ID |
Date |
Author |
Type |
Category |
Subject |
2050
|
Wed Jun 22 13:42:38 2016 |
awade | Misc | | New silicon sample (8616P Tang) arrived: initial unbox and check |
The new Si sample (8616P Tang) has arrived from Stanford. We will do scatter measurements at 1550 nm (and maybe 2 um). Its dimensions are 10x60x96 mm.
The two 10x96 mm surfaces have been high quality parallel polished (this was used for an absorption measurement) and one of the 60x96 mm surfaces is a moderate polish from which we will measure scatter. The sample is one half of a larger bar from which absorption measurements were made. The non-shiny side of the sample is labeled accordingly. Ashot Markosyan has sent us a bulk absorption uniformity scan across the sample, measuring the ppm/cm absorption stepping out from the center of the original bar. Attached is the absorption measured at 1550 nm, one side of our sample (labeled center) corresponds to position = 0 mm. Please note that for these 1550 nm measurements of absorption, the plotted alpha values need to be divided by a factor of 20. This is supposed to be their worst sample
A measurement of the sample at 2 um found absorption to be 2.5 ppm/cm.
I unboxed it from its shipping packaging to inspect the element. Some pictures are attached. It can be very hard to see some of the features as all I have is an old Kodak point and shot and Si is very shiny. I turned off the lights and used the Fiber coupled quartz lamp in the PSL flow cupboard to illuminate the sample a little better. Also attached are my hand sketched illustrations of the markings. There is almost definitely a residue of a finger print on the inner edge (towards the 'middle') of the sample on one of the highly polished sides: you can just make this marking out in the photo, it is clear under room lights.
The edges of the sample have been beveled but there is there are very small chips (<0.25 mm) out from most edges (couldn't get the camera to focus on these). Si is very brittle, so this is not something unexpected and won't affect scatter measurements.
There appear to be minor scratches on the large 60x96 mm surface (these are more obvious under the quartz lamp). There also might be some chemical splotch residue and dust around the edges. But I think a drag and wipe should remedy this. Otherwise I couldn't immediately identify any other potential defects.
Highly polished front face (1 of 2)
Unpolished side
Above: moderately polished face
Unpolished back face.
See center bottom, mark
Some small chips on shaper edges
Under quartz spotlight #1
Under quartz spotlight #2
Under quartz spotlight #3
This might be a finger print ~17 mm in from edge (middle side)
Sketch scanned from log book with hard to see features
|
Attachment 1: image002.png
|
|
Attachment 2: LabbookSketchSiSampleFeatures.pdf
|
|
2049
|
Mon Mar 7 15:47:03 2016 |
KojiN | Misc | PD QE | Back reflection measurement with the chopper |
To reduce the scattering, we did some try and error.
The setup was same as the elog2048.
First, to reduce the scattering from the Iris 1, in front of the Iris 1 the aluminum foil which has the hole for the laser was placed.
We made several foils, different hole sizes and different shapes.
We cannot found the reasonable tendency, but with the best foil the measured scatterd value became 350 uVrms (BRDF = 7.1 * 10^(-4) [1/sr]).
This is the comparable value to the BRDF measured with PBS and QWP (elog2047).
After that, we changed the steering mirror #2 from Newfocus 5104 to Newport 10Q20HE.1.
Then, without aluminum foil, i.e. in completely same setup as elog2048, the measured scatterd value became 11 mVrms.
And we tried to place the best aluminum foil and the measured valued as reduced to 820 uVrms. |
2048
|
Mon Mar 7 15:34:46 2016 |
KojiN | Misc | PD QE | Back reflection measurement with the chopper |
Background:
We are still suffering from excess amount of back reflection compared with the predicted number from the BRDF measurement. Assuming this light is not coming from the target PD, we want to reduce the scattering from other optics. As an attempt, we decided to use a BS instead of the combination of a cube PBS and a QWP. This way we can reduce the number of the optics involved, particularly the PBS which may cause more scattering than others.
What we did:
The PBS and the QWP were removed, and then a 50:50 BS was inserted as shown in Fig. 1.
The distance between the target PD and Iris2 was measured to be 38.1 cm. The diameter of Iris2 was set to be 7 mm.
As a part of the calibration, the incident power on the target PD was measured using PDA100A (Gain 0dB). The measured value was 3.973 V with no chopper blocking the beam.
An HR mirror was then placed before the target PD to measure the reflected light power with the chopper still halted and the iris 1 is oped using PDA100A (Gain 0dB) which is placed at the position as shown in Fig. 1. It was measured to be 1.253 V. This means that the reflectivity of the BS is about 0.315. This reduction is considered to be the clipping at Iris1 as the returning beam has the bigger beam size (i.e. the iris worked as a sort of mode cleaner).
(==> There might be a misunderstanding because of my not clear explanation. So, I edited this paragraph again. KN)
Then, the chopper was started at 253 Hz. The reflected light from the HR mirror was measured using a FFT analyzer to be 642 mVrms (PDA100A gain 0 dB).
(==> mVrms? KA ==> mVrms. I fixed. KN)
Throughout the measurement, the opening of Iris1 was adjusted such that the clip level was less than 0.1% of the incident power (see also elog2047).This yielded the Iris1 diameter of 1.8 mm.
Finally, the back reflection was measured in the same way as elog 2040 , and also with Iris1 opend as much as possible.
Result:
The measured voltage with Iris1 with the diameter of 1.8 mm: 720 uVrms (PDA100A gain 70 dB)
==> This corresponds to the BRDF of 1.4 x 10-3 [1/sr].
The measured voltage with Iris1 completely open: 640 uVrms (PDA100A gain 70 dB)
==> This corresponds to the BRDF of 1.2 x 10-3 [1/sr].
These values are much (x40) larger than the BRDF value expected from the BRDF measurement (5x10-5 [1/sr], see also elog2045).
In fact, these are much (x2) larger than the ones with the PBS & the QWP (8x10-4 [1/sr], see also elog 2047).
(It is not so clear how these numbers were calculated. How these calibrations were used? How did you incorporate the BS transmissivity and reflectivity??? KA)
==> For calculating the BRDF, the fomular as shown elog2034 is used. Seeing the fomular, we need the ratio of the incident light power and the scattered (reflected, in this measurement) light power, the incident angle, and the solid angle. The ratio of the incident light power and the scattered light power can be obtained by taking the ratio of the measured voltage with the chopper as a part of the calibration, Vin (642 mVrms), and the measured voltage with Iris1 with the diameter of 1.8 mm (720 uVrms) or with Iris1 completely open (640 uVrms), Vsca, considering the difference of the PDA100A gain. In this calculation the BS reflectivity and transmissivity do not appear, see also Fig. 2. In addition, using the incident angle, 15 deg, and the solid angle from the PD(C30665GH) to the Iris2, 2.6x10-4 sr, the BRDF can be calculated.
And when the Iris was closed, the value was larger. This means that the Iris may make a scattering.
(==> This didn't make sense compared with the numbers above. KA ==> The numbers were written conversely. I fixed. KN)
Fig. 1 Setup for measuring the back reflection using the BS.
Fig. 2 BRDF calculation with chopper and the BS. The left side shows that the setup of the measurement of the back reflection light. The right side shows that the setup of the measurement of the incident light. The bottom fomular shows how to obtain the ratio between the incident light power and the scatterd light power from the voltage measured with the FFT analyzer.
[Ed.: KA] |
Attachment 1: setup.pdf
|
|
Attachment 2: BRDF_cal.pdf
|
|
2047
|
Sun Mar 6 10:19:14 2016 |
KojiN | Misc | PD QE | Estimation of the back scattering |
(This is the work on Thursday.)
Background:
We observed large discrepancy of the back reflection between the values measured with a BS and estimated from the BRDF. The back-scattering from the PD (C30665GH) without the reflector measured in elog2040 (2*10^(-3) [1/sr]) was about 40 times larger than the value expected from the BRDF measurement (elog2045, 5*10^(-5) [1/sr]). We thought this came from the ampbient light, scattering from other optics (e.g. PBS), and scattering from the chopper. We tried to measure the back reflection again with a refined setup.
What we did:
As the first attempt, for dumping the scattering light from the chopper, a "wall" to dump the light with the aluminum foil was mede and was placed between the lens and PBS as shown in Fig. 1. Before the wall was placed, the back reflection was about 162.6 mVrms (PDA100A gain 70dB). After the wall was placed, the back reflection was improved to about 82.5 mVrms (PDA100A gain 70dB).
Secondly, the chopper was tilted to reduce the direct reflection. Then the back reflection was reduced to about 3.48 mVrms (PDA100A gain 70dB).
Thirdly, we found that the incident beam was clipped at the incident Iris and the level was 0.4% in terms of the transmitted light power. (The Iris had been opened 1.2 mm in diameter.) The clip was small but could not be negligible. Thus we opened the Iris to 1.8 mm in diameter and the clip level was less than 0.1%, which is the detection limit of the degital multimater used to measure the output signal of the PD (C30665GH).
In order to obtain a reference calibration, we placed an HR mirror right before the target PD and measured the reflected power on the measurement PD (PDA100A).
All the numbers were measured with an FFT analyzer to read the peak value at the chopping frequency.
After that, we measured the back reflection with the same way as elog2040.
Measurement:
We measured the back scattering in following conditions:
1. d_ref = 5 cm, theta_ref = 0.8 deg, without the reflector,
2. d_ref = 5 cm, theta_ref = 0.8 deg, with the reflector,
3. d_ref = 5 cm, theta_ref = 1.7 deg, without the reflector,
4. d_ref = 5 cm, theta_ref = 1.7 deg, with the reflector,
5. d_ref = 2 cm, theta_ref = 4.3 deg, without the reflector,
6. d_ref = 2 cm, theta_ref = 4.3 deg, with the reflector,
where the paremeters are explanes in Fig. 2. In conditions 3--6, using the CCD camera, the primary incident beam on the PD aligned was on 1 mm far from the boundary of the PD, i.e. the beam was aligned at 0.5 mm off from the center of the PD as shown in Fig. 2. (I'm sorry but in the condition 1 and 2 the beam position was unknown.) The theta_ref is determined as widely as we can keeping the output signal of the PD (C30665GH) maximum.
For calculating the BRDF. the fomular in elog2034 is used.
(Please note that r_l becoms the radius of the Iris in front of the PDA100A and d becomes the distance between PD (C30665GH) and the Iris)
The Iris diameter was 1.143 cm, the distance between PD (C30665GH) and the Iris, d, was 43.1 cm.
Result:
The measured value with HR mirror in front of the PD (C30665GH) was 2.400 Vrms (PDA100A gain was 0 dB.)
The voltages measured in the conditions were as follows (PDA100A gain was 70 dB).
1: 3.192 +/- 0.007mVrms
2: 27.3 +/- 0.01 mVrms
3: 3.347 +/- 0.004 mVrms
4: 3.786 +/- 0.008 mVrms
5: 3.881 +/- 0.009 mVrms
6: 3.846 +/- 0.007 mVrms.
The BRDF results are shown in Figs. 3 and 4.
The error estimation was not yet be done.
Discussion:
* Primary incident beam effect
When we see the values measured in the condtions 1, 3, and 5, the primary incident beam effect is about 8*10^(-4) [1/sr] which is still larger about 10 times larger than the expected value from the BRDF measurement, 5*10^(-5) [1/sr].
This means that, for observing the back reflection of the primary incident beam, we must reduce the noise, i.e. the scattering which does not come from the PD.
* Secondary reflected beam effect
When we see the the gaps of the value between condition 1 and 2, between 3 and 4, and between 5 and 6, when theta_ref is made larger, the secondary reflection beam effect was reduced.
The difference between condition 3 and 4 is 1.1*10^(-4) [1/sr] and the expected secondary reflection light effect from the BRDF measurement was 0.8*10^(-4) [1/sr].
They look consistent.
The difference between condition 5 and 6 can not be seen in current sensitivity and the expected secondary reflection light effect from the BRDF measurement was 1*10^(-5) [1/sr]
Thus it is reasonable that there is no difference between 5 and 6.
The difference between 3 and 5 is considered to come from the laser power drift.
Therefore, the secondary reflected beam effecte can be explained with the BRDF measurement.
Fig. 1 Setup for measuring the back reflection.
Fig. 2 Geometry of the PD and the reflector
Fig. 3 Measured back scattering in terms of the BRDF in conditions 1--6.
Fig. 4 Measured back scattering in terms of the BRDF in conditions 3--6
|
Attachment 1: setup.pdf
|
|
Attachment 2: ref_ps.pdf
|
|
Attachment 3: Backscat.pdf
|
|
Attachment 4: BackScat2.pdf
|
|
2046
|
Sun Mar 6 02:25:47 2016 |
Koji | Misc | General | PD glass reflections |
OMC_Lab/252 |
2045
|
Sat Mar 5 01:42:02 2016 |
KojiN | Misc | PD QE | Measurement of the BRDF with the Iris in the incident path |
The BRDF with Iris was measured with the chopper and FFT analyzer again because in the previous measuremt it is suspected that there may be a beam clip.
This time, the BRDF was measured after is is confirmed that the are no clipping using a IR view.
The result is shown in Fig. 1.
The error estimation has not yet be done.
The two data shown in Fig. 1 are consistent in the skirt ares by are not consistent arond the peak.
Thus, we checked the clipping effect by making a clip intentionally and measuring the BRDF at several point.
When there is a clip, the previous BRDF is reproduced.
There, we conclude that there was a clip in the previous measurement.
Fig. 1 BRDF with the Iris.
|
Attachment 1: BRDF_with_chopper.pdf
|
|
2044
|
Fri Mar 4 16:18:45 2016 |
Koji | Misc | General | Photos of the experiment |
https://goo.gl/photos/FaqFC7tuxz76m2236 |
2043
|
Wed Mar 2 14:38:31 2016 |
KojiN | Misc | PD QE | Measurement of the reflectivity, EQE, and IQE of the PD without glass window |
Using the improved reflectivity measurement, the IQE is calculated again (see also elog2031).
The QEs are calcurated with the same way as elog2020 and elog2022 and the results are shown in Figs. 1 and 2.
Figure 1 shows that, for p-pol, the not-enhanced EQE and the IQE are 0.90 +/- 0.03 and 0.94 +/- 0.04 at 15 deg, respectively.
Figure 2 shows that, for p-pol, the EQE is enhanced 4.0 +/- 0.4% and, between 30 deg and 50 deg, the EQE is enhanced almost up to the IQE.
Fig. 1 Measured values of the QEs for p-pol and s-pol.
Fig. 2 Improvement ratio of the QEs for p-pol and s-pol.
|
Attachment 1: QE3.pdf
|
|
Attachment 2: QE_ratio2.pdf
|
|
2042
|
Wed Mar 2 13:07:36 2016 |
KojiN | Misc | PD QE | Distance between the 1st incident beam and the 2nd reflected beam |
Using the setup for the measurement of the back scattering (in elog2040), the relative distance between the 1st incident beam and the 2nd reflected beam was measured.
First, the 1st incident beam position was detemined by aligning the reflector and overlapping the 2nd reflected beam on the 1st incidnet beam with the monitor of the value of the multimeter ("worst angle").
After that, the reflector was aligned onto the point that the back scattering is smallest and the relative angle of the knob of the reflector mount was noted ("best angle").
Using the value of the relative angle and some geometrical features of the mount, the relative reflector angle between "worst angle" and "best angle" was determined 0.4 degree, i.e. the incident angle to the reflector was 0.4 deg.
From this value, relative distance between the 1st incident beam and the 2nd reflected beam on the PD was determined 0.8 mm as shown in Fig. 1.
And the 1st incident beam and the 2nd reflected beam on the Iris was determined as 1.6 mm.
This means the main reflected beam was dumped by the Iris. (The Iris apature radius is 0.65 mm.)
The distance on the PD is designed as 1 mm.
The design value and the measured value is consistent.
For obtainning the design value, the 1st incident beam should be miscentered a bit more.
Fig. 1 Beam positions on the PD.
|
Attachment 1: Beam_Position.pdf
|
|
2041
|
Wed Mar 2 10:17:56 2016 |
KojiN | Misc | PD QE | Measurement of the BRDF with the Iris in the incident path |
With the chopper, the BRDF at 15 deg incident angle with the Iris on the laser path was measured again.
This is because the BRDF with the Iris on the laser path was limited by the sensitivity even at the scattering angle very close to the peak (e.g. 35 deg).
The setup up is shown in Fig. 1.
The diameter of the Iris2 was 7 mm at -25, -27, and -33 deg, and was 3.5 mm at -29 and -31 deg.
