  60   Wed Feb 6 02:34:10 2013 KojiOpticsCharacterizationWedge measurement with the autocollimator and the rotation stage

Measurement:

  • A6:   α = 0.665 deg, β = +3.0 arcmin (1.5 div up)
  • A7:   α = 0.635 deg, β =   0.0 arcmin (0.0 div up)
  • A8:   α = 0.623 deg, β = - 0.4 arcmin (-0.2 div up)
  • A9:   α = 0.670 deg, β = +2.4 arcmin (1.2 div up)
  • A10: α = 0.605 deg, β = +0.4 arcmin (0.2 div up)
  • A11: α = 0.640 deg, β = +0.8 arcmin (0.4 div up)
  • A12: α = 0.625 deg, β = - 0.6 arcmin (-0.3 div up)
  • A13: α = 0.630 deg, β = +2.2 arcmin (1.1 div up)
  • A14: α = 0.678 deg, β =   0.0 arcmin (0.0 div up)
  • B1:   α = 0.665 deg, β = +0.6 arcmin (0.3 div up)
  • B2:   α = 0.615 deg, β = +0.2 arcmin (0.1 div up)
  • B3:   α = 0.620 deg, β = +0.9 arcmin (0.45 div up)
  • B4:   α = 0.595 deg, β = +2.4 arcmin (1.2 div up)
  • B5:   α = 0.635 deg, β = - 1.8 arcmin (-0.9 div up)
  • B6:   α = 0.640 deg, β = +1.6 arcmin (0.8 div up)
  • B7:   α = 0.655 deg, β = +2.5 arcmin (1.25 div up)
  • B8:   α = 0.630 deg, β = +2.8 arcmin (1.4 div up)
  • B9:   α = 0.620 deg, β = - 4.0 arcmin (-2.0 div up)
  • B10: α = 0.620 deg, β = +1.2 arcmin (0.6 div up)
  • B11: α = 0.675 deg, β = +3.5 arcmin (1.75 div up)
  • B12: α = 0.640 deg, β = +0.2 arcmin (0.1 div up)

Analysis:

  • \theta_H = ArcSin[Sin(α) * n]
  • \theta_V = ArcSin[Sin(β) / n]/2
     
  • A6:   \theta_H = 0.490 deg, \theta_V =  0.017 deg
  • A7:   \theta_H = 0.534 deg, \theta_V =  0.000 deg
  • A8:   \theta_H = 0.551 deg, \theta_V = -0.0023 deg
  • A9:   \theta_H = 0.482 deg, \theta_V =  0.014 deg
  • A10: \theta_H = 0.577 deg, \theta_V =  0.0023 deg
  • A11: \theta_H = 0.526 deg, \theta_V =  0.0046 deg
  • A12: \theta_H = 0.548 deg, \theta_V = -0.0034 deg
  • A13: \theta_H = 0.541 deg, \theta_V =  0.013 deg
  • A14: \theta_H = 0.471 deg, \theta_V =  0.000 deg
  • B1:   \theta_H = 0.490 deg, \theta_V =  0.0034 deg
  • B2:   \theta_H = 0.563 deg, \theta_V =  0.0011 deg
  • B3:   \theta_H = 0.556 deg, \theta_V =  0.0051 deg
  • B4:   \theta_H = 0.592 deg, \theta_V =  0.014 deg
  • B5:   \theta_H = 0.534 deg, \theta_V = -0.010 deg
  • B6:   \theta_H = 0.526 deg, \theta_V =  0.0091 deg
  • B7:   \theta_H = 0.504 deg, \theta_V =  0.014 deg
  • B8:   \theta_H = 0.541 deg, \theta_V =  0.016 deg
  • B9:   \theta_H = 0.556 deg, \theta_V = -0.023 deg
  • B10: \theta_H = 0.556 deg, \theta_V =  0.0068 deg
  • B11: \theta_H = 0.475 deg, \theta_V =  0.020 deg
  • B12: \theta_H = 0.526 deg, \theta_V =  0.0011 deg

Quote:

Measurement:

  • A1: α = 0.68 deg, β = 0 arcmin (0 div)
  • A2: α = 0.80 deg, β = -6 arcmin (3 div down)
  • A3: α = 0.635 deg, β = -1.6 arcmin (0.8 div down)
  • A4: α = 0.650 deg, β = 0 arcmin (0div)
  • A5: α = 0.655 deg, β = +2.4 arcmin (1.2 div up)

Analysis:

  • \theta_H = ArcSin[Sin(α)*n]
  • \theta_V = ArcSin[Sin(β) / n]/2
     
  • A1: \theta_H = 0.465 deg, \theta_V = 0.000 deg
  • A2: \theta_H = 0.547 deg, \theta_V = -0.034 deg
  • A3: \theta_H = 0.434 deg, \theta_V = -0.009 deg
  • A4: \theta_H = 0.445 deg, \theta_V = 0.000 deg
  • A5: \theta_H = 0.448 deg, \theta_V = 0.014 deg

 

  59   Mon Feb 4 00:39:08 2013 KojiOpticsCharacterizationWedge measurement with the autocollimator and the rotation stage

Method:

  • Mount the tombstone prism on the prism mount. The mount is fixed on the rotation stage.
  • Locate the prism in front of the autocollimator.
  • Find the retroreflected reticle in the view. Adjust the focus if necessary.
  • Confirm that the rotation of the stage does not change the height of the reticle in the view. 

    If it does, rotate the AC around its axis to realize it.
    This is to match the horizontal reticle to the rotation plane.
  • Use the rotation stage and the alignment knobs to find the reticle at the center of the AC.
Make sure the reticle corresponds to the front surface.
  • Record the micrometer reading.
  • Rotate the micrometer of the rotation stage until the retroreflected reticle for the back surface.
  • There maybe the vertical shift of the reticle due to the vertical wedging. Record the vertical shi
  • Record the micrometer reading. Take a difference from the previous value.
     

Measurement:

  • A1: α = 0.68 deg, β = 0 arcmin (0 div)
  • A2: α = 0.80 deg, β = -6 arcmin (3 div down)
  • A3: α = 0.635 deg, β = -1.6 arcmin (0.8 div down)
  • A4: α = 0.650 deg, β = 0 arcmin (0div)
  • A5: α = 0.655 deg, β = +2.4 arcmin (1.2 div up)

Analysis:

  • \theta_H = ArcSin[Sin(α) / n]
  • \theta_V = ArcSin[Sin(β) / n]/2
     
  • A1: \theta_H = 0.465 deg, \theta_V = 0.000 deg
  • A2: \theta_H = 0.547 deg, \theta_V = -0.034 deg
  • A3: \theta_H = 0.434 deg, \theta_V = -0.009 deg
  • A4: \theta_H = 0.445 deg, \theta_V = 0.000 deg
  • A5: \theta_H = 0.448 deg, \theta_V = 0.014 deg

 

Attachment 1: autocollimator_wedge_measurement.pdf
autocollimator_wedge_measurement.pdf
  58   Tue Jan 22 17:56:32 2013 KojiMechanicsGeneralRotary stage selection

Newport UTR80

Newport 481-A (SELECTED)

  • Sensitivity: 15 arcsec
  • Graduations: 1 deg
  • Vernier: 5 arcmin
  • Fine travel range: 5 deg
  • With Micrometer

Newport RS40

  • Sensitivity: 16 arcsec
  • Graduations: 2 deg
  • Vernier: 12 arcmin
  • Fine travel range: 10 deg
  • Micrometer BM11.5

Newport RS65

  • Sensitivity: 11 arcsec
  • Graduations: 2 deg
  • Vernier: 12 arcmin
  • Fine travel range: 10 deg
  • Micrometer SM-06 to be bought separately

Elliot science MDE282-20G

  • Sensitivity: 5 arcsec
  • Graduations: 2 deg
  • Vernier: 10 arcmin
  • Fine travel range: 10 deg
  • Micrometer 2 arcmin/1div
  • Metric

Suruga precision B43-110N

Thorlabs precision B43-110N

  57   Tue Jan 22 11:10:25 2013 KojiFacilityGeneralEyeware storage and hooks for the face shields are installed

A carpenter has come to install the eyeware storage and hooks for the face shields.

Attachment 1: P1223116.JPG
P1223116.JPG
  56   Sat Jan 19 20:47:41 2013 KojiOpticsCharacterizationWedge measurement with the autocollimator

The wedge angle of the prism "A1" was measured with the autocollimator (AC).

The range of the AC is 40 arcmin. This means that the mirror tilt of 40arcmin can be measured with this AC.
This is just barely enough to detect the front side reflection and the back side reflection.

