40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  TCS elog  Not logged in ELOG logo
Entry  Fri May 11 12:32:21 2018, Aidan, Laser, Hartmann sensor, HWS green LED fiber launcher LED_test_01_20180511.pdf
    Reply  Fri May 11 16:18:47 2018, Aidan, Laser, Hartmann sensor, HWS green LED fiber launcher LED_test_02_20180511.pdf
       Reply  Wed May 16 16:47:29 2018, Aidan, Laser, Hartmann sensor, HWS green LED fiber launcher - D1800125-v5_SN01 HWS_LED_D1800125_V5_sn01.pdfHWS_LED_D1800125_V5_sn01.mHWS_D1800125-v5_sn01_characterize_20180516.tar
          Reply  Fri May 18 16:32:37 2018, Aidan, Laser, Hartmann sensor, HWS green LED fiber launcher - D1800125-v5_SN01 
Message ID: 208     Entry time: Fri May 11 12:32:21 2018     Reply to this: 209
Author: Aidan 
Type: Laser 
Category: Hartmann sensor 
Subject: HWS green LED fiber launcher 

 [Aidan, Marie]

We tested the output of the fiber launcher D1800125-v3. We were using a 6mm spacer in the SM1 lens tube and 11mm spacer in the SM05 lens tube and the 50 micron core fiber.

The output of the fiber launcher was projected directly onto the CCD. Images of these are attached (coordinates are in pixels where 100 pixels = 1.2mm)

There is a lot of high-spatial frequency light on the output. It looks like there is core and cladding modes in addition to a more uniform background. There was an indication that we could clear up these annular modes with an iris immediately after the fiber launcher but I didn't get any images. We're going to test this next week when we get an SM1 mountable iris.

Attachment 1: LED_test_01_20180511.pdf  407 kB  Uploaded Fri May 11 13:35:13 2018  | Hide | Hide all
LED_test_01_20180511.pdf
ELOG V3.1.3-