  750   Tue Nov 19 22:29:39 2013 xiaoyueDailyProgressCrackleHysteresis Test Set Up

Attending: EricG, Gabriele, Xiaoyue

We installed two steel blades with different stiffness. By changing the hanging weight we are leveled the two arms, also the two end mirrors roughly. Leveling the whole stage by a more careful balancing of bearing weight, or does it matter much to the noise issue was left as an open question.

Eric walked me through the control system including the matlab program, emdm interfacing, foton, data viewer, striptool, diagnostic test tool, cymac terminal for signal generation. For every new system, a transfer function should be investigated by sweeping the driving frequency; or we can drive with a range of frequencies and extract the information from the output using Fourier analysis, but the coherence function exhibited a poor correlation (<<1, in a random fashion) between input and output signal when we tried this method.

We went through Eric’s python code for hysteresis analysis. We are driving the system on(+) – off – on(-) – off series with the sign indicating the direction of the driving force. The time interval between consequent actions is half an hour. We set it to run two cycles for each of the three levels of driving voltages.

  752   Thu Nov 21 22:58:51 2013 xiaoyueDailyProgressCrackleOptical Setup

Gabriele and I set up a rough Michelson interferometer alignment.

Matlab model -- Gabriele

crackle_setup.jpg

Side view (left) Top view (right)

side_view.jpgtop_view.jpg

where the maroon beams are reflected from end mirrors. 

We tilted both of the 45 degree mirrors to deviate the reflected beam with an angle from incoming beams to bypass mirror 1(M1), to be detected by photodiode (PD).

Also it should be noted that the two end mirrors (ENDM1, ENDM2) have height difference of 1cm approximately, so we differentiate the optical path lengths of the two arms accordingly.

The alignment procedure is listed briefly as below:

- prepare the first mirror with a horizontal beam using irises.

- prepare a vertical reference using two irises aligned with a home-made plumb bob. 

- Align the 45 degrees inclined mirrors for vertical beams.

- Make sure the end mirrors are horizontal by overlapping incoming and reflected beams

- Roughly align the entire setup with beams superimposed

- Tweak the 45 degrees mirrors to separate the beam with a small angle (but large enough to bypass M1)

- Recenter or decenter mirrors to extract the symmetric port 

- Install photodiodes and beam dumps

(We may want to order more visible range mirrors / D-cut mirror / D-cut mirror mounts.)

  753   Tue Nov 26 01:03:45 2013 xiaoyueDailyProgressCrackleSystem Installed

Attending: Eric, Gabriele, Xiaoyue

We overlapped the two arms by first eliminate the fringes as much as possible. However the angular motion of the suspended mirrors kept bringing the fringes in. Then we did finer aligning by maximizing the signal amplitude, making sure a good overlapping is giving magnificent interference (either constructive or destructive). A beating envelope was observed.

Then we installed the system into the chamber. Then we went through the procedures to generate the transfer function for the damping loop. We forgot to turn off the damping which at the beginning raised very strange phase behavior. As people walked around affected the signal a lot, we decided to pump the chamber overnight. 

  772   Wed Feb 12 11:13:01 2014 xiaoyueDailyProgressCrackleseismic noise coupling characterization

In order to study the current setup better to improve the design of the 2nd version experiment, I did some seismic coupling analysis. The plan is to shake the optical table with a white noise (5V, pre-amplified with a bandwidth 10-1k Hz, input 0.5A, 2V) mini-shaker (B&K type 4810), while the motion of the table is monitored by a 3-axis accelerometer. A full description of the seismic noise picture should include analysis of the coupling from table motion to Mich signal, bench (damped by rubbers) inside the chamber to Mich signal, and we also want to characterize different coupling from bench to different optical elements.

The first and easiest thing I tried is to characterize the transfer function between table vibration and Mich signal. The table was shaken in x, y, and z directions separately, with hopefully three linearly independent measurements of a_x, a_y, and a_z measured for each 1hr shaking. In this way we built matrix [a_xx, a_xy, a_xz; a_yx, a_yy, a_yz; a_zx, a_zy, a_zz] where the additional index indicates the shaking direction. With [e_x, e_y, e_z] for each shaking measured by servo, simply solve [a_xx, a_xy, a_xz; a_yx, a_yy, a_yz; a_zx, a_zy, a_zz][T_x, T_y, T_z] = [e_x, e_y, e_z] will give us the transfer function along x, y and z. 

coh8_140128_fit.png

The picture is plotted with coherence between a_ii (where i is the driving direction) and mich signal is larger than 0.8. the result seems to agree with our expectation that the seismic noise transfer function follows the power law trend. 

However, the coherence is very bad. I tried increasing the noise power by limiting bandwidth to 10 - 300 Hz and inputing 0.8A 3.0V to the shaker). From the result of the z-drive result, the coherence in low frequency range is improved a lot, so I am going to finish the three-axial analysis. While the last trial of data is fit by 1/f^2, the later trial is better fit by 1/f^3, but this kind of fit is tricky. There are many resonances and it is difficult to judge which fit is the best. 

driveZ_fit.png

 

I tried superimposing the two and they are similar where both of them have good coherence.

superimpose.png

I also did analysis (keep data with coherence > 0.8) for the coupling from directly the bench motion to optical elements. I mounted another mini-accelerometer on the newport mirror mount and clamp it to the table like what we did inside the chamber.

coh8_140207.png

x has a generally higher level because the accelerometer is mounted along x-direction on the mirror mount. It seems that the transfer function is much smaller than one, which probably indicates a difference in calibration between the two signals. I will at some point mount the accelerometer one next to the other and measure the relative calibration, which should be simply a flat transfer function. 

Another problem here is that I got very bad coherence at low frequency. I have no good explanation why there seems to be a high-pass cutoff around 30 Hz, but we definitely need to push the measurement down to 10 Hz. 

  773   Wed Feb 12 18:02:01 2014 xiaoyueDailyProgressCracklestress-strain info about maraging steel blades

I got a chance to talk with Norna about the strain range we are in for our maraging steel blades:

We typically load the blades to a stress level of 800 MPa to 1000MPa. The upper value there is approx. 55% of the yield stress which is ~1.9 GPa. The Young’s modulus E = 186 GPa.

