ID |
Date |
Author |
Type |
Category |
Subject |
15
|
Sat Aug 11 00:59:14 2012 |
Koji | Facility | General | Laser Safety Barrier | It seemed that a laser safety barrier was installed today!?

|
16
|
Mon Aug 13 16:59:11 2012 |
Koji | Clean | General | Room Cleaning Log | Floor wiped with a wet wiper (Aug 13, 2012)
Floor wiped with a wet wiper (Aug 15, 2012)
Floor wiped with a wet wiper (Sep 25, 2012)
Air conditioning prefilter replaced (Sep 25, 2012)
Floor wiped with a wet wiper (Oct 01, 2012)
Floor wiped with a wet wiper (Nov 06, 2012) / ATF too
Floor wiped with a wet wiper (Jan 04, 2013)
Floor wiped with a wet wiper (Mar 23, 2013)
Floor wiped with a wet wiper (Apr 17, 2013)
Air conditioning prefilter replaced (Apr 17, 2013)
Floor wiped with a wet wiper (Jun 24, 2013)
Removing Vladimir's mess. Floor swept with a broom (Jun 26, 2013)
Completed removing Vladimir's mess. Floor swept with a wet wiper (Jun 27, 2013)
Air conditioning prefilter replaced (Sep 12, 2013)
Floor wiper head replaced. (Dec 10, 2013)
Floor wiped with a wet wiper (Dec 10, 2013)
Floor wiped with a wet wiper (Apr 1, 2014)
Air conditioning prefilter replaced (Dec 30, 2014)
Air conditioning prefilter replaced (some time in 2015...)
Floor wiped with a wet wiper (Dec 1, 2015)
Floor wiped with a wet wiper (Aug 23, 2016)
Air conditioning prefilter replaced (Aug 8, 2017) = 1 stock remains
Air conditioning prefilter replaced (Unkniwn) = no stock remains
Air conditioning prefilter replaced (Jul 25, 2022) = 5 stock remains
Floor wiped with a wet wiper (Mar 7, 2023) |
17
|
Mon Aug 13 17:01:35 2012 |
Koji | Clean | General | Particle Counts | Aug 13, 2012 / 0.5um 1000~2000/(0.1 cu ft) / 0.7um 400-600/(0.1 cu ft) by ATF particle counter (MET ONE 227A)
They are counts/(0.1 ft^3)! These numbers should be multiplied by 10 to know the particle "CLASS". |
18
|
Tue Aug 14 03:29:06 2012 |
Koji | Supply | General | Clean supply rack | Clean supplies & some cleaning tools are located at the right side of the entrance.
The file cabinet there was moved to the left side of the door, but will be removed eventually.

|
19
|
Wed Aug 22 20:16:43 2012 |
Koji | Facility | General | Workbenches have been installed / Clean room stools | Last Friday, new workbenches were installed. Vladimir got a new table and a cleanroom stool.

The other two workbenches were also nicely set.

|
20
|
Tue Sep 25 14:18:14 2012 |
Koji | Clean | General | Particle Counts | Particle counts
Before the prefilter is installed: 0.5um 1191cnts, 0.7um 346cnts
2:20 prefilter installed
2:25 0.5um 650 / 0.7um 255
3:00 0.5um 578 / 0.7um 99
4:00 0.5um 480 / 0.7um 102
5:00 0.5um 426 / 0.7um 76
They are counts/(0.1 ft^3)! These numbers should be multiplied by 10 to know the particle "CLASS". |
21
|
Mon Oct 1 16:06:55 2012 |
Koji | Clean | General | Particle Counts | 1. It turned out that the particle counter MET ONE 227A at ATF shows
(particle count)/(0.1 ft^3)
This means that the numbers I saw previously should be multiplied by 10.
So the nominal class of the room was 5000.
2. As our GT-321s have no diffuser, I borrowed a diffuser from 227A.
The diffuser actually increases the count. We need to buy them.
All the measurments below are performed with the diffuser and calibrated in Count/ft^3.
3. Measured the particle level without the HEPA running.
With diffuser: [cnt/ft^3]
|
GT-321 #1 |
GT-321 #2 |
227A |
0.3um |
152622 |
137511 |
- |
0.5um |
14706 |
14823 |
11860 |
Over Class 10000
4. The two HEPA fans are turned on at the speed "MED".
Basically no particles are detected in the HEPA booth.
With diffuser, inside of the HEPA booth:
|
GT-321 #1 |
GT-321 #2 |
227A |
0.3um |
0 |
0
|
- |
0.5um |
0 |
0 |
0 |
The particle level in the room (outside of the HEPA booth) is also improved
With diffuser, outside of the HEPA booth GT-321 #1:
0.3 um 18612
0.5 um 1728
5. The two HEPA fans are turned on at the speed "LOW".
Particle levels are still zero inside.
With diffuser, inside of the HEPA booth, GT-321 #1:
0.3 um 0
0.5 um 0
The particle level in the room (outside of the HEPA booth) is also improved
but the cleaning power for 0.3um seems degraded.
With diffuser, outside of the HEPA booth, GT-321 #1:
0.3 um 34488
0.5 um 1386
|
22
|
Fri Oct 5 03:39:58 2012 |
Koji | Optics | General | RoC Test setup | Based on Zach's experiment design, I wrote up a bit more detailed optical layout for the mirror test.

Item: Newfocus Fast PD
Qty.: 1
Mirror: Newfocus Fast PD
Mount: Post
Post: Post Holder (Newfocus)
Fork: Short Fork
Item: Thorlabs RF PD
Qty.: 1
Mirror: Thorlabs RF PD
Mount: Post
Post: Post Holder (Newfocus)
Fork: Short Fork
Item: Newfocus Broadband
Qty.: 1
Mirror: Newfocus EOM
Mount: Newfocus
Post: Custom Mount? or Pedestal X"?
Fork: Short Fork
Item: Newfocus Resonant
Qty.: 1
Mirror: Newfocus EOM
Mount: Newfocus
Post: Custom Mount? or Pedestal X"?
Fork: Short Fork
Item: ND Filter
Qty.: 2
Mirror: -
Mount: Thorlabs FIlter Holder
Post: Pedestal X"
Fork: Short Fork
Item: New Port Lens Kit 1"
Qty.: 1
Item: Thorlabs ND Kit
Qty.: 1
Item: Plano Convex Lens
Qty.: f=100, 100, 150, 200
Mirror: New Port (AR)
Mount: Thorlabs
Post: Post Holder (Newfocus)
Fork: Short Fork
Item: Bi-Convex Lens
Qty.: 75
Mirror: New Port (AR)
Mount: Post
Post: Post Holder (Newfocus)
Fork: Short Fork
Item: Flipper Mirror
Qty.: 1
Mirror: CVI Y1-10XX-45P
Mount: New Focus Flipper
Post: Pedestal X"
Fork: Short Fork
Item: Steering Mirror
Qty.: 8
Mirror: CVI Y1-10XX-45P
Mount: Suprema 1inch
Post: Pedestal X"
Fork: Short Fork
Item: PBS
Qty.: 3
Mirror: PBS 1inch BK7
Mount: Newport BS Mount
Post: Pedestal X"
Fork: Short Fork
Item: Knife Edge Beam Dump
Qty.: 4
Mirror: Thorlabs Knife Edge
Mount: Post
Post: Post Holder (Newfocus)
Fork: Short Fork
Item: Half Wave Plate
Qty.: 4
Mirror: CVI QWPO-
Mount: CVI
Post: Pedestal X"
Fork: Short Fork
Item: Quater Wave Plate
Qty.: 3
Mirror: CVI QWPO-
Mount: CVI
Post: Pedestal X"
Fork: Short Fork
Item: OMC Curved Mirror
Qty.: 2
Mirror: -
Mount: Suprema 0.5inch + Adapter
Post: Pedestal X"
Fork: Short Fork
Item: Prism Holder
Qty.: 1
Mirror: OMC Prism
Mount: Newport Prism Mount
Post: Pedestal X"
Fork: Short Fork
Item: CCD
Qty.: 1
Mirror: Thorlabs?