The measured wedge angle of the A1 prism was 0.478 deg.

Ideally a null measurement should be done with a rotation stage.

Attachment 1: autocollimator_wedge_measurement.pdf
autocollimator_wedge_measurement.pdf
  55   Fri Jan 18 13:25:17 2013 KojiOpticsConfigurationAutocollimator calibration

An autocollimator (AC) should show (0,0) if a retroreflector is placed in front of the AC.
However, the AC may have an offset. Also the retroreflector may not reflect the beam back with an exact parallelism.

To calibrate these two errors, the autocollimator is calibrated. The retroreflector was rotated by 0, 90, 180, 270 deg
while the reticle position are monitored. The images of the autocollimator were taken by my digital camera looking at the eyepiece of the AC.

Note that 1 div of the AC image corresponds to 1arcmin.

Basically the rotation of the retroreflector changed the vertical and horizontal positions of the reticle pattern by 0.6mdeg and 0.1mdeg
(2 and 0.4 arcsec). Therefore the parallelism of the retrorefrector is determined to be less than an arcsec. This is negligibly good for our purpose.

The offset changes by ~1div in a slanted direction if the knob of the AC, whose function is unknown, is touched.
So the knob should be locked, and the offset should be recorded before we start the actual work every time.

Attachment 1: autocollimator_calibration.pdf
autocollimator_calibration.pdf autocollimator_calibration.pdf autocollimator_calibration.pdf autocollimator_calibration.pdf
  54   Wed Jan 16 14:10:50 2013 KojiOpticsCharacterizationAutocollimator tests of optics perpendicularity/parallelism

The items:

- Autocollimator (AC) borrowed from Mike Smith (Nippon Kogaku model 305, phi=2.76", 67.8mm)

- Retroreflector (corner cube)

- Two V grooves borrowed from the 40m

Procedure:

- Autocollimator calibration

o Install the AC on a optical table

o Locate the corner cube in front of the AC.

o Adjust the focus of the AC so that the reflected reticle pattern can be seen.

o If the retroreflection and the AC are perfect, the reference reticle pattern will match with the reflected reticle pattern.

o Measure the deviation of the reflected reticle from the center.

o Rotate the retroreflector by 90 deg. Measure the deviation again.

o Repeat the process until total four coordinates are obtained.

o Analysis of the data separates two types of the error:
   The average of these four coordinates gives the systematic error of the AC itself.
   The vector from the center of the circle corresponds to the error of the retroreflector.

- Wedge angle measurement

 

To be continued

  53   Thu Jan 10 18:37:50 2013 KojiOpticsCharacterizationWedging of the PZTs

Yesterday I measured the thickness of the PZTs in order to get an idea how much the PZTs are wedged.

For each PZT, the thickness at six points along the ring was measured with a micrometer gauge.
The orientation of the PZT was recognized by the wire direction and a black marking to indicate the polarity.

A least square fitting of these six points determines the most likely PZT plane.
Note that the measured numbers are assumed to be the thickness at the inner rim of the ring
as the micrometer can only measure the maximum thickness of a region and the inner rim has the largest effect on the wedge angle.
The inner diameter of the ring is 9mm.



The measurements show all PZTs have thickness variation of 3um maximum.

The estimated wedge angles are distributed from 8 to 26 arcsec. The directions of the wedges seem to be random
(i.e. not associated with the wires)



As wedging of 30 arcsec causes at most ~0.3mm spot shift of the cavity (easy to remember),
the wedging of the PZTs is not critical by itself. Also, this number can be reduced by choosing the PZT orientations
based on the estimated wedge directions --- as long as we can believe the measurements.



Next step is to locate the minima of each curved mirror. Do you have any idea how to measure them?

Attachment 1: PZT_wedging.pdf
PZT_wedging.pdf
  52   Sun Jan 6 23:22:21 2013 KojiMechanicsGeneralSolidWorks model of the OMC suspension

D0900295_AdvLIGO_SUS_Output_Mode_Cleaner_Overall_Assembly.png

Attachment 2: D0900295_AdvLIGO_SUS_Output_Mode_Cleaner_Overall_Assembly.easm
  51   Wed Jan 2 07:45:39 2013 KojiOpticsCharacterizationFirst Contact test

Conclusion: Good. First contact did not damage the coating surface, and reduced the loss


- Construct a cavity with A1 and C2

- Measure the transmission and FWHM (of TEM10 mode)

- Apply First Contact on both mirrors

- Measure the values again


Transmission:

2.66 +/- 0.01 V -> 2.83  +/- 0.01 V

==> 6.3% +/- 0.5 % increase

FWHM of TEM10:

Before: (66.1067, 65.4257, 66.1746) +/- (0.40178, 0.38366, 0.47213) [kHz]
After: (60.846, 63.4461, 63.7906) +/- (0.43905, 0.56538, 0.51756) [kHz]

==> 5.1% +/- 2.7% decrease

Question: What is the best way to measure the finesse of the cavity?

Attachment 1: Cav_scan_response_zoom_20121125_C2_before.pdf
Cav_scan_response_zoom_20121125_C2_before.pdf
Attachment 2: Cav_scan_response_zoom_20121125_C2_after.pdf
Cav_scan_response_zoom_20121125_C2_after.pdf
  50   Wed Jan 2 07:35:55 2013 KojiOpticsCharacterizationThickness of a curved mirror

Measured the thickness of a curved mirror:

Took three points separated by 120 degree.

S/N: C2, (0.2478, 0.2477, 0.2477) in inch => (6.294, 6.292, 6.292) in mm

  49   Mon Dec 31 03:11:45 2012 KojiOpticsCharacterizationFurther more RoC measurement

Total (excluding C2, C7, C8): 2.575 +/- 0.005 [m]

New results

C6: RoC: 2.57321 +/− 4.2e-05m

C7: RoC: 2.56244 +/− 4.0e−05m ==> Polaris mount

C8: RoC: 2.56291 +/− 4.7e-05m ==> Ultima mount

C9: RoC: 2.57051 +/− 6.7e-05m

Previous results

C1: RoC: 2.57845 +/− 4.2e−05m

C2: RoC: 2.54363 +/− 4.9e−05m ==> Josh Smith @Fullerton for scattering measurement

C3: RoC: 2.57130 +/− 6.3e−05m   

C4: RoC: 2.58176 +/− 6.8e−05m

C5: RoC 2.57369 +/− 9.1e−05m

Attachment 1: Cav_scan_response_zoom_20121125_C6_9.pdf
Cav_scan_response_zoom_20121125_C6_9.pdf Cav_scan_response_zoom_20121125_C6_9.pdf Cav_scan_response_zoom_20121125_C6_9.pdf Cav_scan_response_zoom_20121125_C6_9.pdf
  48   Mon Dec 31 03:10:09 2012 KojiOpticsGeneralSolidWorks model of the OMC breadboard
Attachment 1: D1201439_aLIGO_Breadboard_layout_assy_121224.png
D1201439_aLIGO_Breadboard_layout_assy_121224.png
Attachment 2: D1201439_aLIGO_Breadboard_layout_assy_130105.easm
  47   Mon Dec 31 01:45:04 2012 KojiGeneralGeneralWork completed in Nov and Dec [!]

Completed work of the previous months: [Jul] [Aug] [Sep] [Oct] [Nov] [Dec]


  • Work done
  • Things delivered
    • The ionized gun used in the clean room at Downs: made by Terra Universal.com (Jeff's room)
      http://www.terrauniversal.com/static-control/ionizing-blow-off-guns.php
    • Flow path: N2 cylinder - Filter - Gun (Jeff's room)
    • Power strips Tripp Lite PS3612 (Ordered Nov. 8, Delivered Nov. 12)
    • Kapton tapes (1in x 6, 1/2in x 12 Delivered Nov. 15)
    • Sticky Mats (VWR 18888-216 Delivered Nov. 12 and 21992-042)
    • Duck tape (PK3) (Delivered Nov. 12)
    • Wipers 12"x12" 2ply x 119 pairs x case15 (Delivered Nov. 12)
    • Syringes (1mL&2mL) & Needles (20G x dozen)
    • Stainless trays with cover (Steve Delivered Nov. 12)
    • Gold Plated allen keys (Steve Delivered Nov. 12)
    • Forceps (Delivered Nov. 12) / Tweezers / Scissors (Delivered Nov. 12)
    • OMC testing optics / opto-mechanics
    • SolidWorks raytracing model
    • Mode design for HAM6 layout [Zach]
       
    • Black Glass / Black Glass holder / AR ==> Some at the 40m, some from LLO
    • Ionized air blow
      • N2 or Air cylinder: 4N - UHP or 5N - Research Grade.  (... steal from Downs)
  46   Wed Dec 26 14:33:33 2012 KojiFacilityGeneralLase Interlock wired

Two switches are connected in series.