—> strain rate = d/dt [ (F sin(wt) + F0) / E ] = (wF/E) cos(wt) —> maximum rate = wF/E = (2 pi 0.125 Hz) (800 counts ~ 1um deflection) / 186 GPa

400 MPa ~ 1mm deflection

4 kPa ~ 1um deflection

—> max rate = 1.689e-08 /second

Also, the triangular shape gives equal stress along the length of the blade when loaded.

In addition, as [1996 Dahmen] "Hysteresis, avalanches, and disorder-induced critical scaling: A renormalization-group approach" shows a relationship between hysteresis and crackle, on page 14 878 they show how a model scaling hysteresis loop area with r, where r is associated with the average avalanche size: A_sing ~ r^(2-alpha) (MFT :alpha =0), I am thinking maybe materials with larger hysteresis loop area would generate crackling noise more “easily”. If this is the case, a good candidate would be the AISI 1085, which is also called music wire; it’s a high carbon steel. 

stress_strain.png

Reference: [1998 Beccaria] "The creep problem in the VIRGO suspensions: a possible solution using maraging steel"

  774   Fri Feb 21 11:36:56 2014 xiaoyueDailyProgressCrackleSeismic noise coupling to optical elements

I took the calibration line for the two accelerometers by mounting the small one rigidly next to the 3-axial seismometer and shake to get calib =  transfer function G (w/ coh<0.8=NAN). I also solved the problem where we lost information in low frequency range by correcting the frequency axis -- I used wrong sampling frequency because I did not notice that DMPA_IN1 has sampling frequency of 1/4*8192 while the ACCX_OUT channel is 1/2*8192.

After all the debugging, the coupling transfer function looks like:

smoother_set1_2.png

with resolution of 0.02 Hz, and 2^8 averages (# of hanning windows)

I characterized two different mirror mount configurations -- setup 1 & 2, see figure below -- for comparison. setup2 seemes less robust as expected — as it has an extra post & post-holder joint.

setup.png

Now the setup is ready for further seismic coupling characterization of different optical elements that we might consider using in the upgraded version of experiment.

collection.jpg

  776   Tue Feb 25 23:16:51 2014 xiaoyueDailyProgressCrackleSeismic coupling to different opto-mechanics

I took some measurement for several combinations of opto-mechanics during the weekend:

Screenshot_2014-02-25_23.07.14.png

mirror_mounts.pngtrans_stage.png

The structure around 20Hz is unwanted, and we suspect it's due to the resonance of optical bench. 
Also, there seems to be a non trivial dependence of the structure there on the post length. For example, the shortest one (0.5”) has a large structure, then there is no data for the 1” one, the 1.5” shows no structure, the 2” is large, the 3" is small again and the 4” is large again. Irene Fiori from Virgo [2013 Fiori] argued that the contact surface between the components might play a role.

In this case we repeat the measurements on the small plate on rubber, instead of shaking the whole table.

setup.png

Use DTT to have a live view of the changes:
1. test on the rubber plate —> structure at 20 Hz disappears, seems like the 20 Hz is the bench resonance frequency
2. test by tightening clamps —> structure around 1kHz moves to higher frequency
3. test by tightening mirror to post —> structure at high frequency becomes smaller

Below is an example for change#2, where after tightening the clamp, the green ref. curve (power spectrum of ACC#2) shifts to the red one.

tighten.png

The transfer function is smooth and flat, up to 300Hz.

  777   Wed Feb 26 16:53:22 2014 xiaoyueDailyProgressCrackleNewport Suprema vs. Thorlabs Polaris

From the latest measurements, it seems that Thorlabs Polaris mirror mount will work better for our purpose:

NS_TP_0224.png

We suspect the better junction between mirror mount and the post be the explanation for the smoother /flatter transfer function of TP (Thorlabs Polaris) series. Also, there is a trend that shorter post works better for the TP series. NS (Newport Suprema) has a similar trend except for .5’ post, which has more bumpy structures probably due to the poor clamping. 

 

  783   Fri Apr 11 01:33:28 2014 xiaoyueDailyProgressCrackleOSEM electronics PD calibration

I wired the electronics up according to the satellite amplifier circuit diagram from document LIGO-T040106-01-K.

satellite_amplifier.png

As the OSEM available have LED SME2470 and PD SMD2420 same as the amplifier is designed for. For simplicity I skipped the voltage regulator part  and supply reverse voltage for PD, current for LED using DC voltage supplies. Also, since I am reading Vout uisng a multimeter, no high-current buffer is included in my testing circuit. Another thing to note is that instead of LT1124ACN8 operated at +/- 15V I used OPA604AP operated at +/- 12V, because it's the op amp I found in lab at hand..

I first used a hand-cut aluminum flag, mounted on a translational stage, for calibration. I set reverse voltage as 10V, LED current 35 mA. 

trial2_3.png

By cutting the beam I expected to obtain a symmetric Erf funciton-like Vout vs. flag position data, but as is shown in the figure, trial 2, I got long nonlinear tail for the other half of the data tr, along with the fact that I cannot completely block the light. I trimmed the flag a little bit to make the edge sharper and collected the data as trial 3.

I chose three different fits to analyze the linear behavior range:

 

linear range (mm)

center V_out (V)

sensitivity (V/mm)

black

0.2

8.9

18

blue

0.3

8.4

3

red

0.6

5.4

16

My first guess for the cause of bad shape is the poor quality of my hand-cut flag. Also Gabriele suggested that probably the op amp was saturated at high V_out range.

I changed the flag from the arbitrarily cut square into a slim metallic post, and I decreased the LED current a little bit to ensure the V_out is below 10V to avoid saturation.

0410.png

In this trial I am able to get a much better — in sense of both symmetry and linear range — response. The maximum linear range I am able to fit is 0.8 mm with a sensitivity of 6V/mm.

  785   Fri Apr 25 15:38:20 2014 xiaoyueDailyProgressCrackleOSEM electronics PD calibration

I measured the noise for the circuit using SR785, at 10Hz 15 avgs to be 16.14 nVpk2 /Hz ~ 1e-08 m /sqrt(Hz) at .1'' flag position, which is the midway of our linear range. However compared to the noise value of 1e-10 m/sqrt(Hz), tested in LIGO-T040106-01-K, our noise is higher by 2 order of magnitude.

The most probable source of the excess noise is the fact that I am not using a stabilized current supply for the LED. I added in an adjustable voltage regulator LM317T (as I didn't find a precision 10V reference at hand) and it pushed the noise down to 15.92 pVpk2 /Hz ~ 1e-10 m /sqrt(Hz).