Mount: Thorlabs?
Post: Post Holder (Newfocus)
Fork: Short Fork |
23
|
Mon Oct 8 11:30:47 2012 |
Koji | Optics | General | EG&G 2mm photodiode angle response | EGE&G 2mm photodiode angle response measured by Sam T1100564-v1 |
24
|
Tue Oct 9 04:59:24 2012 |
Koji | Optics | General | OMC Test Optical Setup | 
|
25
|
Tue Oct 9 05:03:15 2012 |
Koji | Electronics | General | OMC Test Electronics Setup | 
|
26
|
Fri Oct 12 17:15:19 2012 |
Koji | Optics | General | Loan from the 40m / ATF |
HWP set
Optics: CVI QWPO-1064-08-2-R10
Mount: New Focus #9401
Post: Pedestal 2.5inch
- Returned: Oct 19, 2012 by KA
QWP set
Optics: CVI QWPO-1064-05-4-R10
Mount: New Focus #9401
Post: Pedestal 2.5inch
- Returned: Jan 17, 2013 by KA
- Faraday set
Optics: OFR IO-2-YAG-HP Returned: Mar 21, 2013 by KA
Mount: New Focus #9701 Returned: Apr 17, 2013 by KA
- Post: Pedestal (1.5+0.25inch)x2
Steering Mirror 1
Optics: CVI Y1-1037-45S
Mount: Newport Ultima U100-AC
Post: Pedestal 3inch
- Returned: Jan 17, 2013 by KA
Steering Mirror 2
Optics: CVI Y1-1037-45P
Mount: Newport Ultima U100-AC
Post: Pedestal 3inch
- Returned: Jan 17, 2013 by KA
Steering Mirror 3
Optics: New Focus 5104
Mount: Newport Ultima U100-AC
Post: Pedestal 3inch
- Returned: Jan 17, 2013 by KA
- Prism Mount
Mount: Thorlabs KM100P+PM1 2014/7/17
- Post: Pedestal 1.5+1+1/8inch
- 0.5" Mirror Mount
Mount: Newport U50-AReturned: Apr 17, 2013 by KA
Mount: Newport U50-A 2014/7/17
- Post: Pedestal 1.5+2inch
- Black Glass Beam Dump
- Optics: 1" sq. schott glass x3
- Mount: Custom Hexagonal 1"
- Post: Pedestal 3inch
PBS Set
05BC16PC.9 (PBS 1064 1000:1)
Mount: Custom Aluminum
- Returned: Jan 17, 2013 by KA
Lenses
KBX067.AR33 f=125mm
KPX106 f=200mm, KPX109 f=250mm unknown-coat
KPX088.AR33 f=75mm
KPX094.AR33 f=100mm
PLCX-C (BK7) 3863 (f=7.5m), 2060 (f=4.0m), 1545 (f=3.0m), 1030 (f=2.0m) non-coat
PLCX-UV (FS) 30.9 non-coat(!) f=60mm
- Returned: Jan 17, 2013 by KA
- Pedestals
1/4" x5, 1/8" x3, Returned: Jan 17, 2013 by KA
- 0.5" x1, 1.5" x1
Another loan from the 40m on Oct 17th, 2012
Minicircuits
Splitter ZFSC-2-5 x2
Filter SLP-1.9 x2 / BLP-1.9 x1/2 / SLP-5 x1
- Returned: Jan 17, 2013 by KA
- Connectors / Adaptors
SMA TEE x1 / SMA 50Ohm x 1 / BNC T x 10, Returned: Jan 17, 2013 by KA
SMA TEE x1 / SMA 50Ohm x 1Returned: May 20, 2013 by KA
Pomona Box x1, Returned: Jan 17, 2013 by KA
- Pomona Box x1
Power supply for New Focus Fast PD made by Jamie R Returned: Apr 17, 2013 by KA
BS-1064-50-1037-45S / Newport U100-A mount / 1"+2" Pedestal, Returned: Jan 17, 2013 by KA
BS-1064-50-1025-45P / Newport U100-A mount / 3/4" post + Base, Returned: Jan 17, 2013 by KA
BNC cable 21ft x2, Returned: Jan 17, 2013 by KA
- SMA Cable 6ft
Another loan from the 40m on Nov 21th, 2012
- Mounting Base Thorlabs BA-2 x 17
- Mounting Posts (phi=3/4", L=2.65", normal x15, and 1/4"-20 variant x2)
Yet another loan from the 40m on Jan 16th, 2013
V-groove Mounting Bases Custom. Qty.2Returned: Feb 25, 2013 by KA
Loan from ATF
32.7MHz EOM+Tilt aligner
Thorlabs Broadband EOM+Tilt aligner
Forks x 5Returned: Feb 25, 2013 by KA
JWIN Camera x 2 |
28
|
Tue Oct 16 15:50:09 2012 |
Koji | General | General | Work completed in August/September [!] |
- Work done
- Things ordered
- Office Depot
- [delivered] Office Depot(R) Brand Stretch Wrap Film, 20 x 1000 Roll, Clear / 445013
- [delivered] Eveready(R) Gold AA Alkaline Batteries, Pack Of 24 / 158448
- [delivered] Rubbermaid(R) Roller Sponge Mop / 921841
- [delivered] Rubbermaid(R) Roller Sponge Mop Replacement / 921858
- [delivered] Rubbermaid(R) Sanitizing Caddy, 10 Quarts, Yellow / 674125
- [delivered] Glad(R) Tall Kitchen Trash Bags, 13 Gallon, White, Box Of 28 / 269268
- Global Industrial Equipment
- [delivered] Extended Surface Pleated Cartridge Filter Serva-Cell Mp4 Slmp295 12X24X2 Gl WBB431699
- Global Industrial Equipment
- [delivered] Nexel Poly-Z-Brite Wire Shelving 30"W x 21"D x 63"H Nexel Poly-Z-Brite™ Wire Shelving Starter Unit WB189209
- [delivered] Stem Casters Set of (4) 5" Polyurethane Wheel, 2 With Brakes 1200 lb. Capacity WB500592
- Rack Solutions
- [delivered] Open Frame Server Racks
1 x 20" Depth Kit (Ideal for Audio/Video or Networking Racks) P/N: 111-1779
1 x 36U, Rack-111 Post Kit P/N: 111-1728
1 x Caster Kit for Open Frame RACK-111 P/N: 111-1731
- [delivered] 36U Side Panel Kit $199.99 P/N: 102-1775
- Rack shelf
- [delivered] 1 RMS 19 X 15 SINGLE SIDED NON-VENTED SHELF 70121637
- Work bench, Stools
- [not yet] 72"L X 30"W Production Bench - Phenolic Resin Square Edge-Blue Form attached WB237381LBL
- [not yet] 72"W Lower Shelf For Bench - 15"D- Blue Form attached WB606951
- [not yet] ESD-Safe Vinyl Clean Room Stool with Nylon Base with Drag Chain Blue Form attached WBB560852
- P Touch
- [delivered] Brother PT-2030 Desktop Office Labeler Punch-out product 672828
- [delivered] Brother(R) TZe-241 Black-On-White Tape, 0.75 x 26.2 Punch-out product 239384
- [delivered] Brother(R) TZe-231 Black-On-White Tape, 0.5 x 26.2 Punch-out product 239400
- UV light guide
- [delivered] Fiber Optic Single Light Guide 5mm OD X 3mm ID X 1M L Note: This light guide can be used with MKIII UV Cure unit. OLB1081
- Gloves (7.5, 8.0)
- [delivered] GLOVE ACCTCH NR-LTX SZ7.5 PK25 Punch-out product 79999-306
- [delivered] GLOVE ACCTCH NR-LTX SZ8 PK25 Punch-out product 79999-308
- Lab coat (L,XL), Sticky Mat, Shoe Covers (L, XL), Cap, Mask
- [delivered] LAB XP WH EL WR.COLL. NP L30EA Punch-out product 82007-618
- [delivered] LAB XPWH EL WR.COLL. NP XL30EA Punch-out product 82007-620
- [delivered] VWR MAT ADHESIVE 30L 18X36 BLU Punch-out product 21924-110 (This was too small)
- [delivered] VWR SHOECVR NSKID AP 2XL 150PR Punch-out product 414004-651
- [delivered] VWR SHOECVR NSKID AP XL 150PR Punch-out product 414004-650
- [delivered] CAP BOUFFANT 24IN RAYON CS500 Punch-out product 10843-053
- [delivered] MASK VLTC TIES N/STRL PK50 Punch-out product 10869-020
- VWR
- [delivered] FACE SHIELD UVC-803 Supplier: UVP 33007-151
- [Delivered] Laser safety glasses
|
29
|
Tue Oct 16 15:51:01 2012 |
Koji | General | General | Plan Update: October [!] | Completed work of the previous months: [Jul] [Aug] [Sep] [Oct] [Nov] [Dec]
Facility/Supplies
- Work done
- Particle Level measured / HEPA activated [ELOG]
- Particle counter peripherals arrived ~Oct 12.
