Thu Apr 4 01:35:04 2013 |
Koji | Optics | Characterization | Mode matching to the OMC cavity |
The fiber output was matched with the lenses on a small bread board.
The detailed configuration is found in the following elog link.
http://nodus.ligo.caltech.edu:8080/OMC_Lab/105 |
96
|
Thu Apr 4 01:43:06 2013 |
Koji | Optics | Characterization | Mirror T measurement |
[Zach, Koji]
The measurement setup for the transmission measurement has been made at the output of the fiber.
- First, we looked at the fiber output with a PBS. It wasn't P-pol so we rotated the ourput coupler.
What we found was that it wasn't actually linearly polarized.
So the input coupler was rotated to correct it. This terribly misaligned the input coupling.
After some iteration of rotating and aligning the input/output couplers, we obtained reasonable
extiction ratio like 10mW vs 100uW (100:1) with 11mW incidence. (Where is the rest 0.9mW!?)
- The P-pol (transmission) out of PBS goes into the mirror. Here we tested mirror A1.
The mirror is mounted on the prism mount supported by a rotational stage for precise angle adjustment
We limited the input power down to 5mW so that we can remove the attenuator on the power meter.
The reading of the power meter was fluctuating, indeed depending on MY position.
So we decided to turn off the lighting of the room. This made the reading very stable.
The offset of the power meter was -0.58uW
The transmitted power for the normal incidence was 39.7uW with the incident 4.84mW.
[39.7-(-0.58)] / [4.84*1000-(-0.58)] *10^6 = 8320 ppm
The transmitted power for the 4deg incidence was 38.0uW with the incident 4.87mW.
[38.0-(-0.58)] / [4.87*1000-(-0.58)] *10^6 = 7980 ppm
cf. The specification is 7931ppm
|
97
|
Thu Apr 4 23:44:52 2013 |
Koji | Optics | Configuration | Beam launched into fiber |
We had to move our flipper mirror to share the beam between Peter's setup and ours as our flipper is at the place where the ISS PD array base is supposed to be!
There was no place to insert the flipper in the setup. We (Peter and Koji) decided to move the laser back for ~2".
This entirely changed the alignment of the setup. The fiber coupler was my reference of the alignment.
Once the beam is aligned, I check the coupling to the fiber. It was 50%.
I tweaked the lens and eventually the coupling is improved to 83%. (24.7mW incident, 20.4mW obtained.)
Then, I started to check the AOM path. I noticed that the 1st (or -1st) order beam is very weak.
The deflection efficiency is ~0.1%. Something is wrong.
I checked the driver. The driver's coupler output (1:10) show the amplitude ~1V. (good)
I check the main output by reducing the offset. When the coupler output is 100mV, the main output was 1V. (good)
So is the AOM itself broken??? |
99
|
Fri Apr 5 18:18:36 2013 |
Zach | Optics | Configuration | AOM probably broken |
Quote: |
Then, I started to check the AOM path. I noticed that the 1st (or -1st) order beam is very weak.
The deflection efficiency is ~0.1%. Something is wrong.
I checked the driver. The driver's coupler output (1:10) show the amplitude ~1V. (good)
I check the main output by reducing the offset. When the coupler output is 100mV, the main output was 1V. (good)
So is the AOM itself broken???
|
As Koji noticed that the AOM efficiency was very low, I figured I would try looking at it with a fresh set of eyes. The end result is that I have to agree that the AOM appears to be broken.
First, I measured the input impedance of the AOM using the AG4395A with the impedance test kit (after calibrating). The plot is below. The spec sheet says the center frequency is 200 MHz, at which Zin should be ~50 ohms. It crosses 50 ohms somewhere near 235 MHz, which may be reasonable given that the LC circuit can be tuned by hand. However, it does surprise me that the impedance varies so much over the specified RF range of ±50 MHz. Maybe this is an indication that something is bad.

I removed the cover of the modulator (which I think Koji did, as well) and all the connections looked as I imagine they should---i.e., there was nothing obviously broken, physically.
I then tried my hand at realigning the AOM from scratch by removing and replacing it. I was not able to get better than 0.15%, which is roughly what Koji got.
So, perhaps our best course of action is to decide what we expect the Zin spectrum to look like, and whether that agrees with the above measurement. |
100
|
Mon Apr 8 11:11:37 2013 |
Koji | Optics | Characterization | More Mirror T measurement |
More Ts of the mirrors were measured.
A mirror specification:
Request: 8300+/-800 ppm
Data sheet: 7931ppm
C mirror specification:
Request: 50+/-10 ppm
Data sheet: 51.48ppm or 46.40ppm
Mirror | P_Incident P_Trans P_Offset | T_trans
| [mW] [uW] [uW] | [ppm]
-------+------------------------------+---------
A1 | 10.28 82.9 -0.205 | 8.08e3
A2 | ----- ----- ------ | ------
A3 | 10.00 83.2 -0.205 | 8.34e3
A4 | 10.05 80.7 -0.205 | 8.05e3
A5 | 9.94 81.3 -0.205 | 8.20e3
A6 | 10.35 78.1 -0.205 | 7.57e3
A7 | 10.35 77.8 -0.205 | 7.54e3
A8 | 10.30 78.0 -0.205 | 7.60e3
A9 | 10.41 84.1 -0.205 | 8.10e3
A10 | 10.35 77.3 -0.205 | 7.49e3
A11 | 10.33 77.9 -0.205 | 7.56e3
A12 | 10.34 78.7 -0.205 | 7.63e3
A13 | 10.41 85.4 -0.205 | 8.22e3
A14 | 10.34 84.4 -0.205 | 8.18e3
-------+------------------------------+---------
C1 | 10.30 0.279 -0.225 | 48.9
C2 | ----- ----- ------ | ------
C3 | 10.37 0.240 -0.191 | 41.6
C4 | 10.35 0.278 -0.235 | 49.6
C5 | 10.40 0.138 -0.235 | 35.9 => PZT assembly #2
C6 | 10.34 0.137 -0.235 | 36.0 => PZT assembly #1
C7 | 10.37 0.143 -0.229 | 35.9
C8 | 10.41 0.224 -0.237 | 44.3
C9 | 10.36 0.338 -0.230 | 54.8
C10 | 10.39 0.368 -0.228 | 57.4
C11 | 10.38 0.379 -0.209 | 56.6
C12 | 10.28 0.228 -0.238 | 45.3
C13 | 10.36 0.178 -0.234 | 39.8
-------+------------------------------+---------
|
101
|
Mon Apr 8 11:29:08 2013 |
Koji | Optics | Characterization | Mirror/PZT Characterization links |
|
103
|
Mon Apr 8 20:56:52 2013 |
Koji | Optics | Configuration | PZT & Curverd Mirror arrangement |
Assembly #1:
Mounting Prism #16
PZT #26
Mirror C6
Assembly #2:
Mounting Prism #20
PZT #23
Mirror C5 |
104
|
Mon Apr 8 21:11:14 2013 |
Koji | Optics | General | PZT assembly gluing |
[Jeff, Zach, Koji]
PZT assembly gluing
Glue gun -> to be returned to MIT
Fixtures x2
Al bases, spacers
spare screws
mirrors / prisms / PZTs
IPA bottle
clean tools x2
first contact kit
gloves (7.5)
|
105
|
Mon Apr 8 23:42:33 2013 |
Koji | Optics | Configuration | Fake OMC roughly aligned |
Mode matching:
|
106
|
Tue Apr 9 13:56:09 2013 |
Koji | Optics | General | PZT assembly post gluing / pre baking pictures |
|
107
|
Wed Apr 10 00:40:30 2013 |
Zach | Optics | Configuration | fauxMC locked |
[Koji, Zach]
Tonight, we locked the "fauxMC". We obtained a visibility of >99%.
Koji had aligned it roughly last night, but we wanted to have a couple steering mirrors in the path for this practice cavity (the periscope mirrors will serve this function in the real setup), so we marked the alignment with irises and installed two extra mirrors.
