  252   Sun Mar 6 02:13:28 2016 KojiOpticsCharacterizationPD glass reflectionsOn friday, I removed a glass cover of a
G30655 with a PD can cutter.

When a beam shoots a Perkin Elmer/Excelitas
 P3048124.JPGP3048125.JPG 
  253   Sun Mar 13 21:22:27 2016 KojiElectronicsCharacterizationDark current measurement of the HQE PD and other PDsTransfered for RGA scan

B4 (C1-05) -> F1
C1 (C1-07) -> F2
  
  254   Sun Mar 13 22:02:09 2016 KojiOpticsCharacterizationHQEPD QE measurement (direct comaprison)Direct comparison of the PD responsibities

We
can expect 5% increase of the QE with the
  
  255   Sat Mar 26 01:49:48 2016 KojiOpticsCharacterizationHQEPD QECalibration
of the transimpedance

Use KEITHLEY 2450 as a calibrated
 QE1.pngQE2.png 
  256   Sat Mar 26 17:39:50 2016 KojiElectronicsCharacterizationHQEPD dark noiseDark noise measurement for 6 HQEPDs and
1 C30665. All of these showed sufficiently
low dark current noise levels compared with
 PD_dark_current.pdf 
  257   Sat Mar 26 18:22:24 2016 KojiElectronicsCharacterizationBaking / Contamination tests of the PDsFor the production of the aLIGO PDs, the
following transfer of the PDs were carried
out
  
  258   Tue Apr 5 18:14:55 2016 KojiGeneralLoan / LendingQPD Lending CrackleXiaoyue

QPD head
X-Z stage
  
  259   Tue Apr 5 18:22:40 2016 KojiElectronicsCharacterizationBaking / Contamination tests of the PDsPossible
reduction of the QE was observed after air-bake
at 75degC.
 QE_after_air_bake.pdf 
  260   Tue Apr 5 21:20:15 2016 KojiElectronicsCharacterizationMore dark noise measurementAll survived PDs have been measured.  PD_dark_current.pdf 
  261   Fri Jun 10 17:12:57 2016 KojiGeneralConfigurationL1 OMC DCPD replacementNew DCPD(T) = A1-23
DCPD(T) = DCPDB: extracted and accomodated
in CAGE-G SLOT1
  
  262   Fri Jul 22 22:24:05 2016 KojiGeneralGeneralHQEPD inventoryAs of Jul 22, 2016
As of Aug 11, 2016
As of Aug 16, 2016
  
  263   Fri Aug 12 14:58:17 2016 KojiGeneralConfigurationH1 OMC DCPD replacement Preparation of 3rd OMC for the
use in H1

New DCPD(T) = B1-01
  
  264   Mon Aug 15 10:09:10 2016 KojiGeneralGeneralPrev H1 OMC shipped to CITPrevious H1 OMC shipped from LHO to CIT

https://ics-redux.ligo-la.caltech.edu/JIRA/browse/Shipment-8196
  
  265   Mon Aug 22 12:58:16 2016 KojiGeneralGeneralUV bond samples -> Garilynn- FS base + Mounting Prism

- FS or SF2 1/2" piece + FS
or SF2 1/2" piece
  
  266   Tue Aug 23 23:36:54 2016 KojiOpticsCharacterizationInspection of the damaged CM1 (prev H1OMC)1. Calum and GariLynn checking the CM1
defect from the front side.
2. Same as above
 7x 
  267   Thu Aug 25 02:17:09 2016 KojiOpticsCharacterizationInspection of the damaged CM1 (prev H1OMC)Initial inspection results by Calum, et
al.
https://dcc.ligo.org/LIGO-E1600268
  
  268   Fri Sep 9 14:34:31 2016 KojiGeneralGeneralItem lendingTo 40m

First Contact Kit by Calum

Class A Kapton sheets
  
  269   Fri Sep 9 19:43:32 2016 KojiOpticsGeneralD1102211 OMC Diode Mount Glass Block went to DownsD1102211 OMC Diode Mount Glass Block (11pcs)
have been given to Calum@Downs
  
  270   Mon Nov 21 21:19:20 2016 KojiOpticsGeneralLWE NPRO Laser / Input Optics / Fiber Coupling- About 1.5 month ago, an 700mW LWE NPRO
has been brought to OMC Lab.