- Making the OMC optical test setup [ELOG] [ELOG] [ELOG] [ELOG] [ELOG]
- OMC Bread board dimensions / weights measurement by Jeff and Jam [ELOG]
- UV epoxy has arrived - stored in a freezer in the office
- Laser sign installed during my trip by Peter/Eric
- OMC design downselect [DCC Link]
- Things to buy
- Things to be done
- Cavity ref/trans/finesse
- PD Q.E. & Reflectivity measurement vs incident angle
- Work in progress
- RoC measurement
- R&T measurement
- Wedge measurement
- Work to be done
- Replacing a file cabinet next to the south wall by a lockable cabinet
- Additional clean supplies ~ glove 8.5,9,9.5
- Stainless bats
- Ion gun safety issues: https://dcc.ligo.org/cgi-bin/private/DocDB/ShowDocument?docid=88631
- Design
- Test
- Continuous monitoring of the particle level
- Note: Optical Table W96" x D48" x H27"
Beaurocracy
- Laser SOP / HV use? / UV?
- Procedures to be decided
- PZT alignment
- UV glue? (heat) / gluing test
- Balance
- N2 cylinder/lines/filter
- Design
- Mode design for HAM6 layout
- Finalization of scattering paths
- Things to be decided / confirmed
- How to handle optics / assemblies (Talk to the prev people)
- First contact? (Margot: applicable to a short Rc of ~2.5m)
- Gluing templates to be designed (how to handle it?)
- PDs
- Things to buy
- Need to buy a fiber for mode cleaning?
- Mode content of the ELIGO dark beam?
- Jitter noise?
- How to determine the design?
- Why Fused Silica? (How much is the temp fluctuation in the chamber?)
- How to align the cavity mirrors, input mirrors, QPDs, PDs, beam dumps.
- PZTs @LLO
Electronics
- Thorough scrutinization of cabling / wiring / electronics
- ELIGO OMC Wiring diagram D070536-A2
- Occupies 2 DB25s -> They were anchored on the sus cage
- Preamps for DCPDs will be fixed on the ISI table
-> DB25 for the DCPDs will be anchored on the table
- Use longer thin cables for the DCPDs in order to route them through the suspension stages
- Turn the heater cable to the one for the other PZT
- Electronics / CDS electronics / software
- Things to be tested
- QPD/PD pre-selections (QE/noise)
- PD preamp design (Rich)
- Functionality test of QPD/PD/PZT
Shipping, storage etc
|
30
|
Wed Oct 17 20:36:04 2012 |
Koji | Optics | General | RoC test cavity locked | The RoC test setup has been built on the optical table at ATF.
The cavity formed by actual OMC mirrors have been locked.
The modulation frequency of the BB EOM was swept by the network analyzer.
A peak at ~30MHz was found in the transfer function when the input beam was misaligned and clipping was introduced at the transmission PD.
Without either the misalignment or the clipping, the peak disappears. Also the peak requires these imperfections to be directed in the same way
(like pitch and picth, or yaw and yaw). This strongly suggests that the peak is associated with the transverse mode.
The peak location was f_HOM = 29.79MHz. If we consider the length of the cavity is L=1.20m, the RoC is estimated as
RoC = L / (1 - Cos[f_HOM/(c/2/L) * PI]^2)
This formula gives us the RoC of 2.587 m.
I should have been able to find another peak at f_FSR-f_TMS. In deed, there was the structure found at 95MHz as expected.
However, the peak was really weak and the location was difficult to determine as it was coupled with the signal from residual RFAM.
The particle level in the clean booth was occasionally measured. Every measurement showed "zero".
To be improved:
- The trans PD is 1801 which was found in ATF with the label of the 40m. It turned out that it is a Si PD.
I need to find an InGaAs PD (1811, 1611, or my BBPD) or increase the modulation, or increase the detected light level.
(==> The incident power on 1810 increased. Oct 17)
- The BS at the transmission is actually Y1-45P with low incident angle. This can be replaced by 50% or 30% BS to increase the light on the fast PD.
(==> 50% BS is placed. Oct 17)
- I forgot to put a 50ohm terminator for the BB EOM.
(==> 50Ohm installed. Oct 17)
- A directional coupler could be used for the BBEOM signal to enhance the modulaiton by 3dB.
- The mode matching is shitty. I can see quite strong TEM20 mode.
- Use the longer cavity? L=1.8m is feasible on the table. This will move the peak at 27MHz and 56MHz (FSR=83MHz). Very promising.
(==> L=1.8m, peak at 27MHz and 56MHz found. Oct.17)
|
31
|
Thu Oct 18 20:23:33 2012 |
Koji | Optics | Characterization | Improved measurement | Significant improvement has been achieved in the RoC measurement.
- The trans PD has much more power as the BS at the cavity trans was replaced by a 50% BS. This covers the disadvantage of using the a Si PD.
- The BB EOM has a 50Ohm terminator to ensure the 50Ohm termination at Low freq.
- The length of the cavity was changed from 1.2m to 1.8m in order to see the effect on the RoC measurement.
By these changes, dramatic increase of the signal to noise ratio was seen.
Now both of the peaks corresponds to the 1st-order higher-order modes are clearly seen.
The peak at around 26MHz are produced by the beat between the carrier TEM00 and the upper-sideband TEM01 (or 10).
The other peak at around 57MHz are produced by the lower-sideband TEM01 (or 10).
Peak fitting
From the peak fitting we can extract the following numbers:
- Cavity FSR (hence the cavity length)
- Cavity g-factor
- Approximate measure of the cavity bandwidth
Note that the cavity itself has not been touched during the measurement.
Only the laser frequency and the incident beam alignment were adjusted.
The results are calculated by the combination of MATLAB and Mathemaica. The fit results are listed in the PDF files.
In deed the fitting quality was not satisfactory if the single Lorentzian peak was assumed.
There for two peaks closely lining up with different height. This explained slight asymmetry of the side tails
This suggests that there is slight astigmatism on the mirrors (why not.)
The key points of the results:
- FSR and the cavity length: 83.28~83.31MHz / L=1.799~1.800 [m] (surprisingly good orecision of my optics placement!)
- Cavity g-factor: Considering the flatness of the flat mirror from the phase map, the measured g-factors were converted to the curvature of the curved mirror.