 

  786   Mon Apr 28 00:49:09 2014 xiaoyueDailyProgressCrackleOSEM PD firnal circuit and noise characterization

 Zach designed a wonderful circuit for us, and I wired it up according to the schematics:

OSEM_circuit.png

IMG_0369.JPG

For the two OPA140, which I didn’t find in lab I used OP27E instead, and I used two [R = 150, 3W] in series to substitute for [R13 = 270, 1W]. The original PD readout, without the 2-stage whitening, is very low at only 0.166 V with full LED exposure, thus I changed R16, the transimpedance load from 470 to 27k to give a max 9.9V voltage output. The value was determined by a potentiometer but finally substituted with a metal film resistor for noise concern. I also added a 1nF to the feedback of transimpedance for oscillation attenuation. 

I re-calibrated the linear range to be 1.143 mm with sensitivity of 5.5V/mm at center voltage of 5.4245 V. Using AC coupling input configuration and measure two channels with channel 2 shorted as a reference, the RMS noises at a mid-way flag position are measured as below in units of m /sqrt(Hz) :

 

75Hz

10Hz

1Hz

.5Hz

noise floor

3.4612E-12

3.4512E-12

3.5888E-11

8.4109E-07

output RMS

9.1688E-09

1.9909E-08

2.3692E-10

5.7407E-09

RMS before whitening

8.5183E-11

2.2194E-10

7.5755E-10

3.4784E-07

whatever.png

The performance is in expectation expect that low frequency range (<1Hz) is behaving weird. I measured the transfer function of one 0.75-75 Hz whitening stage to confirm the noise amplification after two-stage whitening.  

IMG_0392.JPG 

To make the analysis more organized I should interface the signal analyzer ASAP..

  787   Tue May 13 02:08:49 2014 xiaoyueDailyProgressCracklecrackle1 seismic, laser intensity and shot noise

I measured the typical optical table motion using the wilcoxen accelerometers, and fed it through the transfer function I measured before for the seismic noise estimation.

I also measured the laser intensity noise along with the shot noise. First I opened the chamber and blocked one arm of the beam to decouple the laser noise from other noises that can rise from the asymmetry of Michelson configuration. Also to make sure my data is understandable, I confirmed that the photodiode is not saturated by plotting out the beam power vs. output voltage curve, using a half waveplate + PBS (Thorlabs PBS101A) in front of the fibre input.

PDA100A_linear.png

Then I used SR785 (interfaced with Q's help) to analyze the laser noises, with the experimental shot noise value extrapolated at high frequency level off floor: mean of the floor RIN_ns  = 1.2282e-07/sqrt(Hz). While the incoming beam power P is measured as 0.275 mW we convert to obtain: ns = RIN_ns * P = 1.2282e-07*0.275e-3 =  3.3776e-11 W/sqrt(Hz), which agrees with the theoretical value ns_t = sqrt( 2hvP ) = sqrt(2*6.62e-34 [J*s]*(3e8[m/s]/633e-9[m])*0.275e-3[J/s])  = 1.31e-11 W/sqrt(Hz). 

Then I did the unit conversion for laser noises, from V/sqrt(Hz) to m/sqrt(Hz),

DataViewer PDBOUT = 2720 count  ~ ocilloscope = 1.50V with 1Mohm termination

Gmich = 1.5e11 count /m

(___ V/sqrt(Hz)) * 1/1.50V*2720 count /(1.5e11 count/m) = (___ )*1.2089e-08 m/sqrt(Hz)

nse_nla_nsh_scatter.png 

I took laser noises in frequency intervals for finer line-width resolutions. Also it should be noted that I observed 20kHz (and its harmonics) peaks now and then, and I have no clue where they are from -- I got rid of them one time by short-circuiting A of Channel2, terminating all B's with 50ohm caps, but it's not replicable..

  789   Wed Jun 11 00:40:50 2014 xiaoyueDailyProgressCrackleEBSD map for maraging steel blade sample

I did an Electron Backscatter Diffraction (EBSD, please refer to http://www.ebsd.com for more background information) analysis for the maraging steel (18% Ni Grade 250) blade sample.

Maraging_map_003um_061014.jpg

To generate this map, BCC and FCC structures were used as the model. It was found that BCC structure is the best fit. Then the high resolution map with pixel size of 0.03 um was taken.

We should use the pole figures as a color key to interpret the map -- it tells us how the grains are orientated in the sample. Since the sample is thin and flat we care only about the Z0 direction, which is the direction normal to the blade surface. For example the greens correspond to grains with zone axis (101) facing up along Z0.

Maraging_poleF_003um_061014.jpg

The next step is to choose a good grain with single slip geometry, within which to make a pillar with at least 1 um diameter for compression test.

  801   Thu Jul 3 01:09:38 2014 xiaoyueDailyProgressCrackleSS Blade Response

 Xiaoyue,Eric

The goal is to design a damping loop for the newly installed stainless steel blades.

We measured the transfer function for the blade response:

Jul1SSTFmag.png

Jul1SSTFphase.png

A comparison with the old maraging steel blades measurements can be found in Elog 757 and 587.

We modeled the transfer function and try to compensate the 2nd peak at ~10 Hz using a inverse function drive, however it did not work well. We doubted that the seismic noise is responsible for the problem, because seismic noise can excite the torsion / lateral tilting mode that cannot be damped / controlled by the coil drive, which only takes care of the vertical motion of the blades. To check the guess we took measurement of the coherence function of the seismic signal and the shadow sensor readout while common excitation is on for a sweep sine measurement:

Jul1SS_seismic_cohA.png

In the Coherence Function measurement Z, Y, X table data correspond to the seismic noises measured by the 3-axial Wilcoxon accelerometer. The Z post data was taken by the seismometer mounted directly on the post.

The high coherence ~10 Hz indicates that we might be able to improve the compensator design with better management of the cables -- now they are hanging and touching freely on the vacuum chamber wall -- i.e. routing the cables such that there is no direct path from chamber motion to motion of the second stack, and closing the lid.