After obtaining flashes with the WinCam placed at the output coupler, we removed the WinCam and put a CCD camera at one of the curved mirror transmissions and used this to get a strong TEM00 flash. Then, we installed the REFL PD/CCD, swept the laser PZT and optimized the alignment by minimizing the REFL dips. Finally, we connected the RF electronics and locked the cavity with the LB box. We used whatever cables we had around to trim the RF phase, and then Koji made some nice SMA cables at the 40m.
One thing we noticed was that we don't have enough actuation range to keep the cavity locked for very long---even with the HV amp (100V). We are going to offload to the NPRO temperature using an SR560 or pomona box circuit. We may also make an enclosure for the cavity to protect it from the HEPA blasting.
Tomorrow, after we do the above things, we will practice measuring the transmission, length (FSR) and mode spectrum of the cavity before moving on to the real McCoy.

|
109
|
Fri Apr 12 09:25:31 2013 |
Koji | Optics | Characterization | Alignment of the OMC (without glue) |
[Zach Koji]
The first attempt not to touch the curved mirrors did not work. (Not surprising)
The eigenmode is not found on the mirror surface.
We decided to touch the micrometers and immediately found the resonance.
Then the cavity alignment was optimized by the input steering mirrors.
We got the cavity length L and f_TMS/f_FSR (say gamma, = gouy phase / (2 pi) ) as
L=1.1347 m (1.132m nominal)
gamma_V = 0.219176 (0.21879 nominal)
gamma_H = 0.219418 (0.21939 nominal)
This was already sufficiently good:
- the 9th modes of the carrier is away from the resonance 10-11 times
of the line width (LW)
- the 13th modes of the lower f2 sideband are 9-10 LW away
But
- the 19th modes of the upper f2 sideband are 1-3 LW away
This seems to be the most dangerous ones.
and
- The beam spots on the curved mirrors are too marginal
So we decided to shorten the cavity round-trip 2.7mm (= 0.675mm for each micrometer)
and also use the curved mirrors to move the eigenmode toward the center of the curved mirrors.
After the movement the new cavity length was 1.13209 m.
The spot positions on the curved mirrors are ~1mm too close to the outside of the cavity.
So we shortened the outer micrometers by 8um (0.8 div).
This made the spot positions perfect. We took the photos of the spots with a IR sensor card.
The measured cavity geometry is (no data electrically recorded)
L=1.13207 m (1.132m nominal, FSR 264.8175MHz)
gamma_V = 0.218547 (0.21879 nominal, 57.8750MHz)
gamma_H = 0.219066 (0.21939 nominal, 58.0125MHz)
- the 9th modes of the carrier is 11-13 LW away
- the 13th modes of the lower f2 sideband are 5-8 LW away
- the 19th modes of the upper f2 sideband are 4-8 LW away
The raw transmission is 94.4%. If we subtract the sidebands and
the junk light contribution, the estimated transmission is 97.6%.
Note:
Even if a mirror is touched (i.e. misaligned), we can recover the good alignment by pushing the mirror
onto the fixture. The fixture works pretty well!
|
110
|
Sat Apr 13 21:06:02 2013 |
Koji | Optics | General | OMC Bottom-side: cavity glued |
[Jeff, Zach, Lisa, Koji]
Gluing of the cavity mirrors went very well!!!
Preparation
- Checked if the cavity is still resonating. => Yes.
- Checked the FSR: 264.251MHz => 1.1345m
2.5mm too long => Move each micrometer by 0.625mm backward
- FSR&TMS (I)
Aligned the cavity again and checked the FSR: 264.8485MHz => 1.13194m
TMS(V): 58.0875MHz => gamma_V = 0.219324
TMS(H): 58.1413MHz => gamma_H = 0.219526
the 9th modes of the carrier is 9.7-10.4 line width (LW) away from the carrier resonance
the 13th modes of the lower f2 sideband are 9.2-10.2 LW away
the 19th modes of the upper f2 sideband are 0.3-1.8 LW away
We found that this coincidence of the resonance can be corrected by shortening the cavity round-trip by 0.5mm
- Spot positions (I)
The spots on the curved mirrors were ~1mm too much inside (FM side). In order to translate the cavity axis,
MM2 and MM4 were pushed by θ
θ/2.575 = 1mm ==> θ = 2.6 mrad
The separation of the micrometers are ~20mm
d/20mm = 2.6mrad ==> d = 52um
1div of the micrometer corresponds to 10um => 5div = 50um
- Move the micrometers and adjusted the input steering to recover the alignment.
- In any case we were confident to adjust the FSR/TMS/spot positions only with the micrometers
BS1/FM1/FM2 gluing
- Aligned the cavity
- Glued BS1/FM1/FM2 one by one while the cavity resonance was maintained.
FM2 was slipping as the table is not leveled well and the fixture was not supporting the optic.
- FSR&TMS (II)
FSR: 264.964875MHz => 1.13144m (Exactly 0.5mm shorter!)
TMS(V): 58.0225MHz => gamma_V = 0.218982
TMS(H): 58.1225MHz => gamma_H = 0.219359
the 9th modes of the carrier is 10.3~11.7 LW away
the 13th modes of the lower f2 sideband are 7.4~9.3 LW away
the 19th modes of the upper f2 sideband are 1.5~4.4 LW away
- Spot positions (II)
Looked OK.
CM2 gluing
- Glued CM2. The mirror was supported from the back with allen keys.
- FSR&TMS (III)
FSR: 264.9665625MHz => 1.13144m
TMS(V): 58.1275MHz => gamma_V = 0.219377
TMS(H): 58.0813MHz => gamma_H = 0.219202
the 9th modes of the carrier is 10.2~10.9 LW away
the 13th modes of the lower f2 sideband are 8.5~9.4 LW away
the 19th modes of the upper f2 sideband are 1.4~2.7 LW away
- Spot positions (III)
Looked slightly off at CM2. Pushed MM2 by 4um.
CM1 gluing
- Glued CM1.
- FSR&TMS (IV)
FSR: 264.964875MHz => 1.13144m
TMS(V): 58.06625MHz => gamma_V = 0.219145
TMS(H): 58.08625MHz => gamma_H = 0.219220
the 9th modes of the carrier is 10.8~11.1 LW away
the 13th modes of the lower f2 sideband are 8.2~8.6 LW away
the 19th modes of the upper f2 sideband are 2.6~3.2 LW away
- Spot positions (final confirmation)
Looked OK.
Final measurement
- After everything was finished, more detailed measurement has been done.
- FSR&TMS (final)
FSR: 264.963MHz => 1.13145m
TMS(V): 58.0177MHz => gamma_V = 0.218966
TMS(H): 58.0857MHz => gamma_H = 0.219221
the 9th modes of the carrier is 10.8~11.7 LW away
the 13th modes of the lower f2 sideband are 7.3~8.6 LW away
the 19th modes of the upper f2 sideband are 2.6~4.5 LW away
Final values for the micrometers
- MM1: The one closest to the input mirror (CM1) 0.78mm
- MM2: The other one on CM1 0.89
- MM3: The one closest to the output mirror (CM2) 0.90
- MM4: The other one on CM2 0.90
/------------\
0.90 0.78
\------------/
0.90 0.89
|
111
|
Tue Apr 16 00:40:45 2013 |
Koji | Optics | General | PD/QPD path gluing ~ preparation |
[Jeff Koji]
- Placed the optics on the PD/QPD path
- Checked the alignment of the beam on the dummy PD/QPD mounts
- There is a bit of (~0.5mm) shift of the spot position from the center. Mainly downward. This is well within a ball park of the PD mounts.
- The PD/QPD path gluing will take place tomorrow.
- Went to the 40m and received the DCPDs from Bob's lab.
- Took six ISC QPDs for the sake of the OMCs.
- They are now in the OMC lab.
- Measured the B mirror / E mirror R&Ts.
- Found anomalously high loss (3%) for the B mirrors (BSs)
- Went through the all mirrors. Some mirrors (3 or 4) seemed less lossy (<~1%). They will be used for the DCPD BS. |
112
|
Tue Apr 16 08:12:14 2013 |
Koji | Optics | Characterization | Further More Mirror T measurement |
T&Rs of the B mirrors and some of the E mirrors are measured.