In angles wider than -35 deg, the Irs 2 wad removed for obtaining larger light and the lens diameter is 1.905 cm.
We determined the choppwer frequency as 253 Hz, considering the PD dark noise.
The PD dark noise at 253 Hz is about 100 uV as shown in Fig. 2.
The peak value at 253 Hz is read and the value is calbrated to the voltage of the PD.
The calibration constant is determined cahnging the laser power and measureing the DC value of the PD (without chopper) and the peak value of the FFT analyzer and the constant is determined as 0.80 +/- 0.01. (from FFT to PD)
The result of the BRDF is shown in Fig. 3.
Fig. 1 BRDF measurement setup with chopper.
Fig. 2 PD dark noise.
Fig. 3 BRDF measured with and without the chopper.
|
Attachment 1: BRDF_chopper_setup.pdf
|
|
Attachment 2: BRDF_with_chopper.pdf
|
|
Attachment 3: PD_darknoise.pdf
|
|
2040
|
Tue Mar 1 09:55:28 2016 |
KojiN | Misc | PD QE | Estimation of the back scattering |
In terms of the BRDF, the mount of the back scattering of our technique can be estimated using the result of elog2038 and elog2039.
In our new technique, there are two effect of the back scattering:
1. the first incident light effect,
2. the second reflection light effect.
If the reflector is placed on the 5 cm far from the PD, the two effect is estimated as follows.
The BRDF of the first effect is roughly estimated as 5*10^(-5) [1/st] form the result around 0deg of elog2039.
The BRDF of the second effect is roughly estimated as 1.5*10^(-4) (= 3*10^(-3) * 0.05) [1/st] from the result at 31 deg of elog2039 (2*10^(-3) [1/st]) and the reflectivity at 15 deg of elog2038 (0.05).
The Iris 1 diameter was 1.4 mm.
For confirming this estimation is reasonable, the back scattering effect was measured with the setup as shown in Fig. 1.
In this setup, the PBS and the QWP are used as a isolator.
The Iris 1 is place on the beam waist.
We measured with/without Iris 1 and with/without the reflector, i.e. in four pattern.
For calculating the BRDF, the formula in elog2034 is used.
For r_l, the Iris radius in front of the PDA100A (5.83 mm) is used and, for d, the distance between the PD (C30665GH) and the Iris in front of the PDA100A (35.56 cm) is used.
The result is shown in Fig. 2.
Taking the difference of the result between w/ Iris, w/o reflector and w/ Iris, w/ reflector, we can obtain the effect of the reflector, i.e. the second effect.
The difference is (1.0 +/- 0.8)*10^(-4) [1/st].
This is comparable of the upper estimation.
From these estimations, the second effect is larger by a factor of two or three.
However, considering the BRDF result arond the peak of elog 2039, we can reduce the second effect easily and dramatically by the reflector placed closer to the PD (e.g. 2.5 cm).
We will check this with more precise measurement.
Fig. 1 Setup of the back scattering measurement.
Fig. 2 BRDFs of with/without Iris and with/without reflector.
|
Attachment 1: BackSca_setup.pdf
|
|
Attachment 2: BRDF_enh.pdf
|
|
Attachment 3: BRDF_enh2.txt
|
cond sca(mV) error zero(mV) dist(cm) d(cm)
incident power was 5.44 V (PDA100A gain 0 dB)
1 337.0 0.4 284.1 35.56 1.166
2 324.0 0.2 246.0 35.56 1.166
3 327.6 0.2 251.2 35.56 1.166
4 329.4 0.2 250.8 35.56 1.166
|
2039
|
Tue Mar 1 09:12:51 2016 |
KojiN | Misc | PD QE | Measurement of the BRDF with the Iris in the incident path |
The BRDF at 15 deg incident angle with the Iris in the incident path was measured for p-pol with the same way as elog2034.
(Please note that, in this measurement, d is the distance between the PD (C30665GH) and the Iris in fron of the PDA100A.)
The setup is shown in Fig. 1.
The result is shown in Fig. 2.
Figure 2 indicates that the Iris reduced the BRDF by about 10 times.
Fig. 1 BRDF measurement setup with the Iris.
Fig. 2 BRDF with and without the Iris.
|
Attachment 1: BRDF_setup.pdf
|
|
Attachment 2: BRDF_withIris.pdf
|
|
2038
|
Mon Feb 29 08:19:53 2016 |
KojiN | Misc | PD QE | Measurement of the reflectivity, EQE, and IQE of the PD without glass window |
The reflectivity of the PD (C30665GH) at 15 deg incident angle was measured precisely with PDA100A as shown in Fig. 1.
Fig. 1 Reflectivity of the PD (C30665GH).
Ref_wow_again.txt |
Attachment 1: Reflectivity.pdf
|
|
Attachment 2: Ref_wow_again.txt
|
angle inc(V) zero ref(V,P) ref(V,S) zero
-80 5.40 0.030 2.33 2.89 0.030
-70 5.42 0.030 0.919 1.205 0.029
-60 5.41 0.030 0.335 0.498 0.027
-50 5.42 0.030 0.141 0.262 0.027
-40 5.43 0.028 0.107 0.215 0.024
-30 5.42 0.029 0.153 0.241 0.021
-20 5.42 0.030 0.234 0.282 0.022
-15 5.43 0.030 0.271 0.301 0.023
-10 5.43 0.027 0.302 0.317 0.024
... 9 more lines ...
|
2037
|
Sun Feb 28 15:47:06 2016 |
KojiN | Misc | PD QE | BRDF measurement around the peak |
The result shown in the elog2036 looks not to consistent to the previous result.
Thus we checked the alignment of the Iris.
First, the effect of the tilt of the Iris was checked and it was found that the tilt effect is very small.
Second, we found that the center of the Iris was off from the center of the lens.
This off effect is dominant in the BRDF measurement.
The off was about 1 mm. However, the diameter of the Iris was 3.5 mm and 1 mm is about 30% and large to 3.5 mm.
Thus we re-align the Iris and measure the BRDF of the PD (C30665GH) at the 15 deg incident angle again as shown in Fig. 1.
Figure 1 shows that the new result is consistent to the previous wide range measurement in the error.
The reason why the error at -29 deg and -31 deg is very large is the effect of the main reflected beam.
The result that the new data arond peak and the previous wide range data are combined for p-pol and s-pol is also attached.
Note that the incident beam is at 0 deg and the angle is that we must take care in terms of the back scattering.
Fig. 1 New data and previous data of the BRDF of the PD (C30665GH).
Fig. 2 BRDF of the PD (C30665GH) for p-pol and s-pol (wide range).
Fig. 2 BRDF of the PD (C30665GH) for p-pol and s-pol around the peak.
|
Attachment 1: BRDF_log_compare.pdf
|
|
Attachment 2: BRDF_log_all_wide.pdf
|
|
Attachment 3: BRDF_log_all_narrow.pdf
|
|
Attachment 4: BRDFat30deg.txt
|
angle inc(V) zero sca(mV,P) sca (mV) zero dis(cm)
-70 5.43 0.029 280.8 280.2 276.2 23
-60 5.44 0.030 280.3 282.1 271.8 23
-50 5.44 0.030 283.0 284.2 266.7 23
-45 5.45 0.029 311.5 317.2 267.2 23
-40 5.43 0.029 348.1 350.1 267.1 23
-35 5.44 0.029 468.6 470.1 263.9 23
-25 5.43 0.027 440.3 439.7 270.6 23
-20 5.43 0.028 349.5 349.1 278.6 23
-15 5.45 0.029 329.5 324.3 278.5 23
... 3 more lines ...
|
Attachment 5: BRDFat30deg_again.txt
|
angle inc(V) zero sca(mV,P) sca (mV) zero dis(cm) d(cm)
-35 5.44 0.029 276.5 276.5 261.4 23 0.35
-36 5.44 0.027 288.2 289.1 276.4 23 0.35
-37 5.43 0.028 293.6 291.3 280.5 23 0.35
-23 5.43 0.027 295.2 295.1 277.6 23 0.7
-25 5.43 0.030 290.2 287.2 283.2 23 0.35
|
2036
|
Fri Feb 26 23:13:02 2016 |
KojiN | Misc | PD QE | BRDF measurement around the peak |
The BRDF of the PD (C30665GH) for 15 deg incident angle was measured arond the peak with the same was as elog2035 and elog2034 but with the additaional Iris as shown in Fig. 1.
For calculating the BRDF. the fomular in elog2034 is used.
(Please note that r_l becoms the radius of the Iris in front of the PDA100A and d becomes the distance between PD (C30665GH) and the Iris)
The Iris diameter is 3.5 mm.
There is no difference for polarization as the wider scan.
The result for p-pol is shown in Fig. 2 with the wider scan (the relsult of the elog 2034). This result was wrong.
There is a asymmetry around the -30 deg and the data around -35 deg are not consistent to the previous data.
We think these are come from the slight tilt of the Iris.
We will confirn that after tomorrow.
Fig. 1 Setup for the BRDF measurement arond peak.
Fig. 2 BRDF for p-pol arond peak. Wrong result.
|
Attachment 1: BRDF_log.pdf
|
|
Attachment 2: BRDF_setup_fine.pdf
|
|
2035
|
Fri Feb 26 10:00:18 2016 |
KojiN | Misc | PD QE | BEDF measurement for wide angle |
The BRDFs at 0 deg and 15 deg in incident angle were measured for the wide angle with the same way as elog2034 and Fig. 1 but finer scaning.
The results for P-pol and S-pol are shown in FIgs. 2 and 3.
It looks that there is no difference for polarization.
The tendency that the BRDF increases from -75 deg is consistent to the previous result of the BRDF measurement of the viewport used in LIGO.
Between -40 deg and -70 deg, we face the sensitivity limit of the PD (PDA100A). (In this range, the estimated power is 1 nW and the power range of the PDA100A is 500 pW in the specsheet.)
For comparing the two BRDF, the BRDF plot in which the BRDF at 15 deg incident angle is shifted by 30 deg is shown as Fig. 4.
In this plot, the BRDF for P-pol is picked.
It looks that there is no difference between 0 deg and 15 deg in incident angle from -40 deg to 40 deg.
Fig. 1 BRDF measurement setup. thet_i is the incident angle.
Fig. 2 BRDF at 0 deg incident angle for P-pol and S-pol.
Fig. 3 BRDF at 15 deg incident angle for P-pol and S-pol.
Fig. 4 BRDF at 0 deg incident angle and shiftd BRDF at 15 deg incident angle for P-pol.
|
Attachment 1: BRDF_setup.pdf
|
|
Attachment 2: BRDFat0deg_log.pdf
|
|
Attachment 3: BRDFat30deg_log.pdf
|
|
Attachment 4: BRDF_log_P.pdf
|
|
2034
|
Wed Feb 24 23:51:07 2016 |
KojiN | Misc | PD QE | Test of the BRDF measurement |
For confirming if the BRDF of the PD (C30665GH) can be measured, the setup as shown in Fig. 1 is established.
For now, incident angle to the PD is 0 deg for simplicity.
The beam size on the PD (C30665GH) is estimated as 215 um in the x direction and 210 um in the y direction.
If the reflected light is incident to the lens, the light is focued at the PD (PDA100A).
For determining theta_s and aligning the lens, metal tube, and PD (PDA100A), the incident angle is set to theta_s/2 temporally and seeing the reflected light they are placed without PD.
Specifically, the reflected light is cented to the lens and the output of the tube.
After that, the PD (PDA100A) is placed and the incident angle is set to 0 deg.
The distance between lens and PD is 31 cm for theta_s>0, and 23 cm for theta_s<0.
These distances are determined by space constraint.
And also for space constraint, the BRDF cannot be measured at the theta_s beteween 40 and 80 deg.
The incident power is also measured by the PD (PDA100A).
The gain of the PD (PDA100A) was 0 dB at the incident power measurement and 70 dB at the scattered light power measurement.
The BRDF is obtained with folloing equation,
,
where P_S is the power of the scatterd light, P_i is the power of the incident light, theta_i is the incident angle, dOmega is the solid angle of the detector, r_l is the radius of the lens, and d is the distance between the PD (C30665GH) and the lens.
With the setup and the method, the BRDF is measured for P-pol and S-pol as shown in Figs. 2 and 3.
The error of the Figs. 2 and 3 are determined by the systematic error of the PD and the distance error.
When Figs. 2 and 3 are observed, the BRDF looks to be able to be measured.
However the data quality can be improved.
Thus we are going to scan finer angles and change the incident angle from tomorrow.
Fig. 1 Setup for the BRDF measurement.
Fig. 2 Measured BRDF of the PD (C30665GH) at 0 deg incident angle.
Fig. 3 Semilog version of Fig. 2.
|
Attachment 1: BRDF_mes.pdf
|
|
Attachment 2: BRDFat0deg.pdf
|
|
Attachment 3: BRDFat0deg2.pdf
|
|
Attachment 4: BRDFat0deg.txt
|
angle inc(V) zero sca(mV,P) sca (mV) zero dis(cm)
-90 5.43 0.026 330.0 336.9 328.7 23
-80 5.44 0.027 372.2 375.7 309.1 23
-70 5.44 0.026 295.8 295.1 292.9 23
-60 5.45 0.026 294.1 295.9 288.4 23
-50 5.43 0.027 289.1 288.9 285.2 23
-40 5.44 0.027 294.0 293.6 286.4 23
-30 5.44 0.026 296.3 292.8 282.1 23
-20 5.43 0.028 316.7 309.0 290.3 23
-10 5.44 0.026 366.4 365.7 292.9 23
... 4 more lines ...
|
Attachment 5: BRDF.m
|
% BRDF @ 0deg
filename='BRDFat0deg.txt';
delimiterIn = ' ';
headerlinesIn = 1;
A = importdata(filename,delimiterIn,headerlinesIn);
G0=1.51*10^3; % V/A
EG0= 0.02; % %
G70=4.75*10^6; % V/A
... 25 more lines ...
|
2033
|
Wed Feb 24 08:18:06 2016 |
KojiN | Misc | PD QE | Systematic error of the power meter (RM9, Ophir) |
In the spec sheet, the systematic error of RM9 (power meter, Ophir) is 5%.
This is not very good.
However, the power meter was calibrated in the company at 808 nm wavelength and the deviation was 0.6%
This is very good.
If the 0.6% error could be applied to our power measurement, the systematic error would be improved.
Thus we inquired the error of RM9 to Ophir.
The answer was that the error is not 0.6% but is 5% at 1064 nm. |
Attachment 1: RM9_RM9-PD_0.pdf
|
|
2031
|
Tue Feb 23 00:22:01 2016 |
KojiN | Misc | PD QE | Re-measurement of the QE of the PD without window glass |
The QEs were measured again at every 10 deg incident angle.
The power meter to measure the incident angle was S401C (Thorlabs).
Output voltage was read by the degital multimeter 77 IV (Fluke).
The results are shown in Figs. 1--3.
(About the reflectivities, the data measured in elog2024 are used for now. Those data are obtained with S130C (Thorlabs. error is +/- 7%.))
In Fig. 3, the ratio of the enhanced QE and the not-enhanced QE is shown.
In these figures, the error of the S401C (3%), 77 IV (about 0.4% (depends on the measure value)), and resistances of the readout circuit (assumed to be 1%) are taken into consideration and they are assumed to be independent.
These results are consistent to the data we have measured by last week in the error.
Current errors of the EQEs in Figs. 1 and 2 about +/- 4% and they are mainly determined by the systematic error of the power meter.
Figure 3 claims that the QEs are improved with +/- 0.6% error using our new method.
Fig. 1 Measured QEs.
Fig. 2 Zoomed version of Fig. 1.
Fig. 3 Improvement ratio of the QE.
|
Attachment 1: QE.pdf
|
|
Attachment 2: QE2.pdf
|
|
Attachment 3: QE3.pdf
|
|
Attachment 4: QE_enh.txt
|
angle inc(mW,P) error DC(V,P) error enh(V,P) error inc(mW,S) error DC(V,S) error enh(V,S) error
-10 11.45 0.3 1.914 0.007 2.000 0.007 11.48 0.3 1.915 0.007 2.003 0.007
-15 11.50 0.3 1.936 0.007 2.013 0.007 11.53 0.3 1.922 0.007 2.007 0.007
-20 11.56 0.3 1.950 0.007 2.010 0.007 11.62 0.3 1.939 0.007 2.012 0.007
-30 11.53 0.3 1.984 0.007 2.026 0.007 11.65 0.3 1.954 0.007 2.023 0.007
-40 11.55 0.3 2.001 0.007 2.029 0.007 11.58 0.3 1.968 0.007 2.029 0.007
-50 11.66 0.3 1.996 0.007 2.031 0.007 11.60 0.3 1.953 0.007 2.028 0.007
-60 11.67 0.4 1.938 0.007 2.030 0.007 11.68 0.4 1.888 0.007 2.026 0.007
-70 11.69 0.4 1.768 0.006 2.003 0.007 11.68 0.4 1.680 0.006 1.974 0.007
|
Attachment 5: Ref_QE_wow.txt
|
angle Ref(mW,P) error DC(V,P) error Ref(mW,S) error DC(V,S) error
-10 0.480 0.001 2.02 0.02 0.504 0.001 2.02 0.02
-15 0.434 0.001 2.02 0.02 0.487 0.001 2.02 0.02
-20 0.367 0.001 2.04 0.02 0.454 0.001 2.04 0.02
-30 0.217 0.001 2.06 0.02 0.363 0.001 2.06 0.02
-40 0.136 0.001 2.06 0.02 0.313 0.002 2.04 0.02
-50 0.197 0.002 2.06 0.02 0.408 0.002 2.02 0.02
-60 0.532 0.002 1.98 0.02 .806 0.005 1.94 0.02
-70 1.54 0.01 1.80 0.02 2.04 0.02 1.72 0.02
|
Attachment 6: QE_enh_cal_NK.m
|
%
filename='QE_enh.txt';
delimiterIn = ' ';
headerlinesIn = 1;
A = importdata(filename,delimiterIn,headerlinesIn);
filename='Ref_QE_wow.txt';
B = importdata(filename,delimiterIn,headerlinesIn);
phiP=A.data(:,2);
... 74 more lines ...
|
2030
|
Mon Feb 22 23:51:18 2016 |
KojiN | Misc | PD QE | Beam shape reflected by the PD |
The beam shape reflected by the PD was measured.
The measurement was done 5 cm and 15 cm far from the PD at 15 deg, 45 deg, and 60 deg in incident angle.
(Please note that only at 15 deg, the measurement was done at 5 cm because of the space constraint.)
The results are shown in Figs. 1 and 2 and Tabs. 1 and 2.
When Figs. 1 and 2 are observed, the beam shape reflected by the PD looks still gaussian.
In the simulation, the beam size is estimated as, at 5 cm, 403 um in x-direction and 391 um in y-direction and, at 15 cm, 792 um in x-direction and 766 um in y-direction, respectively.
When Tabs. 1 and 2 are observed, the measured beam size looks not so different from the simulated one.
Fig. 1 Screen of the CCD camera at 15 deg and 15 cm.
Fig. 2 Screen of the CCD camera at 45 deg and 15 cm.
Tab. 1 Beam size in x-direction
|
15 deg |
45 deg |
60 deg |
5 cm (6 cm) |
470 um |
406 um |
411 um |
15 cm |
736 um |
625 um |
608 um |
Tab. 2 Beam size in y-direciton
|
15 deg |
45 deg |
60 deg |
5 cm (6 cm) |
522 um |
489 um |
491 um |
15 cm |
815 um |
811 um |
760 um |
|
Attachment 1: BeamAt15deg15cm.pdf
|
|
Attachment 2: BeamAt45deg15cm.pdf
|
|
2029
|
Sun Feb 21 22:02:01 2016 |
Koji Antonio, | Misc | PD QE | PD window removal few notes |
Rich removed the window from PD on Friday. The basics steps for the removal are the following:
0. PD in a socket, it helps;
1. In the beginning you open the screws on the cutter up so they just hold the photodiode;
2. Make a light little line first at desired height and make sure that it is a circle and not a helix;
3. You do not want to go start right away full force on it, you want to make tiny little incremental cuts. Eventually it just falls apart;
 