  802   Wed Jul 9 17:19:15 2014 xiaoyueDailyProgressCrackleEBSD map for maraging steel blade sample

As the small grain size (max ~ 5 um) measured in the last maraging steel sample was skeptical for well annealed steel, I annealed the other non-annealed maraging steel sample under the condition described in E0900023-v12: 450 C for 100 hrs, however in air but not in inert gas atmosphere. I did another EBSD analysis for the self-annealed sample:

maraging0707.png

Although the data quality is not good mainly because of relatively poor polishing, we should still see that the grain size is actually really large (max ~ 50 um). It indicates that the last sample is not well annealed as it's claimed to be. I plan to double-check by taking a look at the grains of the original non-annealed sample. 

On the side I characterized the carbon steel sample:

whatever.002.jpg

It should be noted that the Z inverse pole figure (color key) stays the same for all measurements (green 101, blue 111, red 001).

Quote:

I did an Electron Backscatter Diffraction (EBSD, please refer to http://www.ebsd.com for more background information) analysis for the maraging steel (18% Ni Grade 250) blade sample.

Maraging_map_003um_061014.jpg

To generate this map, BCC and FCC structures were used as the model. It was found that BCC structure is the best fit. Then the high resolution map with pixel size of 0.03 um was taken.

We should use the pole figures as a color key to interpret the map -- it tells us how the grains are orientated in the sample. Since the sample is thin and flat we care only about the Z0 direction, which is the direction normal to the blade surface. For example the greens correspond to grains with zone axis (101) facing up along Z0.

Maraging_poleF_003um_061014.jpg

The next step is to choose a good grain with single slip geometry, within which to make a pillar with at least 1 um diameter for compression test.

 

  804   Sat Jul 12 01:11:41 2014 xiaoyueDailyProgressCrackle2D FFT for EBSD Map

I made a 2D FFT for the EBSD images using matlab

Quote:

 

 Seems like a rather qualitative analysis. Is there any way you can make a 2D FFT of this so that we can see what the distribution of grain sizes are? What are typical sorts of grain size analysis people do in order to get quantitative comparisons?

 where the left one is for the as-received maraging steel sample, right for the self-annealed sample.

maraging0610_norm.pngmaraging0707_norm.png

and a selected cross-section for direct comparison:

cross_sec_fft2.png 

Yet from the plots I can only infer that a more scattered plot is corresponding to a smaller "grain components". May need more study on typical analysis, and here is an EXAMPLE how other people did it.

  809   Wed Jul 16 12:48:09 2014 xiaoyueDailyProgressCrackleEBSD for SS340, Grain Size Distribution

I did EBSD for Stainless Steel 340:

SS1_reduct.tif

I also did a grain size analysis in their manager-data software, which directly gives us histogram of grain sizes. It should be noted that different statistics would lead to different results, but still it should allow us to compare the grain information quantitatively between different materials.

SS_Grain.png

 

  813   Thu Jul 24 12:35:36 2014 xiaoyueDailyProgressCrackleSingle Crystalline Cu micropillar

Will the tiny modulation bring about crackles is a long lasting question for me. Since driving at force above the pinning barriers is the fundament to generate dislocation slip, I am thinking about using some well-studied material and investigate if, and how the small amplitude perturbation can overcome the threshold with time. My idea is to

1. take some characteristic stress-strain curves -- for literature comparison, basically to make sure my data makes sense

2. load the pillar close to yielding; this may be difficult because the smaller the sample, the more stochastic nature its mechanical property would carry, mainly because of the large statistical fluctuation caused by small ensemble of dislocations.

3. hold the indenter tip at the load, but apply oscillations using a built-in method used for measuring contact stiffness, and look for slip events over time.

I luckily inherited a single crystalline Cu sample. I electropolished the surface using electrolyte consists of 110 mL 85% phosphoric acid, 40 mL nitric acid, 50 mL 99.7% acetic acid, with applied voltage = 1.5V, current = 100 mA, T(heat stage) = 60 C. When reacting the solution gave off terrible yellow smoke and the solution turned from transparent to green and then to blue.. It should be noted that the reacted solution has a much higher temperature than target temperature 60 C because the process is exothermic, but it somehow worked, giving me smooth shiny sample surface. I tried FIBing pillars yesterday but with the current voltage /current I got some pudding-shape pillars... I need to consult people about the right working condition for Cu.

  819   Mon Aug 4 18:00:17 2014 xiaoyueDailyProgressCracklemicro cantilever

I made a micro-cantiliver out of SS340, single grain. Relocating the EBSD grain map in SEM is tricky, I used a self-defined coords system. However I will put the sample back for a EBSD analysis after making full use of this area:

re-locate.jpg

The cantilever is of size 860 nm * 6 um which mimics the real blade in ratio width : length = 1:7.

cantilever.png

I also made a micro-pillar (1 : 2.5 um) in order to compare the nano-pillar compressions to the nano-cantilever bending to see how the signature is different when the strain gradients are present.

1.03um_004.tif

Tests will be conducted soon. 

  820   Tue Aug 5 10:41:05 2014 xiaoyueDailyProgressCrackleMichelson Locked

[Gabriele,Xiaoyue] 

From last Thu we are no longer able to lock the Michelson. Today we tried first with 0.5 gain instead of 1, and we added a resonant gain around 10 Hz, where the error signal showed a large peak. This gave us a reasonable lock. From there we measured and fit the plant transfer function and re-designed the compensators for the structures above 100 Hz. Applying the new compensator we were able to lock the Michelson fairly robust, increasing the gain to 2 (unity gain frequency at about 150 Hz)

plant_oltf3.jpg

It should be noted that if we keep increasing the gain to 3, there appeared to be some unstable structure around unity gain frequency. We also tried to increase gain at the other three resonance frequency peaks but it kicked the system to be unstable; we finally decided to use the simplest version of boost. An allipticlowpass filter at 600 Hz was also applied because high frequency resonances were getting unstable.

Now the lock acquisition procedure is:

1. Switch local damping on

2. Engage filters "lock", "notch1810"

3. Ramp up the gain to 0.5

4. Engage “Resgain”, “compensator1/2/3”

5. Ramp up the gain to 2

6. Engage the low pass and additional resonant gains at 30 Hz: “lowpass” and “boost”

We updated the autolocker accordingly. We started a data taking with common driving at 800 counts

  823   Thu Aug 7 12:13:32 2014 xiaoyueDailyProgressCrackleHunt ghost beam

Yesterday we vented the chamber and located a secondary beam on SYPD. We pushed it away from the sensing region by re-centering the main beams on both PDs. We were thinking of using iris to block the spurious beam but it clipped the reflection beam. Also we took care of two undamped beams that were dumped on the chamber wall. This helped a lot with the bump feature saw around 400-500 Hz.