I found that these BSs have high loss (1%~3%) . As this loss will impact the performance of the squeezer
we should pick the best ones for the DCPD path. B5, B6, and B12 seems the best ones.
Mirror | P_Incident P_Trans P_Refl | T R loss |
| [mW] [mW] [mW] | |
-------+--------------------------------------+-------------------------------------------+
B1 | 13.80+/-0.05 7.10+/-0.05 6.30+/-0.05 | 0.514+/-0.004 0.457+/-0.004 0.029+/-0.005 |
B2 | 14.10+/-0.05 6.50+/-0.05 7.15+/-0.05 | 0.461+/-0.004 0.507+/-0.004 0.032+/-0.005 |
B3 | 13.87+/-0.05 7.05+/-0.05 6.55+/-0.05 | 0.508+/-0.004 0.472+/-0.004 0.019+/-0.005 |
B4 | 13.85+/-0.05 6.78+/-0.05 6.70+/-0.05 | 0.490+/-0.004 0.484+/-0.004 0.027+/-0.005 |
B5 | 13.65+/-0.05 6.93+/-0.05 6.67+/-0.05 | 0.508+/-0.004 0.489+/-0.004 0.004+/-0.005 |
B6 | 13.75+/-0.05 6.70+/-0.05 6.92+/-0.05 | 0.487+/-0.004 0.503+/-0.004 0.009+/-0.005 |
B7 | 13.83+/-0.05 7.00+/-0.05 6.60+/-0.05 | 0.506+/-0.004 0.477+/-0.004 0.017+/-0.005 |
B8 | 13.90+/-0.05 6.95+/-0.05 6.68+/-0.05 | 0.500+/-0.004 0.481+/-0.004 0.019+/-0.005 |
B9 | 13.84+/-0.05 6.95+/-0.05 6.70+/-0.05 | 0.502+/-0.004 0.484+/-0.004 0.014+/-0.005 |
B10 | 13.97+/-0.05 6.98+/-0.05 6.72+/-0.05 | 0.500+/-0.004 0.481+/-0.004 0.019+/-0.005 |
B11 | 13.90+/-0.05 7.05+/-0.05 6.70+/-0.05 | 0.507+/-0.004 0.482+/-0.004 0.011+/-0.005 |
B12 | 13.90+/-0.05 6.98+/-0.05 6.78+/-0.05 | 0.502+/-0.004 0.488+/-0.004 0.010+/-0.005 |
-------+--------------------------------------+-------------------------------------------+
Mirror | P_Incident P_Trans P_Refl | T R loss |
| [mW] [uW] [mW] | [ppm] |
-------+-------------------------------------------+------------------------------------------+
E4 | 13.65+/-0.05 0.0915+/-0.0005 13.50+/-0.05 | 6703+/-44ppm 0.989+/-0.005 0.004+/-0.005 |
E12 | 13.75+/-0.05 0.0978+/-0.0005 13.65+/-0.05 | 7113+/-45 0.993+/-0.005 0.000+/-0.005 |
E16 | 13.90+/-0.05 0.0975+/-0.0005 13.30+/-0.05 | 7014+/-44 0.957+/-0.005 0.036+/-0.005 |
-------+-------------------------------------------+------------------------------------------+
|
113
|
Tue Apr 16 09:43:58 2013 |
Koji | Optics | Configuration | Mirror list for L1OMC |
L1 OMC
Cavity Mirrors
FM1 (input coupler): A8
FM2 (output coupler): A7
CM1 (curved mirror close to FM1): C6
CM2 (curved mirror close to FM2): C5
DCPD path
BS3 (BS for DCPDs): B5 B7
QPD path
BS1 (input steering): E10
SM1 (steering mirror next to BS1): E12
BS2 (BS for QPD path): B3
SM2 (steering mirror next to BS2): E4
SM3 (steering mirror next to SM2): E16

|
114
|
Tue Apr 16 23:26:51 2013 |
Koji | Optics | Characterization | Further More Mirror T measurement |
Since the previous measurement showed too high loss, the optical setup was checked.
It seemed that a PBS right before the T&R measurement setup was creating a lot of scattering (halo) visible with a sensor card.
This PBS was placed to confirm the output polarization from the fiber, so it was ok to remove it.
After the removal, the R&T measurement was redone.
This time the loss distributed from 0.2% to 0.8% except for the one with 1.3%. Basically 0.25% is the quantization unit due to the lack of resolution.
At least B7, B10, B12 seems the good candidate for the DCPD BS.
The AR reflection was also measured. There was a strong halo from the main reflection with an iris and sense the power at ~.5mm distance to separate the AR reflection from anything else. Now they are all somewhat realistic. I'll elog the measurement tonight.
33.6 +/- 0.2 uW out of 39.10+/-0.05 mW was observed. The offset was -0.236uW.
This gives us the AR reflectivity of 865+/-5ppm . This meets the spec R<0.1%
Mirror | P_Incident P_Trans P_Refl | T R loss |
| [mW] [mW] [mW] | |
---------------------------------------------------------------------------------------------
B1 | 39.10+/-0.05 19.65+/-0.05 19.25+/-0.05 | 0.503+/-0.001 0.492+/-0.001 0.005+/-0.002 |
B2 | 39.80+/-0.05 19.90+/-0.05 19.70+/-0.05 | 0.500+/-0.001 0.495+/-0.001 0.005+/-0.002 |
B4 | 39.50+/-0.05 19.70+/-0.05 19.30+/-0.05 | 0.499+/-0.001 0.489+/-0.001 0.013+/-0.002 |
B5 | 39.50+/-0.05 19.70+/-0.05 19.50+/-0.05 | 0.499+/-0.001 0.494+/-0.001 0.008+/-0.002 |
B6 | 39.55+/-0.05 19.50+/-0.05 19.95+/-0.05 | 0.493+/-0.001 0.504+/-0.001 0.003+/-0.002 |
B7 | 40.10+/-0.05 19.80+/-0.05 20.20+/-0.05 | 0.494+/-0.001 0.504+/-0.001 0.002+/-0.002 |
B8 | 40.15+/-0.05 19.80+/-0.05 20.20+/-0.05 | 0.493+/-0.001 0.503+/-0.001 0.004+/-0.002 |
B9 | 40.10+/-0.05 19.90+/-0.05 19.90+/-0.05 | 0.496+/-0.001 0.496+/-0.001 0.008+/-0.002 |
B10 | 40.10+/-0.05 19.70+/-0.05 20.30+/-0.05 | 0.491+/-0.001 0.506+/-0.001 0.002+/-0.002 |
B11 | 40.20+/-0.05 19.80+/-0.05 20.20+/-0.05 | 0.493+/-0.001 0.502+/-0.001 0.005+/-0.002 |
B12 | 40.20+/-0.05 19.90+/-0.05 20.20+/-0.05 | 0.495+/-0.001 0.502+/-0.001 0.002+/-0.002 |
---------------------------------------------------------------------------------------------
|
115
|
Wed Apr 17 07:30:04 2013 |
Koji | Optics | General | QPD path glued |
Yesterday, all of the glass components for the QPD path were glued.
- Check the alignment of the beam with the cavity.
- Placed the prisms
- Placed the QPD mount for the gluing test. An alignment disk instead of a diode was placed on the mount.
- Checked the spot positions at the QPDs. A CCD camera with a lens was used to find the spot.
The spots were ~0.5mm lower on the QPD1, and ~1mm lower on the QPD2.
- Glued the first steering mirror while the spot position was continuously monitored.
- Glued the BS in the QPD path while the spot position was monitored.
- FInally a glass bracket was glued.
- The spot on QPD2 was too low to absorb by the QPD shim.
- Once the steering mirror was clamped by a cantilever spring (to prevent slipping), the spot moved up a bit.
(Or, we should say, the cantilever misaligned the optics a bit in pitch in a preferrable direction.)