Attachment 1: PC263073.jpg
PC263073.jpg
Attachment 2: PC263074.jpg
PC263074.jpg
Attachment 3: PC263075.jpg
PC263075.jpg
  45   Wed Dec 19 18:47:03 2012 KojiCleanGeneralFirst Contact Training with Margot

Steve and I visited Margot to have a training session for application of First Contact on optics.

- Make "thick" layer of first contact. It becomes thin when it gets dried.

- Apply more FC once a peek sheet is placed on the FC

- Wait for drying (~15min)

- Rip off the FC layer by pulling a peek tab. Make sure the ionized N2 is applied during ripping.

- Margot has a Dark Field Microscope. We checked how the dusts are removed from the surface.
There are many dusts on the mirror even if they are invisible. First Contact actually removes
these dusts very efficiently. Margot told us that even carbonhydrates (like finger prints) can be removed by FC.

  44   Tue Dec 18 20:04:40 2012 KojiOpticsCharacterizationPrism Thickness Measurement

The thicknesses of the prism mirrors (A1-A5) were measured with micrometer thickness gauge.
Since the thickness of the thinner side (side1) depends on the depth used for the measurement,
it is not accurate. Unit in mm.

A1: Side1: 9.916, Side2: 10.066 => derived wedge angle: 0.43deg
A2: Side1: 9.883, Side2: 10.065 => 0.52
A3: Side1: 9.932, Side2: 10.062 => 0.38
A4: Side1: 9.919, Side2: 10.060 => 0.40
A5: Side1: 9.917, Side2: 10.058 => 0.40

prism.png

  43   Thu Nov 29 21:18:23 2012 KojiOpticsGeneralOMC Mounting Prisms have come

PB293030.JPG

PB293032.JPG

  42   Mon Nov 26 01:40:00 2012 KojiOpticsCharacterizationMore RoC measurement

C1: RoC: 2.57845 +/− 4.2e−05m

C2: RoC: 2.54363 +/− 4.9e−05m

C3: RoC: 2.57130 +/− 6.3e−05m   

C4: RoC: 2.58176 +/− 6.8e−05m

C5: RoC 2.57369 +/− 9.1e−05m

 

==> 2.576 +/- 0.005 [m] (C2 excluded)

Attachment 1: RoC_measurement.pdf
RoC_measurement.pdf RoC_measurement.pdf RoC_measurement.pdf RoC_measurement.pdf RoC_measurement.pdf
  41   Mon Nov 19 13:33:14 2012 KojiOpticsCharacterizationResuming testing mirror RoCs

In order to resume testing the curvatures of the mirrors, the same mirror as the previous one was tested.
The result looks consistent with the previous measurement.

It seems that there has been some locking offset. Actually, the split peaks in the TF@83MHz indicates
the existence of the offset. Next time, it should be adjusted at the beginning.

Curved mirror SN: C1
RoC: 2.5785 +/- 0.000042 [m]

Previous measurements
=> 2.5830, 2.5638 => sqrt(RoC1*RoC2) = 2.5734 m
=> 2.5844, 2.5666 => sqrt(RoC1*RoC2) = 2.5755 m

Attachment 1: Cav_scan_response_zoom_20121016.pdf
Cav_scan_response_zoom_20121016.pdf
  40   Sat Nov 17 02:31:34 2012 KojiOpticsCharacterizationMirror T test

Mirror T test

The mirror was misaligned to have ~2deg incident (mistakenly...) angle.

C1: Ptrans = 7.58uW, Pinc = 135.0mW => 56.1ppm

C1 (take2): Ptrans = 7.30uW, Pinc = 134.4mW => 54.3ppm

C2: Ptrans = 6.91uW, Pinc = 137.3mW => 50.3ppm

C3: Ptrans = 6.27uW, Pinc = 139.7mW => 44.9ppm

C4: Ptrans = 7.62uW, Pinc = 139.3mW => 54.7ppm

C5: Ptrans = 6.20uW, Pinc = 137.5mW => 45.1ppm

A1: Ptrans = 1.094mW, Pinc = 133.6mW => 8189ppm

  39   Fri Nov 9 00:43:32 2012 KojiOpticsCharacterizationFurther more wedge measurement

Now it's enough for the first OMC (or even second one too).
Today's measurements all distributed in theta>0.5deg. Is this some systematic effect???
I should check some of the compeled mirrors again to see the reproducibility...

A1    Horiz Wedge    0.497039    +/-    0.00420005    deg / Vert Wedge     0.02405210    +/-    0.00420061    deg

A2    Horiz Wedge    0.548849    +/-    0.00419993    deg / Vert Wedge     0.05087730    +/-    0.00420061    deg
A3    Horiz Wedge    0.463261    +/-    0.00420013    deg / Vert Wedge     0.00874441    +/-    0.00420061    deg
A4    Horiz Wedge    0.471536    +/-    0.00420011    deg / Vert Wedge     0.01900840    +/-    0.00420061    deg
A5    Horiz Wedge    0.458305    +/-    0.00420014    deg / Vert Wedge     0.00628961    +/-    0.00420062    deg

B1    Horiz Wedge    0.568260    +/-    0.00419988    deg / Vert Wedge    -0.00442885    +/-    0.00420062    deg
B2    Horiz Wedge    0.556195    +/-    0.00419991    deg / Vert Wedge    -0.00136749    +/-    0.00420062    deg
B3    Horiz Wedge    0.571045    +/-    0.00419987    deg / Vert Wedge     0.00897185    +/-    0.00420061    deg
B4    Horiz Wedge    0.563724    +/-    0.00419989    deg / Vert Wedge    -0.01139000    +/-    0.00420061    deg
B5    Horiz Wedge    0.574745    +/-    0.00419986    deg / Vert Wedge     0.01718030    +/-    0.00420061    deg
E1    Horiz Wedge    0.600147    +/-    0.00419980    deg / Vert Wedge     0.00317778    +/-    0.00420062    deg
E2    Horiz Wedge    0.582597    +/-    0.00419984    deg / Vert Wedge    -0.00537131    +/-    0.00420062    deg
E3    Horiz Wedge    0.592933    +/-    0.00419982    deg / Vert Wedge    -0.01082830    +/-    0.00420061    deg

-------

To check the systematic effect, A1 and B1 were tested with different alignment setup.

A1    Horiz Wedge    0.547056    +/-    0.00419994    deg / Vert Wedge    0.0517442    +/-    0.00420061    deg
A1    Horiz Wedge    0.546993    +/-    0.00419994    deg / Vert Wedge    0.0469938    +/-    0.00420061    deg
A1    Horiz Wedge    0.509079    +/-    0.00420003    deg / Vert Wedge    0.0240255    +/-    0.00420061    deg

B1    Horiz Wedge    0.547139    +/-    0.00419994    deg / Vert Wedge    0.0191204    +/-    0.00420061    deg



 

Attachment 1: wedge_measurement_overall.pdf
wedge_measurement_overall.pdf wedge_measurement_overall.pdf wedge_measurement_overall.pdf wedge_measurement_overall.pdf wedge_measurement_overall.pdf wedge_measurement_overall.pdf wedge_measurement_overall.pdf wedge_measurement_overall.pdf
Attachment 2: 121108a_A1.pdf
121108a_A1.pdf 121108a_A1.pdf 121108a_A1.pdf 121108a_A1.pdf
  38   Thu Nov 8 20:12:10 2012 KojiOpticsConfigurationHow many glass components we need for a plate

Optical prisms 50pcs (A14+B12+C6+E18)
Curved Mirrors 25pcs (C13+D12)

  Qty

Prisms

Curved No BS OMC Wedge tested
Coating A: IO coupler   14 0  2 prisms 5/5
Coating B: BS 45deg   12 0  2 prisms  0/5
Coating C: HR   6 13 2 curved  
Coating D: Asym. output coupler   0 12 -  
Coating E: HR 45deg   18 0  4 prism (1 trans + 3 refl) 0/3
D1102209 Wire Mount Bracket 25      4  
D1102211 PD Mount Bracket 30      8  

 

  37   Thu Nov 8 19:52:57 2012 KojiOpticsGeneralHow to apply UV epoxy

KA's question:

Do you know how to apply this epoxy?
Do we need a plunger and a needle for this purpose?

Nic saids:

When we did it with Sam, I seem to remember just squirting some on some foil then dabbing it on with the needle.