After all of these tweaks the lock still worked! We increased the loop gain at ~11 Hz and ~16 Hz and this suppressed the resonance peaks seen in the error signal spectrum. When we were ready to do more adjustment the cables (taped on the wall) fell and we are going to fix it and pick everything up again today.

Quote:

Xiaoyue, Gabriele 

The data collected during the night were not very good, since the interferometer kept unlocking very often. We can't get any crackle data out of them.

We then tweaked a bit more the interferometer, remeasured and refitted the high frequency compensators. We also cleaned all mirrors. We noticed that the shape of the high frequency structure changed after the cleaning, but they did not change if we moved the beam centering on the end mirrors, by realigning the Michelson to a different point.

We see a strange noise spectrum, in the form of a bump up to 400-500 Hz. This reminds us of spurious interference with a ghost beam. It also seems (but we are not so sure) that the size of the bump changes with the alignment. So, we maybe have some ghost beam interfering with our main beam. We checked that the bump is the same in both AP and SP photodiodes.

We're leaving the autolocker on, to check the lock robustness over night. We also replugged all seismometers, since we suspect that lock losses might eb caused by seismic shocks.

 

  829   Wed Aug 13 13:08:25 2014 xiaoyueDailyProgressCracklemicromechanical analysis of yielding

I compressed a pillar (D = 1.1 um, H = 2.5 um, made out of SS304 single grain) using G200 nanoindenter.

SS304_P1.jpg

Using inflection point I got F_yield  = 0.3 mN, and knowing the pillar diameter to be 1 um we can estimate yield stress ~ 0.3 mN / pi (0.55 um)^2 = 316 MPa

However a more conventional definition for yielding point is the “0.2% offset” where people draw a line with slope of elasticity from 0.2% strain, and find the first cross over to be the yielding ~ 0.4 mN/ pi(0.55um)^2 = 421 MPa

pillar_comress.png 

I would also love to compare the pillar compression data with the indentation data.

Indentation_Plot.png

In order to extrapolate yield stress information I need to convert the load vs. depth data to stress vs. strain ones. It involves a better knowledge of the indenter tip, as so far I got contradicting result from projected area function calibration and the tip radius claimed in the spec sheet (max projected area exceeds the claimed tip area). Also I need to learn more about how to find the actual “contact height” which excludes the non-plastic indentation from the machine loading depth.

  830   Sat Aug 16 17:16:05 2014 xiaoyueDailyProgressCrackleCOMSOL simulation

I used COMSOL to compute the stress distribution on our blades. I set fixed constraint and a boundary load on the block clamp surfaces as demonstrated in the figure. In simulation I used extremely fine free tetrahedral mesh. The first principle stress contour is plotted. We see that we have large amount of blade areas that are bearing stress over 80% of the micro mechanical yielding (~250 MPa).

demo_bound_conditions.jpgstress_contour_fill.png

  832   Wed Aug 20 21:17:34 2014 xiaoyueDailyProgressCrackleIncreasing driving amplitude

We were able to collect seven hours of data using driving amplitude 800 counts last night, and the first analysis of the data in a narrow quiet band does not show statistical difference between driving on and off states. We plan to take data with larger driving amplitude. I compared the error signal power spectrums for several candidate amplitudes and there are no distinctive features showing up. Also the system have stayed locked for ~20 mins with common drive amplitude being 1600 counts, so I updated the code and we are going to collect data with double the driving amplitude tonight. 

Compare_aug20.png

  833   Fri Aug 22 09:50:01 2014 xiaoyueDailyProgressCrackleCrackle2 born

 [Gabriele, Xiaoyue] We checked all the mechanical parts received. Although there are some problems, Crackle2 are officially born!

 photo_1.JPG

 And I updated my wallpaper:

wallpaper.001.jpg

  836   Wed Sep 3 10:28:34 2014 xiaoyueDailyProgressCrackleSign of non-linear noise

Before we had trouble locking the Michelson for long periods of hours. We found that the problem lies in the autolocker script which checks if the IFo is locked with two conditions: RMS of error signal below threshold, and also RMS of correction below threshold. But the threshold of the correction was set to 1500, which is less than our driving amplitude. Therefore the autolocker at some point believes the IFO is unlocked because the correction gets too large. We increased the correction threshold to 5000 and then it stayed locked without problems.

Thus for the past week we have collected data with driving amplitude of 2400, 3000, 4000, and within certain quiet bands, all of them are showing a sign of non-linear noise being modulated at 2FQ, which agrees with our prediction for a crackling noise due to the fast driving change. We are going to take  more data with various of driving amplitudes and see if there's a correlation between the noise and the driving power. For an expedient comparison, I analyzed the three data set in band 548 - 570 Hz. 

 compare.png

We believe that the 1FI noise is due to misalignment. It should be noted that analyzing different bands, different hours, will give different results for the 2FQ modulation. Also we should question why the sign of the noise power could flip.

  840   Wed Sep 10 10:39:46 2014 xiaoyueDailyProgressCrackleCrackle1 driving amplitude calibration

I first double checked if we are driving in a linear range.

 

Blade A

Blade B

Drive Amplitude (counts)

Shadow Sensor Reading (counts)

Oscilloscope Reading of SS output (mV)

Shadow Sensor Reading (counts)

Oscilloscope Reading of SS output (mV)

0

-943

560

-1031

628

1000

-956

 

-1042

 

2000

-969

 

-1053

 

3000

-982

 

-1064

 

4000

-996

592

-1075

656

Then using Eric's calibration (Elog 601) I calculated 1.97 nm /count for blade A and 1.59 nm /count for blade B. It should be noted that we have set DRVA gain = -1, DRVB gain = -1.19, so the effective driving is (0, 1, 2, 3, 4)k * 1.19 counts.