- The other steering mirror is clamped by a cantilever spring (to prevent slipping), the spot moved up a bit.
Or, we should say, the cantilever misaligned the optics a bit in pitch in a preferrable direction.)
- The last steering mirrors was also glued in a same way. As a result the spot is 0.5mm below the center of the alignment disk.
- Once the PD mounting brackets were glued, we can't place the QPD mount on it as the PEEK bar can't be inserted without moving the gluing template.
- The QPD mount with out the glass bracket was used to check the alignment of the beam dumps.
As the beam dumps have a wide aperture and the yaw alignment of the QPD is big, we could accommodate the beams in the dumps easily.
- The dumps were glued. |
120
|
Mon May 6 19:31:51 2013 |
Koji | Optics | Characterization | Spot position measurement on the diode mounts |
Measurement Order: DCPD2->DCPD1->QPD1->QPD2
DCPD1: 1.50mm+0.085mm => Beam 0.027mm too low
DCPD2: 1.75mm+0.085mm => Beam 0.051mm too high (...less confident)
QPD1: 1.25mm+0.085mm => Beam 0.077mm too low
QPD2: 1.25mm+0.085mm => Beam 0.134mm too low
or 1.00mm+0.085mm => Beam 0.116mm too high
|
121
|
Wed May 8 15:08:57 2013 |
Koji | Optics | Characterization | Spot position measurement on the diode mounts |
Remeasured the spot positions:
DCPD1: 1.50mm+0.085mm => Beam 0.084mm too high
DCPD2: 1.50mm+0.085mm => Beam 0.023mm too high
QPD1: 1.25mm+0.085mm => Beam 0.001mm too low
QPD2: 1.25mm+0.085mm => Beam 0.155mm too low
|
132
|
Thu May 30 15:00:28 2013 |
Koji | Optics | General | QPD alignment |
The QPD alignment was adjusted using the aligned beam to the cavity and the 4ch transimpedance amplifier.
As I have a test cable for the QPD, I attached a DB9 connector on it so that I can use the QPD transimpedance
amplifier to read the photocurrent. The transimpedance of the circuit is 1kV/A.
As this board (D1001974) does not have X/Y/SUM output, I quickly made the summing circuit on a universal
board I took from Japan a while ago.
The spot on the QPD1 (shorter arm side) seems too low by ~0.64mm. It seems that the QPD is linearly responding
to the input misalignment, so there is no optical or electrical problem.
As I wonder how much I can improve the situation by replacing the diodes, I swapped the diodes between QPD1 and QPD2.
Now QPD1 and QPD2 have the diode #43 and #38, respectively. It improved the situation a llitle (about 60um).
But the beam is still 0.58mm too low. 95% of the power is on the upper two elements.
Of course this is at the edge of the linear range.
I confirmed we still can observe the cavity is fringing even with the beam is aligned on this QPD. So this may be
sufficient for the initial alignment.
The QPD2 was in a better situation. The spot is about 100um too low but this is still well with in the linear range.
The incident powers on the diodes were also measured. The estimated responsivities and Q.E.s are listed below.
The reflection from the diode is adjusted to hit the beam dump properly.
Here are the raw numbers
QPD# QPD1 QPD2
Diode# #43 #38
-------------------------------------
Power Incident 118.8 uW 115.7uW
Sum Out 78.8 mV 84.6 mV
Vertical Out 69.1 mV 11.9 mV
Horizontal Out 9.8 mV -1.6 mV
SEG1 -1.90 mV -17.8 mV
SEG2 -2.18 mV -17.5 mV
SEG3 -32.0 mV -25.3 mV
SEG4 -42.0 mV -23.8 mV
-------------------------------------
Responsivity[A/W] 0.66 0.73
Q.E. 0.77 0.85
-------------------------------------
Arrangement of the segments
View from the beam
/ 2 | 1 X
|---+---|
\ 3 | 4 /
|
133
|
Fri May 31 05:46:54 2013 |
Koji | Optics | General | QPD alignment |
Peter F suggested to check the bottom surface of the PD housings if there is any protrusion/interference/whatever.
And that was true! It was found that the front side of QPD1 (Left) was lifted by a machining burr.
It seems that this burr consistently exists as the other one also have it (see QPD2 picture (right)) although it is not too terrible compared to the one in QPD1.

Once these burrs were removed, the spots were found on the right position of each diode.
From the measurement of the power on each segment, the positions of the spots were estimated. (listed in the table)
They indicate that the spots are within 0.1mm from the center. This is good enough.
The quantum efficiency was measured from the incident power and the sum output. It seems that there are
some difference between the diodes. The numbers are consistent with the measurement the other day.
QPD# QPD1 QPD2
Diode# #43 #38
-------------------------------------
Power Incident 84.7 uW 86.2 uW
Sum Out 56 mV 61 mV
Vertical Out -6.8 mV 10 mV
Horizontal Out 4.2 mV 8.8 mV
SEG1 -17 mV -15 mV
SEG2 -14.5 mV -11 mV
SEG3 -11 mV -15 mV
SEG4 -13 mV -20 mV
-------------------------------------
Spot position X +25 um +46 um (positive = more power on SEG1 and SEG4)
Spot position Y -42 um +46 um (positive = more power on SEG3 and SEG4)
-------------------------------------
Responsivity[A/W] 0.66 0.71
Q.E. 0.77 0.82
-------------------------------------
Arrangement of the segments
View from the beam
/ 2 | 1 X
|---+---|
\ 3 | 4 /
---------------
I(w,x,y) = Exp[-2 (x^2 + y^2)/w^2]/(Pi w^2/2)
(SEG_A+SEG_B-SEG_C-SEG_D)/(SEG_A+SEG_B+SEG_C+SEG_D) = Erf[sqrt(2) d/w]
d: distance of the spot from the center
w: beam width
|
134
|
Fri May 31 14:07:54 2013 |
Koji | Optics | Characterization | Transverse Mode Spacing measurement afte the baking |
Measurement for pitch
Free Spectral Range (FSR): 264.9703 +/− 0.0007 MHz
Cavity roundtrip length: 1.131419 +/− 0.000003 m
Transverse mode spacing (TMS): 57.9396 +/− 0.0002 MHz
TMS/FSR: 0.218664 +/− 0.000001
Assuming the line width of the cavity 1/400 of the FSR...
- the 9th modes of the carrier is 12.8 line width (LW) away from the carrier resonance
- the 13th modes of the lower f2 sideband are 5.7 LW away
- the 19th modes of the upper f2 sideband are -6.8 LW away
Measurement for yaw
Free Spectral Range (FSR): 264.9696 +/− 0.0004 MHz
Cavity roundtrip length: 1.131422 +/− 0.000002 m
Transverse mode spacing (TMS): 58.0479 +/− 0.0002 MHz
TMS/FSR: 0.219074 +/− 0.000001
- the 9th modes of the carrier is 11.3 line width (LW) away from the carrier resonance
- the 13th modes of the lower f2 sideband are 7.8 LW away
- the 19th modes of the upper f2 sideband are -3.7 LW away
The followings are the previous values before the bake
[from this entry]
- After everything was finished, more detailed measurement has been done.
- FSR&TMS (final)
FSR: 264.963MHz => 1.13145m
TMS(V): 58.0177MHz => gamma_V = 0.218966
TMS(H): 58.0857MHz => gamma_H = 0.219221
the 9th modes of the carrier is 10.8~11.7 LW away
the 13th modes of the lower f2 sideband are 7.3~8.6 LW away
the 19th modes of the upper f2 sideband are 2.6~4.5 LW away |
135
|
Mon Jun 3 18:58:08 2013 |
Koji | Optics | Configuration | OMC final tests |
- QPD mount aligned, QPD output checked
The spots are with 100um from the center of the diodes. [ELOG Entry (2nd photo)]
- TMS/FSR dependence on the PZT V
Shows significant dependence on the PZT voltages
It seems that the curvartures get longer when the voltages are applied to the PZTs.
The effect on these two PZTs are very similar. The dependence is something like
(TMS/FSR) ~ 0.219 - 1e-5 V
May cause resonance of the higher-order modes (like 13th order of the 45MHz sidebands) at a specific range of the PZTs.