RoC = 2.583~4 [m] and 2.564~7 [m]. (Note: This fluctuation can not be explained by the statistical error.)
The mode split is an order of 10kHz. This number also agrees with the measurement taken yesterday.
If the curved mirror had the nominal curvature of 2.5m, the flat mirror should have the curvature of ~20m. This is very unlikely.
- Approximate cavity line width: FWHM = 70~80kHz. This corresponds to the finesse of ~500. The design value is ~780.
This means that the locking offset is not enough to explain the RoC discrepancy between the design and the measurement.
|
32
|
Wed Nov 7 01:28:20 2012 |
Koji | Optics | Characterization | Wedge angle test (A1) | Wedge angle test
Result: Wedge angle of Prism A1: 0.497 deg +/- 0.004 deg
Principle:
o Attach a rail on the optical table. This is the reference of the beam.
o A CCD camera (Wincam D) is used for reading out spot positions along the rail.
o Align a beam path along the rail using the CCD.
o Measure the residual slope of the beam path. (Measurement A)
o Insert an optic under the test. Direct the first surface retroreflectively. (This means the first surface should be the HR side.)
o Measure the slope of the transmitted beam. (Measurement B)
o Deflection angle is derived from the difference between these two measurements.
Setup:

o An Al plate of 10" width was clamped on the table. Four other clamps are located along the rail to make the CCD positions reproducible.
o A prism (Coating A, SN: A1) is mounted on a prism mount. The first surface is aligned so that the reflected beam matches with the incident beam
with precision of +/-1mm at 1660mm away from the prism surface. ==> precision of +/- 0.6mrad
o In fact, the deflection angle of the transmission is not very sensitive to the alignment of the prism.
The effect of the misalignment on the measurement is negligible.
o Refractive index of Corning 7980 at 1064nm is 1.4496
Result:
Without Prism
Z (inch / mm), X (horiz [um] +/-4.7um), Y (vert [um] +/-4.7um)
0” / 0, -481.3, -165.1
1.375" / 34.925, -474.3, -162.8
3" / 76.2, -451.0, -186.0
4.375" / 111.125, -432.5, -181.4
6" / 152.4, -432.5, -181.4
7.375" / 187.325, -330.2, -204.6
9" / 228.6, -376.7, -209.3
With Prism / SN of the optic: A1
Z (inch / mm), X (horiz [um] +/-4.7um), Y (vert [um] +/-4.7um)
0” / 0, -658.3, -156.8
1.375" / 34.925, -744.0, -158.1
3" / 76.2, -930.0, -187.4
4.375" / 111.125, -962.6, -181.4
6" / 152.4, -1190.4, -218.6
7.375" / 187.325, -1250.9, -232.5
9" / 228.6, -1418.3, -232.5
Analysis:
Wedge angle of Prism A1: 0.497 deg +/- 0.004 deg
[Click for a sharper image]
|
33
|
Wed Nov 7 20:21:42 2012 |
Koji | General | General | Work completed in October [!] | Completed work of the previous months: [Jul] [Aug] [Sep] [Oct] [Nov] [Dec]
- Work done
- Particle Level measured / HEPA activated [ELOG]
- Particle counter peripherals arrived ~Oct 12.
- Making the OMC optical test setup [ELOG] [ELOG] [ELOG] [ELOG] [ELOG]
- OMC Bread board dimensions / weights measurement by Jeff and Jam [ELOG]
- UV epoxy has arrived - stored in a freezer in the office
- Laser sign installed during my trip by Peter/Eric
- OMC design downselect [DCC Link]
|
34
|
Wed Nov 7 20:44:11 2012 |
Koji | General | General | Plan Update: November [!] | Completed work of the previous months: [Jul] [Aug] [Sep] [Oct] [Nov] [Dec]
- Work in progress
- R&T measurement
- Wedge measurement
- Work to be done
- QPD/PD pre-selections (QE/noise)
-
-
-
- Misc. / Beaurocracy?
- Continuous monitoring of the particle level
- Replacing a file cabinet next to the south wall by a lockable cabinet
- Ion gun safety issues: https://dcc.ligo.org/cgi-bin/private/DocDB/ShowDocument?docid=88631
- Laser SOP / HV use? / UV?
- Things delivered
- Things ordered
- Power strips Tripp Lite PS3612 (Ordered Nov. 8, Delivered Nov. 12)
- Kapton tapes (1in x 6, 1/2in x 12 Delivered Nov. 15)
- Sticky Mats (VWR 18888-216 Delivered Nov. 12 and 21992-042)
- Duck tape (PK3) (Delivered Nov. 12)
- Wipers 12"x12" 2ply x 119 pairs x case15 (Delivered Nov. 12)
- Syringes (1mL&2mL) & Needles (20G x dozen)
- Stainless trays with cover (Steve Delivered Nov. 12)
- Gold Plated allen keys (Steve Delivered Nov. 12)
- Forceps (Delivered Nov. 12) / Tweezers / Scissors (Delivered Nov. 12)
- Things to buy / get
- OMC testing optics / opto-mechanics
- Black Glass / Black Glass holder / AR ==> Some at the 40m, some from LLO
- Ionized air blow
- N2 or Air cylinder: 4N - UHP or 5N - Research Grade. (... steal from Downs)
- Clean tools, tray, storage
- Supply
- Additional clean supplies ~ glove 8.5,9,9.5
- Stainless bats / Pure solvents (Metha / Aceton / Iso) / Syringes / Lint free cloth / Paper lens tissue
- Lab coats
- ATF
- Tefron tape
- Thorlabs 8-32 screw kit / Thorlabs HW-KIT1
- Pedestal Shims - Newport
- Things to be done
- Cavity ref/trans/finesse
- PD Q.E. & Reflectivity measurement vs incident angle
- Functionality test of QPD/PD (PeterK) /PZT
- Procedures to be decided
- PZT alignment
- UV glue? (heat) / gluing test
- Balance
- N2 cylinder/lines/filter
- Shipping procedure: New shipping cage design on going (Jeff) => Plastic box similar to COC
- Design
- Solidworks raytracing model
- Mode design for HAM6 layout
- Things to be decided / confirmed
- How to handle optics / assemblies (Talk to the prev people)
- First contact? (Margot: applicable to a short Rc of ~2.5m)
- Gluing templates to be designed (how to handle it?)
- Jitter noise?
- How to align the cavity mirrors, input mirrors, QPDs, PDs, beam dumps.
Electronics ==> Rich |
35
|
Thu Nov 8 13:24:53 2012 |
Koji | Optics | Characterization | More wedge measurement | A1
Horiz Wedge 0.497 +/- 0.004 deg
Vert Wedge 0.024 +/- 0.004 deg
A2
Horiz Wedge 0.549 +/- 0.004 deg
Vert Wedge 0.051 +/- 0.004 deg
A3
Horiz Wedge 0.463 +/- 0.004 deg
Vert Wedge 0.009 +/- 0.004 deg
A4
Horiz Wedge 0.471 +/- 0.004 deg
Vert Wedge 0.019 +/- 0.004 deg
A5
Horiz Wedge 0.458 +/- 0.004 deg
Vert Wedge 0.006 +/- 0.004 deg |
36
|
Thu Nov 8 19:47:55 2012 |
Koji | Electronics | Configuration | Solder for PZTs | Rich saids:
I have ordered a small roll of solder for the OMC piezos.
The alloy is: Sn96.5 Ag3.0 Cu0.5 |
37
|
Thu Nov 8 19:52:57 2012 |
Koji | Optics | General | How to apply UV epoxy | KA's question:
Do you know how to apply this epoxy?
Do we need a plunger and a needle for this purpose?