  841   Wed Sep 10 11:34:20 2014 xiaoyueDailyProgressCrackleSS304 cantilever bending

 I did a micro-cantilever bending + oscillation test using pico-indenter, with displacement control.

dl_vs_t.png

To analyze the data I subtracted elastic fit l(t) = k*d(t) + \beta * t, where t is time, l is the load, k is the stiffness, \beta is the drifting coefficient.

fitl_dt.png

The fitted stiffness is 35 N/m, and I calculated the Young's modulus E according to the cantilever geometry (triangular cross section with base b = 2 um, height h = 1.5 um; length of beam l = 14 um ) using equation k = 3*E*I /l^3, where I is the area moment of inertia of the beam = b*h^3 /36). The result E  = 170 GPa falls into reasonable expectation for material SS304.

In order to seek for the plastic kicks, I look at the histogram of the noise residuals:

 hist_residual.png

The analysis didn't show any sign of crackles but there are several things to improve for the test: since this is only a test run and the load is too far from yielding, I am going to increase static load and do more tests, but before that I have to figure out why every time after unloading the tip always crushed into my sample and destroyed my cantilever... Also it's still not clear to me why the drifting effect only shows up in loading curve but not in displacement data. 

On the macroscopic side, I am taking more data with the SS304 blades but with faster driving at 0.5 Hz. More analysis is coming soon.

  848   Wed Sep 17 10:48:01 2014 xiaoyueSummaryCrackleA brief review and recent progress on micro-mechanical model

 I will update some comments later..

  852   Sun Oct 19 23:04:30 2014 xiaoyueDailyProgressCrackleImproved seismic isolation stacks into the chamber

I put up the stacks with a similar configuration as Gabriele built in Elog 851. To integrate them into the chamber, I added another square bread board to the bottom for mounting the posts for horizontal support. it’s now a rubber >> square board >> spring >> round board >> rubber >> round board >> optical setup (from bottom to top) layers of cake. 

stacks.png

However unfortunately, the setup overshoots the chamber celling, but I think we can simply cut the post by 3’’ to solve the problem. 

  857   Mon Nov 17 12:52:07 2014 xiaoyueDailyProgressCrackleCrackle1 reborn

[Eric, Xiaoyue]

The central post has been trimmed by 3 inch. We rebuilt the crackle1 optical layout with the PZT mirror replacing one of the 45 degree mirrors out side of the chamber and got fringes. 

photo_2-1.JPG

I also finished soldering the ribbon cable with shadow sensors and photodiodes connectors on board for the improved re-cabling use.

photo_1-2.JPG 

 We are going to put everything into the chamber this week.

  858   Thu Nov 20 23:03:10 2014 xiaoyueDailyProgressCrackleInto the chamber

[Eric, Xiaoyue]

We reorganized cables with the improved cabling configuration and put everything into the chamber. 

Before taking further steps, I am going to mount an accelerometer onto the post with the chamber pumped, to characterize the seismic noise in this current setup.

crackle1_chamber.JPG

  880   Mon Jan 12 10:38:16 2015 xiaoyueDailyProgressCrackleCrackle2 optical layout

During the holidays I built the optical layout referring to our former design, but soon we found that because of the shorter actual blade length, the input beam has to be sent through the clearance between the cage edge and the suspended block, which will lead possible clipping problem. 

Gabriele redesigned the layout accordingly (Elog 874) . Last week we re-built the layout and suspended the blocks using our custom-built positioner, which consists of x, y, z tranlational stages and a rotating plate. We hope to align the magnets to the OSEMs with desired precision, but everytime the suspension wire is being clamped, the block will rotate away from the aligned angle. This might be a result of the screwed clamping surface.

  905   Mon Feb 2 18:38:18 2015 xiaoyueDailyProgressCrackleOSEMs working

[Gabriele, Xiaoyue] For the past week we finished soldering all the OSEM electronic components and cables, and now the 6 OSEMs for damping the two blades are all mounted.

Today we built the model X1KR2 for sensing and controlling,

and the interfacing medm screen kr2_damping,

We fitted the whitening TFs for all 12 circuit boards and uploaded the dewhitening filters. Finally we checked the OSEMs and they are all working properly now after some debuggingsmiley  Our next step is to do calibration, and start to design the damping loop.

  910   Wed Feb 4 14:44:17 2015 xiaoyueDailyProgressCrackleOSEMs damped

[Xiaoyue, Gabriele]

We calibrated the shadow sensor /coil driver from counts to volt and then to micron / N according to Enrico’s calibration [Enrico's report (actually mN would be a better scale but we didn’t realize it later), and uploaded the conversion to the sensing filter to OSEMX/Y/Z and COILX/Y/Z. Then we designed the damping loop, with DAMP being a differentiator to reduce the resonance Q, ELP100 being the low pass 100 Hz to suppress high frequency noise, and int being the integrator to enhance the control in low frequency range. We compared the damped spectrum with the free-running spectrum and the control works very well.

However we found that the amplitude y-direction low frequency oscillation is higher than x, and it’s corresponding to the 100mHz peak appearing in the spectrum. The problem was solved by stiffening the board using two posts to clamp down the board in y direction, but the trade-off is an increase of some high frequency noises. Unfortunately we didn’t save the  comparison right away. We have to do more investigation on if the stiffening helps with this low frequency oscillation. 

We also tried to measure the transfer function but the coherence was not good enough with a quick noise injection and then a quick try of sine sweep.

Then we realigned the interferometer and see the fringes on both photodiodes. The slow oscillating fringes indicates that the layout is now very robust with its arm difference confined in few wavelengths.

We made the servo model by creating the LOCK mask and added the lock input/filter to the blade block. 

The next step is to lock the interferometer. 

  929   Mon Feb 23 12:00:29 2015 xiaoyueSummaryCracklecrackle model up to date

I have been using single crystalline copper nanopillar to study metal’s pre-yield behavior. I chose copper because it’s FCC (face-centered-cubic) crystal, that has well-defined slip systems so the dislocation dynamics is relatively simple to understand. Pillars with diameter D = 500 nm, aspect ratio D: L = 1:3 are FIB (focused ion beam) milled. We want small system to further simplify our first investigation as the “mechanical annealing” theory and experiment have shown that in small enough system  there’s no dislocation multiplication but only slipping. I first did compression tests to measured the yield stress to be ~ 400 MPa and young’s modulus to be 143 GPa. They are very close to what’s been reported in literatures [1, 2].

 

Then I carried out nanoDMA tests in which, I prescribed an increasing “elastic” load in steps with a small sinusoidal oscillation on top. With the displacement data being taken, I first plotted the histogram of the displacement residual rate. Some outliers are observed as possible crackle events.