We can't change anything any more, but the impact needs to be assessed
- DC response of the PZTs [ELOG Entry]
PZT voltages were swept. Observed multiple fringes during the sweep.
The data to be analyzed.
- AC response of the PZTs [ELOG Entry]
PZT1 and PZT2 well matched. The first resonance at 10kHz.
- Open loop TF of the servo
The UGF more than ~30kHz.
- Cleaning of the main optics with First Contact
Done. Visible scattering seen with an IR was reduced, but still exist.
All four cavity mirrors have about the same level of scattering.
Each scattering is a group of large or small bright spots.
It's actually a bit difficult to resolve the bright spots with the IR viewer.
- Raw transmission: i.e. Ratio between the sum of the DCPD paths and the incident power
May 8th (before the baking): 0.918
May 8th (First Contact applied): 0.940 (improved)
Jun 2nd (after the baking): 0.927 (worse)
Jun 2nd (First Cotact applied): 0.964 (improved)
Date |
2013/6/2 |
2013/6/2 |
2013/6/2 |
Condition |
Before the cleaning |
After the FC cleaning |
After drag wiping |
Input Power [mW] |
39.8 |
38.4 |
38.4 |
REFLPD dark offset [V] |
-0.0080 |
-0.0080 |
-0.0080 |
REFLPD locked [V] |
0.048 |
0.0437 |
0.046 |
REFLPD unlocked [V] |
6.41 |
6.39 |
6.37 |
|
|
|
|
Transmitted Power to DCPD1 (T) [mW] |
18.8 |
18.8 |
18.8 |
Transmitted Power to DCPD2 (R) [mW] |
18.1 |
18.2 |
18.2 |
FM2 transmission [mW] |
- |
- |
- |
CM1 transmission [mW] |
0.200 |
0.193 |
0.198 |
CM2 transmission [mW] |
0.204 |
0.204 |
0.205 |
Input BS transmission [mW] |
0.260 |
0.228 |
0.245 |
|
|
|
|
Cavity Finesse |
396.9 |
403.79 |
403.79 |
|
|
|
|
Junk Light Power (Pjunk) [mW] |
0.303 |
0.302 |
0.317 |
Coupled beam power (Pcouple) [mW] |
39.50 |
38.10 |
38.08 |
Mode Matching (Pcouple/Pin) [mW] |
0.992 |
0.992 |
0.992 |
Cavity reflectivity in power |
0.00112 |
0.000211 |
0.000206 |
Loss per mirror [ppm] |
111 |
35.9 |
34.8 |
Cavity transmission for TEM00 carrier
|
0.934 |
0.971 |
0.972 |
- TMS/FSR/Finesse change before/after cleaning [ELOG Entry]
Just a small change from the parameters before the bake.
No quantitative difference.
Method:
BB EOM produces the AM sidebands together with the PM sidebands.
Ideally, the PM sidebands does not produce the signal at the transmission, the output is dominated by the AM component.
This is only true when there is no lock offset. In reality the curve is contaminated by the PM-AM conversion by the
static offset or dynamic deviation of the locking point. So I had to take the central part of the TF and check the
dependence of the fit region and the finesse.
Before the cleaning: Finesse 396.9
After the cleaning: Finesse 403.8
To Do
- Placement of the DCPD housings
- Through-put test with DCPDs
- Transmission dependence on the incident power
(although the max incident is limited to ~35mW)
- Application of the first contact for the surface protection |
145
|
Tue Jun 18 10:01:11 2013 |
Koji | Optics | Characterization | Cavity Finesse analysis |
This is the analysis of the cavity finesse data taken on Apr/13/2013 (before baking), May/30/2013 (after baking), and Jun/02/2013 (after cleaning).
If we believe this result, baking contaminated the cavity, and the first contact removed it. That agrees with the power measurement of the transmitted light. |
151
|
Fri Aug 16 15:31:17 2013 |
Koji | Optics | Configuration | Mirror list for OMC(002) |
OMC(002)
Cavity Mirrors
FM1 (input coupler): A9
FM2 (output coupler): A13
CM1 (curved mirror close to FM1): C9 (PZT ASSY #6 / M6 /PZT21/C9)
CM2 (curved mirror close to FM2): C4 (PZT ASSY #4 / M11/PZT25/C4)
DCPD path
BS3 (BS for DCPDs): B10
QPD path
BS1 (input steering): E3
SM1 (steering mirror next to BS1): E5
BS2 (BS for QPD path): B9
SM2 (steering mirror next to BS2): E1
SM3 (steering mirror next to SM2): E2

|
152
|
Fri Aug 16 16:36:19 2013 |
Koji | Optics | General | Optics List |
Link to the "Mirror/PZT Characterization links"
Breadboard
BB1 OMC(001) OMC
BB2 OMC(002) OMC
BB3 -
BB4 OMC(003) OMC
BB5 -
BB6 -
Mounting Prisms:
M01
M02
M06 OMC(002) CM1 (PZT ASSY #6)
M07
M10 OMC(003) CM1 (PZT ASSY #5)
M11 OMC(002) CM2 (PZT ASSY #4)
M12
M13 OMC(003) CM2 (PZT ASSY #3)
M14
M15
M16 OMC(001) CM1 (PZT ASSY #1)
M17
M20 OMC(001) CM2 (PZT ASSY #2)
M21
M22
Mirror A:
A1 fOMC FM1
A2 Fullerton for the scattering measurement
A3 fOMC FM2
A4
A5
A6 OMC(003) FM2
A7 OMC(001) FM2
A8 OMC(001) FM1
A9 OMC(002) FM1
A10
A11
A12 OMC(003) FM1
A13 OMC(002) FM2
A14
Mirror B:
B1
B2
B3 OMC(001) BS2 (QPD)
B4
B5 OMC(003) BS2 (QPD)
B6
B7 OMC(001) BS3 (DCPD)
B8
B9 OMC(002) BS2 (QPD)
B10 OMC(002) BS3 (DCPD)
B11
B12 OMC(003) BS3 (DCPD)
Mirror C:
C1 OMC(003) CM1 (PZT ASSY #5)
C2 Fullerton for the scattering measurement
C3 OMC(003) CM2 (PZT ASSY #3)
C4 OMC(002) CM2 (PZT ASSY #4)
C5 OMC(001) CM2 (PZT ASSY #2)
C6 OMC(001)
CM1 (PZT ASSY #1)
C7 fOMC CM1
C8 fOMC CM2 -> OMC(002) CM1 (PZT ASSY #6)
C9 OMC(002) CM1 (PZT ASSY #6) -> BURNT
C10 (Liyuan tested)
C11 (Liyuan tested)
C12 curvature untested, faux OMC CM2
C13 curvature untested
Mirror E:
E1 OMC(002) SM2
E2 OMC(002) SM3
E3 OMC(002) BS1
E4 OMC(001) SM2
E5 OMC(002) SM1
E6
E7 OMC(003) BS1
E8 OMC(003) SM1
E9
E10 OMC(001) BS1
E11
E12 OMC(001) SM1
E13 OMC(003) SM2
E14
E15
E16 OMC(001) SM3
E17 OMC(003) SM3
E18
PZT:
PZT11
PZT12
PZT13
PZT14 OMC(003) CM1 (PZT ASSY #5)
PZT15 OMC(003) CM2 (PZT ASSY #3)
PZT21 OMC(002) CM1 (PZT ASSY #6)
PZT22
PZT23 OMC(001) CM2 (PZT ASSY #2)
PZT24
PZT25 OMC(002) CM2 (PZT ASSY #4)
PZT26 OMC(001) CM1 (PZT ASSY #1)
|
154
|
Wed Aug 21 08:31:21 2013 |
Koji | Optics | Characterization | H1 OMC cavity alignment |
Alignment of the H1 OMC cavity mirrors
- The cavity mirrors as well as the first steering mirror were aligned on the cavity side template.