|
Attachment 2: Screen_Shot_2016-02-21_at_8.34.43_PM.png
|
|
2028
|
Sun Feb 21 21:53:04 2016 |
KojiN | Misc | PD QE | Measurement of the enhanced QE of the PD without glass window |
If we want to check soon the EQE with a different beam size, we may want just to add (for a kind of quick check) an additional lens in the path without changing what we already have.
Quote: |
> My interpretation of this is that we get ~2% increase in the peak EQE for p-pol by doing the extra bounce. Is that correct? If so, its a pretty good result.
Your interpretation for the current result is collect. However, there are still several errors which aren't took into consideration, such as the error of the resistances and so on. Thus, for saying that the EQE is increased by about 2%, we must obtain more accurate data or analyse the results more preciselly.
> I wonder if you can do some more angles to see if there are any features in the angular dependence.
We plan to do a finer angular scan with the more accurate setup.
> It would also be interesting to modify the beam size to see if there is any change in the EQE. What is the estimated beam size now in the x & y directions? Does the reflected beam overlap the first beam? It would be good if the 2nd reflection from the PD can be steered to not go back into the fiber.
Actually, it is not easy to change the beam size immediatelly because the lenses and the PD are fixed. However, we plan to change the beam size for observing the BRDF. Thus in that time we will measure the EQE changing the beam size.
The estimated beam size of the first incident light on the PD is 215 um in the x direction and 210 um in the y direction, and the size of the second reflected light on the PD is 436 um in the x direction and 422 um in the y direction.
The second reflected light is aligned not to overlap the first incident light. However, so far, the distance of the two beams cannot not be measured because the second light is so weak that we cannot see.
> Is there a diagram with the reflector in place?
This is the diagram with the reflector.