Attachment 1: UVepoxy.jpg
UVepoxy.jpg
  36   Thu Nov 8 19:47:55 2012 KojiElectronicsConfigurationSolder for PZTs

Rich saids:

I have ordered a small roll of solder for the OMC piezos. 
The alloy is: Sn96.5 Ag3.0 Cu0.5

  35   Thu Nov 8 13:24:53 2012 KojiOpticsCharacterizationMore wedge measurement

A1
Horiz Wedge    0.497    +/-    0.004 deg
Vert Wedge      0.024    +/-    0.004 deg

A2
Horiz Wedge    0.549    +/-    0.004 deg
Vert Wedge      0.051    +/-    0.004 deg

A3
Horiz Wedge    0.463    +/-    0.004 deg
Vert Wedge      0.009    +/-    0.004 deg

A4
Horiz Wedge    0.471    +/-    0.004 deg
Vert Wedge      0.019    +/-    0.004 deg

A5
Horiz Wedge    0.458    +/-    0.004 deg
Vert Wedge      0.006    +/-    0.004 deg

Attachment 1: wedge_measurement_overall.pdf
wedge_measurement_overall.pdf wedge_measurement_overall.pdf wedge_measurement_overall.pdf wedge_measurement_overall.pdf wedge_measurement_overall.pdf
  34   Wed Nov 7 20:44:11 2012 KojiGeneralGeneralPlan Update: November [!]

Completed work of the previous months: [Jul] [Aug] [Sep] [Oct] [Nov] [Dec]


  • Work done
    • Wedge measurement (1st trial) [ELOG]
    • RoC measurement [ELOG]
  • Work in progress
    • R&T measurement
    • Wedge measurement
  • Work to be done
    • QPD/PD pre-selections (QE/noise)
    •  
    •  
    •  
    • Misc. / Beaurocracy?
      • Continuous monitoring of the particle level
      • Replacing a file cabinet next to the south wall by a lockable cabinet
      • Ion gun safety issues: https://dcc.ligo.org/cgi-bin/private/DocDB/ShowDocument?docid=88631
      • Laser SOP / HV use? / UV?
  • Things delivered
  • Things ordered
    • Power strips Tripp Lite PS3612 (Ordered Nov. 8, Delivered Nov. 12)
    • Kapton tapes (1in x 6, 1/2in x 12 Delivered Nov. 15)
    • Sticky Mats (VWR 18888-216 Delivered Nov. 12 and 21992-042)
    • Duck tape (PK3) (Delivered Nov. 12)
    • Wipers 12"x12" 2ply x 119 pairs x case15 (Delivered Nov. 12)
    • Syringes (1mL&2mL) & Needles (20G x dozen)
    • Stainless trays with cover (Steve Delivered Nov. 12)
    • Gold Plated allen keys (Steve Delivered Nov. 12)
    • Forceps (Delivered Nov. 12) / Tweezers / Scissors (Delivered Nov. 12)
  • Things to buy / get
    • OMC testing optics / opto-mechanics
    • Black Glass / Black Glass holder / AR ==> Some at the 40m, some from LLO
    • Ionized air blow
      • N2 or Air cylinder: 4N - UHP or 5N - Research Grade.  (... steal from Downs)
    • Clean tools, tray, storage
    • Supply
      • Additional clean supplies ~ glove 8.5,9,9.5
      • Stainless bats / Pure solvents (Metha / Aceton / Iso) / Syringes / Lint free cloth / Paper lens tissue
      • Lab coats
    • ATF
      • Tefron tape
      • Thorlabs 8-32 screw kit / Thorlabs HW-KIT1
      • Pedestal Shims - Newport
  • Things to be done
    • Cavity ref/trans/finesse
    • PD Q.E. & Reflectivity measurement vs incident angle
    • Functionality test of QPD/PD (PeterK) /PZT
       
  • Procedures to be decided
    • PZT alignment
    • UV glue? (heat) / gluing test
    • Balance
    • N2 cylinder/lines/filter
    • Shipping procedure: New shipping cage design on going (Jeff) => Plastic box similar to COC
  • Design
    • Solidworks raytracing model
    • Mode design for HAM6 layout
       
  • Things to be decided / confirmed
    • How to handle optics / assemblies (Talk to the prev people)
    • First contact? (Margot: applicable to a short Rc of ~2.5m)
    • Gluing templates to be designed (how to handle it?)
  • Jitter noise?
  • How to align the cavity mirrors, input mirrors, QPDs, PDs, beam dumps.

Electronics ==> Rich

Attachment 1: OMC_Project_Schedule.pdf
OMC_Project_Schedule.pdf
  33   Wed Nov 7 20:21:42 2012 KojiGeneralGeneralWork completed in October [!]

Completed work of the previous months: [Jul] [Aug] [Sep] [Oct] [Nov] [Dec]


 

  32   Wed Nov 7 01:28:20 2012 KojiOpticsCharacterizationWedge angle test (A1)

Wedge angle test

Result: Wedge angle of Prism A1: 0.497 deg +/- 0.004 deg


 

Principle:

o Attach a rail on the optical table. This is the reference of the beam.

o A CCD camera (Wincam D) is used for reading out spot positions along the rail.

o Align a beam path along the rail using the CCD.

o Measure the residual slope of the beam path. (Measurement A)

o Insert an optic under the test. Direct the first surface retroreflectively. (This means the first surface should be the HR side.)

o Measure the slope of the transmitted beam. (Measurement B)

o Deflection angle is derived from the difference between these two measurements.

Setup:

wedge_measurement_setup.png

o An Al plate of 10" width was clamped on the table. Four other clamps are located along the rail to make the CCD positions reproducible.

o A prism (Coating A, SN: A1) is mounted on a prism mount. The first surface is aligned so that the reflected beam matches with the incident beam
with precision of +/-1mm at 1660mm away from the prism surface. ==> precision of +/- 0.6mrad

o In fact, the deflection angle of the transmission is not very sensitive to the alignment of the prism.
The effect of the misalignment on the measurement is negligible.

o Refractive index of Corning 7980 at 1064nm is 1.4496

Result:

Without Prism
Z (inch / mm), X (horiz [um] +/-4.7um), Y (vert [um] +/-4.7um)
0” / 0, -481.3, -165.1
1.375" / 34.925, -474.3, -162.8
3" / 76.2, -451.0, -186.0
4.375" / 111.125, -432.5, -181.4
6" / 152.4, -432.5, -181.4
7.375" / 187.325, -330.2, -204.6
9" / 228.6, -376.7, -209.3

With Prism / SN of the optic: A1
Z (inch / mm), X (horiz [um] +/-4.7um), Y (vert [um] +/-4.7um)
0” / 0, -658.3, -156.8
1.375" / 34.925, -744.0, -158.1
3" / 76.2, -930.0, -187.4
4.375" / 111.125, -962.6, -181.4
6" / 152.4, -1190.4, -218.6
7.375" / 187.325, -1250.9, -232.5
9" / 228.6, -1418.3, -232.5

Analysis:

Wedge angle of Prism A1: 0.497 deg +/- 0.004 deg

wedge_measurement.png[Click for a sharper image]

  31   Thu Oct 18 20:23:33 2012 KojiOpticsCharacterizationImproved measurement

Significant improvement has been achieved in the RoC measurement.

  • The trans PD has much more power as the BS at the cavity trans was replaced by a 50% BS. This covers the disadvantage of using the a Si PD.
  • The BB EOM has a 50Ohm terminator to ensure the 50Ohm termination at Low freq.
  • The length of the cavity was changed from 1.2m to 1.8m in order to see the effect on the RoC measurement.

By these changes, dramatic increase of the signal to noise ratio was seen.

Now both of the peaks corresponds to the 1st-order higher-order modes are clearly seen.
The peak at around 26MHz are produced by the beat between the carrier TEM00 and the upper-sideband TEM01 (or 10).
The other peak at around 57MHz are produced by the lower-sideband TEM01 (or 10).

 

Peak fitting

From the peak fitting we can extract the following numbers:

  • Cavity FSR (hence the cavity length)
  • Cavity g-factor
  • Approximate measure of the cavity bandwidth

Note that the cavity itself has not been touched during the measurement.
Only the laser frequency and the incident beam alignment were adjusted.

The results are calculated by the combination of MATLAB and Mathemaica. The fit results are listed in the PDF files.
In deed the fitting quality was not satisfactory if the single Lorentzian peak was assumed.

There for two peaks closely lining up with different height. This explained slight asymmetry of the side tails

This suggests that there is slight astigmatism on the mirrors (why not.)