- The locking of the cavity was not so stable as before. Some high freq (several hundreds Hz) disturbance makes the cavity
deviate from the linear range. This can be mitigated by turning off the HEPA units but this is not an ideal condition.
- FSR and TMS were measured.
FSR: 264.305MHz
TMS(V): 58.057MHz
TMS(H): 58.275MHz
These suggest the cavity length L and f_TMS/f_FSR (say gamma, = gouy phase / (2 pi) ) as
L=1.1343 m (1.132m nominal)
gamma_V = 0.219659 (0.21879 nominal)
gamma_H = 0.220484 (0.21939 nominal)
- the 9th modes of the carrier is away from the resonance 6-9 times of the line width (LW)
- the 13th modes of the lower f2 sideband are 11-15 LW away
- the 19th modes of the upper f2 sideband are 0.6-7 LW away
We still need precise adjustment of the gouy phase / cavity length, this was enough for the gluing of the flat mirrors |
155
|
Thu Aug 22 15:34:03 2013 |
Koji | Optics | General | OMC Cavity side gluing |
[Koji Jeff]
o BS1, FM1, FM2 prisms were glued
=> This fixed the unstability of the OMC locking
o Checked the spot position on the curved mirrors.
The height of the template was measured to be 6.16mm.
Using a sensor card, the heights of the spots on the curved mirrors were measured to be 7.4mm (CM1) and 7.9mm (CM2).
This means that the beam is ~1.5mm too low.
When the post clamps were applied to the PZT assemblies, the spot positions moved up a little bit (7.9mm - CM1, 8.2mm - CM2).
This is still ~1mm too low.
We can accommodate this level of shift by the curved mirror and the prisms.
We'll try other PZT assemblies to see if we can raise the beam height. |
159
|
Thu Aug 29 02:52:50 2013 |
Koji | Optics | Characterization | H1OMC Curved Mirror Alignment |
Cavity parameter was measured with 50V bias on PZT1 (CM1)
- PZT combination was changed: PZT1 #21 (PZT ASSY#6) / PZT2 #25 (PZT ASSY #4)
- 19th HOMs of the USB makes accidental resonance with the nominal cavity length.
Because of the mirror astigmatism, HOMs spreads more than the design.
In order to avoid these modes, the cavity length had to be moved from the nominal value (1.134m).
- The clearance between the fixture and the prism was limited. This prevents to shorten the cav length.
The cavity length was made longer about 10mm.
-----
Cavity parameter obtained from the pitch misalignment
Free Spectral Range (FSR): 261.777947 +/− 0.000299 MHz
Cavity roundtrip length: 1.145217 +/− 0.000001 m
Lock offset: 1.636183 +/− 0.238442 kHz
Transverse mode spacing (TMS): 57.581950 +/− 0.000163 MHz
TMS/FSR: 0.219965 +/− 0.000001
Cavity pole (1st order modes, avg and stddev): 353.465396 +/− 0.657630 kHz
Finesse (1st order modes, avg and stddev): 370.302940 +/− 0.688585
Carrier 9th-order HOM: -8.1 line width away
Upper Sideband 13th-order HOM: 13.3 LW away
Lower Sideband 19th-order HOM: 2.2 LW away
-----
Cavity parameter obtained from the pitch misalignment
Free Spectral Range (FSR): 261.777106 +/− 0.000226 MHz
Cavity roundtrip length: 1.145220 +/− 0.000001 m
Lock offset: 0.215937 +/− 0.183434 kHz
Transverse mode spacing (TMS): 57.875622 +/− 0.000116 MHz
TMS/FSR: 0.221087 +/− 0.000000
Cavity pole (1st order modes, avg and stddev): 356.862001 +/− 0.448102 kHz
Finesse (1st order modes, avg and stddev): 366.776766 +/− 0.460598
Carrier 9th-order HOM: -4.1 line width away
Upper Sideband 13th-order HOM: 19.1 LW away
Lower Sideband 19th-order HOM: 10.8 LW away
-----
We could avoid hitting the 19th modes of the 45MHz sidebands.
First accidental hit is the 28th order modes of the lower sideband.
Red: Carrier
Blue: Upper sideband (45MHz)
Green: Lower sideband (45MHz)

|
161
|
Fri Aug 30 12:14:50 2013 |
Koji | Optics | General | H1 OMC Cavity length adjustment |
Short conclusion:
The roundtrip cavity length for the H1 OMC was adjusted to be 1.145m
instead of 1.132m such that the 19th HOMs of the lower sideband do not get resonant together with the carrier.
Background:
The purpose of the OMC is to transmit the carrier TEM00 mode while anything else is rejected.
As the optical cavity has infinite numbers of resonant modes, what we practically do is to select
the roundtrip accumulated gouy phase so that low order higher order modes for the carrier
as well as the sidebands (including the TEM00 modes).
The nominal round trip length of the OMC is 1.132m. The curvature of the mirror is 2.575m.
The nominal ratio between the TMS and FSR is 0.218791 and 0.219385 (TMS_V/TMS_H= 0.9973)
for the vertical and horizontal modes. This split comes from the non-zero angle (~4deg) of incidence on the curved mirrors.
In reality, the TMS/FSR ratio depends on the true curvature of the mirror. More importantly, astigmatism
of the mirror changes the difference of the ratios for the vertical and horizontal modes.
The mirror astigmatism can either reduce or increase the split. between the TMSs. For example,
the L1 OMC showed the TMS/FSR ratio of (0.218822, 0.219218) for the vertical and horizontal modes.
TMS_V/TMS_H is 0.9982 which is 0.18% from the unity. This suggests, roughly to say, that 0.27% of the
astigmatism coming from the AOI of 4deg was partially compensated by the mirror astigmatism. This was lucky.
Something unlucky happened to the case for the first choice of the H1OMC curved mirrors.
TMS_V/TMS_H is 0.990 which is indeed 1% away from the unity. This actually caused some problem:
As the modes spreads too wide, the 19th modes became unavoidable. (see the picture below)

Red - carrier, Blue - upper sideband (+45MHz), Green - lower sideband
After the replacing one of the PZT assembly with another one, 1-TMS_V/TMS_H went down to 6%.
But still the 19th mode is on resonance. In order to shift the 19th mode from the resonance, the cavity length
had to be changed more than the range of the micrometer.
Simple simulation:
Attached Mathematica file calculates expected mode structure when the curved mirror position is
moved by DL (then the total roudtrip length changes 4*DL). This tells us that the 19th mode is
moved from the resonance by giving DL=-0.003 or DL=0.0025.
It was impossible to make the cavity short enough as the gluing fixture interferes with the curved mirror.
In fact, it was also impossible to make the cavity long enough as it was. Therefore PEEK shims with
the thickness of 1.5mm was inserted.

Result:
The FSR and TMS were measured with the longer cavity. 50V was applied to PZT1.
FSR: 261.775MHz
TMS_V: 57.575MHz
TMS_H: 57.880MHz
=> Cavity round trip length of 1.1452m
=> TMS/FSR = {0.219941, 0.221106}
The 19th modes for the lower sidebands are successfully moved from the carrier resonance.
The first accidental resonance is the lower sideband at the 28th order modes.

|
162
|
Fri Aug 30 12:22:56 2013 |
Koji | Optics | General | H1 OMC Cavity side UV gluing |
H1 OMC Cavity side optics was glued on the breadboard
Curved mirror gluing
- Applied the UV glues to CM1/CM2 prisms.
- Checked the spot positions on the curved mirrors
- Apply 50V to CM1
- Measure the FSR and TMS while the cavity was locked.
FSR: 261.70925MHz
TMS_V: 57.60500MHz
TMS_H: 57.94125MHz
=> Cavity round trip length of 1.1455m
=> TMS/FSR = {0.220111, 0.221395}
First accidental resonance is the lower sideband at 28th order modes.

Carrier 9th-order HOM: 2.9~7.6 line width away
Upper Sideband 13th-order HOM: 14.1-20.7 LW away
Lower Sideband 19th-order HOM: 3.3-13.1 LW away
- As this result was satisfactory, the UV illumination was zapped. It did not change the alignment. The cavity was kept locked during the illumination.