Nic saids:
When we did it with Sam, I seem to remember just squirting some on some foil then dabbing it on with the needle. |
38
|
Thu Nov 8 20:12:10 2012 |
Koji | Optics | Configuration | How many glass components we need for a plate | Optical prisms 50pcs (A14+B12+C6+E18)
Curved Mirrors 25pcs (C13+D12)
|
Qty |
Prisms
|
Curved |
No BS OMC |
Wedge tested |
Coating A: IO coupler |
|
14 |
0 |
2 prisms |
5/5 |
Coating B: BS 45deg |
|
12 |
0 |
2 prisms |
0/5 |
Coating C: HR |
|
6 |
13 |
2 curved |
|
Coating D: Asym. output coupler |
|
0 |
12 |
- |
|
Coating E: HR 45deg |
|
18 |
0 |
4 prism (1 trans + 3 refl) |
0/3 |
D1102209 Wire Mount Bracket |
25 |
|
|
4 |
|
D1102211 PD Mount Bracket |
30 |
|
|
8 |
|
|
39
|
Fri Nov 9 00:43:32 2012 |
Koji | Optics | Characterization | Further more wedge measurement | Now it's enough for the first OMC (or even second one too).
Today's measurements all distributed in theta>0.5deg. Is this some systematic effect???
I should check some of the compeled mirrors again to see the reproducibility...
A1 Horiz Wedge 0.497039 +/- 0.00420005 deg / Vert Wedge 0.02405210 +/- 0.00420061 deg
A2 Horiz Wedge 0.548849 +/- 0.00419993 deg / Vert Wedge 0.05087730 +/- 0.00420061 deg
A3 Horiz Wedge 0.463261 +/- 0.00420013 deg / Vert Wedge 0.00874441 +/- 0.00420061 deg
A4 Horiz Wedge 0.471536 +/- 0.00420011 deg / Vert Wedge 0.01900840 +/- 0.00420061 deg
A5 Horiz Wedge 0.458305 +/- 0.00420014 deg / Vert Wedge 0.00628961 +/- 0.00420062 deg
B1 Horiz Wedge 0.568260 +/- 0.00419988 deg / Vert Wedge -0.00442885 +/- 0.00420062 deg
B2 Horiz Wedge 0.556195 +/- 0.00419991 deg / Vert Wedge -0.00136749 +/- 0.00420062 deg
B3 Horiz Wedge 0.571045 +/- 0.00419987 deg / Vert Wedge 0.00897185 +/- 0.00420061 deg
B4 Horiz Wedge 0.563724 +/- 0.00419989 deg / Vert Wedge -0.01139000 +/- 0.00420061 deg
B5 Horiz Wedge 0.574745 +/- 0.00419986 deg / Vert Wedge 0.01718030 +/- 0.00420061 deg
E1 Horiz Wedge 0.600147 +/- 0.00419980 deg / Vert Wedge 0.00317778 +/- 0.00420062 deg
E2 Horiz Wedge 0.582597 +/- 0.00419984 deg / Vert Wedge -0.00537131 +/- 0.00420062 deg
E3 Horiz Wedge 0.592933 +/- 0.00419982 deg / Vert Wedge -0.01082830 +/- 0.00420061 deg
-------
To check the systematic effect, A1 and B1 were tested with different alignment setup.
A1 Horiz Wedge 0.547056 +/- 0.00419994 deg / Vert Wedge 0.0517442 +/- 0.00420061 deg
A1 Horiz Wedge 0.546993 +/- 0.00419994 deg / Vert Wedge 0.0469938 +/- 0.00420061 deg
A1 Horiz Wedge 0.509079 +/- 0.00420003 deg / Vert Wedge 0.0240255 +/- 0.00420061 deg
B1 Horiz Wedge 0.547139 +/- 0.00419994 deg / Vert Wedge 0.0191204 +/- 0.00420061 deg
|
40
|
Sat Nov 17 02:31:34 2012 |
Koji | Optics | Characterization | Mirror T test | Mirror T test
The mirror was misaligned to have ~2deg incident (mistakenly...) angle.
C1: Ptrans = 7.58uW, Pinc = 135.0mW => 56.1ppm
C1 (take2): Ptrans = 7.30uW, Pinc = 134.4mW => 54.3ppm
C2: Ptrans = 6.91uW, Pinc = 137.3mW => 50.3ppm
C3: Ptrans = 6.27uW, Pinc = 139.7mW => 44.9ppm
C4: Ptrans = 7.62uW, Pinc = 139.3mW => 54.7ppm
C5: Ptrans = 6.20uW, Pinc = 137.5mW => 45.1ppm
A1: Ptrans = 1.094mW, Pinc = 133.6mW => 8189ppm |
41
|
Mon Nov 19 13:33:14 2012 |
Koji | Optics | Characterization | Resuming testing mirror RoCs | In order to resume testing the curvatures of the mirrors, the same mirror as the previous one was tested.
The result looks consistent with the previous measurement.
It seems that there has been some locking offset. Actually, the split peaks in the TF@83MHz indicates
the existence of the offset. Next time, it should be adjusted at the beginning.
Curved mirror SN: C1
RoC: 2.5785 +/- 0.000042 [m]
Previous measurements
=> 2.5830, 2.5638 => sqrt(RoC1*RoC2) = 2.5734 m
=> 2.5844, 2.5666 => sqrt(RoC1*RoC2) = 2.5755 m |
42
|
Mon Nov 26 01:40:00 2012 |
Koji | Optics | Characterization | More RoC measurement | C1: RoC: 2.57845 +/− 4.2e−05m
C2: RoC: 2.54363 +/− 4.9e−05m
C3: RoC: 2.57130 +/− 6.3e−05m
C4: RoC: 2.58176 +/− 6.8e−05m
C5: RoC 2.57369 +/− 9.1e−05m
==> 2.576 +/- 0.005 [m] (C2 excluded) |
43
|
Thu Nov 29 21:18:23 2012 |
Koji | Optics | General | OMC Mounting Prisms have come | 

|
44
|
Tue Dec 18 20:04:40 2012 |
Koji | Optics | Characterization | Prism Thickness Measurement | The thicknesses of the prism mirrors (A1-A5) were measured with micrometer thickness gauge.
Since the thickness of the thinner side (side1) depends on the depth used for the measurement,
it is not accurate. Unit in mm.
A1: Side1: 9.916, Side2: 10.066 => derived wedge angle: 0.43deg
A2: Side1: 9.883, Side2: 10.065 => 0.52
A3: Side1: 9.932, Side2: 10.062 => 0.38
A4: Side1: 9.919, Side2: 10.060 => 0.40
A5: Side1: 9.917, Side2: 10.058 => 0.40

|
45
|
Wed Dec 19 18:47:03 2012 |
Koji | Clean | General | First Contact Training with Margot | Steve and I visited Margot to have a training session for application of First Contact on optics.
- Make "thick" layer of first contact. It becomes thin when it gets dried.
- Apply more FC once a peek sheet is placed on the FC
- Wait for drying (~15min)
- Rip off the FC layer by pulling a peek tab. Make sure the ionized N2 is applied during ripping.
- Margot has a Dark Field Microscope. We checked how the dusts are removed from the surface.