The key points of the results:

- FSR and the cavity length: 83.28~83.31MHz / L=1.799~1.800 [m] (surprisingly good orecision of my optics placement!)

- Cavity g-factor: Considering the flatness of the flat mirror from the phase map, the measured g-factors were converted to the curvature of the curved mirror.
RoC = 2.583~4 [m] and 2.564~7 [m]. (Note: This fluctuation can not be explained by the statistical error.)
The mode split is an order of 10kHz. This number also agrees with the measurement taken yesterday.

If the curved mirror had the nominal curvature of 2.5m, the flat mirror should have the curvature of ~20m. This is very unlikely.

- Approximate cavity line width: FWHM = 70~80kHz. This corresponds to the finesse of ~500. The design value is ~780.
This means that the locking offset is not enough to explain the RoC discrepancy between the design and the measurement.

 

Attachment 1: Cav_scan_response_zoom_20121017.pdf
Cav_scan_response_zoom_20121017.pdf
Attachment 2: detailed_RoC_setup.pdf
detailed_RoC_setup.pdf
  30   Wed Oct 17 20:36:04 2012 KojiOpticsGeneralRoC test cavity locked

The RoC test setup has been built on the optical table at ATF.

The cavity formed by actual OMC mirrors have been locked.

The modulation frequency of the BB EOM was swept by the network analyzer.
A peak at ~30MHz was found in the transfer function when the input beam was misaligned and clipping was introduced at the transmission PD.
Without either the misalignment or the clipping, the peak disappears. Also the peak requires these imperfections to be directed in the same way
(like pitch and picth, or yaw and yaw). This strongly suggests that the peak is associated with the transverse mode.

The peak location was f_HOM = 29.79MHz. If we consider the length of the cavity is L=1.20m, the RoC is estimated as

RoC = L / (1 - Cos[f_HOM/(c/2/L) * PI]^2)

This formula gives us the RoC of 2.587 m.

I should have been able to find another peak at f_FSR-f_TMS. In deed, there was the structure found at 95MHz as expected.
However, the peak was really weak and the location was difficult to determine as it was coupled with the signal from residual RFAM.

The particle level in the clean booth was occasionally measured. Every measurement showed "zero".

To be improved:

  • The trans PD is 1801 which was found in ATF with the label of the 40m. It turned out that it is a Si PD.
    I need to find an InGaAs PD (1811, 1611, or my BBPD) or increase the modulation, or increase the detected light level.
    (==> The incident power on 1810 increased. Oct 17)
  • The BS at the transmission is actually Y1-45P with low incident angle. This can be replaced by 50% or 30% BS to increase the light on the fast PD.
    (==> 50% BS is placed. Oct 17)
  • I forgot to put a 50ohm terminator for the BB EOM.
    (==> 50Ohm installed. Oct 17)
  • A directional coupler could be used for the BBEOM signal to enhance the modulaiton by 3dB.
  • The mode matching is shitty. I can see quite strong TEM20 mode.
  • Use the longer cavity? L=1.8m is feasible on the table. This will move the peak at 27MHz and 56MHz (FSR=83MHz). Very promising.
    (==> L=1.8m, peak at 27MHz and 56MHz found. Oct.17)
Attachment 1: detailed_RoC_setup.pdf
detailed_RoC_setup.pdf detailed_RoC_setup.pdf
Attachment 2: omc_cav_lock_SCRN_SHOT.png
omc_cav_lock_SCRN_SHOT.png
Attachment 3: Cav_scan_response_20121016.pdf
Cav_scan_response_20121016.pdf
  29   Tue Oct 16 15:51:01 2012 KojiGeneralGeneralPlan Update: October [!]

Completed work of the previous months: [Jul] [Aug] [Sep] [Oct] [Nov] [Dec]


Facility/Supplies

  • Work in progress
    • RoC measurement
    • R&T measurement
    • Wedge measurement
  • Work to be done
    • Replacing a file cabinet next to the south wall by a lockable cabinet
    • Additional clean supplies ~ glove 8.5,9,9.5
    • Stainless bats
    • Ion gun safety issues: https://dcc.ligo.org/cgi-bin/private/DocDB/ShowDocument?docid=88631
        
  • Design
    • Laser SOP
       
  • Test
    • Continuous monitoring of the particle level
       
  • Note: Optical Table W96" x D48" x H27"

Beaurocracy

  • Laser SOP / HV use? / UV?
  • Procedures to be decided
    • PZT alignment
    • UV glue? (heat) / gluing test
    • Balance
    • N2 cylinder/lines/filter
  • Design
    • Mode design for HAM6 layout
    • Finalization of scattering paths
       
  • Things to be decided / confirmed
    • How to handle optics / assemblies (Talk to the prev people)
    • First contact? (Margot: applicable to a short Rc of ~2.5m)
    • Gluing templates to be designed (how to handle it?)
       
  • PDs
  • Things to buy
  • Need to buy a fiber for mode cleaning?
  • Mode content of the ELIGO dark beam?
  • Jitter noise?
  • How to determine the design?
  • Why Fused Silica? (How much is the temp fluctuation in the chamber?)
  • How to align the cavity mirrors, input mirrors, QPDs, PDs, beam dumps.
  • PZTs @LLO
     

Electronics

  • Thorough scrutinization of cabling / wiring / electronics
    • ELIGO OMC Wiring diagram D070536-A2
      • Occupies 2 DB25s -> They were anchored on the sus cage
      • Preamps for DCPDs will be fixed on the ISI table
        -> DB25 for the DCPDs will be anchored on the table
      • Use longer thin cables for the DCPDs in order to route them through the suspension stages
      • Turn the heater cable to the one for the other PZT
  • Electronics / CDS electronics / software
  • Things to be tested
    • QPD/PD pre-selections (QE/noise)
    • PD preamp design (Rich)
    • Functionality test of QPD/PD/PZT

Shipping, storage etc

  28   Tue Oct 16 15:50:09 2012 KojiGeneralGeneralWork completed in August/September [!]
  • Work done
  • Things ordered
    • Office Depot
      • [delivered] Office Depot(R) Brand Stretch Wrap Film, 20 x 1000 Roll, Clear / 445013
      • [delivered] Eveready(R) Gold AA Alkaline Batteries, Pack Of 24 / 158448
      • [delivered] Rubbermaid(R) Roller Sponge Mop / 921841
      • [delivered] Rubbermaid(R) Roller Sponge Mop Replacement / 921858
      • [delivered] Rubbermaid(R) Sanitizing Caddy, 10 Quarts, Yellow / 674125
      • [delivered] Glad(R) Tall Kitchen Trash Bags, 13 Gallon, White, Box Of 28 / 269268
    • Global Industrial Equipment
      • [delivered] Extended Surface Pleated Cartridge Filter Serva-Cell Mp4 Slmp295 12X24X2 Gl    WBB431699
    • Global Industrial Equipment
      • [delivered] Nexel Poly-Z-Brite Wire Shelving 30"W x 21"D x 63"H Nexel Poly-Z-Brite™ Wire Shelving Starter Unit WB189209
      • [delivered] Stem Casters Set of (4) 5" Polyurethane Wheel, 2 With Brakes 1200 lb. Capacity WB500592    
    • Rack Solutions
      • [delivered] Open Frame Server Racks
        1 x 20" Depth Kit (Ideal for Audio/Video or Networking Racks) P/N: 111-1779
        1 x 36U, Rack-111 Post Kit P/N: 111-1728
        1 x Caster Kit for Open Frame RACK-111 P/N: 111-1731
      • [delivered] 36U Side Panel Kit $199.99 P/N: 102-1775
    • Rack shelf
      • [delivered] 1 RMS 19 X 15 SINGLE SIDED NON-VENTED SHELF 70121637
    • Work bench, Stools
      • [not yet] 72"L X 30"W Production Bench - Phenolic Resin Square Edge-Blue Form attached WB237381LBL    
      • [not yet] 72"W Lower Shelf For Bench - 15"D- Blue Form attached WB606951    
      • [not yet] ESD-Safe Vinyl Clean Room Stool with Nylon Base with Drag Chain Blue Form attached WBB560852    
    • P Touch
      • [delivered] Brother PT-2030 Desktop Office Labeler Punch-out product 672828    
      • [delivered] Brother(R) TZe-241 Black-On-White Tape, 0.75 x 26.2 Punch-out product 239384    
      • [delivered] Brother(R) TZe-231 Black-On-White Tape, 0.5 x 26.2 Punch-out product 239400    
    • UV light guide
      • [delivered] Fiber Optic Single Light Guide 5mm OD X 3mm ID X 1M L Note: This light guide can be used with MKIII UV Cure unit. OLB1081
    • Gloves (7.5, 8.0)
      • [delivered] GLOVE ACCTCH NR-LTX SZ7.5 PK25 Punch-out product 79999-306
      • [delivered] GLOVE ACCTCH NR-LTX SZ8 PK25 Punch-out product 79999-308
    • Lab coat (L,XL), Sticky Mat, Shoe Covers (L, XL), Cap, Mask
      • [delivered] LAB XP WH EL WR.COLL. NP L30EA Punch-out product 82007-618
      • [delivered] LAB XPWH EL WR.COLL. NP XL30EA Punch-out product 82007-620    
      • [delivered] VWR MAT ADHESIVE 30L 18X36 BLU Punch-out product 21924-110 (This was too small)
      • [delivered] VWR SHOECVR NSKID AP 2XL 150PR Punch-out product 414004-651    
      • [delivered] VWR SHOECVR NSKID AP XL 150PR Punch-out product 414004-650    
      • [delivered] CAP BOUFFANT 24IN RAYON CS500 Punch-out product 10843-053    
      • [delivered] MASK VLTC TIES N/STRL PK50 Punch-out product 10869-020
    • VWR
      • [delivered] FACE SHIELD UVC-803 Supplier: UVP 33007-151
         