> Is it possible to use the camera to take an image of the reflection from the PD? I wonder if its Gaussian or messy.
It's possible. We will measure the beam profile of the reflection light from the PD for confirming if the PD can be regarded to be well-polished and flat.
> What about repeating this experiment with the 1.5 micron laser now? Or maybe a HeNe where the IQE is lower?
For now, we don't have the plan to change the wavelength. Before the wavelength is changed, we would like to do the whole measurement includidng the noise estimation such as the back scattering noise. Once we establish the whole experimental method, we will change the wavelength and the PD.
|
|
2027
|
Sun Feb 21 21:16:10 2016 |
KojiN | Misc | PD QE | Measurement of the enhanced QE of the PD without glass window |
> My interpretation of this is that we get ~2% increase in the peak EQE for p-pol by doing the extra bounce. Is that correct? If so, its a pretty good result.
Your interpretation for the current result is collect. However, there are still several errors which aren't took into consideration, such as the error of the resistances and so on. Thus, for saying that the EQE is increased by about 2%, we must obtain more accurate data or analyse the results more preciselly.
> I wonder if you can do some more angles to see if there are any features in the angular dependence.
We plan to do a finer angular scan with the more accurate setup.
> It would also be interesting to modify the beam size to see if there is any change in the EQE. What is the estimated beam size now in the x & y directions? Does the reflected beam overlap the first beam? It would be good if the 2nd reflection from the PD can be steered to not go back into the fiber.
Actually, it is not easy to change the beam size immediatelly because the lenses and the PD are fixed. However, we plan to change the beam size for observing the BRDF. Thus in that time we will measure the EQE changing the beam size.
The estimated beam size of the first incident light on the PD is 215 um in the x direction and 210 um in the y direction, and the size of the second reflected light on the PD is 436 um in the x direction and 422 um in the y direction.
The second reflected light is aligned not to overlap the first incident light. However, so far, the distance of the two beams cannot not be measured because the second light is so weak that we cannot see.
> Is there a diagram with the reflector in place?
This is the diagram with the reflector.