There are many dusts on the mirror even if they are invisible. First Contact actually removes
these dusts very efficiently. Margot told us that even carbonhydrates (like finger prints) can be removed by FC.
|
46
|
Wed Dec 26 14:33:33 2012 |
Koji | Facility | General | Lase Interlock wired | Two switches are connected in series. |
47
|
Mon Dec 31 01:45:04 2012 |
Koji | General | General | Work completed in Nov and Dec [!] | Completed work of the previous months: [Jul] [Aug] [Sep] [Oct] [Nov] [Dec]
- Things delivered
- The ionized gun used in the clean room at Downs: made by Terra Universal.com (Jeff's room)
http://www.terrauniversal.com/static-control/ionizing-blow-off-guns.php
- Flow path: N2 cylinder - Filter - Gun (Jeff's room)
- Power strips Tripp Lite PS3612 (Ordered Nov. 8, Delivered Nov. 12)
- Kapton tapes (1in x 6, 1/2in x 12 Delivered Nov. 15)
- Sticky Mats (VWR 18888-216 Delivered Nov. 12 and 21992-042)
- Duck tape (PK3) (Delivered Nov. 12)
- Wipers 12"x12" 2ply x 119 pairs x case15 (Delivered Nov. 12)
- Syringes (1mL&2mL) & Needles (20G x dozen)
- Stainless trays with cover (Steve Delivered Nov. 12)
- Gold Plated allen keys (Steve Delivered Nov. 12)
- Forceps (Delivered Nov. 12) / Tweezers / Scissors (Delivered Nov. 12)
- OMC testing optics / opto-mechanics
- SolidWorks raytracing model
- Mode design for HAM6 layout [Zach]
- Black Glass / Black Glass holder / AR ==> Some at the 40m, some from LLO
- Ionized air blow
- N2 or Air cylinder: 4N - UHP or 5N - Research Grade. (... steal from Downs)
|
48
|
Mon Dec 31 03:10:09 2012 |
Koji | Optics | General | SolidWorks model of the OMC breadboard | |
49
|
Mon Dec 31 03:11:45 2012 |
Koji | Optics | Characterization | Further more RoC measurement | Total (excluding C2, C7, C8): 2.575 +/- 0.005 [m]
New results
C6: RoC: 2.57321 +/− 4.2e-05m
C7: RoC: 2.56244 +/− 4.0e−05m ==> Polaris mount
C8: RoC: 2.56291 +/− 4.7e-05m ==> Ultima mount
C9: RoC: 2.57051 +/− 6.7e-05m
Previous results
C1: RoC: 2.57845 +/− 4.2e−05m
C2: RoC: 2.54363 +/− 4.9e−05m ==> Josh Smith @Fullerton for scattering measurement
C3: RoC: 2.57130 +/− 6.3e−05m
C4: RoC: 2.58176 +/− 6.8e−05m
C5: RoC 2.57369 +/− 9.1e−05m |
50
|
Wed Jan 2 07:35:55 2013 |
Koji | Optics | Characterization | Thickness of a curved mirror | Measured the thickness of a curved mirror:
Took three points separated by 120 degree.
S/N: C2, (0.2478, 0.2477, 0.2477) in inch => (6.294, 6.292, 6.292) in mm |
51
|
Wed Jan 2 07:45:39 2013 |
Koji | Optics | Characterization | First Contact test | Conclusion: Good. First contact did not damage the coating surface, and reduced the loss
- Construct a cavity with A1 and C2
- Measure the transmission and FWHM (of TEM10 mode)
- Apply First Contact on both mirrors
- Measure the values again
Transmission:
2.66 +/- 0.01 V -> 2.83 +/- 0.01 V
==> 6.3% +/- 0.5 % increase
FWHM of TEM10:
Before: (66.1067, 65.4257, 66.1746) +/- (0.40178, 0.38366, 0.47213) [kHz]
After: (60.846, 63.4461, 63.7906) +/- (0.43905, 0.56538, 0.51756) [kHz]
==> 5.1% +/- 2.7% decrease
Question: What is the best way to measure the finesse of the cavity? |
52
|
Sun Jan 6 23:22:21 2013 |
Koji | Mechanics | General | SolidWorks model of the OMC suspension | 
|
53
|
Thu Jan 10 18:37:50 2013 |
Koji | Optics | Characterization | Wedging of the PZTs | Yesterday I measured the thickness of the PZTs in order to get an idea how much the PZTs are wedged.
For each PZT, the thickness at six points along the ring was measured with a micrometer gauge.
The orientation of the PZT was recognized by the wire direction and a black marking to indicate the polarity.
A least square fitting of these six points determines the most likely PZT plane.
Note that the measured numbers are assumed to be the thickness at the inner rim of the ring
as the micrometer can only measure the maximum thickness of a region and the inner rim has the largest effect on the wedge angle.
The inner diameter of the ring is 9mm.
The measurements show all PZTs have thickness variation of 3um maximum.
The estimated wedge angles are distributed from 8 to 26 arcsec. The directions of the wedges seem to be random
(i.e. not associated with the wires)
As wedging of 30 arcsec causes at most ~0.3mm spot shift of the cavity (easy to remember),
the wedging of the PZTs is not critical by itself. Also, this number can be reduced by choosing the PZT orientations
based on the estimated wedge directions --- as long as we can believe the measurements.
Next step is to locate the minima of each curved mirror. Do you have any idea how to measure them? |
54
|
Wed Jan 16 14:10:50 2013 |
Koji | Optics | Characterization | Autocollimator tests of optics perpendicularity/parallelism | The items:
- Autocollimator (AC) borrowed from Mike Smith (Nippon Kogaku model 305, phi=2.76", 67.8mm)
- Retroreflector (corner cube)
- Two V grooves borrowed from the 40m
Procedure:
- Autocollimator calibration
o Install the AC on a optical table
o Locate the corner cube in front of the AC.
o Adjust the focus of the AC so that the reflected reticle pattern can be seen.
o If the retroreflection and the AC are perfect, the reference reticle pattern will match with the reflected reticle pattern.
o Measure the deviation of the reflected reticle from the center.
o Rotate the retroreflector by 90 deg. Measure the deviation again.
o Repeat the process until total four coordinates are obtained.
o Analysis of the data separates two types of the error:
The average of these four coordinates gives the systematic error of the AC itself.
The vector from the center of the circle corresponds to the error of the retroreflector.
- Wedge angle measurement
To be continued |
55
|
Fri Jan 18 13:25:17 2013 |
Koji | Optics | Configuration | Autocollimator calibration | An autocollimator (AC) should show (0,0) if a retroreflector is placed in front of the AC.
However, the AC may have an offset. Also the retroreflector may not reflect the beam back with an exact parallelism.
To calibrate these two errors, the autocollimator is calibrated. The retroreflector was rotated by 0, 90, 180, 270 deg
while the reticle position are monitored. The images of the autocollimator were taken by my digital camera looking at the eyepiece of the AC.
Note that 1 div of the AC image corresponds to 1arcmin.
Basically the rotation of the retroreflector changed the vertical and horizontal positions of the reticle pattern by 0.6mdeg and 0.1mdeg
(2 and 0.4 arcsec). Therefore the parallelism of the retrorefrector is determined to be less than an arcsec. This is negligibly good for our purpose.
The offset changes by ~1div in a slanted direction if the knob of the AC, whose function is unknown, is touched.
So the knob should be locked, and the offset should be recorded before we start the actual work every time. |
56
|
Sat Jan 19 20:47:41 2013 |
Koji | Optics | Characterization | Wedge measurement with the autocollimator | The wedge angle of the prism "A1" was measured with the autocollimator (AC).
The range of the AC is 40 arcmin. This means that the mirror tilt of 40arcmin can be measured with this AC.
This is just barely enough to detect the front side reflection and the back side reflection.
The measured wedge angle of the A1 prism was 0.478 deg.