Peripheral optics gluing
- QPD path BS/Steering Mirrors were glued
- DCPD path BS was glued
The UV glue was applied to the optics.
Then the optics were placed on the breadboard along with the fixture.
Placed the dummy QPD/DCPD mount with the alignment disks.
The horizontal positions of the spots were well with in the horizontal range of the mounts.
The UV illumination was zapped. Checked the alignment again and no problem was found. |
163
|
Fri Aug 30 12:24:28 2013 |
Koji | Optics | Characterization | H1OMC Spot positions |
Beam heights on the diodes
DCPD1: 14.459mm -> With 1.5mm shim, the beam will be 0.038mm too low.
DCPD2: 14.221mm -> With 1.25mm shim, the beam will be 0.026mm too low.
QPD1: 14.691mm -> With 1.75mm shim, the beam will be 0.056mm too low.
QPD2: 14.379mm -> With 1.5mm shim, the beam will be 0.118mm too low. |
169
|
Mon Oct 14 13:40:16 2013 |
Koji | Optics | Characterization | H1 OMC Optical testing |
Since the middle of September, the optical tests of H1 OMC were took place.
Here is summary of the progress.
TEST1: FSR/FINESSE measurement before applying First Contact
TEST2: Power budget
MIrror cleaning with First Contact
TEST3: FSR/FINESSE measurement after First Contact application
TEST4: Power budget
TEST5: N/A
TEST6: HOM measurement @PZT V=0
TEST7: HOM measurement @PZT V=0-200
TEST8: DC response of the PZT
TEST9: AC response of the PZT
TEST10: PD/QPD alignment / output check
|
170
|
Mon Oct 14 15:50:55 2013 |
Koji | Optics | Characterization | H1 OMC Power budget |
LHO OMC power budget
Date |
2013/9/17 |
2013/9/17 |
2013/10/16 |
2013/10/22 |
Condition |
Before the cleaning |
After the cleaning |
Confirmation |
Confirmation |
Input Power [mW] |
35.2 |
35.4 |
34.54 |
34.9 |
REFLPD dark offset [V] |
-0.00763 |
-0.00763 |
-0.00772 |
-0.000759 |
REFLPD unlocked [V] |
0.0749 +/- 0.0005 |
0.067+/- 0.0005 |
0.0640+/-0.0005 |
0.0530+/-0.0001 |
REFLPD locked [V] |
5.49 +/- 0.01 |
5.55+/-0.01 |
5.28+/-0.01 |
5.26+/-0.01 |
|
|
|
|
|
Transmitted Power to DCPD1 (T) [mW] |
16.5 |
16.4 |
16.1 |
16.0 |
Transmitted Power to DCPD2 (R) [mW] |
15.9 |
16.2 |
15.55 |
15.55 |
FM2 transmission [mW] |
32.4 |
32.9+/-0.1 |
- |
- |
CM1 transmission [mW] |
0.166 |
0.169 |
0.164 |
0.165 |
CM2 transmission [mW] |
0.165 |
0.169 |
0.158 |
0.162 |
Input BS transmission [mW] |
0.234 |
0.218 |
0.230 |
0.227 |
|
|
|
|
|
Cavity Finesse |
373.114 |
373.114 |
373.114 |
373.114 |
|
|
|
|
|
Junk Light Power (Pjunk) [mW] |
0.489 |
0.434 |
0.422 |
0.332 |
Coupled beam power (Pcouple) [mW] |
34.71 |
34.97 |
34.12 |
34.57 |
Mode Matching (Pcouple/Pin) [mW] |
0.986 |
0.988 |
0.988 |
0.990 |
Cavity reflectivity in power |
0.00115 |
0.00119 |
0.00136 |
0.00199 |
Loss per mirror [ppm] |
122 |
124 |
134 |
167 |
Cavity transmission for TEM00 carrier
|
0.933 |
0.932 |
0.927 |
0.913 |
|
171
|
Tue Oct 15 18:50:08 2013 |
Koji | Optics | Characterization | QPD alignment |
1) Deburr the bottom surfaces of the QPD housings
2) Aligned the QPDs
QPD# QPD1 QPD2
Housing# #004 #008
Diode# #44 #46
Shim 1.75mm 001 1.25mm 001
-------------------------------------
Power Incident 125.7 uW 126.4 uW
Sum Out 80.1 mV 78.9 mV
Vertical Out + 3.4 mV 0 mV
Horizontal Out -23.7 mV -26 mV
SEG1 -15.6 mV -13.2 mV
SEG2 -13.1 mV -13.3 mV
SEG3 -29.0 mV -26.4 mV
SEG4 -23.2 mV -26.3 mV
-------------------------------------
Spot position X -13 um - 0.8 um (positive = more power on SEG1 and SEG4)
Spot position Y +93 um +107 um (positive = more power on SEG3 and SEG4)
-------------------------------------
Responsivity[A/W] 0.64 0.62
Q.E. 0.74 0.73
-------------------------------------
Arrangement of the segments
View from the beam
/ 2 | 1 X
|---+---|
\ 3 | 4 /
---------------
I(w,x,y) = Exp[-2 (x^2 + y^2)/w^2]/(Pi w^2/2)
(SEG_A+SEG_B-SEG_C-SEG_D)/(SEG_A+SEG_B+SEG_C+SEG_D) = Erf[sqrt(2) d/w]
d: distance of the spot from the center
w: beam width
|
172
|
Wed Oct 16 19:16:29 2013 |
Koji | Optics | Characterization | PD alignment |
shim 1.5mm 001/002 |
181
|
Tue Mar 25 17:10:10 2014 |
Koji | Optics | Characterization | OMC spot position estimation |
Spot positions were inferred from the photos |
182
|
Thu Apr 17 21:39:25 2014 |
Koji | Optics | General | More alignment |
STORY:
- The cavity mirrors have scattering spots. The cavity alignment should have been scanned to find a cavity mode to have lowest loss possible.
BTW, We only have horizontal dof for the alignment scan.
- After some struggle nice cavity mode was found. The cavity transmission was 96% for the ideally matched TEM00 carrier.
- It turned out that this imposed too much beam shift in the input beam (~2mm).
- This big shift induces a lot of trouble for the peripheral optics (PDs, QPDs, sterring mirrors).
- What should we do???
Analysis:
- The beam needed to go up between CM1 and CM2 to have the right spots on them. ("UP" is the input side of the OMC).
- This imposed the beam between FM1 and FM2 moved up. In other word, for the given alignment of the FMs by the template,
We needed to hit the upper part of the FMs to have the spots on the CMs up.
Solution:
- The above argument suggets that the nominal beam will give us the right spots on the CMs if we rotate the FMs.
Of course this induces the spot move on the FMs. But this should not be the issue as the most of the loss seems to come from the CMs.
- How much misalignment show we give to the FMs? We want to shift the beam by 2mm on the CMs.
The length of the optical lever is ~0.25m. Therefore the mialignment angle should be
theta = 2e-3/2/0.25 = 4e-3 rad = 4mrad.
The template pad has ~20mm separation. The thickness of the shim should be 20mm*4mrad = 80um
- Our aluminum foil seems to have the thickness of 30-40um. We can't have this minimum thickness on the template pad as there is not enough compression pressure
=> Just use a single layer of Al piece to shim the FMs.
Attempt:
- The shims were inserted at the upper pads of the FMs.
- Aligned the input beam and the CMs so that the spots on the CMs are approximately recovered.
- Measure the cavity power budget
Pin: 34.7mW
Refl PD: offset = -7.5mV, unlock = 6.07V, inlock = 89.7mV
Ptrans = 32.5mW
Ptrans(CM2) = 0.181mW
Ptrans(CM2) = 0.184mW
Assume finesse of 400
==>
Pin: 34.7mW
Pjunk: 0.534mW
Pcoupled: 34.1mW
Mode matching: 98.5%
Cavity reflectivity in power: 0.00061
Cavity transmission in power: 0.951 (This is not a best number but acceptable.)