> Is it possible to use the camera to take an image of the reflection from the PD? I wonder if its Gaussian or messy.
It's possible. We will measure the beam profile of the reflection light from the PD for confirming if the PD can be regarded to be well-polished and flat.
> What about repeating this experiment with the 1.5 micron laser now? Or maybe a HeNe where the IQE is lower?
For now, we don't have the plan to change the wavelength. Before the wavelength is changed, we would like to do the whole measurement includidng the noise estimation such as the back scattering noise. Once we establish the whole experimental method, we will change the wavelength and the PD.
|
Attachment 1: 2016_02_19_setup.pdf
|
|
2026
|
Sun Feb 21 14:27:24 2016 |
rana | Misc | PD QE | Measurement of the enhanced QE of the PD without glass window |
My interpretation of this is that we get ~2% increase in the peak EQE for p-pol by doing the extra bounce. Is that correct? If so, its a pretty good result.
I wonder if you can do some more angles to see if there are any features in the angular dependence.
It would also be interesting to modify the beam size to see if there is any change in the EQE. What is the estimated beam size now in the x & y directions? Does the reflected beam overlap the first beam? It would be good if the 2nd reflection from the PD can be steered to not go back into the fiber.
For the circuit, the R=20 Ohms makes it nicely so that the PD bias voltage doesn't change that much, which is good.
Is there a diagram with the reflector in place?
Is it possible to use the camera to take an image of the reflection from the PD? I wonder if its Gaussian or messy.
What about repeating this experiment with the 1.5 micron laser now? Or maybe a HeNe where the IQE is lower? |
2025
|
Fri Feb 19 20:25:14 2016 |
KojiN | Misc | PD QE | Measurement of the enhanced QE of the PD without glass window |
The enhanced QE of the PD without the glass window was measured in the same setup as elog2023.
Fig. 1 Determined enhanced and non-enhanced QEs of the PD without the glass window.
Fig. 2 Zoomed versioin of Fig. 1.
QE_enh.txt QE_enh_ref.txt QE_enh_cal_NK.m |
Attachment 1: QE_enh.pdf
|
|
Attachment 2: QE_enh2.pdf
|
|
Attachment 3: QE_enh.txt
|
angle inc(mW,P) error DC(V,P) error enh(V,P) error inc(mW,S) error DC(V,S) error enh(V,S) error
-15 11.22 0.02 1.96 0.01 2.03 0.01 11.18 0.02 1.94 0.01 2.03 0.01
-30 11.12 0.02 1.98 0.01 2.02 0.01 11.12 0.02 1.96 0.01 2.02 0.01
-45 11.20 0.02 2.01 0.01 2.04 0.01 11.22 0.02 1.97 0.01 2.04 0.01
-60 11.29 0.01 1.94 0.01 2.04 0.01 11.20 0.02 1.88 0.01 2.03 0.01
-70 11.23 0.02 1.75 0.01 2.01 0.01 11.31 0.02 1.67 0.01 1.98 0.01
|
Attachment 4: QE_enh_ref.txt
|
angle inc(mW,P) error ref(mW,P) error inc(mW,S) error ref(mW,S) error
-15 11.42 0.01 0.434 0.001 11.44 0.01 0.487 0.001
-30 11.42 0.01 0.222 0.001 11.45 0.01 0.371 0.001
-45 11.40 0.01 0.143 0.001 11.46 0.01 0.348 0.001
-60 11.32 0.01 0.569 0.001 11.47 0.01 0.870 0.001
-70 11.23 0.01 1.58 0.01 11.27 0.01 2.06 0.01
|
Attachment 5: QE_enh_cal_NK.m
|
% EQE calculation
%V=input('input voltage (V): ');
filename='QE_enh.txt';
delimiterIn = ' ';
headerlinesIn = 1;
A = importdata(filename,delimiterIn,headerlinesIn);
filename='QE_enh_ref.txt';
delimiterIn = ' ';
headerlinesIn = 1;
... 77 more lines ...
|
2024
|
Fri Feb 19 20:15:03 2016 |
KojiN | Misc | PD QE | Measurement of the reflectivity, EQE, and IQE of the PD without glass window |
The glass window of the PD was removed and is placed in our shelf.
The reflectivity, EQE, and IQE of the PD were measured without the glass window as shown in following figures in the same setup as the measurements of the PD with glass window (see also elog2019, elog2020, and elog2022).
The laser power incident to the PD was 11.4 +/- 0.1 mW.
Fig. 1 Measured reflectivities of the PD without the galss window for P-pol and S-pol.
Fig. 2 Determined EQEs and IQEs of the PD without the galss window for P-pol and S-pol.
Fig. 3 Zoomed version of Fig. 2.
Ref_QE_wow.txt QE_cal3_NK.m |
Attachment 1: Ref.pdf
|
|
Attachment 2: QE.pdf
|
|
Attachment 3: QE2.pdf
|
|
Attachment 4: Ref_QE_wow.txt
|
angle Ref(mW,P) error DC(V,P) error Ref(mW,S) error DC(V,S) error
-85 5.82 0.01 0.70 0.02 7.04 0.03 0.60 0.02
-80 4.05 0.02 1.34 0.02 4.72 0.01 1.22 0.02
-70 1.54 0.01 1.80 0.02 2.04 0.02 1.72 0.02
-60 0.532 0.002 1.98 0.02 .806 0.005 1.94 0.02
-50 0.197 0.002 2.06 0.02 0.408 0.002 2.02 0.02
-40 0.136 0.001 2.06 0.02 0.313 0.002 2.04 0.02
-30 0.217 0.001 2.06 0.02 0.363 0.001 2.06 0.02
-20 0.367 0.001 2.04 0.02 0.454 0.001 2.04 0.02
-10 0.480 0.001 2.02 0.02 0.504 0.001 2.02 0.02
... 11 more lines ...
|
Attachment 5: QE_cal3_NK.m
|
% EQE calculation
%V=input('input voltage (V): ');
filename='Ref_QE_wow.txt';
delimiterIn = ' ';
headerlinesIn = 1;
A = importdata(filename,delimiterIn,headerlinesIn);
% filename1 = 'QE_Ref_P.txt';
% filename2 = 'QE_Ref_S.txt';
% filename3 = 'QE_enh.txt';
... 59 more lines ...
|
2023
|
Fri Feb 19 00:39:24 2016 |
KojiN | Misc | PD QE | First QE measurement with the reflector |
QEs (with the glass window) with the reflector were measured as shown in Fig. 1.
The radius of curvature of the reflector is 25 cm and the reflector was placed at a distance of 5 cm from the PD.
At the reflector position, the beam size is 403 um in x-direction and 391 um in y-direction.
And, at the PD position, the beam size of the light reflected by the reflector is 435 um in x-direction and 422 um in y-direction.
When Fig. 1 is observed, the QEs in S-pol are less enhanced than that in P-pol.
This is because there are a few lights which are reflected by the PD and have almost the same power, in other words in this measurement only one or two beams are re-input to the PD by reflector.
In fact, we can see the two peak output voltages of the PD when the reflector are rotated in the yaw direction.
Fig. 1 Measured enhenced and not enhanced QEs.
QE_enh.txt |
Attachment 1: QE_enh.txt
|
# angle(deg) enh.vol(V,P) error vol(V,P) error enh.vol(V.S) error vol(V,S) error
-15. 2.12 0.01 1.94 0.01 2.12 0.01 1.94 0.01
-30. 2.12 0.01 2.00 0.01 2.08 0.01 1.88 0.01
-45. 2.12 0.01 2.08 0.01 1.92 0.01 1.76 0.01
-60. 2.10 0.01 2.04 0.01 1.62 0.01 1.46 0.02
|
Attachment 2: QE_enh.pdf
|
|
2022
|
Fri Feb 19 00:01:36 2016 |
KojiN | Misc | PD QE | Re-measurement of the reflectivities and the EQEs of the PD for S-pol and P-pol |
I calculated IQEs using the measured EQEs and reflectivities and the following formula:
,
where R is the reflectivity of the PD, and T is the transmittance of the PD.
Here T is assumed to be zero and the scattering loss is ignored as you said.
The obtained IQEs are shown in Fig. 1.
Also the EQE, IQE, and reflectivity at a few angles near the angle where the EQE goes maximum for the P pol were measured as shown in Fig. 2 and Fig. 3.
The angle where the EQE goes maximum for the P pol is -51 deg.
So far, in these plots, the systematic errors because of the power meter are not considered.
And I replaced the figures with the figures having larger fonts.
Fig. 1 Measured EQEs and IQEs.
Fig. 2 Measured reflectivities around -51 deg.
Fig. 3 Measured EQEs and IQEs around -51 deg.
QE_Ref_P.txt QE_Ref_S.txt
AroundPeak.txt |
Attachment 1: IQE.pdf
|
|
Attachment 2: Ref.pdf
|
|
Attachment 3: QE_Ref_P.txt
|
# angle(deg) 3beams(mW) error DCout(V) error
# incident power is 11.7 +/- 0.2 mW
-80. 4.6 0.1 0.14 0.02
-75. 3.52 0.02 1.02 0.02
-70. 1.24 0.02 1.58 0.02
-60. 0.627 0.002 2.04 0.02
-50. 0.244 0.002 2.12 0.02
-40. 0.410 0.002 2.10 0.02
-30. 0.512 0.002 2.04 0.02
-20. 0.851 0.002 1.98 0.02
... 13 more lines ...
|
Attachment 4: QE_Ref_S.txt
|
# angle(deg) 3beams(mW) error DCout(V) error
# incident power is 11.7 +/- 0.2 mW
-80. 6.61 0.03 0.04 0.02
-75. 6.17 0.03 0.52 0.02
-70. 5.29 0.02 0.98 0.02
-60. 3.40 0.02 1.42 0.02
-50. 1.85 0.02 1.70 0.02
-40. 1.78 0.02 1.84 0.02
-30. 1.372 0.004 1.90 0.02
-20. 1.293 0.004 1.92 0.02
... 13 more lines ...
|
Attachment 5: AroundPeak.txt
|
angle(deg) inc.pow.(mW,P) error Ref(mW,P) error DCout(V,P) error inc.pow.(mW,S) error Ref(mW,S) error DCout(V,S) error
-47 11.64 0.03 0.173 0.003 2.02 0.01 11.54 0.03 1.92 0.03 1.70 0.01
-49 11.46 0.02 0.375 0.004 1.99 0.01 11.60 0.03 2.12 0.04 1.66 0.01
-51 11.67 0.03 0.282 0.003 2.04 0.01 11.61 0.03 2.36 0.03 1.64 0.01
-53 11.56 0.02 0.325 0.003 2.00 0.01 11.35 0.03 2.59 0.04 1.56 0.01
-55 11.62 0.02 0.518 0.003 1.97 0.01 11.53 0.02 2.86 0.04 1.51 0.02
|
Attachment 6: QE2.pdf
|
|
2021
|
Thu Feb 18 13:39:33 2016 |
Koji | Misc | PD QE | Re-measurement of the reflectivities and the EQEs of the PD for S-pol and P-pol |
Can you calculate the IQE from EQE and reflectivity (by ignoring scatter loss)?
Can you measure the EQE/IQE/reflectivity at the angle where the EQE goes maximum for the P pol?
Can you use larger fonts in the plots? |
2020
|
Thu Feb 18 08:12:53 2016 |
KojiN | Misc | PD QE | Re-measurement of the reflectivities and the EQEs of the PD for S-pol and P-pol |
The reflecrivities and the EQEs of the PD (C30665GH, with the glass window) for S-pol and P-pol were measured at every 10 deg in incident angle as shown in Fig. 1 and Fig. 2.
In this measurement, for alignmnet the steering mirror was used and it was confirmed that the beam was not clipped.
The EQE is obtained from the DC output voltage which was read from the oscilloscope using the following equation (see also elog2019):
,
where I_PD is the photocurrent, V_out is the DC output voltage, h is Planck's constant, c is the speed of light, phi is the incident power, n is the index of refraction of air, e is the elementary electronic charge, and lambda is the wavelength of the laser.
Fig. 1 Measured reflectivities for P-pol and S-pol.
Fig. 2 Determined EQEs for P-pol and S-pol.
|
Attachment 1: QE3.pdf
|
|
Attachment 2: QE_ll.pdf
|
|
2019
|
Tue Feb 16 23:51:31 2016 |
KojiN | Misc | PD QE | Re-measurement of the PD efficiencies for P-pol. and S-pol. |
We aligned the single mode fiber again and We obtained the linear polarization without QWP as shown in Fig. 1.
P-pol. and S-pol. are able to be obtined at 34.8 deg and 81.0 deg in angle of HWP #3, respectively.
Then, We checked the PD read-out circuit and the circuit diagram is shown in Fig. 2.
Important resistanve values are as follows:
R4 = 20.2 Ohm, R5 = open, R11 = 1.000 kOhm, R16 = 9.82 kOhm, R7 = 50.2 Ohm.
In this circuit, the current generated by PD (I_PD) can be calibrated from the DC output voltage (V_out) using the following equation:

(In other words, the trans-impedance of the readout circuit is 219 Ohm.)
After that, we measured the DC output voltage at every 10 deg in the incident angle with the window glass as shown in Fig. 3.
The PD position is 5 cm after the beam waist and, at this position, the beam sizes in x-direction and y-direction are 215 um and 210 um, respectively.
The incident angle is determined with +/- 0.5 deg error by following way:
step1 (in this step, PD is alomost perpendicular to the beam) seeing reflected light from the PD, we centerd the beam to the PD using the steering mirror.
step2 We miscenterd the beam to the border of the PD, and checked the incident angle seeing the reflected light. The incident angle was 0.5 deg (-0.5 deg in the opposite side.) Then the beam was re-centerd to the PD by the steering mirror.
step3 The PD is tilted by the rotational stage (the angle of the stage is called "alpha"), and, if the beam is miscenterd, we centerd the beam to the PD using steering mirror (the angle caused by the steering mirror is called "beta"). Thus the incident angle is "alpha + beta"
step4 Without touching the steering mirror, the PD is rotated to be perpendicular to the beam (i.e. "alpha" became 0 deg).
step5 We checked whether the beam is still on the PD or not (in today's measurement, the beam is on the PD in this step). If the beam is on the PD, the incident angle at step3 is estimated as follows,
alpha - 0.5 < alpha + beta (incident angle) < alpha + 0.5 (degree)
Therefore, the incident angle is determined with +/- 0.5 deg error.
We are going to measure again the reflectivity tomorrow with this method.
Fig. 1 Current setup.
Fig. 2 Readout circuit diagram.
Fig. 3 Measured DC output and EQE.
D980454-00(QE).pdf |
Attachment 3: D980454-00(QE).pdf
|
|
Attachment 4: D980454-00(QE)_2.pdf
|
|
Attachment 5: D980454-00(QE).pdf
|
|
2018
|
Sun Feb 14 01:04:07 2016 |
KojiN | Misc | PD QE | Measurement of PD reflectivities and PD efficiencies for P-pol. and S-pol. |
According to the polariztion measurement, we found that the laser is not linear polarization.
Thus the QWP #2 was placed after collimator #2 as shown in Fig. 1 and the linear polarization laser was obtained.
Then the PD, C30665GH, with the glass window was set on the PD mount and placed at the center of the rotational stage as shown in Fig. 1.
The PD position is 5 cm after the beam waist and, at this position, the beam sizes in x-direction and y-direction are 215 um and 210 um, respectively.
In this setup, there are two reflected lights which may be come from the glass window and the PD.
Changing the laser incident angle to the PD, I measured the PD reflectivities and the PD efficiencies for P-pol. and S-pol.
The results are shown in Fig. 2 and Fig. 3.
The reason why the measurement end before the tasty part like ~40deg is that the beam is clipped.
Possibly, the light may already be clipped at ~25deg.
The alignment, i.e. the centering of the light to the PD, was optimized before the measurement as possible as we could.
The measurement was done only by rotating the rotational stage, i.e. I didn't use the steering mirror.
(Perhaps, I should use the steering mirror for optimizing the alignment at every angle.)
The incidnet power was 11.4 +/- 0.1 mW.
I meausred the powers of the one light, the other light (with the Iris #3), and the two lights (without the Iris #3) at all incident angles.
The errors in Fig. 2 is determined by the error of the measured incident and reflected light power. The position dependency on the power meter is not took into consideration.
When the PD responsivity for 1064 nm and the PD calibration constant (W/V) are 0.78 A/W and 5.8*10^(-3) W/V (see elog:2017), respectively, is assumed, the PD EQE is about 0.9 between -30 degree and 40 degree in the incident angle, according to
,
where I is the photocurrent, h is Planck's constant, c is the speed of light, phi is the incident power, n is the index of refraction of air, e is the elementary electronic charge, and lambda is the wavelength of the laser.
For calibrating the output voltage to EQE, the electric current must be known but, so far, I don't know the resistance in the photo detector circuit because the circuit is not made by myself. Thus I must investigate the circuit.
This is why I used "0.78 A/W" and "5.8*10^(-3) W/V" to estimate the current for a moment. The calibration constant "5.8*10^(-3) W/V" is determined without taking the errors, such as the position dependency on the power meter, the fluctuation of the laser light and so on, into consideration. (Of course, this is a problem.)
The errors in Fig. 3 is determined by the error of the reading error of the oscilloscope and the incident power. The position dependency on the power meter is not took into consideration.
Fig. 1 Current setup.
Fig. 2 Reflectivities for P-pol. and S-pol. between -25 degree and 30 degree in the incident angle. "Beam 1+2" means the measured power of the both reflected light and "Beam1+Beam2" means the sum of "Beam1" and "Beam2".
Fig. 3 PD output voltage for P-pol. and S-pol. in the wide incident angle.
|
2017
|
Thu Feb 11 23:54:48 2016 |
KojiN | Misc | PD QE | Polarization measurement |
For measuring the polarization, the setup as shown in Fig. 1 was prepared.
The angle of the HWP #2 was 30 degree.
Rotating the angle of the HWP #3, I measured the laser power with a power meter and a PD.
And I fitted the measured data to the function, .
Here \theta is the angle of the HWP #3.
The result was shown in Fig. 2 and the paremeters were determined as
(with the power meter) a = 7.965 +/- 0.0005 mW, b = -0.002 +/- 0.003 mW, phi = -40.65 +/- 0.01,
(with the PD) a = 1373 +/- 3, b = -1 +/ 2, phi = 40.52 +/- 0.03.
Accoding to this result, the S-pol. and the P-pol are obtained at 40.6 degree and 85.6 degree of the angle of the HWP #2, respectively.
And the calbration constant of the PD from voltage to power is determined roughly as 5.8*10^(-3) W/V. (Systematic errors have not yet been concerned.)
Fig. 1 Current setup.
Fig. 2 Measure data and fitted curves.
|
2016
|
Thu Feb 11 00:03:24 2016 |
KojiN | Misc | PD QE | Beam Adjustment |
For adjusting the beam, the setup was prepared as shown in Fig. 1.
Our taget is 80 um beam waist in radius as shown in Fig. 2.
(Please note that the axes of Fig. 1 and Fig. 2 do not have the same zero point.)
The waist position is not much took care to if the position can be accessed easily.
In our simulation using the beam measured yesterday, when the lens #2 (f=-200 mm) and the lens #3 (f=150 mm) are placed at z = 0.260 m and z = 0.474 m, respectively, the target radius could be obtained at z = 0.74 m.
However, in this setup we cannot obtain the target beam and the waist size seemed to be larger and the waist position seemed to be farther.
Thus the lenses were moved iteratively.
Finally, the lens #2 and #3 are placed at z = 0.114 m and z = 0.483 m, respectively, and the measured beam is as shown in Fig. 3.
(The beam profiler seemed to be saturated. Thus the additional ND filter (O.D.=4.0) is put and the saturation seemed to be extinguished as shown in Fig. 4.)
As a result, the beam waist radius and the beam waist position are 86 um and z = 0.750 m (for the x-direction) and 89 um and z = 0.750 m (for the y-direction), respectively.
These parameters can still be improved by refining the positions of the lens #2 and #3.
Fig. 1 Current setup.
Fig. 2 Target beam.
Fig. 3 Measured data and fitted curves of the beam profile.
Fig. 4 Screen of the beam profiler.
|
2015
|
Wed Feb 10 00:17:31 2016 |
KojiN | Misc | PD QE | Alignment of the single mode fiber and re-measurement of the beam profile |
The beam measued yesterday was very elliptic.
Thus, for obtaining a round beam, we decided to use a single mode fiber.
The single mode fiber was aligned with KojiA's instruction.
(There is the specific method in the last of this log)
As a result, the obtained mode matching ratio was 64.5%. (Input power for the fiber was 18.7 mW, and output power of the fiber was 12.07 mW.)
Then, the setup as shown in Fig. 1 was prepared.
In this setup, the beam profile after the single mode fiber was measured as shown in Fig. 2.
When Fig. 2 is observed, there is still a kind of lines in the y-direction and we consider it comes from the problem of camera.
The measured data and fitted curves are shown in Fig. 3.(Zero point of the distance was set at the position of the collimator #2)
Specific method to align the single mode fiber
1. place the collimator #1 at the focal point of the laser
2. align the laser light (IR light) to the collimator #1 roughly with steering mirror
3. unset the collimator #2 and set a fiber illuminator instead of the collimator #2
4. at just after the Iris #2, align the laser from the fiber illuminator (Red light) to the IR light with knobs of the collimator #1 holder.
5. at just before the collimator #1, align the IR light to the Red light with the steering mirror #1
6. repeat step 4. and 5. a few times
7. unset the fiber illuminator and confirm the IR laser is ouput from the collimator #2. (If the IR light cannot be observed, go back to step 3.)
8. set a power meter after the collimator #2
9. maximize the IR light power using yaw knobs of the collimator #1 holder and the steering mirror #1
10. maximize the IR light power using pitch knobs of the collimator #1 holder and the steering mirror #1
11. repeat step 9. and 10.
12. If a good mode matching ratio can be obtained, this fiber alignment is finished. If the good mode matching ratio cannot be obtained, change the Lens #1 and re-start from step 1.
Fig. 1 current setup.
Fig. 2 Screen of the beam profiler.
Fig. 3 Measured data and fitted curves of the beam profile.
|
2014
|
Tue Feb 9 00:00:11 2016 |
KojiN | Misc | PD QE | Measurement of a beam profile |
For measuring a current beam profile, the setup as shown in Fig. 1 was prepared.
At first, laser powers at Point 1 and 2 were measured and the results were
* at Point 1: 17.71 mW
* at Point 2: 16.15 mW
The difference may be come from the loss of HWP #2 and the beam damped by Iris.
(Laser power just in front of the Iris cannot be measured because of the space constraint.)
Then the beam profile was measured as shown in Fig. 2.
The measurement was perfomed from 0 cm to 25.4 cm (11 points) and in the measurement the angles of HWP #1 and #2 were 48.1 degree and 22.4 degree, repsectively.
(The zero point is set after the PBS as shown in Fig. 1.)
As shown in Fig. 2, the beam profile, in especially y-direction, seems to be a bit strange.
This is condiered to be an effect of a kind of saturation, but this is a just guess.
The measured data and the fitted curves are shown in Fig. 3.
Fig. 1 Current setup.
Fig. 2 Beam profile measurement.
Fig. 3. Measured data and fitted curves.
|
Attachment 3: BeamAfterPBS.pdf
|
|
2013
|
Fri Feb 5 21:06:14 2016 |
KojiN | Misc | PD QE | Initial alignment of optics |
To pick off adequate laser power for our PD QE enhancement experiment, the first HWP in front of the laser was rotated.
The inititial angle of the first HWP was 48 degree.
For measuring the laser power, some optics were aligned as shown in the following figure using a CCD camera.
In this figure, the beam dump #1 is placed for dummping the laser for the other expetiment that is made on the same optical bench as the PD QE experiment, and the Irises are placed for dumping unnecessary light.
I plan to place HWPs and PBSs additionally and to measure the polarization and the laser power at the beginning of next week.
Current optical setup.
|
2012
|
Thu Feb 4 21:23:31 2016 |
KojiN | Misc | PD QE | Preparation of some components |
I prepared some basic optics for a PD QE enhancement experiment.
Specifically, two half wave plates, a PBS, a BS, a PD mount, and a stage for the PD mount are prepared.
The PD mount has a glued connector for PDs for replacing them easily.
The sage for the PD mount has three micrometers for moving PDs accurately to three axes.
A male pin assignment for a DC power supply of a circuit for the PD is confirmed.
As shown in the following image, #1 pin, #2 pin, and #5 or #9 pin should be connected to +15 V, -15 V, and GND, respectively.
Male pin assignment for the DC power supply of the circuit for the PD
In addition, for aligning the optics, a CCD camera and a lens for the camera are also prepared.
All things are placed on an optical bench without being aligned.
I will align the optics and test the PD circuit and the camera with laser light. |
2011
|
Tue Jan 5 14:48:08 2016 |
rana | SafetyLab Infrastructure | HVAC | heating and cooling |
As noticed by Kate a few times last year, the north side of the lab has hot air comiing out of the HEPA vents and the south side has cold air. This seems to be a problem with the setpoints for the sensors or the hot water actuators.
Let's remember to call physical plant after the current roof leaking situation settles down. |
2010
|
Wed Sep 16 04:43:35 2015 |
Koji | Misc | Seismometer | IP test at the 40m |
Alessandra and I have constructed the IP at the 40m.
The lowest resonant freq I could obtain was 132mHz when 1613g additional mass is placed on the top plate.
It is very tilt sensitive and the bottom plate has three screws to adjust the leveling.
|
Attachment 1: IMG_20150914_181922982.jpg
|
|
Attachment 2: IMG_20150914_182024785_HDR.jpg
|
|
Attachment 3: IMG_20150914_182037693.jpg
|
|
Attachment 4: IMG_20150914_182550377.jpg
|
|
Attachment 5: IMG_20150914_181825903_HDR.jpg
|
|
2009
|
Mon Aug 31 20:32:57 2015 |
Alessandra | Misc | Seismometer | IP final parameters |
Final parameters estimate
Using the method described in the previous entry, I measured the spring constant ( ) of three inverted pendulums with different flexure dimentions. The flexures where all in 302/304 stainless steel and had a 3.2 cm length, while their diameter was different:
1) Diameter: 0.98 mm
: 0.295 Nm
2) Diameter: 1.40 mm
: 1.188 Nm
3) Diameter: 1.99 mm
: 1.507 Nm
Using those values for the sping constant I calculated the value needed for the top mass of the final IP to reach the desired 40 mHz resonant frequency.
In order to do this I used the resonant frequency expression in elog entry 2004 with . Let's see why.
If we have a single flexure structure of height and we apply a force F on top as in picture:

in the steady-state the torque equation is:

thus, using , we obtain:

If we have a double flexure symmetric structure of height where the top part can only roll (as in our final IP) and we apply the same force F as in picture:

In the steady-state the torque equation is:

and 
As I said in a previous entry, a double flexure three-legs IP with spring constant is equivalent to a double flexure one-leg IP with spring constant . As the sping constant for a double flexure symmetric IP is the double of the spring constant of a single flexure IP (wich is the one measured) we obtain that the sping constant of our IP is .
I obtained that the values for the top mass needed to reach a resonant frequency of 40 mHz in the three cases above are:
1) 
2) 
3)
The case 1 can be excluded because the rod is too narrow and buckling would certainly occur. The other two cases are acceptable.
A test to verify if this model is correct will be done on a prototype three-legs IP. |
2008
|
Thu Aug 27 15:09:12 2015 |
Alessandra | Misc | Seismometer | IP measurements |
Spring constant and resonant frequency
In order to verify my calculations for the IP resonant frequency and Comsol results some experimental test has been done at the 40m lab.
With the help of Koji I built an inverted pendulum as in picture below:

It is made of a single flexure in stainless steel of 3.2 cm length and 0.98 mm diameter, a pin-vise and a top alluminium mass.
In order to measure the resonant frequency of the IP we used a LED light and two photosensors. The LED light hit the alluminium mass as in picture and the reflected light was detected by the photosensors. We could read the photosensors output on an oscilloscope:

As we made the alluminum mass oscillate by appliyng a small force on the top mass, we could read the mass displacements on the oscilloscope and measure the resonant frequency.

Expected resonant frequency, calculated with an analytical model and the use of Comsol (as explained in a previous elog) was 3.58 Hz while measured was 2.56 Hz.
To test if there was an error in Comsol estimate of the IP spring constant ( ) I measured it. The set-up used is shown in picture below:

I measured the displacement of the top mass as it was subject only to the weight force and deduced . The value expected from Comsol was 0.504 Nm while measured was 0.263 Nm. So probably the material used on Comsol is not exactly the same stainless steel we're using. This means that previous models on Comsol can't be trusted.
Using the measured I calculated the resonant frequency and it resulted 2.50 Hz (measured 2.56 Hz), which is a good result, considering the approximations made in the analytical model to describe the IP structure.
Inverted pendulum Q factor
As we can see from the oscilloscope picture above, the IP is subjected to damping, which is mostly structural damping. In order to calculate this damping we can approximate our IP impulse response by:

the times at which the peak amplitude occurs are given by:

From these expressions we obtain that

Thus, measuring the amplitude of the oscillations peaks I could deduce Q, which results:

|
2007
|
Sat Aug 22 18:36:50 2015 |
Alessandra | Misc | Seismometer | Temperature stabilization |
Temperature stabilization
Yesterday some measurements where taken to see how the termal enclosure's temperature stabilzes.
I took measurements for decreasing temperature: Initial temperature was 23 C and target temperature was set to 22 C. Results are shown in this plot:

Koji took measurements for increasing temperature using the continuous data acquisition. Initial temperature was 22 C and target temperature was 23 C. Here are the results:

|
2006
|
Fri Aug 21 22:59:00 2015 |
Koji | Misc | Seismometer | Started continuous data taking |
This work was done at the 40m.
On Wednesday:
Alessandra and I jiggled the PID parameters in order to try some temperature stabilization. But the temperature kept going below the set temp and we were confused.
On Thursday (yesterday):
Alessandra and I succeeded to stabilize the box temperature. We used the PID setting of 250 250 0. Previously, the control gain was way too low and the final temp had significant deviation from the set temp. Now with the max I gain, the controller squishes the DC deviation better.
Today
In order to try to take the continuous temperature measurement, I picked up the Rapsberry Pi at the Yend.
It has the IP of 192.168.113.166 accoding to the log http://nodus.ligo.caltech.edu:8080/40m/8745
Initially I could not login to the host. After struggling with the HDMI cable and connected the display/keyboard/mouse to the host, I found that the user name is not "controls" but "pi". How can I know!
I could manage to change the password to the nominal 40m controls' one, and also created "controls" account by "adduser" command.
After played with Gautam's code a bit, I understood how to acess to the serial port.
Some tips
- python library "serial" is installed only for python3.
- In order to use the USB port, the user must be in "dialout" group
I made a little bit of coding to have constant but slow sampling rate. Now I started to log the temperature data with the rate of 10sec.
This program changes the set temp from 22degC to 23 degC and continuously take the data.
controls@raspberrypi ~ $ cat 150821_222610.log
1440221170 21.9 : 08/21/15 22:26:10
1440221180 21.9 : 08/21/15 22:26:20
1440221190 21.9 : 08/21/15 22:26:30
1440221200 21.9 : 08/21/15 22:26:40
1440221210 21.9 : 08/21/15 22:26:50
1440221220 21.9 : 08/21/15 22:27:00
1440221230 21.9 : 08/21/15 22:27:10
1440221240 21.9 : 08/21/15 22:27:20
1440221250 21.9 : 08/21/15 22:27:30
1440221260 21.9 : 08/21/15 22:27:40
1440221270 21.9 : 08/21/15 22:27:50
controls@raspberrypi ~ $ cat 150821_222802.log
1440221290 21.9 : 08/21/15 22:28:10
1440221300 21.9 : 08/21/15 22:28:20
1440221310 22.0 : 08/21/15 22:28:30
1440221320 22.0 : 08/21/15 22:28:40
1440221330 22.0 : 08/21/15 22:28:50
1440221340 22.0 : 08/21/15 22:29:00
1440221350 22.0 : 08/21/15 22:29:10
1440221360 22.1 : 08/21/15 22:29:20
1440221370 22.1 : 08/21/15 22:29:30
1440221380 22.1 : 08/21/15 22:29:40
1440221390 22.1 : 08/21/15 22:29:50
1440221400 22.1 : 08/21/15 22:30:00 |
Unfortunately this program stopped when I logged out. It was fixed by running the command via nohup (no hangup):
nohup ./TC200scan.py &
The previous attempt injected the heat to the system, I'll let the system cooled down for several hours and then run the step rise to log the data.
BTW: TC200 doesn't allow us to observe the data below 0.1 degC level. The other temp controller we brought from WB read the thermister resistance with the precision of 0.001 Ohm. |
2005
|
Wed Aug 19 19:14:03 2015 |
Alessandra | Misc | Seismometer | Thermal enclosure tests |
Thermal enclosure tests
Some tests have been done to see how the thermal enclosure behaves.
- The first was made on Monday by Megan and Koji.
Inside the thermal enclosure there are two heaters on opposite sides and they are connected to the temperature controller box.
In order to determine the behavior of the enclosure, the target temperature was set to 35 °C, and measurements of the temperature where taken starting from room temperature (20.1 °C) using a thermistor attached near to one of the heaters.
Results are shown in this plot:

