Message ID: 119     Entry time: Tue Mar 1 17:05:45 2011
Author: Aidan 
Type: Electronics 
Category: Hartmann sensor 
Subject: Dalsa 1M60 current draw 

Steve and I measured the current drawn by the Dalsa 1M60 by connecting it to the BK Precision 1735 lab power supply that display current and voltage supplied. We tried the camera at a variety of different voltages. The results are presented below:

Voltage  Current(t<5s)  Current(5s<t<10s)   Current(t>10s)

12.7V       0.6A              0.8A              1.11A

15.0V       0.55A             0.69A             0.91A

18.0V       0.41A             0.57A             0.75A

20.0V       0.42A             0.52A             0.67A

Additionally, we tried running the other camera with the lab power supply. I varied the exposure mode and exposure time and checked the current drawn. The supplied voltage was 18.0V.

Exposure Mode 4: current = 0.67A

Exposure Mode 2: 58Hz, exposure time = 16 ms, current = 0.70A

Exposure Mode 2: 58Hz, exposure time = 100 us, current = 0.72A

Exposure Mode 2: 1Hz, exposure time = 998 ms, current = 0.68A

Exposure Mode 2: 1Hz, exposure time = 16 us, gm 0, current = 0.69A

Exposure Mode 2: 1Hz, exposure time = 16 us, gm 2, current = 0.69A