I also analyzed the storage and loss modulus, defined as the real and imaginary parts of the input-output transfer function. Interestingly there seems to be non-zero peaks as the static load is approaching the yielding!

To deal with the data more carefully, I thought over the calibration for machine noise. Since the resonance frequency of pillar is in MHz range and the machine resonance is ~ 130 Hz. We have the simple relationship that k_calibr = k_raw - k_air. 

I took several more data and chose the “good” ones based on the criteria that their storage stiffness saturates to near E ~ 19uN/nm. After taking absolute values of the loss moduli and plotting modulus vs. stress we see features at ~ 180 and ~ 320 MPa load.

My interpolation of the “good” data shows promising consistency.

However for now I don’t quite understand why I see negative loss moduli in my analysis. I keep doubting my error analysis but so far I don’t see how it could be wrong: I fit the data with sinusoids and use the fitted paramaters and their associated standard errors to do monte-carlo simulation. The means also agree with my FFT analysis. I am attaching the sample matlab codes for both my monte-carlo and FFT. 

 

Attachment 9: monte_carlo_B8P6.m
%% load data
filename = 'B8P6_QS.txt';
data = importdata(filename);s
fs = 1000;
tag = 'B8P6';
% work for P4-7

time = data.data(:,1);
disp  = data.data(:,2);
load = data.data(:,3);
... 133 more lines ...
Attachment 10: monte_carlo.m
function [kr0, ki0, ekr, eki, kr_raw, ki_raw] = monte_carlo(A0, B0, dA, dB, yr0, yi0, dyi, dyr)

A = A0 + randn(1e6, 1).*dA;
B = B0 + randn(1e6, 1).*dB;
yr = yr0 + randn(1e6, 1).*dyr;
yi = yi0 + randn(1e6, 1).*dyi;

H = (A+1i*B)./(yr+1i*yi);
kr_raw = real(H);
ki_raw = imag(H);
... 4 more lines ...
Attachment 11: B3P8_1Hz_FFT.m
%% load data
filename = 'CuB3P8_QS.txt';
data = importdata(filename);
fs = 1000;

time = data.data(:,1);
disp  = data.data(:,2);
load = data.data(:,3);
load0 = data.data(:,4);
dispV = data.data(:,5);
... 70 more lines ...
  932   Thu Feb 26 01:49:28 2015 xiaoyueDailyProgressCracklecrackle2 realigned

Today I realigned crackle2 optical layout, obtaining fringes with better contrast defect (AP ~98%, SP ~94%); however I have to check it the contrast defect is due to unbalanced BS reflectivity, as the reflectivity and transmission depends on the polarization, and can be as low as 40% according to Thorlabs specsheet for BSW17. Note also that our AOI is not 45 degree.

  

The damping loop is working ok without modification, but from time to time, the z2, y2 corrections are large. I doubted it's due to y2 magnet touching in z direction but I have to check if all filter banks are engaged. Neverthless the position readouts all have small variance.

  

I started soldering the breakout boards for suspending the optical setup. We are having problem with not enough spacing between the 2-pin connectors. I am going to live with it by soldering one another connectors to the back for now. Meanwhile we are going to redesign the PCB board.

Attachment 1: fringes_150225.png
fringes_150225.png
Attachment 2: BSW11_Transmission_780.jpg
BSW11_Transmission_780.jpg
Attachment 3: damping_block1.png
damping_block1.png
Attachment 4: damping_block2.png
damping_block2.png
  933   Wed Mar 4 03:29:30 2015 xiaoyueDailyProgressCrackleBreakout cables

I soldered the breakout PCBs according to Gabriele's design. In addition to cabling all the 2-pin connectors, I cut the translational stage cable and resolder it to the 4-pin connector.

  

I mounted the board to the back of the optical breadboard and connected it to another board on the table using the 26-pin ribbon cable. I tested that the translational stage is working properly. The damping loop are also working, except that Z1 is reading ~1700 (max) no matter how I change the position of OSEM. The LED's are all on so there must be something wrong with Z1 PD. I still need to cut and solder the power supply for the AP/SP PDs.

  

 

  934   Thu Mar 5 21:43:30 2015 xiaoyueDailyProgressCrackleDMA test on metallic glass

In Elog 929 when I was trying to relate the loss modulus G'' to the energy dissipation I naively took the absolute value, which turns out to be wrong. By directly integrate the stress over strain the energy dissipation is proportional to sin(loss angle), which should be 0 < loss angle < pi/2 for the system to be physical, otherwise my pillar is doing work to the machine.. which makes no sense. Thus the negative G'' indicates some hidden machine anomaly. Following Rana's suggestion we want to do similar test on "loss-less" materials. 

At first I was thinking about using fused silica, but it turned out to be difficult to image and FIB fused silica pillars since the material is non-conductive. I will suffer from charing problem if I don't deposite Au or Cr. Before moving on to KNI I borrowed a sputtered metallic glass (MG) substrate, (composition around 50 at. % Zr, 30 at. % Ni, 20 at. % Al) from my labmate since it's also amorphous and known to have low internal damping. I FIBed 500 nm pillars out of the MG sample and did 1Hz DMA tests as I did on the Cu pillars:

Compared to those of the Cu pillars the values of MG G'' are very small, though not always zero. Also the two-peak trend and the presence of negative values should call further cautions to the machine reliability. We need to understand if we can use the MG data to "calibrate" Cu DMA measurements. 

Attachment 1: MG_B1_2.002.jpg
MG_B1_2.002.jpg
Attachment 2: MGB1P4_B2P1359.png
MGB1P4_B2P1359.png
Attachment 3: MGB1P4_B2P1359.png
MGB1P4_B2P1359.png
Attachment 5: MG_B1_2.002.jpg
MG_B1_2.002.jpg
  935   Mon Mar 9 01:21:06 2015 xiaoyueDailyProgressCrackle1Z PD fixed

I fixed the Z1 PD circuit — it was simply due a poor connection. I also soldered the AP, SP connections. Now everything works out pretty well with the breakout cables.

  

  940   Tue Mar 24 17:07:22 2015 xiaoyueDailyProgressCrackleElectronics good to go

I fixed the connections broken during the rearrangment, and tested the electronics. At first the PD's were all working properly but DAC channels were outputing 0 or 2V voltage no matter what commands were sent, but this problem was solved after rebooting CyMAC. Everything is now in place, working as normal. The boards are good to go for suspension!