Loss per mirror: 75.4ppm
FM power refl/trans: 0.9923 / 7630ppm
CM1 power refl/trans: 0.999882 / 42.8ppm
CM2 power refl/trans: 0.999881 / 43.5ppm
Total roundtrip loss of the cavity (Loss + CM leakage): 388ppm
Result:
How much the input beam is away from the left wall of the OMC breadboard?
40.88mm from the template edge
8.36mm between the template edge and the bread board
=> 32.52mm
How much should this number be? 32.94mm from the solidworks model => With in 0.5mm! Nice!
Next:
- Just in case plce all of the optics and check if the beam is delivered within the alignment range of the optics
|
183
|
Mon May 12 22:43:02 2014 |
Koji | Optics | Characterization | Measured FSR/TSM of the OMC cavity |
Data analysis of the FSR/TSM measruement last week.
1. FSR was measured with "the golden arches" technique.
FSR = 263.0686 MHz +/- 900Hz
Lcav = 1.1396 m --> 7.6 mm too long! (nominal 1.132m)
2. Transverse mode spacings for the vertical and horizontal modes were measured.
TMS/FSR = 0.219366 (V) / 0.220230 (H) (Predicted value with the current cavity length 0.2196/0.2202 very close!)
We want to make this to be ~0.219 (~3% less)
With the current parameters, the 19th-order lower sideband make the coincident resonance.

|
184
|
Wed May 14 02:15:15 2014 |
Koji | Optics | Characterization | FSR/TSM adjustment of the OMC cavity |
1. FSR was adjusted and measured with "the golden arches" technique again.
FSR = 264.8412 MHz +/- 1400Hz => Lcav = 1.13197 m. (nominal 1.132m)
2. Transverse mode spacings for the vertical and horizontal modes were measured.
TMS/FSR = 0.218144 (V) / 0.219748 (H)
This is almost perfect!
The 19th-order lower sideband hit the resonance. Next step is to glue some of the flat mirrors.

|
185
|
Fri May 16 00:13:36 2014 |
Koji | Optics | Characterization | Cavity mirror gluing part 1 |
BS1/FM1/FM2 for I1OMC were glued.
FM1 had to be intentionally rotated.
FM1 had to be intentionally shifted to avoid scattering spot.
Pin: 36.3 / Ptrans: 33.7 = Raw transmission 92.8%
Vunlock = 6.30 / Vlock = 0.120
Mode matching (estim) 0.98
Loss per mirror 84ppm
Cavity transmission 0.947
ummm
Tomorrow:
- Transmission needs to be optimized
- Apply 50V to a PZT
- Cavity FSR/HOM should be optimized
- gluing
Put a cover
Return power meter / DC supply |
186
|
Sat May 17 07:40:14 2014 |
Koji | Optics | Characterization | I1OMC cavity mirrors glued |
I1OMC cavity mirrors were glued.
FSR = 264.82MHz => Lcav = 1.132m (nominal 1.132m)
TMS/FSR for Vpzt1=Vpzt2=0: 0.2185 (V) and 0.2196 (H) (nominal 0.219)
aLIGO OMC: Power Budget 2014/5/16
<<<Measured Values>>>
Input Power: 35.7 [mW]
Transmitted Power through FM2: 33.5 [mW]
Transmitted Power through CM1: 0.188 [mW]
Transmitted Power through CM2: 0.192 [mW]
Reflection PD DC output (Unlocked): 6.2 [V]
Reflection PD DC output (Locked): 0.096 [V]
Reflection PD DC output (Dark Offset): -0.00745 [V]
Assumed cavity finesse : 400.
<<<Results>>>
Input Power: 35.7 [mW]
Uncoupled light Power (Junk light + sidebands): 0.575698 [mW]
Input TEM00 Carrier Power: 35.1243 [mW] (Ratio: 0.983874)
Cavity reflectivity (in power): 548.319 ppm
Cavity transmission (in power): 0.953756
Loss per mirror: 70.1183 ppm
FM1 power transmission: 7640.17 ppm
FM2 power transmission: 7640.17 ppm
CM1 power transmission: 43.2093 ppm
CM2 power transmission: 44.1337 ppm
|
187
|
Thu Jun 19 23:16:50 2014 |
Koji | Optics | General | All of the prisms have been glued |
- All of the PRISM mirrors have been glued
- 4 out of 5 beam dumps have been glued
TODO
EP30-2 gluing of the INVAR blocks for the PDs
PDs/QPDs need to be slightly lower -> order more shims
Remove the templates
Glue the last beam dump
Vibration test?
Bring the OMC to the bake lab
Vacuum baking
Bring it back to the OMC lab
Cabling / Wiring
Optical tests
Backscattering test
Packing / Shipping

|
188
|
Fri Jun 20 18:59:12 2014 |
Koji | Optics | General | All of the invar blocks have been glued |
All of the INVAR blocks have been glued.
I found thinner shims in the stock.
On Monday, the template will be removed.
EP30-2 7g mixed with 0.35g of 75-90um sphere
TODO
EP30-2 gluing of the INVAR blocks for the PDs
PDs/QPDs need to be slightly lower -> order more shims
Remove the templates
Glue the last beam dump
Vibration test?
Bring the OMC to the bake lab
Vacuum baking
Bring it back to the OMC lab
Cabling / Wiring
Optical tests
Backscattering test
Packing / Shipping |
189
|
Mon Jun 23 21:54:16 2014 |
Koji | Optics | General | All of the gluing completed |
The bottom-side templates were removed.
The last beam dump was removed
TODO
ICS entry
Bring the OMC to the bake lab
Vacuum baking
Bring it back to the OMC lab
Cabling / Wiring
VIbratin test
Optical tests
Backscattering test
Packing / Shipping
|
196
|
Sun Jul 6 02:45:56 2014 |
Koji | Optics | General | FSR Measurement |
3rd OMC FSR / Finesse measurement
RF AM was injected by detuning a HWP. |
197
|
Sun Jul 6 02:46:20 2014 |
Koji | Optics | Characterization | OMC power budget |
3rd OMC power budget (2014/7/2)
Input power: 34.8mW
REFLPD dark offset: -7.57mV
REFLPD unlocked: 6.22 V
REFLPD locked: 110mV
Transmitted Power: 16.8mW (T) and 15.9mW (R)
CM1 transmission: 0.176mW
CM2 transmission: 0.181mW
Cavity Finesse: 399.73
Junk light: 0.64mW (out of 34.8mW)
Coupled beam: 34.16 mW (out of 34.8mW)
Mode Matching: 0.982
Cavity reflectivity: 467ppm
Loss per mirror in ppm: 63.8ppm
Cavity transmission (for TEM00 carrier): 0.957
FM1: R = 0.992277, T = 7659.46
FM2: R = 0.992277, T = 7659.46
CM1: R = 0.999895, T = 41.5461
CM2: R = 0.999893, T = 42.7309
Compare the above number with the best result obtained during the alignment trials
Input power: 34.4mW
REFLPD dark offset: -7.5mV
REFLPD unlocked: 5.99 V
REFLPD locked: 104mV
Transmitted Power: Total 32.7mW (T+R)
CM1 transmission: 0.194mW
CM2 transmission: 0.194mW
Cavity Finesse: 400
Junk light: 0.631mW (out of 34.4mW)
Coupled beam: 33.77 mW (out of 34.4mW)
Mode Matching: 0.982
Cavity reflectivity: 255ppm
Loss per mirror in ppm: 39.7ppm
Cavity transmission (for TEM00 carrier): 0.968
|
198
|
Sun Jul 6 03:56:40 2014 |
Koji | Optics | Characterization | HOM measurement with PZT vol swept |
Cavity FSR/TMS measurement (2014/7/5) with PZT voltages swept from 0V to 200V (50V step) |
199
|
Sun Jul 6 08:31:14 2014 |
Koji | Optics | Characterization | HOM measurement with PZT vol swept |
3rd OMC, HOM diagram at PZT1=0V and PZT2=50V.
First coincidence with the carrier is the 32nd-order carrier mode. Very good. |