- The SOP can be found here.
  
  271   Wed Dec 7 19:18:10 2016 KojiOpticsGeneralLWE NPRO Laser / Input Optics / Fiber CouplingFIber Input Mount 132deg
Fiber output mount 275deg
-> 525mW P: 517mW S: 8mW extinction
  
  272   Wed Dec 7 21:18:35 2016 KojiGeneralGeneralOMC placed on the table / the beam roughly alignedThe OMC mode matching sled was fixed on
the nominal part of the table. Then the OMC
was located at the nominal position marked
 DSC_0082.jpgDSC_0084.jpg 
  273   Thu Dec 8 21:17:09 2016 KojiGeneralGeneralOMC placed on the table / the beam roughly alignedThe OMC cavity was locked. The alignment
was precisely adjusted. The mode matching
was optimized by the lens positions. The
  
  274   Thu Jan 19 20:57:53 2017 KojiSupplyGeneralPurchaseOrdered:

Office Depot
v AA battery Qty. 24
  
  275   Thu Feb 16 17:23:12 2017 KojiSupplyGeneralPurchase
BNC Patch Panel

8x
Amphenol Connex 112443 / Isolated BNC FEMALE
TO FEMALE BULKHEAD ISOLATED ADAPTER - 50
  
  276   Tue Mar 28 21:04:27 2017 KojiElectronicsCharacterizationPDH ampAttachment 1: PDH amp
RF part (before the preamp was installed)

Attachment 2:
 6x 
  277   Tue May 16 19:05:18 2017 KojiOpticsConfigurationOMC SN002 fix - temporary opticsWorking on the SN002 OMC fix. Checked the
inventory. I think I am using C8 mirror as
the new temporary CM1 and PZT24 as the new
  
  278   Fri May 26 21:53:20 2017 KojiGeneralConfigurationTrans RF PD setupRecent work

- DC output of the trans RF PD
was connected to the BNC patch panel. =>
  
  279   Tue Jun 6 00:49:48 2017 KojiGeneralConfigurationTrans RF PD setupLast week, I further worked on the RF system
to install 20dB coupler on the agilent unit
and setup the R channel. This allowed me
  
  280   Tue Jun 6 22:00:36 2017 KojiGeneralConfigurationTrans RF PD setup- Replaced the PZT with the one used from
the beginning. This must be PZT #21. After
the replacement, the spot positions look
  
  281   Fri Jun 23 01:58:11 2017 KojiOpticsGeneralOMC #002 Repair - CM1 gluing[Alena, Koji]

Jun 21: Alena and Koji
worked on gluing of the CM1 mirror on the
 6x 
  282   Fri Jun 23 10:55:07 2017 KojiOpticsGeneralDust layer on black glass beam dumps?I wondered why the black glass beam dumps
looked not as shiny as before. It was in
fact a layer of dusts (or contaminants) accumulated
 black_glass_dust.JPG 
  283   Sat Jul 1 15:29:57 2017 KojiOpticsGeneralBlack glass cleaning / Final bonding for the emergency repair for OMC #002[Alena, Koji] 

Report of the work on June
30.
 DSC_0347.jpgDSC_0348.jpgDSC_0350.jpgDSC_0352.jpgDSC_0354.jpg 
  284   Sat Jul 1 21:33:18 2017 KojiGeneralGeneralSome purchase notes- Forgot to close the cylinder valve...

v HEPA prefilter (20"x20"x1"
MERV 7)
  
  285   Wed Jul 5 16:59:44 2017 KojiGeneralGeneralThe OMC #002 was packed[Stephen Koji]

The OMC #002 was packed for the
transportation to Downs.
 DSC_0360.JPG 
  286   Sat Jul 29 18:44:38 2017 ranaElectronicsCharacterizationPDH ampattachment 6: DCPD preamp looks like the
opamp is wired for positive feedback?
  
