Message ID: 1839     Entry time: Mon Dec 9 22:06:19 2019
Author: Duo 
Type: DailyProgress 
Category:  
Subject: Two boards tested 

I finished testing S1900294 and S1900297 and plan to ship them to the sites.

I got stuck at a couple things. I think it is good to make a note of these stuff.

1. Diaggui gives "unable to start excitation".

Solution: restart diaggui

2. Drive signal does not go in

Solution: check the status of the digital system - it crashed and needs restart.

3. FASTIMON gives too much noise that the transfer function looks like junk.

Solution: I use 50ohm resistors to replace the coils and 500 counts noise to measure the transfer function. If the resistance is large and the input signal is small, the current will be too small.