    • [Delivered] Laser safety glasses

 

  27   Tue Oct 16 14:50:54 2012 jamie, jeffGeneralGeneralOMC breadboard/plate measurement dimensions

We have measured the dimensions and mass of the OMC glass plates/breadboards:

S/N Mass (g) Length (mm) Width (mm) Height (mm) Notes
01 6146 449.66 149.85 41.42, 41.42  for LLO
02 6126 449.66 149.97 41.32, 41.32  for LHO
03 6143 449.76 149.98 41.39, 41.43  
04 6139 449.78 149.81 41.40, 41.40  for 3IFO
05 6132 449.76 150.03 41.27, 41.31 corner chip, front-bottom-left*
06 6138 449.84 149.71 41.42, 41.42  
  • * orientation is relative to "front" face, i.e. long-short face with S/N on it, with S/N upright.
  • Height measurements were made twice, once at each end.
  • TMeasurements of 03, 05, 02, and 06 were done in the open in the OMC lab.  This was not thought to be too much of an issue since the plates
    are already covered with particulate matter from the tissue paper that they were wrapped in. 
    Measurements of 04 and 01 were done on the optics table, under the clean room enclosure.

Note by Koji:

  • The scale of the bake lab was used. (Max 60kg, Min resolution of 1g)
  • The dimensions were measured by a huge caliper which Jeff brought from Downs.
  • S/N 01, 03, 04 look pretty similar. They should be the primary candidates.
  26   Fri Oct 12 17:15:19 2012 KojiOpticsGeneralLoan from the 40m / ATF
  • HWP set
    • Optics: CVI QWPO-1064-08-2-R10
    • Mount: New Focus #9401
    • Post: Pedestal 2.5inch
    • Returned: Oct 19, 2012 by KA
  • QWP set
    • Optics: CVI QWPO-1064-05-4-R10
    • Mount: New Focus #9401
    • Post: Pedestal 2.5inch
    • Returned: Jan 17, 2013 by KA
  • Faraday set
    • Optics: OFR IO-2-YAG-HP Returned: Mar 21, 2013 by KA
    • Mount: New Focus #9701 Returned: Apr 17, 2013 by KA
    • Post: Pedestal (1.5+0.25inch)x2
  • Steering Mirror 1
    • Optics: CVI Y1-1037-45S
    • Mount: Newport Ultima U100-AC
    • Post: Pedestal 3inch
    • Returned: Jan 17, 2013 by KA
  • Steering Mirror 2
    • Optics: CVI Y1-1037-45P
    • Mount: Newport Ultima U100-AC
    • Post: Pedestal 3inch
    • Returned: Jan 17, 2013 by KA
  • Steering Mirror 3
    • Optics: New Focus 5104
    • Mount: Newport Ultima U100-AC
    • Post: Pedestal 3inch
    • Returned: Jan 17, 2013 by KA
  • Prism Mount
    • Mount: Thorlabs KM100P+PM1 2014/7/17
    • Post: Pedestal 1.5+1+1/8inch
  • 0.5" Mirror Mount
    • Mount: Newport U50-AReturned: Apr 17, 2013 by KA
    • Mount: Newport U50-A 2014/7/17
    • Post: Pedestal 1.5+2inch
  • Black Glass Beam Dump
    • Optics: 1" sq. schott glass x3
    • Mount: Custom Hexagonal 1"
    • Post: Pedestal 3inch
  • PBS Set
    • 05BC16PC.9 (PBS 1064 1000:1)
    • Mount: Custom Aluminum
    • Returned: Jan 17, 2013 by KA
  • Lenses
    • KBX067.AR33 f=125mm
    • KPX106 f=200mm, KPX109 f=250mm unknown-coat
    • KPX088.AR33 f=75mm
    • KPX094.AR33 f=100mm
    • PLCX-C (BK7) 3863 (f=7.5m), 2060 (f=4.0m), 1545 (f=3.0m), 1030 (f=2.0m) non-coat
    • PLCX-UV (FS) 30.9 non-coat(!) f=60mm
    • Returned: Jan 17, 2013 by KA
  • Pedestals
    • 1/4" x5, 1/8" x3, Returned: Jan 17, 2013 by KA
    • 0.5" x1, 1.5" x1

Another loan from the 40m on Oct 17th, 2012

  • Minicircuits
    • Splitter ZFSC-2-5 x2
    • Filter SLP-1.9 x2 / BLP-1.9 x1/2 / SLP-5 x1
    • Returned: Jan 17, 2013 by KA
  • Connectors / Adaptors
    • SMA TEE x1 / SMA 50Ohm x 1 / BNC T x 10, Returned: Jan 17, 2013 by KA
    • SMA TEE x1 / SMA 50Ohm x 1Returned: May 20, 2013 by KA
  • Pomona Box x1, Returned: Jan 17, 2013 by KA
  • Pomona Box x1
  • Power supply for New Focus Fast PD made by Jamie R Returned: Apr 17, 2013 by KA
  • BS-1064-50-1037-45S / Newport U100-A mount / 1"+2" Pedestal, Returned: Jan 17, 2013 by KA
  • BS-1064-50-1025-45P / Newport U100-A mount / 3/4" post + Base, Returned: Jan 17, 2013 by KA
  • BNC cable 21ft x2, Returned: Jan 17, 2013 by KA
  • SMA Cable 6ft

 


Another loan from the 40m on Nov 21th, 2012

  • Mounting Base Thorlabs BA-2 x 17
  • Mounting Posts (phi=3/4", L=2.65", normal x15, and 1/4"-20 variant x2)

Yet another loan from the 40m on Jan 16th, 2013

  • V-groove Mounting Bases Custom. Qty.2Returned: Feb 25, 2013 by KA

Loan from ATF

32.7MHz EOM+Tilt aligner
Thorlabs Broadband EOM+Tilt aligner
Forks x 5Returned: Feb 25, 2013 by KA
JWIN Camera x 2

  25   Tue Oct 9 05:03:15 2012 KojiElectronicsGeneralOMC Test Electronics Setup

electronics_setup.png

Attachment 2: electronics_setup.pdf
electronics_setup.pdf
  24   Tue Oct 9 04:59:24 2012 KojiOpticsGeneralOMC Test Optical Setup

OMC_test_setup.png

 

Attachment 2: OMC_test_setup.pdf
OMC_test_setup.pdf
  23   Mon Oct 8 11:30:47 2012 KojiOpticsGeneralEG&G 2mm photodiode angle response

EGE&G 2mm photodiode angle response measured by Sam T1100564-v1

  22   Fri Oct 5 03:39:58 2012 KojiOpticsGeneralRoC Test setup

Based on Zach's experiment design, I wrote up a bit more detailed optical layout for the mirror test.