  287   Sat Jul 29 21:42:51 2017 KojiElectronicsCharacterizationPDH ampThe polarities indicated in the right circuits
were opposite, obviously.
  
  288   Fri Sep 8 15:14:05 2017 KojiFacilityGeneralPreparation for the plumbing work[Steve, Aaron, Koji]

We've finished the preparation
for the forthcoming plumbing work on (nominally)
 7x 
  289   Mon Nov 27 20:24:24 2017 KojiGeneralGeneralA former LHO PD (Trans) removed from the OMC #002 for the shipment to StockholmAttachment 1: The PD was
removed from the transmission side of the
OMC #002 (former LHO OMC - the one blasted
 6x 
  290   Thu Nov 30 12:18:41 2017 StephenGeneralGeneralPreparation for Modal Testing on 4 DecemberNorna Robertson, Stephen Appert || 
29 Nov 2017, 2 pm to 4 pm  || 
227 Downs, CIT
  
  291   Thu Feb 22 20:21:02 2018 KojiOpticsCharacterizationaLIGO EOM testPOSTED
to 40m ELOG
  
  292   Mon Apr 2 17:27:04 2018 KojiOpticsCharacterizationaLIGO EOM test2nd optical test http://nodus.ligo.caltech.edu:8080/40m/13725   
  293   Thu May 3 21:45:58 2018 awadeGeneralLoan / LendingBorrowed toaster ovenI’ve borrowed the black and decker
toaster oven to dry some sonicated parts.
It is temporarly located in the QIL lab. 
 9CE80545-7A58-4236-B7E3-1EE6C4042DAA.jpeg 
  294   Sat May 5 22:51:04 2018 KojiOpticsGeneral3IFO EOM Optical testThe 3IFO EOM test performed at
the 40m. Result: 40m
ELOG 13819
  
  295   Tue May 15 19:53:45 2018 KojiOpticsGeneralEOM Q comparisonQs' were estimated with a lorentzian
function (eye fit)

  
  296   Wed May 30 16:40:38 2018 KojiMechanicsCharacterizationEOM mount stability testhttps://awiki.ligo-wa.caltech.edu/wiki/EOM_Mount_Stability   
  297   Wed May 30 17:44:23 2018 KojiOpticsCharacterization3IFO EOM surface check3IFO EOM dark microscope images courtesy
by GariLynn and Rich

Attachment1/2: Hole #1
 IMG_5756.JPGIMG_5757.JPGIMG_5758.JPGIMG_5759.JPGLooking_at_Hole2.png 
  298   Mon Jul 2 11:30:22 2018 KojiElectronicsCharacterization3IFO EOM impedance measurement[Rich Koji]

3IFO EOM (before any modification)
was tested to measure the impedance of each
 impedance_eom.pdfEOM_Z_DATA.zip 
  299   Mon Jul 2 12:29:01 2018 KojiElectronicsCharacterizationImpedances of individual components (3IFO EOM)[Rich Koji]

The impedances of the individual
components from the 3IFO EOM (before modification)
 impedance_coils.pdfimpedance_xtals.pdfq_coils.pdfcomponent_models.pdfliso_models.zip 
  300   Mon Jul 2 15:27:31 2018 Rich AbbottElectronicsGeneralWork on EOM (3rd IFO unit)Koji, Rich

We took apart the unit removed
from the 3rd IFO (Unit serial number aLIGO
 EOM3_aLIGO_3rdIfo.JPG 
  301   Tue Jul 3 12:07:47 2018 Rich AbbottElectronicsCharacterizationNotes on 3rd IFO EOMAttached please see my notes summarizing
the models for the electrodes and inductors
within the 3rd IFO EOM
 EOM_Analysis2.pdf 