Thu Oct 21 15:32:32 2021, Aidan, Radhika, Laser, 2um Photodiodes, Improved measurement of QE on photodiodes ~89% at 140K
|
Thu Oct 21 16:52:26 2021, rana, Laser, 2um Photodiodes, Improved measurement of QE on photodiodes ~89% at 140K
|
Thu Oct 21 17:48:33 2021, Aidan, Laser, 2um Photodiodes, Improved measurement of QE on photodiodes ~89% at 140K
|
Fri Oct 22 09:20:13 2021, Aidan, Laser, 2um Photodiodes, Improved measurement of QE on photodiodes ~89% at 140K
|
Fri Oct 22 12:04:13 2021, rana, Laser, 2um Photodiodes, Improved measurement of QE on photodiodes ~89% at 140K
|
Fri Oct 22 13:31:36 2021, Chris, Laser, 2um Photodiodes, Improved measurement of QE on photodiodes ~89% at 140K
|
|
Message ID: 2683
Entry time: Fri Oct 22 09:20:13 2021
In reply to: 2682
|
Author: |
Aidan |
Type: |
Laser |
Category: |
2um Photodiodes |
Subject: |
Improved measurement of QE on photodiodes ~89% at 140K |
|
|
Some of the data recorded during the current/micrometer scanning yesterday.
- Distance between fiber/lens housings = Micrometer + 22.9mm
- QE = Response / (1000Ohm*9.3E-4W/V) *(h*c/[e*L]) = Response * 0.667
Highlighted change to 25mA and also highest QE.
Time |
Temperature (K) |
Laser current (mA) |
Micrometer (mm) |
Peak Response (JPL/REF) |
QE |
2:07PM |
131.4 |
30 |
9.10 |
1.070 |
71.0% |
2:14PM |
133.5 |
30 |
9.10 |
1.090 |
73.0% |
2:20PM |
134.8 |
30 |
9.00 |
1.100 |
75.0% |
2:25PM |
136.4 |
30 |
8.80 |
1.134 |
75.6% |
2:29PM |
137.6 |
30 |
8.70 |
1.140 |
76.0% |
2:31PM |
138.1 |
30 |
8.60 |
1.120 |
74.7% |
2:35PM |
138.8 |
25 |
8.60 |
1.288 |
85.9% |
2:42PM |
140.5 |
25 |
8.80 |
1.246 |
83.1% |
2:45PM |
|
25 |
8.60 |
1.285 |
85.7% |
2:249PM |
142.2 |
25 |
8.40 |
1.310 |
87.4% |
2:53PM |
143 |
25 |
8.20 |
1.322 |
88.0% |
3:01PM |
144 |
25 |
8.00 |
1.328 |
88.6% |
3:10PM |
|
25 |
8.00 |
1.334 |
89.0% |
3:14PM |
147.5 |
25 |
7.80 |
1.314 |
87.6% |
3:16PM |
148.05 |
25 |
7.60 |
1.316 |
87.8% |
3:19PM |
|
25 |
8.00 |
1.328 |
88.6% |
3:23PM |
149.3 |
25 |
8.00 |
1.320 |
88.0% |
3:30PM |
151 |
25 |
8.00 |
1.315 |
87.7% |
3:47PM |
154 |
25 |
8.00 |
1.315 |
87.7% |
4:02PM |
157 |
25 |
8.00 |
1.300 |
86.7% |
4:14PM |
159.6 |
25 |
8.00 |
1.305 |
87.0% |
|