  944   Fri Mar 27 23:26:56 2015 xiaoyueDailyProgressCrackleFlat cable in place

[Gabriele, Xiaoyue]

In order to prevent the cable short circuiting seismic noise, we clamped the flat cable to the second isolation stage and then to the first stage. I glued threads to serve as bolt holes for clamping. Rubber is underneath to give necessary support. The cable was loosely rolled in U shape in each stage to reduce the stiffness. 

    

Attachment 2: IMG_3064.jpg
IMG_3064.jpg
  948   Tue Mar 31 10:58:27 2015 xiaoyueDailyProgressCrackleFused quartz nanopillars fabrication

Using KNI Kurt Lesker Ebeam evaporator, Ottman (grad student in Julia group) helped me deposit 40 nm thickness film of gold to the fused quartz sample [specsheet HERE, GO-FS100-1 family, gift from Matt]. I FIBed two batches of ~ 500 nm diameter nanopillars. 

Since evaporated gold does not have great adhesion so it's not on top of the pillar after FIBing -- you can see it gradually go away, but there's enough on the whole sample for it to be still effectively conductive. 

The next step is to do same DMA test on FQ pillars as the ones done on Cu pillars to calibrate for machine nonlinearity. 

  950   Wed Apr 1 12:29:18 2015 xiaoyueDailyProgressCrackleDMA test on fused quartz

I did several 1Hz oscillation DMA tests on fused quartz using the same loading function used for Cu pillars. Notice there are four branches of storage moduli and the lower end correponds to a consistent, non-crazy loss modulus sweep. This calls a review on the MG data (Elog 934), where the storage modulus seem to exibit two-branch trend, and the higher-end corresponds to the loss modulus curves with larger fluctuation and presence of negative values. I have no clue where this branching in storage modulus comes from and how it's related to the loss modulus insanity, but undertanding it might be crucial to resolve the nagative phase lag issue. 

If zoom in only those good ones, we see: 1. the loss modulus are all postive, although definitely not zero.. 2. there's no obvious two-peak feature showing up; however I want to take more data in 100 ~ 500 MPa range to confirm, as the red and black "peaks" are suspicious.

 

  954   Sun Apr 5 23:52:49 2015 xiaoyueDailyProgressCrackleBreakout cables 2

I finished connecting all OSEM input /output cables to the breakout cable. 

Attachment 1: IMG_3323.jpg
IMG_3323.jpg
  955   Mon Apr 6 00:49:03 2015 xiaoyueSummaryCrackleAggregate Cu, MG, FQ DMA measurements

Aggregating all good measurements for MG, FQ based on the criteria for "good" discussed in Elog 950, we plot to have visual comparison for dynamic modulus amongst the three samples tested so far. From a rough look, the loss modulus of amorphous materials (FQ, MG) are more "linear" than that of Cu in static stress sweep. Also notice that there seems to be a consistent increase in storage modulus in all samples. From the basic linear fitting, amplitude measurements of FQ, MG, Cu_saturate all have a slope of ~ 0.15 e3. If we understand why "good" is good, and taking enough measurements on FQ to confirm the consistency, we might be able to use it as a calibration line.

 

  964   Thu Apr 16 00:06:24 2015 xiaoyueDailyProgressCrackleMichelson aligned on suspended board

I estimated the optical path length based on Yuji's work (LIGO_T1400600-v1), and set up the input steering mirrors accordingly as shown in the figure. The beam needs to travel 3.11 meter to achieve ~ 300 um beam waist at the end mirrors. Unfortunately I under-calculated the length on breadboard at the begining (and didn't notice it until just now..), so the path is now 10 cm longer than expected.

After hours of painful alignement on this 12 d.o.f. flying board, knocking off only one magnet (A), I finally overlapped the splitting beams and saw fringes on both photodiodes, with everything dampped except for A, the breadboard horizontal in plane direction. 

  965   Mon Apr 20 16:28:02 2015 xiaoyueDailyProgressCrackleMichelson poorly locked

I was having a hard time last week locking the Michelson. Yesterday with remote help from Gabriele, we noticed that the first problem is the wrong sign. By inputing a 1000 count offset at the Coil input the two blades were driven in opposite direction, so the servo is loop actually not doing anything to cancel the displacement difference. We resolved the issue by changing the sign of Z2 coil output, while also changing the sign of Z2 ctrl output to compensate for the sign of the entire damping loop. Gabriele also pointed out the second problem where the lock filter had double pole at 10 Hz that limited the bandwidth, eating up too much phase, so we moved the double pole up to 100 Hz. Then we were probably sensing too much correction due to the high frequency noise, so we moved it down to 50 Hz, and the Michelson is locked with gain of 10. 

    

I further pushed the zero down to 40Hz (red curve) because it seems to give more stable lock (50Hz green).

Since the Michelson is very sensitive to environmental noise i.e. seismic, acoustic, air current noise, I left it locked overnight for a better investigation of the error signal spectrum.

  966   Thu Apr 23 02:32:11 2015 xiaoyueDailyProgressCrackleFirst suspended Michelson error signal spectra

Late Elog but on Monday (Elog 965) I left around 6:30pm and the Michelson lost lock in 1.5 hr).

I plotted three spectrum 6:30 to 7:00pm (purple) 6:30 to 7:30pm (blue) and 12:00 to 12:30am (green), calibrated with the non-linear inversion of the Michelson fringe: 

  968   Mon May 4 18:31:38 2015 xiaoyueDailyProgressCrackleBoard sagging down

[Gabriele, Xiaoyue]

The fringes were becoming faster and the former lock scheme no longer works. Using the good part of single blade transfer function I measured, we fitted for the differential blade model, in unit of um/uN. Optical gain is calculated to be 5.61 mW/um using peak to peak SERV_IN1 reading of 0.95 mW. Using plant model -- multiplication of differential blade model and optical gain, in unit of mW/uN -- we designed in sisotool the new compensator. However using this lock filter we are still not able to lock the interferometer.

We start to question if there's anything really wrong with the alignment. By comparing the shadow sensor signals (blue curve damped, red undamped) with the locked ones (green) we clearly see a difference. Then we noticed the board is badly sagged probably due to creep, and is definitely touching down. We are going to re-adjust the OSEM positions to try solving this problem.