Ideally a null measurement should be done with a rotation stage. |
57
|
Tue Jan 22 11:10:25 2013 |
Koji | Facility | General | Eyeware storage and hooks for the face shields are installed | A carpenter has come to install the eyeware storage and hooks for the face shields. |
58
|
Tue Jan 22 17:56:32 2013 |
Koji | Mechanics | General | Rotary stage selection | Newport UTR80
Newport 481-A (SELECTED)
- Sensitivity: 15 arcsec
- Graduations: 1 deg
- Vernier: 5 arcmin
- Fine travel range: 5 deg
- With Micrometer
Newport RS40
- Sensitivity: 16 arcsec
- Graduations: 2 deg
- Vernier: 12 arcmin
- Fine travel range: 10 deg
- Micrometer BM11.5
Newport RS65
- Sensitivity: 11 arcsec
- Graduations: 2 deg
- Vernier: 12 arcmin
- Fine travel range: 10 deg
- Micrometer SM-06 to be bought separately
Elliot science MDE282-20G
- Sensitivity: 5 arcsec
- Graduations: 2 deg
- Vernier: 10 arcmin
- Fine travel range: 10 deg
- Micrometer 2 arcmin/1div
- Metric
Suruga precision B43-110N
Thorlabs precision B43-110N
|
59
|
Mon Feb 4 00:39:08 2013 |
Koji | Optics | Characterization | Wedge measurement with the autocollimator and the rotation stage | Method:
- Mount the tombstone prism on the prism mount. The mount is fixed on the rotation stage.
- Locate the prism in front of the autocollimator.
- Find the retroreflected reticle in the view. Adjust the focus if necessary.
- Confirm that the rotation of the stage does not change the height of the reticle in the view.
If it does, rotate the AC around its axis to realize it.
This is to match the horizontal reticle to the rotation plane.
- Use the rotation stage and the alignment knobs to find the reticle at the center of the AC.
Make sure the reticle corresponds to the front surface.
- Record the micrometer reading.
- Rotate the micrometer of the rotation stage until the retroreflected reticle for the back surface.
- There maybe the vertical shift of the reticle due to the vertical wedging. Record the vertical shi
- Record the micrometer reading. Take a difference from the previous value.
Measurement:
- A1: α = 0.68 deg, β = 0 arcmin (0 div)
- A2: α = 0.80 deg, β = -6 arcmin (3 div down)
- A3: α = 0.635 deg, β = -1.6 arcmin (0.8 div down)
- A4: α = 0.650 deg, β = 0 arcmin (0div)
- A5: α = 0.655 deg, β = +2.4 arcmin (1.2 div up)
Analysis:
- \theta_H = ArcSin[Sin(α) / n]
- \theta_V = ArcSin[Sin(β) / n]/2
- A1: \theta_H = 0.465 deg, \theta_V = 0.000 deg
- A2: \theta_H = 0.547 deg, \theta_V = -0.034 deg
- A3: \theta_H = 0.434 deg, \theta_V = -0.009 deg
- A4: \theta_H = 0.445 deg, \theta_V = 0.000 deg
- A5: \theta_H = 0.448 deg, \theta_V = 0.014 deg
|
60
|
Wed Feb 6 02:34:10 2013 |
Koji | Optics | Characterization | Wedge measurement with the autocollimator and the rotation stage | Measurement:
- A6: α = 0.665 deg, β = +3.0 arcmin (1.5 div up)
- A7: α = 0.635 deg, β = 0.0 arcmin (0.0 div up)
- A8: α = 0.623 deg, β = - 0.4 arcmin (-0.2 div up)
- A9: α = 0.670 deg, β = +2.4 arcmin (1.2 div up)
- A10: α = 0.605 deg, β = +0.4 arcmin (0.2 div up)
- A11: α = 0.640 deg, β = +0.8 arcmin (0.4 div up)
- A12: α = 0.625 deg, β = - 0.6 arcmin (-0.3 div up)
- A13: α = 0.630 deg, β = +2.2 arcmin (1.1 div up)
- A14: α = 0.678 deg, β = 0.0 arcmin (0.0 div up)
- B1: α = 0.665 deg, β = +0.6 arcmin (0.3 div up)
- B2: α = 0.615 deg, β = +0.2 arcmin (0.1 div up)
- B3: α = 0.620 deg, β = +0.9 arcmin (0.45 div up)
- B4: α = 0.595 deg, β = +2.4 arcmin (1.2 div up)
- B5: α = 0.635 deg, β = - 1.8 arcmin (-0.9 div up)
- B6: α = 0.640 deg, β = +1.6 arcmin (0.8 div up)
- B7: α = 0.655 deg, β = +2.5 arcmin (1.25 div up)
- B8: α = 0.630 deg, β = +2.8 arcmin (1.4 div up)
- B9: α = 0.620 deg, β = - 4.0 arcmin (-2.0 div up)
- B10: α = 0.620 deg, β = +1.2 arcmin (0.6 div up)
- B11: α = 0.675 deg, β = +3.5 arcmin (1.75 div up)
- B12: α = 0.640 deg, β = +0.2 arcmin (0.1 div up)
Analysis:
- \theta_H = ArcSin[Sin(α) * n]
- \theta_V = ArcSin[Sin(β) / n]/2
- A6: \theta_H = 0.490 deg, \theta_V = 0.017 deg
- A7: \theta_H = 0.534 deg, \theta_V = 0.000 deg
- A8: \theta_H = 0.551 deg, \theta_V = -0.0023 deg
- A9: \theta_H = 0.482 deg, \theta_V = 0.014 deg
- A10: \theta_H = 0.577 deg, \theta_V = 0.0023 deg
- A11: \theta_H = 0.526 deg, \theta_V = 0.0046 deg
- A12: \theta_H = 0.548 deg, \theta_V = -0.0034 deg
- A13: \theta_H = 0.541 deg, \theta_V = 0.013 deg
- A14: \theta_H = 0.471 deg, \theta_V = 0.000 deg
- B1: \theta_H = 0.490 deg, \theta_V = 0.0034 deg
- B2: \theta_H = 0.563 deg, \theta_V = 0.0011 deg
- B3: \theta_H = 0.556 deg, \theta_V = 0.0051 deg
- B4: \theta_H = 0.592 deg, \theta_V = 0.014 deg
- B5: \theta_H = 0.534 deg, \theta_V = -0.010 deg
- B6: \theta_H = 0.526 deg, \theta_V = 0.0091 deg
- B7: \theta_H = 0.504 deg, \theta_V = 0.014 deg
- B8: \theta_H = 0.541 deg, \theta_V = 0.016 deg
- B9: \theta_H = 0.556 deg, \theta_V = -0.023 deg
- B10: \theta_H = 0.556 deg, \theta_V = 0.0068 deg
- B11: \theta_H = 0.475 deg, \theta_V = 0.020 deg
- B12: \theta_H = 0.526 deg, \theta_V = 0.0011 deg
Quote: |
Measurement:
- A1: α = 0.68 deg, β = 0 arcmin (0 div)
- A2: α = 0.80 deg, β = -6 arcmin (3 div down)
- A3: α = 0.635 deg, β = -1.6 arcmin (0.8 div down)
- A4: α = 0.650 deg, β = 0 arcmin (0div)
- A5: α = 0.655 deg, β = +2.4 arcmin (1.2 div up)
Analysis:
- \theta_H = ArcSin[Sin(α)*n]
- \theta_V = ArcSin[Sin(β) / n]/2
- A1: \theta_H = 0.465 deg, \theta_V = 0.000 deg
- A2: \theta_H = 0.547 deg, \theta_V = -0.034 deg
- A3: \theta_H = 0.434 deg, \theta_V = -0.009 deg
- A4: \theta_H = 0.445 deg, \theta_V = 0.000 deg
- A5: \theta_H = 0.448 deg, \theta_V = 0.014 deg
|
|
61
|
Thu Feb 7 21:35:46 2013 |
Koji | General | General | Dmass's loan of LB1005 / A2&C2 sent to Fullerton / First Contact @40m | Dmass borrowed the LB1005 servo amp from the OMC lab.
It happened this week although it seems still January in his head. Got it back on Mar 24th
The A2 and C2 mirrors have been sent to Josh Smith at Fullerton for the scatterometer measurement.
