Entry  Thu Feb 18 08:12:53 2016, KojiN, Misc, PD QE, Re-measurement of the reflectivities and the EQEs of the PD for S-pol and P-pol QE3.pdfQE_ll.pdf
    Reply  Thu Feb 18 13:39:33 2016, Koji, Misc, PD QE, Re-measurement of the reflectivities and the EQEs of the PD for S-pol and P-pol 
       Reply  Fri Feb 19 00:01:36 2016, KojiN, Misc, PD QE, Re-measurement of the reflectivities and the EQEs of the PD for S-pol and P-pol 6x
Message ID: 2022     Entry time: Fri Feb 19 00:01:36 2016     In reply to: 2021
Author: KojiN 
Type: Misc 
Category: PD QE 
Subject: Re-measurement of the reflectivities and the EQEs of the PD for S-pol and P-pol 

I calculated IQEs using the measured EQEs and reflectivities and the following formula:

{\rm IQE} = \frac{{\rm EQE}}{1-R-T},

where R is the reflectivity of the PD, and T is the transmittance of the PD.

Here T is assumed to be zero and the scattering loss is ignored as you said.

The obtained IQEs are shown in Fig. 1.

 

Also the EQE, IQE, and reflectivity at a few angles near the angle where the EQE goes maximum for the P pol were measured as shown in Fig. 2 and Fig. 3.

The angle where the EQE goes maximum for the P pol is -51 deg.

So far, in these plots, the systematic errors because of the power meter are not considered.

 

And I replaced the figures with the figures having larger fonts.

Fig. 1 Measured EQEs and IQEs.
Fig. 2 Measured reflectivities around -51 deg.
Fig. 3 Measured EQEs and IQEs around -51 deg.

QE_Ref_P.txt QE_Ref_S.txt

AroundPeak.txt

Attachment 1: IQE.pdf  9 kB  | Show | Show all
Attachment 2: Ref.pdf  3 kB  | Show | Show all
Attachment 3: QE_Ref_P.txt  821 Bytes  | Show | Show all
Attachment 4: QE_Ref_S.txt  821 Bytes  | Show | Show all
Attachment 5: AroundPeak.txt  463 Bytes  | Show | Show all
Attachment 6: QE2.pdf  4 kB  Uploaded Fri Feb 19 09:45:15 2016  | Show | Show all