40m QIL Cryo_Lab CTN SUS_Lab TCS_Lab OMC_Lab CRIME_Lab FEA ENG_Labs OptContFac Mariner WBEEShop
  OMC elog  Not logged in ELOG logo
Entry  Thu Oct 18 20:23:33 2012, Koji, Optics, Characterization, Improved measurement Cav_scan_response_zoom_20121017.pdfdetailed_RoC_setup.pdf
    Reply  Mon Nov 19 13:33:14 2012, Koji, Optics, Characterization, Resuming testing mirror RoCs Cav_scan_response_zoom_20121016.pdf
Message ID: 41     Entry time: Mon Nov 19 13:33:14 2012     In reply to: 31
Author: Koji 
Type: Optics 
Category: Characterization 
Subject: Resuming testing mirror RoCs 

In order to resume testing the curvatures of the mirrors, the same mirror as the previous one was tested.
The result looks consistent with the previous measurement.

It seems that there has been some locking offset. Actually, the split peaks in the TF@83MHz indicates
the existence of the offset. Next time, it should be adjusted at the beginning.

Curved mirror SN: C1
RoC: 2.5785 +/- 0.000042 [m]

Previous measurements
=> 2.5830, 2.5638 => sqrt(RoC1*RoC2) = 2.5734 m
=> 2.5844, 2.5666 => sqrt(RoC1*RoC2) = 2.5755 m

Attachment 1: Cav_scan_response_zoom_20121016.pdf  189 kB  | Hide | Hide all
Cav_scan_response_zoom_20121016.pdf
ELOG V3.1.3-