RoC_test_setup.png

Item: Newfocus Fast PD
Qty.: 1
Mirror: Newfocus Fast PD
Mount: Post
Post: Post Holder (Newfocus)
Fork: Short Fork

Item: Thorlabs RF PD
Qty.: 1
Mirror: Thorlabs RF PD
Mount: Post
Post: Post Holder (Newfocus)
Fork: Short Fork

Item: Newfocus Broadband
Qty.: 1
Mirror: Newfocus EOM
Mount: Newfocus
Post: Custom Mount? or Pedestal X"?
Fork: Short Fork

Item: Newfocus Resonant
Qty.: 1
Mirror: Newfocus EOM
Mount: Newfocus
Post: Custom Mount? or Pedestal X"?
Fork: Short Fork

Item: ND Filter
Qty.: 2
Mirror: -
Mount: Thorlabs FIlter Holder
Post: Pedestal X"
Fork: Short Fork

Item: New Port Lens Kit 1"
Qty.: 1

Item: Thorlabs ND Kit
Qty.: 1

Item: Plano Convex Lens
Qty.: f=100, 100, 150, 200
Mirror: New Port (AR)
Mount: Thorlabs
Post: Post Holder (Newfocus)
Fork: Short Fork

Item: Bi-Convex Lens
Qty.: 75
Mirror: New Port (AR)
Mount: Post
Post: Post Holder (Newfocus)
Fork: Short Fork

Item: Flipper Mirror
Qty.: 1
Mirror: CVI Y1-10XX-45P
Mount: New Focus Flipper
Post: Pedestal X"
Fork: Short Fork

Item: Steering Mirror
Qty.: 8
Mirror: CVI Y1-10XX-45P
Mount: Suprema 1inch
Post: Pedestal X"
Fork: Short Fork

Item: PBS
Qty.: 3
Mirror: PBS 1inch BK7
Mount: Newport BS Mount
Post: Pedestal X"
Fork: Short Fork

Item: Knife Edge Beam Dump
Qty.: 4
Mirror: Thorlabs Knife Edge
Mount: Post
Post: Post Holder (Newfocus)
Fork: Short Fork

Item: Half Wave Plate
Qty.: 4
Mirror: CVI QWPO-
Mount: CVI
Post: Pedestal X"
Fork: Short Fork

Item: Quater Wave Plate
Qty.: 3
Mirror: CVI QWPO-
Mount: CVI
Post: Pedestal X"
Fork: Short Fork

Item: OMC Curved Mirror
Qty.: 2
Mirror: -
Mount: Suprema 0.5inch + Adapter
Post: Pedestal X"
Fork: Short Fork

Item: Prism Holder
Qty.: 1
Mirror: OMC Prism
Mount: Newport Prism Mount
Post: Pedestal X"
Fork: Short Fork

Item: CCD
Qty.: 1
Mirror: Thorlabs?
Mount: Thorlabs?
Post: Post Holder (Newfocus)
Fork: Short Fork

Attachment 1: RoC_test_setup.pdf
RoC_test_setup.pdf
  21   Mon Oct 1 16:06:55 2012 KojiCleanGeneralParticle Counts

1. It turned out that the particle counter MET ONE 227A at ATF shows
(particle count)/(0.1 ft^3)


This means that the numbers I saw previously should be multiplied by 10.
So the nominal class of the room was 5000.

2. As our GT-321s have no diffuser, I borrowed a diffuser from 227A.
The diffuser actually increases the count. We need to buy them.
All the measurments below are performed with the diffuser and calibrated in Count/ft^3.

3. Measured the particle level without the HEPA running.

With diffuser: [cnt/ft^3]

  GT-321 #1 GT-321 #2   227A
0.3um 152622 137511 -
0.5um  14706 14823   11860

Over Class 10000

4. The two HEPA fans are turned on at the speed "MED".

Basically no particles are detected in the HEPA booth.

With diffuser, inside of the HEPA booth:

  GT-321 #1 GT-321 #2  227A 
0.3um 0 0
-
0.5um 0 0 0

The particle level in the room (outside of the HEPA booth) is also improved

With diffuser, outside of the HEPA booth GT-321 #1:
0.3 um 18612
0.5 um   1728

5. The two HEPA fans are turned on at the speed "LOW".

Particle levels are still zero inside.

With diffuser, inside of the HEPA booth, GT-321 #1:
0.3 um 0
0.5 um 0

The particle level in the room (outside of the HEPA booth) is also improved
but the cleaning power for 0.3um seems degraded.

With diffuser, outside of the HEPA booth, GT-321 #1:
0.3 um 34488
0.5 um   1386

 

  20   Tue Sep 25 14:18:14 2012 KojiCleanGeneralParticle Counts

Particle counts

Before the prefilter is installed: 0.5um 1191cnts, 0.7um 346cnts

2:20 prefilter installed
2:25 0.5um 650 / 0.7um 255
3:00 0.5um 578 / 0.7um 99
4:00 0.5um 480 / 0.7um 102
5:00 0.5um 426 / 0.7um 76

They are counts/(0.1 ft^3)! These numbers should be multiplied by 10 to know the particle "CLASS".

  19   Wed Aug 22 20:16:43 2012 KojiFacilityGeneralWorkbenches have been installed / Clean room stools

Last Friday, new workbenches were installed. Vladimir got a new table and a cleanroom stool.

P8171968.jpg

The other two workbenches were also nicely set.

P8171969.jpg

  18   Tue Aug 14 03:29:06 2012 KojiSupplyGeneralClean supply rack

Clean supplies & some cleaning tools are located at the right side of the entrance.
The file cabinet there was moved to the left side of the door, but will be removed eventually.

P8131959.JPG    P8131961.JPG

  17   Mon Aug 13 17:01:35 2012 KojiCleanGeneralParticle Counts

Aug 13, 2012 / 0.5um 1000~2000/(0.1 cu ft) / 0.7um   400-600/(0.1 cu ft) by ATF particle counter (MET ONE 227A)

They are counts/(0.1 ft^3)! These numbers should be multiplied by 10 to know the particle "CLASS".

  16   Mon Aug 13 16:59:11 2012 KojiCleanGeneralRoom Cleaning Log

Floor wiped with a wet wiper (Aug 13, 2012)
Floor wiped with a wet wiper (Aug 15, 2012)
Floor wiped with a wet wiper (Sep 25, 2012)
Air conditioning prefilter replaced (Sep 25, 2012)
Floor wiped with a wet wiper (Oct 01, 2012)
Floor wiped with a wet wiper (Nov 06, 2012) / ATF too
Floor wiped with a wet wiper (Jan 04, 2013)
Floor wiped with a wet wiper (Mar 23, 2013)
Floor wiped with a wet wiper (Apr 17, 2013)
Air conditioning prefilter replaced (Apr 17, 2013)
Floor wiped with a wet wiper (Jun 24, 2013)
Removing Vladimir's mess. Floor swept with a broom (Jun 26, 2013)
Completed removing Vladimir's mess. Floor swept with a wet wiper (Jun 27, 2013)
Air conditioning prefilter replaced (Sep 12, 2013)
Floor wiper head replaced. (Dec 10, 2013)
Floor wiped with a wet wiper (Dec 10, 2013)
Floor wiped with a wet wiper (Apr 1, 2014)

Air conditioning prefilter replaced (Dec 30, 2014)
Air conditioning prefilter replaced (some time in 2015...)

Floor wiped with a wet wiper (Dec 1, 2015)
Floor wiped with a wet wiper (Aug 23, 2016)
Air conditioning prefilter replaced (Aug 8, 2017) = 1 stock remains

  15   Sat Aug 11 00:59:14 2012 KojiFacilityGeneralLaser Safety Barrier

It seemed that a laser safety barrier was installed today!?

P8131960.JPG

 

  14   Wed Aug 1 19:35:00 2012 KojiFacilityGeneralFloor cleaned / Workbench being built / Table top defect

- The floor of the room was cleaned and waxed!

- Sticky mats are placed! Now we require shoe covers!

P8011949.JPG

- Work benches are being built. One unit is done.

P8011948.JPG

- The other is half done because the table top has chippings.