First Contact kit (incl. Peek Sheets)
Manasa borrowed the kit on Feb 7. Got it back to the lab.
|
62
|
Thu Feb 7 23:01:45 2013 |
Koji | Optics | Characterization | UV epoxy gluing test | [Jeff, Yuta, Koji]
Gluing test with UV-cure epoxy Optocast 3553-LV-UTF-HM
- This glue was bought in the end of October (~3.5 months ago).
- The glue was taken out from the freezer at 1:20pm.
- Al sheet was laid on the optical table. We made a boat with Al foil and pour the glue in it (@1:57pm)
- We brought two kinds of Cu wires from the 40m. The thicker one has the diameter of 1.62mm.
The thinner one has the diameter of 0.62mm. We decided to use thinner one being cut into 50mm in length.
- The OMC glass prisms have the footprint of 10mmx20mm = 200mm^2. We tested several combinations
of the substrates. Pairs of mirrors with 1/2" mm in dia. (127mm) and a pair of mirrors with 20mm in dia. (314mm).
- Firstly, a pair of 1/2" mirrors made of SF2 glass was used. A small dub on a thinner Cu wire was deposited on a mirror.
We illuminated the glue for ~10sec. When the surfaces of the pair was matched, the glue did not spread on the entire
surface. The glue was entirely spread once the pressure is applied by a finger. Glue was cured at 2:15pm. 12.873mm
thickness after the gluing.
Some remark:
1. We should be careful not to shine the glue pot by the UV illuminator.
2. The gluing surface should be drag wiped to remove dusts on the surface.
- Secondly, we moved onto 20mm mirror pair taken from the remnant of the previous gluing test by the eLIGO people.
This time about 1.5 times more glue was applied.
- The third trial is to insert small piece of alminum foil to form a wedge. The thickness of the foil is 0.041mm.
The glue was applied to the pair of SF2 mirror (1/2" in dia.). A small dub (~1mm in dia) of the glue was applied.
The glue filled the wedge without any bubble although the glue tried to slide out the foil piece from the wedge.
So the handling was a bit difficult. After the gluing we measured the thickness of the wedge by a micrometer gauge.
The skinny side was 12.837mm, and the thicker side was 12.885mm. This is to be compared with the total thickness
12.823mm before the gluing. The wedge angle is 3.8mrad (0.22deg). The glue dub was applied at 2:43, and the UV
illumination was applied at 2:46.
- At the end we glued a pair of fused silica mirrors. The total thickness before the gluing was 12.658 mm.
The glue was applied at 2:59pm. The thickness after the gluing is 12.663 mm.
This indicates the glue thickess is 5um.
|
63
|
Thu Feb 21 18:44:18 2013 |
Koji | Optics | Configuration | Perpendicularity test | Perpendicularity test of the mounting prisms:
The perpendicularity of the prism pieces were measured with an autocollimator.
Two orthogonally jointed surfaces forms a part of a corner cube.
The deviation of the reflected image from retroreflection is the quantity measured by the device.
When the image is retroreflected, only one horizontal line is observed in the view.
If there is any deviation from the retroreflection, this horizontal line splits into two
as the upper and lower halves have the angled wavefront by 4x\theta. (see attached figure)
The actual reading of the autocollimator is half of the wavefront angle (as it assumes the optical lever).
Therefore the reading of the AC times 30 gives us the deviation from 90deg in the unit of arcsec.
SN / measured / spec
SN10: 12.0 arcsec (29 arcsec)
SN11: 6.6 arcsec (16 arcsec)
SN16: 5.7 arcsec (5 arcsec)
SN20: -17.7 arcsec (5 arcsec)
SN21: - 71.3 arcsec (15 arcsec)
|
64
|
Wed Feb 27 18:18:48 2013 |
Koji | Optics | Configuration | More perpendicularity test | Mounting Prisms:
(criteria: 30arcsec = 145urad => 0.36mm spot shift)
SN Meas.(div) ArcSec Spec.
10 0.3989 11.97 29 good
11 0.2202 6.60 16 good
16 0.1907 5.72 5
good
20 -0.591 -17.73 5
good
21 -2.378 -71.34 15
21 -1.7 -51. 15
01 -0.5 -15. 52
02 -2.5 -75. 48
06 -1.0 -30. 15
good
07 1.7 51. 59
12 -2.2 -66. 40
13 -0.3 - 9. 12
good
14 -2.8 -84. 27
15 -2.5 -75. 50
17 0.7 21. 48
22 2.9 87.
63
Mirror A:
A1 -0.5 -15. NA good
A3 0.5 15. NA
good
A4 0.9 27. NA
good
A5 0.4 12. NA
good
A6 0.1 3.
NA good
A7 0.0 0.
NA good
A8 0.0
0.
NA good
A9 0.0
0.
NA good
A10 1.0
30.
NA good
A11 0.3
9.
NA good
A12 0.1
3.
NA good
A13 0.0
0.
NA good
A14 0.6
18.
NA good
Mirror B:
B1 -0.9
-27.
NA good
B2 -0.6
-18.
NA good
B3 -0.9
-27.
NA good
B4 0.7
21.
NA good
B5 -1.1
-33.
NA
B6 -0.6
-18.
NA good
B7 -1.8
-54. NA
B8 -1.1
-33.
NA
B9 1.8
54.
NA
B10 1.2
36.
NA
B11 -1.7
-51.
NA
B12 1.1
33.
NA
|
65
|
Fri Mar 1 23:06:15 2013 |
Koji | Optics | Configuration | More perpendicularity test final | Perpendicularity of the "E" mirror was measured.
Mounting Prisms:
(criteria: 30arcsec = 145urad => 0.36mm spot shift)
SN Meas.(div) ArcSec Spec.
10 0.3989 11.97 29 good
11 0.2202 6.60 16 good
16 0.1907 5.72 5
good
20 -0.591 -17.73 5
good
21 -2.378 -71.34 15
21 -1.7 -51. 15
01 -0.5 -15. 52
02 -2.5 -75. 48
06 -1.0 -30. 15
good
07 1.7 51. 59
12 -2.2 -66. 40
13 -0.3 - 9. 12
good
14 -2.8 -84. 27
15 -2.5 -75. 50
17 0.7 21. 48
22 2.9 87.
63
Mirror A:
A1 -0.5 -15. NA good
A3 0.5 15. NA
good
A4 0.9 27. NA
good
A5 0.4 12. NA
good
A6 0.1 3.
NA good
A7 0.0 0.
NA good
A8 0.0
0.
NA good
A9 0.0
0.
NA good
A10 1.0
30.
NA good
A11 0.3
9.
NA good
A12 0.1
3.
NA good
A13 0.0
0.
NA good
A14 0.6
18.
NA good
Mirror B:
B1 -0.9
-27.
NA good
B2 -0.6
-18.
NA good
B3 -0.9
-27.
NA good
B4 0.7
21.
NA good
B5 -1.1
-33.
NA
B6 -0.6
-18.
NA good
B7 -1.8
-54. NA
B8 -1.1
-33.
NA
B9 1.8
54.
NA
B10 1.2
36.
NA
B11 -1.7
-51.
NA
B12 1.1
33.
NA
Mirror E:
E1 -0.8 -24. NA
good
E2 -0.8 -24.
NA
good
E3 -0.25 - 7.5
NA
good
E4 -0.5 -15. NA
good
E5 0.8 24. NA
good
E6 -1.0 -30. NA
good
E7 -0.2 - 6. NA
good
E8 -0.8 -24. NA
good
E9 -1.0 -30. NA
good
E10 0.0 0. NA
good
E11 -1.0 -30. NA
good
E12 -0.3 - 9. NA
good
E13 -0.8 -24. NA
good
E14 -1.0 -30. NA
good
E15 -1.2 -36. NA
E16 -0.7 -21. NA
good
E17 -0.8 -24. NA
good
E18 -1.0 -30. NA
good
|
|