The fit was made with:

and fit results are:

We see that the temperature of 35 °C is not reached. That is probably due to leakage: the equilibrium temperature is reached when the given heat and leakage heat balance.
- Yesterday I measured values of decreasing temperature starting from 24.3 °C. Results are shown in this plot:

The fit was made with:

and fit results are:

- Today we wanted to choose the target temperature at 23 °C, which is at the middle of the two equilibrium temperatures (25 °C with the heater on and 20 °C with the heater off), reach the linear control range, and test the closed loop response by changing the setpoint by 1 °C.
Unfortunately what happened is that in about four hours the temperature reached 23.8 °C and kept increasing. So we tried to use different approaches to reach a constant temperature of 23 °C.
- We tried to change the temperature controller settings from P=250; I=0; D=0 to P=50; I=0; D=0 and we set a target temperature of 23 °C. Starting from 23.4 °C, in about 20 min we reached 22.4 °C.
- We chose P=100; I=0; D=0 with target temperature 23.5 °C but, starting from 22.4 °C, a temperature of 22.3 °C was reached in about 5 min and didn't change anymore.
- We chose P=100; I=10; D=0 with target temperature 23 °C but, starting from 22.2 C, instead of increasing, temperature started dropping and became 22.1 °C.
|
2004
|
Thu Aug 13 18:55:27 2015 |
Alessandra | Misc | Seismometer | Parameters for the new shape of the IP |
Parameters for the new shape of the IP
The final shape for the IP is made of three legs having a block on top. To make the calculations simpler I considered a one-leg structure as following:

Sizes for the different parts composing the IP are:
- bottom cylinder and top cylinder:

- middle cylider:

- first block (width, depth, height):

- second block:

The three cylinders are made in Steel AISI 4340.1 (density: 7850 kg/m^3) while the two blocks are in aluminum (density: 2700 kg/m^3).
Resonant frequency
The resonant frequency for this structure is given by:
![f_{IP}=\frac{1}{2\pi}\sqrt{\frac{k}{I}-\frac{g}{I}[m_1(2l_1+l_2)+m_2(l_1+\frac{l_2}{2})+m_3(2l_1+l_2+\frac{c_3}{2})+m_4(2l_1+l_2+c_3+\frac{c_4}{2})]}](http://latex.codecogs.com/gif.latex?f_%7BIP%7D%3D%5Cfrac%7B1%7D%7B2%5Cpi%7D%5Csqrt%7B%5Cfrac%7Bk%7D%7BI%7D-%5Cfrac%7Bg%7D%7BI%7D%5Bm_1%282l_1+l_2%29+m_2%28l_1+%5Cfrac%7Bl_2%7D%7B2%7D%29+m_3%282l_1+l_2+%5Cfrac%7Bc_3%7D%7B2%7D%29+m_4%282l_1+l_2+c_3+%5Cfrac%7Bc_4%7D%7B2%7D%29%5D%7D)
Where is the spring constant and the moment of inertia of the structure.
Moment of inertia
depends on the axis of rotation chosen:
- For an axis perpendicular to and as in picture

we obtain:

- For an axis perpendicular to and as in picture

we obtain:

Spring constant
We are interested in calculating the resonant frequency for a three legs structure.
In order to do this we can think of a one-leg IP of lenght as a mass on a spring of constant . The relation between the IP spring constant, ,and is:

Now we can think of the three legs of our IP as three springs in parallel, which are equivalent to a single spring of constant and consequently the three legs-IP can be studied as an analogous one-leg IP of spring constant .
Using Comsol (as explained in a previous entry) I calculated and obtained:

Thus

Top mass values
- The following plot shows how
changes as increases for oscillations with respect to the axis perpendicular to and :

- This plot is related to the oscillations with respect to the axis perpendicular to
and :

From the two plots we can deduce that a resonant frequency of 40 mHz is reached for in both cases.
Anyway, using a fixed mass on top of the structure there will be a small difference between the resonant frequencies in the two directions of oscillation due to the difference between the two moments of inertia. |
2003
|
Thu Aug 13 13:12:48 2015 |
Rich Abbott | Electronics | FSS | TTFSS Fieldbox, Mysteries Revealed |
Link to a block diagram to aid in the setup of the PDH loops TTFSS_Fieldbox.pdf |
Attachment 1: TTFSS_Fieldbox.pdf
|
|
2002
|
Wed Aug 12 16:28:20 2015 |
Megan | Misc | Seismometer | Michelson layout diagram - Final (?) version |
Attached is my final (?) version of the layout for the seismometer's Michelson.
Features:
- All of the bases are Newport bases from Koji's most recent order: two 9914 bases, and five 9912 bases (three of which are machined).
- The machining of three of the 9912 bases is cutting off 1.25" of the length, and milling the slot in closer towards the mounting hole. Cutting off some of the length allows elements to be placed closer together, and milling the slot further inwards allows for two connection points to the rhomboid.
- The two bases on to the inverted pendulum (one is to hold an end mirror, and the other is a couterweight) can each be attached with 1/4-20's in four places.
- The rest of the bases (all on the rhomboid) can be attached with 1/4-20's in two places. Two connection points ensures that the optics can't rotate/cause noise. The exception is the base for the photodiode; it is fastened in just one place because it is in the breadboard's corner. I'm not sure if just one connection for the photodiode will be an issue; I will check with Koji how important it is to have more than one connection point.
- All of the mounts leave some room in the slots for movement of the optics. The beamsplitter's base and end mirror 2's base have just 1/8" leeway when moving away from each other; they are in close proximity so this limits their range of motion.
- The whole lower half of the rhomboid breadboard is empty, so there is plenty of room for any counterweights that may be needed.
- There is ~2" of space after the fiber coupler for a lens (if needed).
|
Attachment 1: michelson_layout_diagram_3.pdf
|
|
2001
|
Wed Aug 12 16:04:22 2015 |
Megan | Misc | Seismometer | Light through fiber + beam scans |
Yesterday (08/11/15) Koji helped me get some light through the fiber that will eventually bring light to the seismometer's Michelson. A fiber illuminator was attached on the output end of the fiber, and this produced a beam bright enough to see on the input end of the fiber. This beam was aligned on all of the optics between the fiber and the premode cleaner. The laser was then turned on, and the IR beam was then aligned from the PMC to the fiber. Both beams could be seen on the detector cards, so we were able to co-align the two beams through many small adjustments. Once the beams were co-aligned, the illuminator was taken off of the output end of the fiber, and the fiber was replaced into its mount. Then, using a power meter, we were able to take readings on either end of the fiber to see how much of the light was actually making it through the fiber. Before the fiber, the beam was at ~64mW, and after the fiber, the beam was at ~18mW, which is about 28% throughput.
After getting some light through the fiber, we set up the beam scan CCD in front of the output end of the fiber. A long rail was fastened to the table parallel to the beam, so that the CCD could slide in a straight path down the table (see Attachment 1).
After making sure the beam spot remained on the CDD's face over the whole length of the rail, the width of the beam was recorded at every inch interval down the length of the rail. Measuring the distance from the fiber in inches was convenient, because the holes in the table are placed at one-inch intervals. The value shown by the beam scan software is the Gaussian diameter of the beam (the point at which the beam intensity is ~13.5% of its peak value), so when plotting the data the recorded value was divided by 2 to give the Gaussian radius. Attachment 2 is the plot of the beam profile, beam radius as a function of distance from the fiber's collimating lens. The plot shows the clear dispersion of the beam as it gets further away from the beam.
The next beam scan that was done was after a lens that is upstream from the input of the fiber. The setup to do this scan was similar, but I used a ruler rather than the larger rail used in the previous scan (see Attachment 3).
Attachment 4 shows this beam's profile.
This part of the beam may need to be scanned further out than just 12", and that can be done next. Another task is to measure the Y values of the beam as well. The next step for the exisitng data is to fit a Gaussian to them, which will give the size of the waist and the position of the waist. These values can then be used in a la mode to find the optimal lens placement to maximize the amount of light that makes it through the fiber. |
Attachment 1: IMG_5463.jpg
|
|
Attachment 2: beam_scan_1.pdf
|
|
Attachment 3: IMG_5464.jpg
|
|
Attachment 4: beam_scan_2.pdf
|
|
2000
|
Wed Aug 12 14:04:47 2015 |
Alessandra | Misc | Seismometer | Breadboard and IP top mass dimentions |
Here are the breadboard and IP top mass dimentions resulting from measurements on the objects.
The objects are presented in a top view and the bottom part is the thickness. The "o" sign on top left is actually present on the objects to distinguish their dimentions. |
Attachment 1: breadboard.ps
|
|
Attachment 2: ip_mass.ps
|
|