P8011947.JPG

  13   Tue Jul 31 21:33:17 2012 KojiGeneralGeneralPlan Update: August [!]

Completed work of the previous months: [Jul] [Aug] [Sep] [Oct] [Nov] [Dec]


Facility/Supplies

  • Work done
  • Things ordered
    • Office Depot
      • [delivered] Office Depot(R) Brand Stretch Wrap Film, 20 x 1000 Roll, Clear / 445013
      • [delivered] Eveready(R) Gold AA Alkaline Batteries, Pack Of 24 / 158448
      • [delivered] Rubbermaid(R) Roller Sponge Mop / 921841
      • [delivered] Rubbermaid(R) Roller Sponge Mop Replacement / 921858
      • [delivered] Rubbermaid(R) Sanitizing Caddy, 10 Quarts, Yellow / 674125
      • [delivered] Glad(R) Tall Kitchen Trash Bags, 13 Gallon, White, Box Of 28 / 269268
    • Global Industrial Equipment
      • [delivered] Extended Surface Pleated Cartridge Filter Serva-Cell Mp4 Slmp295 12X24X2 Gl    WBB431699
    • Global Industrial Equipment
      • [delivered] Nexel Poly-Z-Brite Wire Shelving 30"W x 21"D x 63"H Nexel Poly-Z-Brite™ Wire Shelving Starter Unit WB189209
      • [delivered] Stem Casters Set of (4) 5" Polyurethane Wheel, 2 With Brakes 1200 lb. Capacity WB500592    
    • Rack Solutions
      • [delivered] Open Frame Server Racks
        1 x 20" Depth Kit (Ideal for Audio/Video or Networking Racks) P/N: 111-1779
        1 x 36U, Rack-111 Post Kit P/N: 111-1728
        1 x Caster Kit for Open Frame RACK-111 P/N: 111-1731
      • [delivered] 36U Side Panel Kit $199.99 P/N: 102-1775
    • Rack shelf
      • [delivered] 1 RMS 19 X 15 SINGLE SIDED NON-VENTED SHELF 70121637
    • Work bench, Stools
      • [not yet] 72"L X 30"W Production Bench - Phenolic Resin Square Edge-Blue Form attached WB237381LBL    
      • [not yet] 72"W Lower Shelf For Bench - 15"D- Blue Form attached WB606951    
      • [not yet] ESD-Safe Vinyl Clean Room Stool with Nylon Base with Drag Chain Blue Form attached WBB560852    
    • P Touch
      • [delivered] Brother PT-2030 Desktop Office Labeler Punch-out product 672828    
      • [delivered] Brother(R) TZe-241 Black-On-White Tape, 0.75 x 26.2 Punch-out product 239384    
      • [delivered] Brother(R) TZe-231 Black-On-White Tape, 0.5 x 26.2 Punch-out product 239400    
    • UV light guide
      • [delivered] Fiber Optic Single Light Guide 5mm OD X 3mm ID X 1M L Note: This light guide can be used with MKIII UV Cure unit. OLB1081
    • Gloves (7.5, 8.0)
      • [delivered] GLOVE ACCTCH NR-LTX SZ7.5 PK25 Punch-out product 79999-306
      • [delivered] GLOVE ACCTCH NR-LTX SZ8 PK25 Punch-out product 79999-308
    • Lab coat (L,XL), Sticky Mat, Shoe Covers (L, XL), Cap, Mask
      • [delivered] LAB XP WH EL WR.COLL. NP L30EA Punch-out product 82007-618
      • [delivered] LAB XPWH EL WR.COLL. NP XL30EA Punch-out product 82007-620    
      • [delivered] VWR MAT ADHESIVE 30L 18X36 BLU Punch-out product 21924-110 (This was too small)
      • [delivered] VWR SHOECVR NSKID AP 2XL 150PR Punch-out product 414004-651    
      • [delivered] VWR SHOECVR NSKID AP XL 150PR Punch-out product 414004-650    
      • [delivered] CAP BOUFFANT 24IN RAYON CS500 Punch-out product 10843-053    
      • [delivered] MASK VLTC TIES N/STRL PK50 Punch-out product 10869-020
    • VWR
      • [delivered] FACE SHIELD UVC-803 Supplier: UVP 33007-151
         
    • [Delivered] Laser safety glasses
  • Work in progress

    •  
  • Work to be done
    • Replacing a file cabinet next to the south wall by a lockable cabinet
    • Laser sign
    • Safety glass holder/rack/shelf
    • Prepare clean supplies ~ glove 8.5,9,9.5
    • Ion gun safety issues: https://dcc.ligo.org/cgi-bin/private/DocDB/ShowDocument?docid=88631
        
  • Design
    • Optical layout - Laser SOP
    • Additional HEPA stage
       
  • Test
    • Confirm particle level
       
  • Note: Optical Table W96" x D48" x H27"

Beaurocracy

  • Laser SOP
  • HV use?
  • UV?

Mechanics

  • Ongoing Work 
    • Cone-shaped wire clamp design (at the OMC end) - Jeff
       
  • Design
    • Wire preparation fixture - Jeff
    • How do we hold the PDs, QPDs, and black glass - we put 2 PDs and 2 QPDs on the PD mounting blacket. - Jeff
    • Integrated solidwork model - Derek
      • Q: How the wires are clamped at the top side?
      • Q: How much the length of the wire should be?
      • Q: Locations of the wire mounts on the plate
    • Cabling investigation:
      • Where do the cables from the feed-thrus anchored?
      • List of the current internal cables and their lengths
      • List of the required internal cables and their lengths
      • Can we route the intermediate stage of the suspension? Do we need new cables?
    • Dummy intermediate stage structure
       
    • Metal templates
      • First, decide an optical design
      • takes at least a month
         
    • Weights how heavy / how many
       
    • Earthquake stop design (Sam B)
       
  • Test
    • Cone-shaped wire clamp test - Jeff/Koji 
    • Balancing the plates
      • The Faraday isolator cage isn't clean
    • Dummy metal payload test at the sites???
       
  • Procedures to be decided
    • PZT alignment
    • Prototyping with metal parts?
    • UV glue? (heat) / gluing test
    • Balance

Optics

  • Things ordered
    • Newport LB servo
    • Halogen Lamp
    • N2 cylinder/lines/filter
  • Ongoing Work 
    • Mirrors to be delivered ~Aug
       
    • Design down select - Between "Single output & BS" vs "Two outputs & no BS"
    • Down selecting procedure:
      1. Assume ELIGO beam component
      2. Assume amount of 9MHz / 45MHz sidebands at the OMC input
      3. Calculate transmitted power
      4. Require HOM to be smaller than the TEM00 offset 
         
    • UV cured epoxy (Quate obtained)
  • Design
    • Mode design for HAM6 layout
    • Finalization of scattering paths
       
  • Tests
    • Measurement of PD angles
    • R&T of each mirror
    • Curvature of the curved mirrors
    • Cavity ref/trans/finesse
    • PD Q.E. & Reflectivity measurement vs incident angle
       
  • Things to be decided / confirmed
    • How to handle optics / assemblies (Talk to the prev people)
    • First contact? (Margot: applicable to a short Rc of ~2.5m)
    • Gluing templates to be designed (how to handle it?)
       
  • PDs
  • Misc
    • CCD beam analyzer (Zach: It is fixed.)
    • Are two PZTs used?
      • YES, for redundancy, range, upconversion tests.
         
  • Things to buy
  • Need to buy a fiber for mode cleaning?
  • Mode content of the ELIGO dark beam?
  • Jitter noise?
  • How to determine the design?
  • Why Fused Silica? (How much is the temp fluctuation in the chamber?)
  • How to align the cavity mirrors, input mirrors, QPDs, PDs, beam dumps.
  • PZTs @LLO
     

Electronics

  • Thorough scrutinization of cabling / wiring / electronics
    • ELIGO OMC Wiring diagram D070536-A2
      • Occupies 2 DB25s -> They were anchored on the sus cage
      • Preamps for DCPDs will be fixed on the ISI table
        -> DB25 for the DCPDs will be anchored on the table
      • Use longer thin cables for the DCPDs in order to route them through the suspension stages
      • Turn the heater cable to the one for the other PZT
  • Electronics / CDS electronics / software
  • Things to be tested
    • QPD/PD pre-selections (QE/noise)
    • PD preamp design (Rich)
    • Functionality test of QPD/PD/PZT

Shipping, storage etc


Jun/July
    - Lab renovation
Aug

    - Mechanics design
    - Mirror delivery
Sept

    - Basic optics test
    - Glue training
Oct
    - Cavity test
Nov

    - Suspending test
Dec
    - Shipping to LLO

Open questions
    Two optical designs
    Procedure
    Modeling
    Clamp design / stencil design
    gluing-installation procedure

  12   Tue Jul 31 21:29:43 2012 KojiGeneralGeneralWork completed in July [!]

 

  11   Tue Jul 24 11:41:29 2012 KojiGeneralGeneralUseful references

Nicolas Smith,
LIGO Document T0900383-v1
3mm Photodiode Characterization for Enhanced LIGO
https://dcc.ligo.org/cgi-bin/private/DocDB/ShowDocument?docid=4498

Tobin Fricke,
LIGO Document P1000010-v1
Homodyne detection for laser-interferometric gravitational wave detectors
https://dcc.ligo.org/cgi-bin/private/DocDB/ShowDocument?docid=8443

Nicolas Smith,
LIGO Document P1200052-v1
Techniques for Improving the Readout Sensitivity of Gravitational Wave Antennae
https://dcc.ligo.org/cgi-bin/private/DocDB/ShowDocument?docid=90498