able to take some data using the 120 V DC power supply. The reflectors cut at the focal point and radius were both tested; the semi-circle cut proved to give a better focus, likely because roughly half the heat is lost using the focal-point reflectors. For upcoming tests, the semicircle reflectors will be used. We varied the surface shine by using the dull and reflective side of Al foil, as well as using the machined Al itself. The best result was given by using the more reflective side of Al foil.

Figure 1 shows the steady-state surface deformation profile detected by the HWS. The heaters don't have a uniform distribution along the wire, so more heat is radiated in the center of it, thus more of it is being focused to the center of the test optic. The data needs to be analyzed to determine the radius of the focus. Our rough estimate is about ~1.5 - 2 cm. We cannot collect any more data until we get a new power supply (AC 120 V).

Today, I came up with a new design for mounting the reflectors. I used a big 3-axis stage and a small 4-axis stage. This provides 5 degrees of freedom: 3 translational and 2 rotational, which is what we need for fine-tuning the focus and directing it at different angles incident to the test optic. The only problem with this design is that the 3-axis stage is too tall for the box, so the lid won't close.There is a smaller one available, but I have to figure out a way to increase its height, since the screw size is different from the ones on the pedestals available.

Additionally, Chub used metal-to-metal epoxy to glue a screw to the back of a reflector. I will wait until tomorrow to test it, because it is a slow acting epoxy. If it works, I have the necessary tools to do the same with the other reflectors. With the current deisgn the reflector wil be screwed in to where the round screw is in the stage. If it heats up a lot and affects the material of the stages, a small optical post (top of stage) will be used to make up for the absorbed heat.

 

Attachment 1: 20190731_170920.jpg  4.517 MB  Uploaded Thu Aug 1 16:41:17 2019  | Show | Hide all | Show all
Attachment 2: 20190801_145836.xcf  45.956 MB  Uploaded Thu Aug 1 16:41:36 2019
Attachment 3: 20190801_145836.jpg  4.751 MB  Uploaded Thu Aug 1 16:41:53 2019  | Hide | Hide all | Show all
20190801_145836.jpg
Attachment 4: 20190801_150635.jpg  4.484 MB  Uploaded Thu Aug 1 16:42:11 2019  | Hide | Hide all | Show all
20190801_150635.jpg
Attachment 5: 20190801_145842.jpg  4.468 MB  Uploaded Thu Aug 1 16:42:22 2019  | Hide | Hide all | Show all
20190801_145842.jpg
ELOG V3.1.3-